Search Results
Found 17 results
510(k) Data Aggregation
(169 days)
HANSEN MEDICAL, INC.
The Hansen Medical Magellan™ Robotic Catheter 9Fr is intended to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices.
The Magellan Robotic Catheter 9Fr Long Length (80 cm) is also indicated for use as a conduit for manual placement of therapeutic devices in the neuro vasculature.
The Magellan™ Robotic Catheter 9Fr is intended to be used with the Hansen Medical Magellan™ M Robotic System and accessories.
The proposed change to the Hansen Medical Magellan™ Robotic Catheter (MRC) 9Fr is to the Indications for Use only; there is no change to the fit, form, or function of the device or to any of its components and accessories cleared under K132369.
The MRC 9Fr is a telescoping design comprised of a steerable Leader within a steerable Guide and is designed to be used with the Hansen Medical Magellan™ Robotic System (Magellan System). The distal sections of each Hansen Catheter component articulate individually and the distal end of the Leader can extend up to 21cm beyond the distal end of the Guide.
The Leader is a tubular catheter with four pull wires articulating the distal 3cm. The Leader has a 6Fr outer diameter and 3Fr inner diameter, and is available in 125cm, 141cm, and 158cm lengths. The Leader fits inside the Guide and is only used in conjunction with the Guide. The distal end of the Leader has a 3cm flexible articulating section. The Leader has a radiopaque marker band between the articulating section and the atraumatic tip. The tip extends 4mm distal to the marker band.
The Guide is a tubular catheter with four pull wires articulating the distal 2.8cm. The Guide has a 6Fr inner diameter and 9Fr outer diameter. The Guide is available in 50cm, 65cm, or 80cm lengths. The Guide fits around the Leader and may be used with or without the Leader. When the Leader is removed from the Guide, a 6Fr compatible percutaneous catheter may be delivered through the Guide.
The Hansen Catheter family consists of three different lengths of Leader and Guide pairs, and all sizes are sterilized via EtO for single use only. However, the MRC 9Fr Short (50cm) and MRC 9Fr Medium (65cm) catheters are not included for the expanded indication as they are not long enough to access the target anatomy.
This document is a 510(k) premarket notification for the Hansen Medical Magellan Robotic Catheter 9Fr, which is intended to facilitate navigation to anatomical targets and provide a conduit for therapeutic devices. The submission focuses on expanding the indications for use of the 80cm long version of the catheter to include neurovasculature.
Here's an analysis of the provided text with respect to your request:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" in the traditional sense of a numerically defined threshold for performance for a new AI/medical device. Instead, it relies on demonstrating substantial equivalence to a predicate device. This is a common approach for Class II medical devices seeking 510(k) clearance in the US. The "performance" assessment is a comparison of technological characteristics and intended use.
Here's a table based on the "Performance Attribute and Characteristic Comparison" table (on page 8) for the expanded indication compared to the predicate device:
Performance Specification | Predicate Device: MicroVention Chaperon Guide (6Fr) | Predicate Device: MicroVention Chaperon Leader (5Fr) | Hansen Medical Magellan Robotic Catheter 9Fr Guide (Long) | Hansen Medical Magellan Robotic Catheter 9Fr Leader |
---|---|---|---|---|
Outer Diameter (OD) | 6Fr (0.084 in) | 9Fr (0.128 in) | 6Fr (0.084 in) | |
Inner Diameter (ID) | 5Fr (0.071 in) | 6Fr (0.084 in) | 3Fr (0.039 in) | |
Guidewire Compatibility | 0.035 in, 0.018 in, 0.014 in | 0.035 in | 0.035 in, 0.018 in, 0.014 in | 0.035 in, 0.018 in, 0.014 in |
Total Insertion Length | 95 cm | 117 cm | 80 cm | 101 cm |
Distal Flex Length | 7 cm | 7 cm | 2.7 cm | 3.0 cm |
Hydrophilic Coating Length | N/A | 15 cm | 45 cm | 85 cm |
Tip Configuration | STR, MP2, BUR | VTR, SIM2, JB2 | Configurable bend to any angle up to 90° in all directions | Configurable bend to any angle up to 180° in all directions |
Intended Use (Neurovasculature) | Yes (inclusive of neurovasculature) | Yes (inclusive of neurovasculature) | Yes (expanded to neurovasculature for 80cm length) | Yes (expanded to neurovasculature for 80cm length) |
Reported Device Performance (against the predicate): The document states: "The minor differences in the characteristics between the two products do not alter the performance and do not present any new issues of safety or effectiveness." It concludes that the device "is substantially equivalent with respect to intended use and technological characteristics as the predicate device, and is expected to perform equivalently in accessing the neuro vasculature." The "performance" is demonstrated through design validation and verification activities described in Section 18.0 Performance Testing (which is not included in the provided text but is referenced).
2. Sample size used for the test set and the data provenance
The document explicitly states: "Clinical evaluation is not required for this device." Therefore, there is no "test set" in the sense of patient data for clinical evaluation, nor is there information on sample size or data provenance for such a test. The evaluation relies on comparing technological characteristics and performance data from internal design validation and verification.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable, as no clinical test set was required or used.
