Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K130582
    Date Cleared
    2014-02-04

    (336 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    PVC films form a barrier to body fluids and bloodborne Pathogens. The PVC rubber is water tight under normal conditions of use. Its tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. Poly (vinyl chloride) glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements.

    AI/ML Overview

    The provided text details the 510(k) summary for "Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)". This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing a detailed study description of an AI-powered device. Therefore, much of the requested information regarding AI device studies (like sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for AI) is not applicable or present in this document.

    However, I can extract the acceptance criteria and reported device performance for these medical gloves:

    1. Table of Acceptance Criteria and Reported Device Performance:

    CharacteristicsStandard (Acceptance Criteria)Device Performance
    DimensionASTM standard D 5250-06 (Reapproved 2011)Meets
    Physical PropertiesASTM standard D 5250-06 (Reapproved 2011)Meets
    Freedom from Pinholes21 CFR 800.20, ASTM D5151-06 (Reapproved 2011)Meets (AQL 2.5)
    Powder ResidualASTM D 5250-06 (Reapproved 2011) and D6124-06 (Reapproved 2011)Meets (
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1