Search Results
Found 3 results
510(k) Data Aggregation
(46 days)
Fuji Systems Corporation
The CELLO II Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of intravasular catheters into a selected blood vessel in the peripheral and neuro vasculature systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures.
The CELLO II Balloon Guide Catheter is a coaxial-lumen, braid-reinforced, variable stiffness catheter with one radiopaque ring marker on the distal end of the balloon and a bifurcated Luer hub on the proximal end. A compliant urethane balloon is mounted on the distal end. Balloon Guide Catheter dimensions and recommended balloon volumes are indicated on the product label. Each catheter is supplied with inserters, a Y-Connector, 3-way stopcocks and syringes.
The CELLO II Balloon Guide Catheters are intended for use in hospitals or other health care facilities which are equipped with trained personnel and specialized equipment to perform peripheral and neurovascular procedures.
The CELLO II Balloon Guide Catheter is intended for facilitating the insertion and guidance of intravascular catheters into selected blood vessels in the peripheral and neuro vasculature systems. The tip of the catheter features a balloon of urethane rubber. Radiopaque ring marker identifies the distal end of the balloon. The shaft is a dual lumen type with coaxial structure featuring a built-in braided stainless steel coil.
The materials of construction are similar to those used in many other similar catheters. The shaft is made of polyurethane, polyamide, stainless steel, and fluorine resin; the balloon is made of urethane. Patient contact is of limited duration, less than 24 hours.
Here's a breakdown of the acceptance criteria and study information for the CELLO II Balloon Guide Catheter, based on the provided document:
This document describes a medical device, a catheter, and the tests performed to demonstrate its safety and effectiveness. It is not an AI/ML device, so many of the requested fields regarding AI performance, expert review, training sets, and comparative effectiveness studies are not applicable.
1. Table of Acceptance Criteria and Reported Device Performance
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Nonclinical Tests: CELLO II Balloon Guide Catheter | ||
Surface | Free from extraneous matter, process and surface defects. | Pass |
Force at Break | Tip and all joints meet reliability/confidence requirements in statistical confidence limits test. | Pass |
Freedom from Leakage | Shall not leak liquid when pressurized per methods in ISO 10555-1, Annex C. Shall not leak air when aspirated per methods in ISO 10555-1, Annex D. | Pass |
Tip Configuration | Distal tip smooth, rounded, tapered or similarly finished. | Pass |
Hubs | Comply with ISO 80369-7. | Pass |
Freedom from Leakage and Damage upon Inflation | No leakage or evidence of damage, such as herniation or bursting of the shaft or balloon. | Pass |
Dimensional Verification | All dimensions meet reliability/confidence requirements in statistical confidence limits test. | Pass |
Balloon Preparation, Deployment and Retraction | Catheters can be advanced to intended sites within a tortuous 3D vessel model where other devices can be deployed distally and retracted; all devices could be retracted without damage. | Pass |
Balloon Rated Burst Volume | All balloons met reliability/confidence requirements in statistical confidence limits test. | Pass |
Balloon Fatigue | All balloons withstand 20 cycles of inflation. | Pass |
Balloon Compliance | All balloons have predictable change in size with volume of contrast solution. | FIO (For Information Only) |
Balloon Inflation-Deflation Time | All balloons inflate and deflate predictably. | FIO (For Information Only) |
Flexibility and Kink | All catheters meet reliability/confidence requirements in statistical confidence limits test. | Pass |
Torque Strength | All models can be torqued at least 135° when the distal tip is fixed in a tortuous 3D vessel model. | FIO (For Information Only) |
Radiopacity | Radiopaque markers are visible during angiography under a variety of conditions. | FIO (For Information Only) |
Particulate testing | All catheters meet the particle generation criteria and exhibit similar size and quantity of particulates in comparison to the predicate device. | Pass |
Tip Flexibility | The tips of catheters meet the tip flexibility criteria. | Pass |
Contrast Agent Durability | There was no damage after application of pressure of 300 PSI. | Pass |
Biocompatibility: CELLO II Balloon Guide Catheters | ||
Cytotoxicity | Non-Cytotoxic based on ISO 10993-5:2009 L929 MEM Elution Test. | Non-Cytotoxic (Pass) |
Sensitization | Non-Sensitizing based on ISO10993-10:2010 Guinea Pig Maximization Test. | Non-Sensitizing (Pass) |
Irritation/Intracutaneous Reactivity | Non-Irritant based on ISO10993-10:2010 Intracutaneous Injection Test. | Non-Irritant (Pass) |