4. Adjudication method for the test set
Not applicable, as no clinical test set was required or used.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a robotic steerable catheter, not an AI-assisted diagnostic or interpretative device that augments human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical robotic catheter, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The document does not detail specific "ground truth" as would be used in an AI or diagnostic imaging study. Instead, the "truth" for this device's performance relies on engineering and performance testing data (referenced as "design validation and verification activities described in Section 18.0 Performance Testing" on page 7) to ensure physical characteristics (dimensions, flexibility, guidewire compatibility, etc.) meet specifications and are comparable to the predicate device for the intended use in neurovasculature.
8. The sample size for the training set
Not applicable. This is not an AI/machine learning device that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This is not an AI/machine learning device that requires a training set.
Ask a specific question about this device
(77 days)
Hansen Medical, Inc.
The Hansen Medical Magellan™ Robotic Catheter eKit is intended to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices.
The Magellan™ Robotic Catheter eKit is intended to be used with the Hansen Medical Magellan™ Robotic System and accessories.
The Hansen Medical Magellan™ Robotic Catheter eKit (MRC eKit) incorporates Microcatheter Driver components to be used in conjunction with the Magellan Robotic Catheter 6Fr (MRC 6Fr ) cleared under K515463. Both the predicate device (MRC 6Fr) and the MRC eKit are comprised of a Guide (Outer Catheter) with dual bend articulating sections (distal and proximal) paired with a non-articulating Leader. The devices are both provided in two lengths (60cm and 95cm) and have been designed to be used with Hansen Medical Magellan Robotic System. Both the MRC 6Fr and the MRC eKit are intended to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices.
The MRC eKit is comprised of the Hansen Medical 6Fr Guide catheter that is found in the commercialized product MRC 6Fr, and Microcatheter Driver components. The Microcatheter Driver components include a Microcatheter Support, a Microcatheter Driver base, a Connector Tube, a Support Tube packaged with and for use with the Leader, and a Microcatheter Valve. These components function the same as the Wire Support in the MRC 6Fr device to robotically insert and retract the Leader. Modifications have been made to now allow for robotic insertion and retraction of third-party inner catheters / microcatheters compatible with 5F Guiding catheters (
Here's an analysis of the provided text regarding the acceptance criteria and study for the Hansen Medical Magellan Robotic Catheter eKit:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Test/Performance Metric | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Mechanical Performance | Dimensional Analysis | Not specified | All pre-determined acceptance criteria were met. |
Burst Pressure Testing | Not specified | All pre-determined acceptance criteria were met. | |
Joint Separation Force Testing | Not specified | All pre-determined acceptance criteria were met. | |
Functional Performance | Simulated Use Testing | Not specified | All pre-determined acceptance criteria were met. Specifically, the reliability of inserting and retracting a third-party microcatheter was confirmed. |
Missing Information: The document states that "All of the pre-determined acceptance criteria were met," but it does not explicitly list the specific numerical or qualitative acceptance criteria for each test. This is a significant gap in the provided information.
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample sizes used for any of the tests (Dimensional Analysis, Burst Pressure Testing, Joint Separation Force Testing, Simulated Use Testing).
The data provenance is from in-house design verification testing conducted by Hansen Medical, Inc. based in Mountain View, CA, USA. The studies appear to be prospective as they were conducted as part of the device's design verification.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
This information is not applicable to the type of testing performed. The studies described are engineering/mechanical and functional performance tests of a medical device, not clinical studies requiring expert ground truth for interpretation of medical data (e.g., image analysis, diagnosis). The "ground truth" here is the adherence to design specifications and mechanical performance standards, which would be assessed by engineers and quality control personnel.
4. Adjudication Method for the Test Set
This information is not applicable as the described tests are objective measurements of physical properties and functional operation, rather than subjective assessments requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. The provided text indicates that "Clinical evaluation is not required for this device." The studies focused on confirming the device's physical and functional performance against pre-determined criteria.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
This question is not applicable as the device is a physical robotic surgical catheter, not an AI algorithm.
7. The Type of Ground Truth Used
The "ground truth" for the tests performed was based on engineering design specifications and performance standards. For example, burst pressure would be measured against a specified maximum pressure tolerance dictated by the design. Similarly, joint separation force would need to meet a minimum threshold. The "reliability of inserting and retracting a third party microcatheter" in simulated use testing would be assessed against a defined successful operation criterion.
8. The Sample Size for the Training Set
This question is not applicable. The device is a physical medical instrument, not an AI model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This question is not applicable as there is no training set for this device.
Ask a specific question about this device
(78 days)
HANSEN MEDICAL, INC.
The Hansen Medical Magellan Robotic Catheter 6Fr is intended to be used to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices.
The Magellan Robotic Catheter 6Fr is intended to be used with the Hansen Medical Magellan Robotic System and accessories.
The Magellan Robotic Catheter 6Fr v1.2 and accessories are a modification of the predicate Magellan Robotic Catheter 6Fr (MRC 6Fr) and accessories cleared under K133552. Both the predicate device and the modified Magellan Robotic Catheter 6Fr are comprised of a Guide (Outer Catheter) with dual bend articulating sections (distal and proximal) paired with a nonarticulating Leader (Inner Catheter). The devices are provided in two lengths (60cm and 95cm). Like the predicate device, the MRC 6Fr v1.2 is designed to be used with Hansen Medical Magellan Robotic System and is intended to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices. Both the MRC 6Fr v1.2 and the predicate may also be used for delivery of diagnostic contrast agents. The modified MRC 6Fr is compatible with power contrast injection systems up to a maximum of 600 psi. Whereas, the predicate device is rated for a maximum pressure of 100 psi. The device is provided sterile and is intended for single use only. The catheter is one of several compatible devices available for use with Hansen Medical's Magellan Robotic System cleared under K111004, K132369 and K141614.
