Search Results
Found 5 results
510(k) Data Aggregation
(35 days)
HEYUAN HONGLI INDUSTRIES, INC.
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
POWDERED VINYL EXAM GLOVES, WHITE, Class I Device, 80LYZ
This is a 510(k) Summary for Powdered Vinyl Exam Gloves.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Standard/Test) | Reported Device Performance |
---|---|
ASTM D 5250-00E4 | Met or exceeded standard |
ASTM D 5151-99 | Met or exceeded standard |
ASTM D 6124-01 | Met or exceeded standard |
ISO 2859 | Met or exceeded standard |
MIL Std 105E | Met or exceeded standard |
Bio-Compatibility: Dermal Sensitization | Met or exceeded standard |
Bio-Compatibility: Primary Skin Irritation | Met or exceeded standard |
Bio-Burden (Procedure Method 8315) | Met or exceeded standard |
2. Sample Size for the Test Set and Data Provenance
The provided document does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective).
3. Number of Experts Used to Establish Ground Truth and Qualifications
This information is not applicable as the evaluations were based on standardized performance tests rather than expert interpretation of medical data or images.
4. Adjudication Method for the Test Set
This information is not applicable as the evaluations were based on standardized performance tests rather than human interpretation requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted as this device is a medical glove, not an AI-assisted diagnostic tool.
6. Standalone (Algorithm Only) Performance Study
A standalone performance study was not conducted as this device is a medical glove, not an algorithm.
7. Type of Ground Truth Used
The ground truth used was based on the objective measurements and pass/fail criteria defined by the listed industry and regulatory standards (ASTM, ISO, MIL Std, and bio-compatibility tests).
8. Sample Size for the Training Set
This information is not applicable as the device is not an AI/ML algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable as the device is not an AI/ML algorithm and therefore does not have a training set. The "ground truth" for demonstrating performance was established by the predefined specifications within the referenced ASTM, ISO, and MIL standards, and bio-compatibility testing procedures.
Ask a specific question about this device
(23 days)
HEYUAN HONGLI INDUSTRIES, INC.
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
POWDER-FREE POLY VINYL EXAM GLOVES, COLOR: WHITE
The document describes the acceptance criteria and the study for "POWDER-FREE POLY VINYL EXAM GLOVES, COLOR: WHITE."
Here's the breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Standard/Test) | Reported Device Performance |
---|---|
ASTM D 5250-00 | Met or exceeded |
ASTM D 5151-00 | Met or exceeded |
ASTM D 6124-00 | Met or exceeded |
ISO 2859 | Met or exceeded |
Bio Burden | Met or exceeded |
Bio-Compatibility: Dermal Sensitization | Met or exceeded |
Bio-Compatibility: Primary Skin Irritation | Met or exceeded |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for the specific tests (ASTM D 5250-00, D 5151-00, D 6124-00, ISO 2859, Bio Burden, Dermal Sensitization, Primary Skin Irritation). The data provenance is implied to be from tests performed on the device by Heyuan Hongli Industries, Inc., which is based in China. The submission date is July 11, 2002, suggesting these were retrospective tests conducted on the manufactured gloves.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document. The tests performed are standardized and do not typically involve human experts establishing a "ground truth" in the way a diagnostic AI device would. Instead, these are material and biological safety tests with predefined methodologies and pass/fail criteria.
4. Adjudication Method for the Test Set
This information is not applicable as the tests are standardized and have objective pass/fail criteria. There isn't an adjudication process involving multiple human reviewers to determine a "ground truth" for these types of tests.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC comparative effectiveness study was not done. This device is a medical glove, not an AI-powered diagnostic tool. Therefore, the concept of human readers improving with or without AI assistance is not relevant to this device.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This is not applicable. The device is a physical product (medical glove), not an algorithm or AI.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance is established by objective measurements and assessments based on the specified ASTM and ISO standards, as well as bio-compatibility tests. These standards define the acceptable physical properties (e.g., strength, dimensions), absence of holes, and biological safety (e.g., absence of skin irritation, allowable bioburden). There is no "expert consensus," "pathology," or "outcomes data" ground truth in the context of this device's testing.
8. The Sample Size for the Training Set
This information is not applicable. The device is a physical product and does not involve a training set as a machine learning model would. Production quality control would involve sampling from manufacturing batches, but this is distinct from "training data."
9. How the Ground Truth for the Training Set was Established
This information is not applicable as there is no "training set" for a medical glove in the context of AI/algorithm development.
Ask a specific question about this device
(28 days)
HEYUAN HONGLI INDUSTRIES, INC.
Ask a specific question about this device
(28 days)
HEYUAN HONGLI INDUSTRIES, INC.
Ask a specific question about this device
(23 days)
HEYUAN HONGLI INDUSTRIES, INC.
Ask a specific question about this device
Page 1 of 1