Acute Systemic Toxicity | No Acute Toxicity based on ISO10993-11:2017 Systemic Injection Test. | No Acute Toxicity (Pass) |
Material-Mediated Pyrogenicity | Non-Pyrogenic based on ISO10993-11:2017 Rabbit Pyrogen Test. | Non-Pyrogenic (Pass) |
Hemocompatibility (Hemolysis) | Non-Hemolytic based on ASTM F756-17 Hemolysis-Rabbit Blood (Direct and Indirect contact). | Non-Hemolytic (Pass) |
Hemocompatibility (PTT) | Hemocompatible based on ASTM F2382-18 Partial Thromboplastin Time Test (Direct contact). | Hemocompatible (Pass) |
Hemocompatibility (Platelet & Leukocyte Binding) | Hemocompatible based on ISO10993-4:2017 Platelet and leukocyte Binding- (Direct contact). | Hemocompatible (Pass) |
Hemocompatibility (Complement Activation) | Non-Complement Activation based on ISO10993-4:2017 Complement Activation Assay (Direct contact). | Non-Complement Activation (Pass) |
Hemocompatibility (Thrombogenicity) | Non-Thrombogenic based on ISO10993-4:2017 Thrombogenicity Study in Dog. | Non-Thrombogenic (Pass) |
Biocompatibility: Y-Connector | ||
Cytotoxicity | Non-Cytotoxic based on ISO 10993-5:2009 L929 MEM Elution Test. | Non-Cytotoxic (Pass) |
Sensitization | Non-Sensitizing based on ISO10993-10:2010 Guinea Pig Maximization Test. | Non-Sensitizing (Pass) |
Irritation/Intracutaneous Reactivity | Non-Irritant based on ISO10993-10:2010 Intracutaneous Injection Test. | Non-Irritant (Pass) |
Acute Systemic Toxicity | No Acute Toxicity based on ISO10993-11:2017 Systemic Injection Test. | No Acute Toxicity (Pass) |
Material-Mediated Pyrogenicity | Non-Pyrogenic based on ISO10993-11:2017 Rabbit Pyrogen Test. | Non-Pyrogenic (Pass) |
Hemocompatibility (Hemolysis) | Non-Hemolytic based on ASTM F756-17 Hemolysis-Rabbit Blood (Direct and Indirect contact). | Non-Hemolytic (Pass) |
2. Sample size used for the test set and the data provenance
The document does not explicitly state a sample size for each individual non-clinical or biocompatibility test. These tests are typically performed on a statistically significant number of manufactured units or material samples to ensure representativeness and reliability of the results. The provenance is likely from the manufacturer's internal testing facilities, as these are in vitro and in vivo (animal studies for biocompatibility) tests, not human data. The document implies these tests were conducted as part of the regulatory submission (premarket notification).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is a medical device (catheter) and not an AI/ML device that requires expert ground truth for image interpretation or diagnosis. The "ground truth" for these tests comes from established scientific methods, standards (e.g., ISO, ASTM), and laboratory measurements.
4. Adjudication method for the test set
Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or AI/ML evaluations where multiple human readers assess cases and a consensus or tie-breaker is needed. These are laboratory and material tests.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a physical medical device, not an AI/ML diagnostic or assistive tool. No MRMC studies were performed.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. No AI algorithm is involved.
7. The type of ground truth used
The ground truth for the nonclinical and biocompatibility tests is based on:
- Established industry standards: Such as ISO 10555-1, ISO 80369-7, ISO 10993 series, and ASTM F756-17, ASTM F2382-18.
- Quantitative measurements: Measuring physical properties (e.g., force, dimensions, burst volume, inflation/deflation time, torque, particle count).
- Qualitative observations: Visual inspection (e.g., surface, damage, tip configuration, radiopacity).
- Biological assays/animal studies: For biocompatibility endpoints (e.g., cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility experiments using rabbit blood and in vivo thrombogenicity study in dogs).
8. The sample size for the training set
Not applicable. This is not an AI/ML device.
9. How the ground truth for the training set was established
Not applicable. This is not an AI/ML device.
Ask a specific question about this device
(147 days)
FUJI SYSTEMS CORP.
The CELLO Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of intravascular catheters into a selected blood vessel in the peripheral and neuro vasculature systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures.
The CELLO Balloon Guide Catheter is a coaxial-lumen, braid-reinforced, variable stiffness catheter with two radiopaque markers on both the distal and proximal ends of the balloon and a bifurcated luer hub on the proximal end. A compliant silicone balloon is mounted on the distal end. Balloon Guide Catheter dimensions and recommended inflation volumes are indicated on the product label. Each catheter is supplied with an appropriately-size Dilator.