This document, an FDA 510(k) summary for the Hansen Medical Magellan Robotic Catheter 6Fr v1.2, is a premarket notification for a medical device. It focuses on demonstrating "substantial equivalence" to a previously cleared predicate device, rather than proving performance against specific acceptance criteria for a novel AI device or diagnostic. Therefore, the information required to answer your specific questions about acceptance criteria, a study proving the device meets those criteria, and details about a test set, expert involvement, and ground truth is largely not present in this document.
The document mainly describes modifications to an existing device (increased pressure rating for contrast injection) and uses design verification testing to show that the performance remains substantially equivalent to the predicate device.
However, I can extract the following information that is tangentially related or directly stated:
1. A table of acceptance criteria and the reported device performance
The document states:
- "All of the pre-determined acceptance criteria were met."
- "Testing performed on the Magellan Robotic Catheter 6Fr v1.2 included the following:
- Tensile Strength Testing
- Simulated Use Testing
- Pressure Burst Testing
- ISO MEM Elution Cytotoxicity Testing
- ASTM Hemolysis
- Chemical Characterization Testing"
However, the specific numerical or qualitative acceptance criteria themselves and the reported device performance values are NOT detailed in this public summary. This level of detail would typically be found in the full test reports submitted to the FDA, not in the summary document.
Acceptance Criteria | Reported Device Performance |
---|---|
(Not specified in document) | (Not specified in document) |
Tensile Strength acceptable limits | Met |
Simulated Use successful operation | Met |
Pressure Burst resistance to 600 psi | Met |
ISO MEM Elution Cytotoxicity within limits | Met |
ASTM Hemolysis within limits | Met |
Chemical Characterization within limits | Met |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample size: Not specified. The document only mentions "testing" and "design verification testing."
- Data provenance: Not specified, but given it's a 510(k) for a US market device, the testing would typically be conducted under US regulatory standards. It's pre-market testing, so it's inherently prospective for the device modifications being assessed.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable / Not specified. This type of information relates to clinical studies, particularly for diagnostic devices or AI, which is not the focus of this 510(k) given its scope is modification of a previously cleared Class II catheter. The document explicitly states: "Clinical evaluation is not required for this device."
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable / Not specified. See point 3.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is not an AI device or a diagnostic device. It's a robotic catheter used for navigation and delivery of therapeutic devices. MRMC studies are not relevant here.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a physical medical device (catheter) used with a robotic system, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For the engineering tests conducted (Tensile Strength, Pressure Burst, etc.), the "ground truth" would be the engineering specifications and validated test methods. For example, a pressure burst test would have a defined pass/fail pressure, which is the ground truth. There is no expert consensus, pathology, or outcomes data used as "ground truth" for these types of mechanical and material tests.
8. The sample size for the training set
- Not applicable. This isn't an AI/machine learning device that requires a training set.
9. How the ground truth for the training set was established
- Not applicable. This isn't an AI/machine learning device.
In summary: This document is a regulatory submission for a minor modification to an existing robotic catheter. It confirms that the modified device remains substantially equivalent to its predicate. The proof of meeting acceptance criteria lies in standard engineering and material compatibility testing, not in clinical studies involving human interpretation or AI performance that would require the detailed information you are seeking.
Ask a specific question about this device
(27 days)
HANSEN MEDICAL, INC.
The Hansen Medical Magellan Robotic System and accessory components are intended to be used to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices.
The Magellan Robotic System is intended to be used with compatible Hansen Medical robotically steerable catheters.
The Hansen Medical Magellan Robotic System and Accessory Components are designed to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices. The fundamental concept of the system is based on a master/slave control system that enables and visualizes positioning of a steerable catheter tip at a desired point inside the vasculature, while enabling a physician to remain seated and away from the x-ray radiation source. The modification to the Magellan Robotic System is software update referred to as Magellan v1.9.1.
This document is a 510(k) premarket notification for the Hansen Medical Magellan Robotic System and Accessory Components (K151730). It primarily details a software update (Magellan v1.9.1) to an already cleared device (K141614). The core argument for substantial equivalence relies on the fact that the modifications do not change the intended use, fundamental scientific technology, or operating principles.
As such, this submission does not describe a study to prove a device meets acceptance criteria in the way a new or significantly modified device submission might. Instead, it aims to demonstrate that a software update to an existing device does not degrade performance and maintains substantial equivalence.
Therefore, many of the requested elements (like sample size for test sets, number of experts for ground truth, MRMC studies, standalone performance details, training set size, etc.) are not applicable or not provided in this type of submission because the focus is on maintaining existing safety and effectiveness rather than establishing new performance benchmarks.
However, I can extract information related to the acceptance criteria and the study type that was mentioned:
1. Table of acceptance criteria and the reported device performance
Acceptance Criteria Type | Reported Device Performance |
---|---|
Software Verification Testing | All pre-determined acceptance criteria were met. |
System Validation Testing | All pre-determined acceptance criteria were met. |
Important Note: The document states that "All of the pre-determined acceptance criteria were met," but it does not explicitly list what those specific acceptance criteria were (e.g., specific thresholds for accuracy, reliability, or safety metrics). The document focuses on confirming that the updated software did not introduce new risks or deviations from the predicate device's expected performance.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size: Not specified.
