Search Results
Found 3 results
510(k) Data Aggregation
K Number
K202433Device Name
Aqueduct 200 Cervical Dilation Balloon Catheter
Manufacturer
GTIMD LLC
Date Cleared
2021-10-15
(416 days)
Product Code
PON
Regulation Number
884.4260Why did this record match?
Applicant Name (Manufacturer) :
GTIMD LLC
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Aqueduct 200 Cervical Dilation Balloon Catheter is intended to be used whenever cervical softening and dilation is desired. Some examples are: treatment of cervical stenosis, IUD placement and removal, Radium placement, drainage of uterine cavity, endometrial biopsy, uterine curettage, suction cannula aspiration, operative hysteroscopy.
Device Description
The Aqueduct 200 Cervical Dilation Balloon Catheter is a balloon cervical dilation catheter which enables the simultaneous dilation of both sides of the cervical canal. The subject device is an updated version of the Aqueduct 100 Cervical Dilation Balloon Catheter cleared under K160664 that merges two separate dilation balloons into one balloon, removes the anchor balloon, replaces the four-way hub with a three-way hub, and removes the stylet from the shaft. The catheter consists of a 2-lumen shaft. One lumen inflates a cylindrical dilation balloon. A second lumen is for infusion of saline solution and also contains a camera and camera cable. In use, the catheter is inserted through the vagina and cervical canal and into the uterus. Using visual feedback from the forward-looking camera and LED light source the physician advances the catheter through the cervical canal and identifies the end of the canal when entering the internal orifice of the uterus. When the physician identifies the orifice, the catheter is positioned inside the cervix. This aligns the cylindrical dilation balloon at the internal cervical os and external cervical os. The balloon is inflated with 12.5 ml of saline, providing gradual mechanical dilation of the cervix. The camera cable on the catheter connects to a camera module that provides power to the camera and allows connection to an HDMI port on a monitor to assess proper device placement during a clinical procedure. After 3 minutes of dilation of the internal and external orifices of the uterus, a controlled injection of 1-2.5 ml of saline may be made through the catheter infusion lumen which exits on the distal end of the catheter. Evidence that cervical dilation is complete can be determined once droplets of the saline injection are observed exiting through the external opening of the cervix. Optimal dilation of 8-9 mm within the cervical canal is typically achieved following 5 minutes of balloon dilation. The entire procedure from catheter insertion to removal is completed in 6-7 minutes. The deflation of the balloon is conducted by attaching a luer lock syringe to the swabable valve and removing saline from the balloon. The subject device is packaged in a mylar/Tyvek pouch and ethylene oxide (EO) sterilized to a SAL 10-6. The packaged device has a shelf-life of 15 months.
Ask a Question
K Number
K202427Device Name
Aqueduct 100 Plus Cervical Dilation Balloon Catheter
Manufacturer
GTIMD LLC
Date Cleared
2020-11-06
(73 days)
Product Code
PON
Regulation Number
884.4260Why did this record match?
Applicant Name (Manufacturer) :
GTIMD LLC
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Aqueduct 100 Plus Cervical Dilation Balloon Catheter is intended to be used whenever cervical softening and dilation is desired. Some examples are: treatment of cervical stenosis, IUD placement and removal, Radium placement, drainage of uterine cavity, endometrial biopsy, uterine curettage, suction cannula aspiration, operative hysteroscopy.
Device Description
The Aqueduct 100 Plus Cervical Dilation Balloon Catheter is a balloon cervical dilation catheter which enables the simultaneous dilation of both sides of the cervical canal. The subject device is an updated version of the Aqueduct 100 Cervical Dilation Balloon Catheter cleared under K160664 that merges two separate dilation balloons into one balloon, removes the anchor balloon, replaces the four-way hub with a three-way hub, and removes the stylet from the shaft. The catheter consists of a 2-lumen shaft. One lumen inflates a cylindrical dilation balloon. A second lumen is for infusion of saline solution. In use, the catheter is inserted through the vagina and cervical canal and into the uterus. The catheter is positioned inside the cervix. The user aligns the cylindrical balloon at the internal cervical os and external cervical os. The balloon is inflated with 12.5 ml of saline, providing gradual mechanical dilation of the cervix. After 3 minutes of dilation of the internal and external orifices of the uterus, a controlled injection of 1-2.5 ml of saline may be made through the catheter infusion lumen which exits on the distal end of the catheter. Evidence that cervical dilation is complete can be determined once droplets of the saline injection are observed exiting through the external opening of the cervix. Optimal dilation of 8-9 mm within the cervical canal is typically achieved following 5 minutes of balloon dilation. The entire procedure from catheter insertion to removal is completed in 6-7 minutes. The deflation of the balloon is conducted by attaching a luer lock syringe to the swabable valve and removing saline from the balloon. The subject device is packaged in a mylar/Tyvek pouch and ethylene oxide (EO) sterilized to a SAL 10-6. The packaged device has a shelf-life of 15 months.
Ask a Question
K Number
K160664Device Name
Aqueduct 100 Cervical Dilator
Manufacturer
GTIMD LLC
Date Cleared
2016-08-01
(145 days)
Product Code
PON
Regulation Number
884.4260Why did this record match?
Applicant Name (Manufacturer) :
GTIMD LLC
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Aqueduct 100 Cervical Dilation Balloon Catheter is intended to be used whenever cervical softening and dilation is desired. Some examples are: treatment of cervical stenosis, IUD placement and removal, Radium placement, drainage of uterine cavity, endometrial biopsy, uterine curettage, suction cannula aspiration, operative hysteroscopy.
Device Description
The Aqueduct 100 Cervical Dilator is a triple-balloon cervical dilation catheter which enables the simultaneous dilation of both sides of the cervical canal. The catheter consists of a 3-lumen shaft. One lumen inflates an anchoring balloon. A second lumen inflates the 2 cylindrical dilation balloons. The third lumen is for infusion of saline solution and also contains a fixed stiffening stylet to add rigidity to the catheter. The stylet is permanently sealed within the infusion lumen. In use, the catheter is inserted through the vagina and cervical canal and into the uterus where the distal spherical balloon is inflated. The catheter is then withdrawn until the spherical balloon comes into contact with the internal orifice of the uterus, which positions and anchors the catheter within the cervix. This aligns the two cylindrical balloons at the internal cervical os and external cervical os. The balloons are inflated with saline providing gradual mechanical dilation of the cervix.
After 3 minutes of dilation of the internal and external orifices of the uterus, a gentle, controlled injection of 1ml, and up to 2.5ml, of saline can be made through the catheter infusion lumen which exits between the two balloons (and inside the cervical canal). Evidence that cervical dilation is complete can be determined once droplets of the saline injection are observed entering through the external opening of the cervix. Optimal dilation of 8-9mm within the cervical canal is typically achieved following 5 minutes of balloon dilation. The entire procedure from catheter insertion to removal is completed in 6-7 minutes.
The Aqueduct 100 Cervical Dilator is packaged in a mylar/Tyvek pouch and EO sterilized to SAL 10-6.