Here's an analysis of the provided 510(k) Summary, focusing on acceptance criteria and the study proving the device meets them:
Acceptance Criteria and Device Performance for CELLO Balloon Guide Catheter (K120781)
1. Table of Acceptance Criteria and Reported Device Performance
The device under review is primarily mechanical, and the "acceptance criteria" are embodied in the results of the non-clinical and biocompatibility tests conducted. The reported device performance is indicated by whether each test passed, failed, or was for informational purposes only.
Test Category | Specific Test | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|---|
Nonclinical Tests (CELLO Balloon Guide Catheters) | |||
Surface | Free from extraneous matter, process and surface defects. | Acceptable surface finish, no defects. | Pass |
Force at Break | All joints meet reliability/confidence requirements in statistical confidence limits test | Joints must withstand specific force without breaking, within statistical limits. | Pass |
Freedom from Leakage | Shall not leak liquid when inflated. Air shall not leak into the hub during aspiration. | No leakage of liquid when inflated, no air leakage into hub during aspiration. | Pass |
Tip Configuration | Distal tip smooth, rounded, tapered or similarly finished. | Distal tip must be atraumatic. | Pass |
Hubs | Comply with ISO 594-1 and ISO 594-2 | Hubs must conform to specified ISO standards. | Pass |
Freedom from Leakage and Damage on Inflation | No leakage or evidence of damage, such as herniation or bursting of the shaft or balloon. | No leakage or damage (herniation, bursting) during inflation. | Pass |
Dimensional Verification | All dimensions meet reliability/confidence requirements in statistical confidence limits test | All dimensions must be within specified tolerances, within statistical limits. | Pass |
Balloon Preparation, Deployment and Retraction | Catheters can be advanced to intended sites within a realistic tortuous path model where other devices can be deployed distally and retracted; all devices could be retracted without damage. | Successful navigation through tortuous path, deployment and retraction of other devices, and full device retraction without damage. | Pass |
Balloon Rated Burst Pressure (RBP) | All balloons meet reliability/confidence requirements in statistical confidence limits test | Balloons must withstand RBP without bursting, within statistical limits. | Pass |
Balloon Fatigue | All balloons withstand 20 cycles of inflation | Balloons must withstand 20 inflation cycles without failure. | Pass |
Balloon Compliance | All balloons have predictable change in size with pressure | Balloons should exhibit consistent and predictable size changes with pressure application. | FIO* |
Balloon Inflation-Deflation Time | All balloons inflate and deflate predictably | Balloons should inflate and deflate within acceptable, predictable timeframes. | FIO* |
Flexibility and Kink | All catheters meet reliability/confidence requirements in statistical confidence limits test | Catheters must maintain integrity and function without kinking, within statistical limits. | Pass |
Torque Strength | All models can be torqued at least 180° when the distal tip is fixed in a tortuous path model | Catheters must transmit torque effectively in a tortuous path without failure. | FIO* |
Radiopacity | Radiopaque markers are visible during angiography under a variety of conditions | Radiopaque markers must be clearly visible during angiography in various conditions. | FIO* |
Nonclinical Tests (Dilator) | |||
Surface | Free from extraneous matter, process and surface defects. | Acceptable surface finish, no defects. | Pass |
Force at Break | All joints meet reliability/confidence requirements in statistical confidence limits test | Joints must withstand specific force without breaking, within statistical limits. | Pass |
Freedom from Leakage | Shall not leak liquid when inflated. Air shall not leak into the hub during aspiration. | No leakage of liquid when inflated, no air leakage into hub during aspiration. (Likely refers to the dilator's hub functionality, if applicable to aspiration) | Pass |
Tip Configuration | Distal tip smooth, rounded, tapered or similarly finished. | Distal tip must be atraumatic. | Pass |
Hubs | Comply with ISO 594-1 and ISO 594-2 | Hubs must conform to specified ISO standards. | Pass |
Freedom from Leakage and Damage on Inflation | No leakage or evidence of damage, such as herniation or bursting of the shaft or balloon. | No leakage or damage (herniation, bursting) during inflation. (Likely refers to any balloon on the dilator, if applicable, or general structural integrity) | Pass |
Biocompatibility (CELLO Balloon Guide Catheters & Dilator) | |||
L29 MEM Elution Test-ISO | Non-Cytotoxic | No evidence of cytotoxicity. | Pass |
Kligman Maximization Test | No Sensitization | No evidence of sensitization. | Pass |
Intracutaneous Injection Test | Non-Irritant | No evidence of irritation. | Pass |
Systemic Injection Test-ISO | No acute toxicity | No evidence of acute toxicity. | Pass |
Rabbit Pyrogen Test (Material Mediated) - ISO | Non-pyrogenic | No evidence of pyrogenicity. | Pass |
Hemolysis- Direct Contact | Non-hemolytic | No evidence of hemolysis (direct contact). | Pass |
Hemolysis-Rabbit Blood-ASTM Indirect Contact | Non-hemolytic | No evidence of hemolysis (indirect contact). | Pass |
Complement Activation-Direct Contact | Non-Complement Activation | No evidence of complement activation (direct). | Pass |
Complement Activation Assay-ISO Indirect Contact | Non-Complement Activation | No evidence of complement activation (indirect). | Pass |
In-vivo Thrombogenicity (Catheter) | Patent vessels, No thrombus | Better thrombo-resistance than control. | Pass (Better than control) |
In-vivo Thrombogenicity (Dilator) | Less thrombus than control | Better thrombo-resistance than control. | Pass (Better than control) |
*FIO: For Information Only - These tests were performed, but the results were provided for informational purposes and not necessarily as pass/fail acceptance criteria in the same way as the others. However, the presence of these results implies an expected functional characteristic.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not explicitly state the exact sample sizes used for each nonclinical or biocompatibility test. It mentions "All joints," "All dimensions," and "All balloons" meeting reliability/confidence requirements in "statistical confidence limits test," which implies that a statistically significant number of samples were tested. However, the specific number of units tested for each criterion is not provided.