- Data Provenance: Not specified. Given the nature of software verification and system validation, these would typically be internal laboratory tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable/Not specified. This type of information is typically relevant for clinical studies or studies involving human interpretation of data, which was not the focus here. The performance was assessed through engineering and system-level tests.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable/Not specified. Adjudication methods are typically used in clinical trials where multiple human readers assess cases, which is not the case for this software update submission.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. An MRMC comparative effectiveness study was not done. This submission concerns a software update to a robotic navigation system, not an AI-assisted diagnostic tool for human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, in a sense. The "Software Verification Testing" and "System Validation Testing" represent standalone evaluations of the updated software and system to confirm they perform as expected without human intervention impacting the robotic movement or calculations directly. However, the performance is evaluated against the system's designed specifications, not necessarily an "algorithm only" in the AI sense.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For "Software Verification Testing" and "System Validation Testing," the ground truth would be based on the design specifications, functional requirements, and safety standards established for the device. For example, a navigation system's ground truth could be its ability to accurately move the catheter to a programmed position within a defined tolerance. It does not involve medical ground truth like pathology or expert consensus on a diagnosis.
8. The sample size for the training set
- Not applicable/Not specified. This document pertains to a software update to an existing robotic control system, not a machine learning or AI model that uses a training set in the conventional sense.
9. How the ground truth for the training set was established
- Not applicable. (See #8)
Ask a specific question about this device
(233 days)
HANSEN MEDICAL, INC.
Ask a specific question about this device
(52 days)
Hansen Medical
The Hansen Medical Magellan Robotic System and accessory components are intended to be used to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices.
The Magellan Robotic System is intended to be used with compatible Hansen Medical robotically steerable catheters.
The Hansen Medical Magellan Robotic System and accessory components are designed to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices. The fundamental concept of the system is based on a master/slave control system that enables and visualizes positioning of a steerable catheter tip at a desired point inside the vasculature, while enabling a physician to remain seated and away from the x-ray radiation source. The modification adds a mechanical joint between the Magellan Robotic System Rail and Adapter Plate to facilitate removal.
The Magellan Robotic System and accessory components are compatible with Hansen Medical robotically steerable catheters approved under previous 510(k)s.
This document describes a 510(k) premarket notification for the Hansen Medical Magellan Robotic System, which is a steerable catheter control system. The submission is for a modification to an already cleared device (K132369).
Here's the breakdown of the acceptance criteria and the study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Reported Device Performance |
---|---|
Visual and Dimensional Verification Testing | All pre-determined acceptance criteria were met. |
IEC 606061-1 Testing | All pre-determined acceptance criteria were met. |
Table Compatibility | All pre-determined acceptance criteria were met. |
Life cycle Testing | All pre-determined acceptance criteria were met. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not detail specific sample sizes for the visual, dimensional, IEC, table compatibility, or life cycle testing. It simply states that testing was performed and all pre-determined acceptance criteria were met. The data provenance is not explicitly mentioned (e.g., country of origin or retrospective/prospective). Given that the testing categories are related to engineering and regulatory standards, it's highly likely this was internal, prospective testing performed by the manufacturer or a contracted lab.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
Not applicable. The testing described (visual, dimensional, IEC, table compatibility, life cycle) are engineering and technical verification tests, not clinical performance tests requiring expert ground truth establishment for a test set in the traditional sense of AI/clinical studies.
4. Adjudication Method for the Test Set
Not applicable. This is not a study involving human readers or clinical adjudication. The acceptance criteria were met based on the results of the specified engineering and technical tests.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was performed or described. This submission is for a robotic system for facilitating navigation in the peripheral vasculature, not an AI-assisted diagnostic or interpretative device that would typically involve human readers. The document explicitly states: "No additional clinical evaluation of the Magellan Robotic System is required as a result of these changes."
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This is not directly applicable in the context of an algorithm's standalone performance for diagnostic or interpretive tasks. The "device" itself (the Magellan Robotic System) operates in a master/slave control system with a physician-in-the-loop for navigation. The testing performed was "Design verification testing," which focuses on the device's technical specifications and adherence to standards rather than an algorithm's standalone clinical accuracy.
7. The Type of Ground Truth Used
The "ground truth" for the engineering and technical tests would be the established specifications, standards (e.g., IEC 606061-1), and design requirements for the device. For example, for dimensional verification, the ground truth would be the engineering drawings and specified tolerances. For IEC testing, the ground truth would be the pass/fail criteria defined by the standard.
8. The Sample Size for the Training Set
Not applicable. This device is not an AI/ML product developed using a training set in the typical sense. It is a robotic system with physical and operational characteristics verified through engineering tests.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no training set for this type of device and submission.
Ask a specific question about this device
(80 days)
Hansen Medical
The Hansen Medical Magellan Robotic Catheter 6Fr is intended to be used to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices.
The Magellan Robotic Catheter 6Fr is intended to be used with the Hansen Medical Magellan Robotic System and accessories.
