K Number
K964219Device Name
ZIP CONDOM CATHETER
Manufacturer
Date Cleared
1997-08-11
(293 days)
Regulation Number
876.5250Why did this record match?
Applicant Name (Manufacturer) :
GOULTER MEDICAL, INC.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
1. For Penile Attachment Only
2. Non-indwelling Urinary Continence Aid
3. Management of Urinary Incontinence
4. Urine Collection Device
Device Description
The ZIP™ Condom Catheter has four basic elements: 1) the condom/sheath with integral urine collection compartment; 2) the Velcro-like band and support strap; 3) the Application ring; and 4) the specially designed male underpants brief.
The condom/sheath with integral urine collection compartment forms a bulbous tube approximately 10 ½ inches long. The tube is divided into two sections, the condom/sheath and the urine collection compartment. Non-return flow valves separate the condom/sheath portion from the urine collection compartment. The distal end of the urine collection compartment is affixed with a twist-open twistclosed stop-cock drain valve. The urine collection capacity is approximately 450 ml. Typical human urine output thus yields a time related capacity of about 4 to 6 hours.
A series of 4 sizes differing in sheath diameter and length will be offered. Specific name designators for the sizes are vet to be determined.
Design: The ZIP™ Condom Catheter expands upon existing male urine continence devices by first approaching the device as a "total" or "system" concept rather than just a collection of items originally developed for other urine collection purposes.
Only the essential elements of the ideal male urine continence system were considered. First, the system had to be easy to apply and use. Second, it should allow use of traditional male urination methods, i.e., the stand-up urinals found in public restrooms. Third, the system should be small enough to go unnoticed during the typical activities of daily living. Fourth, the system should have adequate capacity so that urine disposal intervals are not unreasonably short. Lastly, the system should be reusable and easily cleaned and sanitized.
This is a natural latex containing product. The ZIP™ Condom Catheter is made from the exact same natural latex formulation as the Everyday - Self-Adhesive Urinary External Catheter made by Hollister. Inc. The use of pure latex to construct the ZIP™ Condom Catheter requires the inclusion of certain FDA required cautionary statements. Latex containing devices must legibly bear an appropriate allergic reaction statement on the device labeling. In conformance with this FDA requirement, Goulter Medical, Inc. attaches a label containing the following cautionary statement to the device. "This product is made from natural rubber latex which may cause allergic reactions in some individuals".