- Data Provenance: The studies are nonclinical (benchtop and in vitro/in vivo biocompatibility) and were conducted by the manufacturer, Fuji Systems Corporation, or their designated testing laboratories. The country of origin for the data is not explicitly stated, but the manufacturer is based in Japan. These are retrospective data collected as part of the device development and regulatory submission process.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This submission pertains to a mechanical medical device (catheter) and focuses on engineering performance and biocompatibility. There is no "ground truth" established by human experts in the context of diagnostic accuracy found in AI/imaging device submissions. The "ground truth" for these tests is defined by the objective measurement criteria and standards outlined in the test protocols (e.g., ISO standards, ASTM standards, internal reliability requirements).
4. Adjudication Method for the Test Set
Not applicable. This is a nonclinical study involving objective measurements and standardized tests, not subjective interpretations requiring adjudication by multiple readers/experts.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Not applicable. This is a submission for a mechanical medical device, not a diagnostic imaging or AI-assisted device. Therefore, no MRMC study, human reader improvement, or AI assistance is relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This device is not an algorithm or AI system.
7. The Type of Ground Truth Used
The "ground truth" for the nonclinical tests is based on:
- Objective Measurement: Adherence to specified dimensions, force limits, inflation pressures, deployment characteristics, and physical properties.
- Standardized Protocols: Compliance with international standards (e.g., ISO 594-1, ISO 594-2, ISO biocompatibility tests, ASTM methods).
- Pre-defined Pass/Fail Criteria: Each test had distinct criteria that determined a "Pass" or "Fail" outcome (or "For Information Only").
8. The Sample Size for the Training Set
Not applicable. This is a hardware medical device; it does not involve machine learning algorithms requiring a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(70 days)
FUJI SYSTEMS CORP.
The Silbroncho® is used to isolate the left or the right lung of a patient for surgery, one lung ventilation or one lung anesthesia.
The Silbroncho® double lumen tube is made of silicone and is available in sizes 33 to 39 French. They are designed as a double lumen tube with 2 cuffs and separate 15 mm connectors for isolating and ventilating one lung during surgical procedures.
The provided text describes a 510(k) premarket notification for the "Silbroncho® tubes," a double lumen tube. The submission focuses on demonstrating substantial equivalence to predicate devices rather than directly presenting performance studies against acceptance criteria. Therefore, most of the requested information regarding acceptance criteria and study details cannot be extracted from the provided text.
Here is the information that can be extracted, and where the requested details are not available, it is noted:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state quantitative acceptance criteria or detailed performance results from a study. The submission is based on demonstrating substantial equivalence to pre-existing predicate devices.
Acceptance Criteria (Not Explicitly Stated) | Reported Device Performance |
---|---|
(Not provided in the text) | The device (Silbroncho®) is substantially equivalent to predicate devices (Mallinckrodt Broncho-Cath K771219 and Vitaid wire reinforced ET tube K932647) in terms of indications for use, environments of use, patient population, technology, design features, and materials (with some noted differences in material and sizes). |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
Not applicable. The document does not describe a performance study with a test set. This is a 510(k) submission focused on substantial equivalence.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
Not applicable. No performance study using a ground truth is described.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. No performance study with adjudication is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a medical tube, not an AI-assisted diagnostic tool.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is a medical tube, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable. No performance study requiring a ground truth is described.
8. The sample size for the training set
Not applicable. The document does not describe a machine learning algorithm or a training set.
9. How the ground truth for the training set was established
Not applicable. The document does not describe a machine learning algorithm or a training set.
Ask a specific question about this device
Page 1 of 1