The Magellan Robotic Catheter 6Fr (MRC 6Fr) and accessories are a modification of the predicate Magellan Robotic Catheter 9Fr (MRC 9Fr) and accessories cleared under K132369. The Magellan Robotic Catheter 6Fr is a smaller size catheter comprised of a Guide (Outer Catheter) with dual bend articulating sections (distal and proximal) paired with a non-articulating Leader (Inner Catheter). The device is provided in two lengths (60cm and 95cm). Like the predicate device, the MRC 6Fr is designed to be used with Hansen Medical Magellan Robotic System and is intended to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices. The MRC 6Fr may also be used for delivery of diagnostic contrast agents. The device is provided sterile and is intended for single use only. The catheter is designed to expand Hansen Medical's market offering of compatible devices available for use with the Magellan Robotic System cleared under K132369.
The provided text describes a 510(k) summary for the Hansen Medical Magellan Robotic Catheter 6Fr. It is a submission for a modified device, primarily focusing on its substantial equivalence to a predicate device (Magellan Robotic Catheter 9Fr).
Therefore, the acceptance criteria and study information pertain to demonstrating this substantial equivalence through non-clinical testing.
Here's the breakdown of the information requested, based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document states: "All of the pre-determined acceptance criteria were met." However, it does not explicitly list the quantitative acceptance criteria for each test or the specific numerical performance results. It only lists the types of tests performed.
Test Category | Acceptance Criteria | Reported Device Performance |
---|---|---|
Material & Physical Properties | (Not explicitly stated, but implied to be within acceptable ranges for safety and function) | All pre-determined acceptance criteria were met. |
Visual and Dimensional | (Not explicitly stated) | Met. |
Pressure Leak Testing | (Not explicitly stated) | Met. |
Vacuum Testing | (Not explicitly stated) | Met. |
Tensile Strength Testing | (Not explicitly stated) | Met. |
Flush Testing | (Not explicitly stated) | Met. |
Articulation/Bending Stiffness Testing | (Not explicitly stated) | Met. |
Fatigue Testing | (Not explicitly stated) | Met. |
Guidewire Testing | (Not explicitly stated) | Met. |
Simulated Use Testing | (Not explicitly stated) | Met. |
Biocompatibility Testing | (According to ISO10993-1 standards) | Determined to be biocompatible. All pre-determined acceptance criteria were met. |
ISO MEM Elution Cytotoxicity Test | (Not explicitly stated for specific test, but aligned with ISO 10993-1) | Met. |
ISO Kligman Maximization Test | (Not explicitly stated for specific test, but aligned with ISO 10993-1) | Met. |
ISO Intracutaneous Reactivity Study | (Not explicitly stated for specific test, but aligned with ISO 10993-1) | Met. |
ISO Acute Systemic Toxicity Study | (Not explicitly stated for specific test, but aligned with ISO 10993-1) | Met. |
ASTM Hemolysis | (Not explicitly stated for specific test, but aligned with ISO 10993-1) | Met. |
ISO Thrombogenicity Study | (Not explicitly stated for specific test, but aligned with ISO 10993-1) | Met. |
Prothrombin Time (PT) Assay ISO | (Not explicitly stated for specific test, but aligned with ISO 10993-1) | Met. |
ISO Complement Activation Assay | (Not explicitly stated for specific test, but aligned with ISO 10993-1) | Met. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the sample sizes used for any of the non-clinical tests. It also does not provide information on the country of origin of the data or whether the data was retrospective or prospective. Given these are non-clinical (bench and lab) tests, these details are typically less relevant than for clinical studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable and not provided. The study performed consists of non-clinical, benchtop and lab-based engineering, and biocompatibility tests, not clinical studies requiring expert ground truth for interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable and not provided. As noted above, the study involved non-clinical tests, which do not typically involve adjudication methods for "ground truth" in the way clinical diagnostic image interpretation studies would. The "ground truth" for these tests would be the established scientific and engineering standards and methods.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done. The document explicitly states: "Clinical evaluation is not required for this device." This is a submission for a modified steerable catheter, demonstrating substantial equivalence through non-clinical testing.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable. The device is a robotic catheter, not an algorithm or AI system for diagnostic interpretation that would have "standalone" performance in the context of diagnostic accuracy. The performance testing described is for the physical and biological characteristics of the catheter itself.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the non-clinical tests described would be the established scientific and engineering principles, specifications, and regulatory standards to which the device was tested. For example, for biocompatibility, the ground truth is defined by ISO 10993-1, and for physical tests, it's defined by engineering specifications and industry standards for catheter performance (e.g., specific thresholds for tensile strength, leak rates, articulation angles, etc.). These are not "expert consensus" or "pathology" in the diagnostic sense.
8. The sample size for the training set
This information is not applicable and not provided. This device is a physical medical device (catheter) and not an AI/ML algorithm that requires a "training set."
9. How the ground truth for the training set was established
This information is not applicable and not provided for the same reasons as point 8.
Ask a specific question about this device
(59 days)
HANSEN MEDICAL, INC.
Magellan Robotic System: The Hansen Medical Magellan Robotic System and accessory components are intended to be used to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices.
The Magellan Robotic System is intended to be used with compatible Hansen Medical robotically steerable catheters.
Magellan Robotic Catheter 9Fr. The Hansen Medical Magellan Robotic Catheter 9Fr is intended to be used to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices.
The Magellan Robotic Catheter 9Fr is intended to be used with the Hansen Medical Magellan Robotic System and accessories.
The Hansen Medical Magellan Robotic System and Magellan Robotic Catheter 9Fr and Accessory Components are designed to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices. The fundamental concept of the system is based on a master/slave control system that enables and visualizes positioning of a steerable catheter tip at a desired point inside the vasculature, while enabling a physician to remain seated and away from the x-ray radiation source. The modifications to the Magellan System were made to expand system-quide wire compatibility, provide alternate quide wire navigation using the Master Input Device in addition to the existing Workstation and Bedside Controllers and to provide support for future compatible Hansen vascular catheters.
This document describes a Special 510(k) submission for modifications to the Hansen Medical Magellan Robotic System and Magellan Robotic Catheter 9Fr. This type of submission focuses on demonstrating that the modified device remains substantially equivalent to the previously cleared predicate device, rather than proving new clinical effectiveness. Therefore, the information typically associated with studies proving device performance against acceptance criteria (like sample sizes, ground truth establishment, or clinical effectiveness studies) is largely absent or stated as not required.
Here's an analysis based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list quantitative acceptance criteria in a table format with specific performance metrics (e.g., sensitivity, specificity, accuracy) for an AI/algorithm-based device. This is because the submission is for a robotic catheter system, not an AI diagnostic tool.
Instead, the "acceptance criteria" are implied by the comprehensive design verification and validation testing performed to ensure the modified system continues to meet design specifications and customer requirements. The reported device performance is that it successfully passed these tests and was found to be substantially equivalent to the predicate device.
Acceptance Criteria Category (Implied) | Reported Device Performance |
---|---|
System Set Up | Test Passed |
Catheter Installation | Test Passed |
System-Guide Wire Compatibility | Test Passed |
Procedure Simulation (Catheter & Guide Wire Navigation) | Test Passed |
Magellan Catheters Drive Modes | Test Passed |
User Interface | Test Passed |
3D Controller | Test Passed |
System Status and Error Message Handling | Test Passed |
System Power Cycle | Test Passed |
Emergency and Configuration | Test Passed |
Electrical Safety | Test Passed |
Risk analysis (ISO 14971 compliance) | Activities Completed |
Substantial Equivalence to Predicate Device | Demonstrated through non-clinical testing and comparison of characteristics, indicating successful performance in all modified aspects. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a "test set" in the context of an AI algorithm with data samples. The testing described is verification and validation of a physical/software system. Therefore, details like country of origin or retrospective/prospective data are not applicable. The testing would involve a combination of simulated scenarios, bench testing, and potentially animal/cadaveric studies (though not detailed here for this specific submission).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not applicable. The "ground truth" for a robotic catheter system's performance is typically established through engineering specifications, regulatory standards, and objective measurements during verification and validation testing, not through expert consensus on medical images or diagnoses.
4. Adjudication Method for the Test Set
This information is not applicable. Adjudication methods are relevant for studies involving human interpretation or subjective assessments, which is not the primary focus of this submission for a robotic system.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
No MRMC study was done, nor would it typically be expected for this type of device modification submission. The device is a robotic system to facilitate navigation, not a diagnostic tool that humans would use to interpret cases. The focus is on the functional equivalence of the system, not on comparing human reading performance with and without AI assistance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
This is not applicable. The Magellan system is a master/slave control system, inherently designed for human-in-the-loop operation. It is not an autonomous algorithm operating without human intervention for diagnostic or treatment decisions.
7. The Type of Ground Truth Used
The "ground truth" for the verification and validation tests performed for this robotic system would be based on:
- Engineering Specifications: Whether the system performs according to its design parameters (e.g., catheter tip movement, guide wire compatibility).
- Customer Requirements: Whether the system meets the functional needs for which it was designed.
- Regulatory Standards: Compliance with relevant standards like electrical safety (implied by "Electrical Safety Test") and risk management (ISO 14971).
- Performance against Predicate Device: Demonstration of substantial equivalence in intended use, technological characteristics, and safety/effectiveness.
8. The Sample Size for the Training Set
This is not applicable. The Magellan system is not an AI/machine learning model in the sense of requiring a "training set" of data samples. Its software modifications relate to system functionality, user interface, and compatibility, not learning from a dataset.
9. How the Ground Truth for the Training Set Was Established
This is not applicable, as there is no "training set." The software modifications were likely developed and tested against established engineering requirements and functional specifications based on the device's intended use and the predicate device's performance.
Ask a specific question about this device
(30 days)
Hansen Medical
The Hansen Medical Artisan Extend Control Catheter is intended to facilitate manipulation, positioning and control, for collecting electrophysiological data within the heart atria with electro-anatomic mapping and recording systems, using the following percutaneous mapping catheters: the Polaris-Dx ™ Steerable Diagnostic catheters made by Boston Scientific Corporation and the Livewire™ Electrophysiology catheters made by St. Jude Medical.
The Artisan Extend Control Catheter is intended to be used with the Hansen Medical Sensei® X Robotic Catheter System.
The Hansen Medical Artisan Extend Control Catheter is designed to facilitate manipulation, positioning and control of mapping percutaneous catheters within the atria of the heart. The Artisan Extend Control Catheter consists of a Steerable Guide Catheter (SGC) and Sheath. The Control Catheter is designed to be used with the Hansen Medical Sensei® X Robotic Catheter System.
The Hansen Medical Artisan Extend Control Catheter is a Class II steerable guide catheter, a modification of an existing device (K073225). The device is intended to facilitate manipulation, positioning and control of mapping percutaneous catheters within the atria of the heart, specifically with the Hansen Medical Sensei® X Robotic Catheter System and certain Boston Scientific and St. Jude Medical catheters.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The submission states, "All of the pre-determined acceptance criteria were met with passing results." However, it does not explicitly list the quantitative acceptance criteria for each test. Instead, it lists the types of tests performed.
Test Category | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Functional Tests | Device performs as intended, no leaks, proper flush, adequate tensile strength, compatibility with Sensei X. | Met all pre-determined criteria and passed. |
Pressure Leak Test | (Presumably no leaks or within acceptable leak rate) | Passed |
Vacuum Test | (Presumably no failure under vacuum) | Passed |
Tensile Strength Test | (Presumably maintains structural integrity under specified load) | Passed |
Flush Test | (Presumably allows proper flushing) | Passed |
Compatibility with Sensei X Test | (Presumably proper integration and function with the robotic system) | Passed |
Biocompatibility Tests | No adverse biological reactions beyond accepted levels | Met all pre-determined criteria and passed. |
ISO MEM Elution Cytotoxicity Test | (Presumably non-cytotoxic) | Passed |
ISO Kligman Maximization Test | (Presumably non-sensitizing) | Passed |
ISO Intracutaneous Reactivity Study | (Presumably non-irritating) | Passed |
ISO Acute Systemic Toxicity Study | (Presumably non-toxic systemically) | Passed |
ASTM Hemolysis | (Presumably acceptable levels of hemolysis) | Passed |
ISO Thrombogenicity Study | (Presumably low thrombogenic potential) | Passed |
ISO Complement Activation Assay | (Presumably low complement activation) | Passed |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample sizes used for each of the performance and biocompatibility tests. It only states that design verification testing was performed.
Regarding data provenance: The tests were conducted by the manufacturer, Hansen Medical, Inc., as part of a 510(k) submission to the FDA. This indicates the data is from an internal testing environment rather than a clinical setting. Given the nature of these engineering and biocompatibility tests, they are inherently prospective for the device being evaluated. There is no information regarding the country of origin of the data beyond the manufacturer's location in Mountain View, CA, USA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable to the type of testing performed. The "ground truth" for these engineering and biocompatibility tests is established by industry standards (e.g., ISO, ASTM) and internal design specifications, not by expert consensus on clinical findings. For example, a leak test has an objective pass/fail criterion based on measurable parameters, not on an expert's interpretation.
4. Adjudication Method for the Test Set
This is not applicable. Adjudication methods (like 2+1, 3+1) are used in studies involving subjective human interpretation of data, typically in clinical evaluations or image analysis. The tests described are objective, pass/fail engineering and laboratory tests.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical evaluation was not required for this device." This device is a modification of a previously cleared predicate device, and the changes (simplified proximal flush assembly) were determined not to affect the intended use or fundamental scientific technology, precluding the need for clinical studies.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This concept is not applicable here. This is a medical device (a catheter), not a software algorithm or AI model. The performance is assessed through physical and biological testing, not through standalone algorithm evaluation.
7. The Type of Ground Truth Used
The "ground truth" for the tests performed is based on:
- Industry Standards: Many tests refer to ISO (International Organization for Standardization) and ASTM (American Society for Testing and Materials) standards, which define methodologies and acceptable ranges for performance and biocompatibility.
- Manufacturer Design Specifications: The "pre-determined acceptance criteria" would have been set by Hansen Medical based on the functional requirements of the device and comparison to the predicate device.
- Objective Measurement: For tests like Pressure Leak, Vacuum, and Tensile Strength, the ground truth is based on measurable physical properties and compliance with a defined range or threshold.
8. The Sample Size for the Training Set
This is not applicable. This device is not an AI/ML algorithm that requires a "training set." It is a physical medical device.
9. How the Ground Truth for the Training Set was Established
This is not applicable as there is no training set for a physical medical device.
Ask a specific question about this device
(414 days)
HANSEN MEDICAL, INC.
The Hansen Medical Vascular Catheter Control System, Hansen Catheter and accessory components are intended to be used to facilitate navigation to anatomical targets in the peripheral vasculature and subsequently provide a conduit for manual placement of therapeutic devices.
The Hansen Medical Magellan Robotic System and NorthStar Robotic Catheter and Accessory Components are designed to facilitate remote navigation to anatomical targets in the peripheral vasculature. Subsequent to navigation, the system provides a conduit for manual placement of therapeutic devices. The Magellan Robotic System's operating principles are similar to those of the Sensei X Robotic Catheter System. The fundamental concept of the robotic system is based on a master/slave control system that enables and visualizes positioning of a steerable catheter tip at a desired point inside the vasculature, while enabling a physician to remain seated and away from the x-ray radiation source.
Here's a breakdown of the acceptance criteria and study information for the Hansen Medical Magellan™ Robotic System, NorthStar™ Robotic Catheter, and Accessory Components, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The provided text does not explicitly state numerical acceptance criteria in a table format with corresponding performance results. However, it describes the types of tests conducted and concludes that the device successfully met established specifications and did not raise new questions of safety or effectiveness.
Therefore, the table below reflects what was stated as satisfactorily completed or demonstrated success rather than explicit numerical acceptance criteria.
Test Type | Acceptance/Performance Statement |
---|---|
Bench/Preclinical Testing | |
Visual and Dimensional Verification | Successfully executed; meet established specifications. |
Tensile Testing | Successfully executed; meet established specifications. |
Articulation Force Testing | Successfully executed; meet established specifications. |
Articulation Angle Testing | Successfully executed; meet established specifications. |
Guidewire insertion force | Successfully executed; meet established specifications. |
Tracking in simulated anatomy | Successfully executed; meet established specifications. |
Device Leak Testing | Successfully executed; meet established specifications. |
Device Bending Stiffness Testing | Successfully executed; meet established specifications. |
System Verification Testing | Successfully executed; meet established specifications. |
Electromagnetic Test (EMC) | Successfully executed; meet established specifications. |
Biocompatibility Testing | NorthStar Robotic Catheter successfully passed all tests (Cytotoxicity, Sensitization, Maximization, Acute System Toxicity, Intracutaneous Reactivity/Irritation, Complement Activation, Thrombogenicity, Pyrogenicity, Hemolysis, In Vitro Platelet Aggregation, Prothrombin time assay, Partial Thromboplastin time assay). |
Package Testing | Successfully executed; meet established specifications. |
Device Fatigue Testing | Successfully executed; meet established specifications. |
Shipping/Distribution Testing | Successfully executed; meet established specifications. |
Sterility Testing | Validated sterility assurance level (SAL) is 10⁻⁶ for NorthStar Robotic Catheter (EtO) and Accessory Components (gamma irradiation). |
Shelf Life and Life Cycle Testing | Successfully executed; meet established specifications. |
Coating Particulate Testing | Successfully executed; meet established specifications. |
Device Evaluation in Porcine model | Successfully executed; meet established specifications. |
System Electrical (IEC) Testing | Successfully executed; meet established specifications. |
Software Validation Test | Successfully executed; meet established specifications. |
Animal Studies (GLP) | |
5-day survival study | System functionality and performance met system design requirements in vivo. Resulted in fewer and less severe vascular injuries compared to manual control. |
30-day survival study | System functionality and performance met system design requirements in vivo. Resulted in fewer and less severe vascular injuries compared to manual control. |
Clinical Study | |
Navigation effectiveness | Robotic navigation successfully cannulated 20 of 20 target vessels (100% success rate). |
Angiographic assessment | Selective angiography of the target vessel completed in 20 cases. |
Robotic guidance of catheter/guidewire | System able to remotely navigate a robotically deflectable catheter and guide-wire into the target vessel in all 20 cases. NorthStar Robotic Catheter robotically navigated to targeted lesion in all 20 cases. |
Safety | No device-related adverse events. All patients free from access site complications at discharge. |
2. Sample Size Used for the Test Set and Data Provenance
The clinical study involved:
- Sample Size: 20 procedures (15 enrolled subjects).
- Data Provenance: The study was a "prospective, single center" study. The country of origin is not explicitly stated, but given the applicant's US address and the FDA submission, it is highly likely to be a US-based study. It was prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The provided text describes a clinical study focused on the device's performance in navigating and delivering therapeutic devices rather than an image-based diagnostic task requiring ground truth established by experts interpreting images. Therefore, this information is not applicable in the context of the provided text. The "ground truth" for the clinical study was the actual successful cannulation of vessels and completion of procedures, along with safety outcomes.
4. Adjudication Method for the Test Set
As this was a performance study of a robotic system in a clinical setting (not an image interpretation study), an "adjudication method" in the sense of reconciling differing expert opinions on a diagnosis is not applicable. The primary outcome measures (successful cannulation, completion of angiography, navigation to lesion, absence of adverse events) would have been observed and recorded by the operating physicians and study staff according to the study protocol.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There was no MRMC comparative effectiveness study described in the provided text. The clinical study was a single-arm study evaluating the device's performance, not comparing human readers with and without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
The provided text describes the device as a "master/slave control system" where the physician inputs commands to control the robotic catheter. This inherently describes a human-in-the-loop system. Therefore, a standalone (algorithm only) performance, in the sense of a fully autonomous AI system making decisions without human input, was not performed nor is it relevant to this device's intended use or design.
7. The Type of Ground Truth Used
- Bench/Preclinical Testing: Engineering specifications, ISO standards, and established biological/material science outcomes (e.g., successful visual verification, tensile strength values, biocompatibility panel results, sterility levels, tracking pathways in simulated anatomy).
- Animal Studies: Observed in-vivo functionality, performance against system design requirements, comparison of vascular injury rates (fewer and less severe injuries compared to manual control).
- Clinical Study: Direct observation of successful robotic navigation (cannulation, guidance to target), successful completion of angiography, absence of device-related adverse events, and ultrasound confirmation of no access site complications. This is essentially outcomes data and direct procedural success.
8. The Sample Size for the Training Set
The provided text describes the Magellan Robotic System as having operating principles "nearly identical" to the predicate Sensei X Robotic Catheter System (K102168). It is a robotic control system for a catheter, not a machine learning or AI algorithm that requires a "training set" in the conventional sense used for diagnostic imaging AI. Therefore, the concept of a "training set" with a specific sample size for a machine learning model is not applicable to this device as described. The system's "training" would have been through engineering design, software development, and iterative testing/refinement.
9. How the Ground Truth for the Training Set Was Established
As explained above, the concept of a "training set" for a machine learning model is not applicable here. The "ground truth" for the development of such a robotic system would be its adherence to engineering specifications, safety standards, and functional requirements established through design, simulation, and rigorous testing against known physical and mechanical principles.
Ask a specific question about this device
Page 1 of 2