Search Results
Found 3 results
510(k) Data Aggregation
(377 days)
The Finapres NOVA is intended to be used with patients who have a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals, such as heart rate variability (HRV) and Baroreflex sensitivity (BRS). Cardiac output derived from the flow signal requires a calibration with thermal dilution.
The Finapres NOVA has the option to include additional modules to extend its functionality with ECG and SpO2 measurements and blood pressure calibration.
When the SpO2 module is present, the Finapres NOVA can additionally monitor the functional oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate.
When the ECG module is present, the Finapres NOVA can additionally monitor the ECG parameters of a patient and their pulse rate. Alarms concerning the pulse rate will be available from the monitor.
When the blood pressure calibration module is present, the Finapres NOVA can additionally provide an upper arm non-invasive blood pressure measurement to determine the blood pressure value for calibration.
The Finapres NOVA is intended to be used for subjects above 18 years of age.
The Finapres NOVA is intended for use in hospitals, clinics and research institutions.
The Finapres NOVA is an instrument to noninvasively monitor blood pressure and hemodynamic parameters. The Finapres NOVA provides a characterization of the arterial circulation and its beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals.
The Finapres NOVA has the option to include four additional modules to extend its functionality with ECG and SpO2 measurements, blood pressure calibration and data transfer to and from the device.
The measurement of blood pressure in a finger is based on the arterial volume-clamp method of the Czech physiologist J. Peñáz, and the Physiocal - physiological calibration criteria for the proper unloading of the finger arteries of K.H. Wesseling. With this method, finger arterial pressure is measured using a finger cuff and an inflatable bladder in combination with an infrared plethysmograph, which consists of an infrared light source and detector.
The SpO2, upper arm calibration and ECG modules used in the Finapres NOVA are commercially available OEM modules that are used in FDA approved systems. The finger blood pressure measurement module used in the system is similar to other Finapres Medical Systems B.V. devices available on the market. The analog input/output module has been developed by Finapres during the Finapres NOVA development.
The embedded software in the device provides computation of real-time and beat-to-beat blood pressure as well as hemodynamic parameters from the non-invasely measured blood pressure waveform. Hemodynamic parameters include cardiac output based on the modelflow method and total peripheral resistance.
The addition to the Finapres NOVA software covered in this application is mostly related to the addition of a software module called ANS (Autonomous Nervous System). (Para)sympathetic function can be assessed by a physician through arterial Baro Reflex Sensitivity (BRS) testing and Heart Rate Variability (HRV) analysis. The ANS software module calculates additional parameters, derived from pressure and ECG signals measured with the Finapres NOVA, which are related to Baro Reflex Sensitivity and to Heart Rate Variability.
A baroreceptor is a sensory nerve ending in de wall of the aortic arch and carotid bulbus that is sensitive to changes in blood pressure. These baroreceptors act as receptors of central reflex mechanisms that regulate the blood pressure. The blood pressure is regulated by altering heart rate (baroreflex), cardiac contractility, vasoactivity and humoral activity. The main function of the baroreflex is to maintain a stable blood pressure. To quantify whether the baroreflex is functioning properly the linear regression line of the relation between systolic blood pressure and resulting interbeat interval is estimated. The slope of this linear regression line is defined as the Baro Reflex Sensitivity (BRS) and is to be used in the assessment of the (dys)function of the baroreflex.
Heart rate variability (HRV) is the physiological phenomenon of variation in the time interval between heartbeats. It is determined using the beat-to-beat interval.
Using this additional software module, the Finapres NOVA only presents parameters derived from the interbeat interval (HRV) and blood pressure (BRS) to the user. It is up to the user to draw conclusions on whether these values are normal or abnormal.
The additional Remote Control Module provides the possibility to monitor and control a Finapres Nova from a PC. This can be achieved by establishing a network connection from a PC to the Finapres Nova. Two types of connections can be made: one that is restricted to viewing and one that allows full control of the Finapres Nova.
The additional Nova Scope PC application is used to view and review measurement files recorded with the Finapres Nova on a PC instead of on the NOVA instrument itself.
Here's an analysis of the acceptance criteria and study information for the Finapres NOVA Noninvasive Hemodynamic Monitor, based on the provided text.
It's important to note that the provided text is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed report of a new clinical study with specific acceptance criteria and detailed performance metrics of the device as a standalone product. Therefore, some of the requested information (like specific quantitative acceptance criteria for this device and a detailed study proving this device meets those criteria) is not explicitly present in the format typically found in a full clinical trial report. Instead, the document focuses on demonstrating that the new software module (ANS) performs equivalently to algorithms in a predicate device.
Acceptance Criteria and Study Information for Finapres NOVA Noninvasive Hemodynamic Monitor (ANS software module)
This document describes the 510(k) submission for the addition of an Autonomic Nervous System (ANS) software module to the Finapres NOVA, along with a Remote Control module and Nova Scope PC application. The core of the submission for the ANS module is demonstrating substantial equivalence to the algorithms in a predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not provide a table of specific quantitative acceptance criteria for the Finapres NOVA's ANS module's performance metrics (e.g., specific accuracy thresholds for HRV or BRS values). Instead, the acceptance criterion for the ANS module's performance was to be substantially equivalent to the performance of the algorithms in the secondary predicate device, the TASK FORCE MONITOR 3040.
| Feature/Parameter | Acceptance Criteria | Reported Device Performance | Comments |
|---|---|---|---|
| ANS Software Module Algorithms (HRV, BRS) | Substantially equivalent in performance to algorithms in the predicate device (TASK FORCE MONITOR 3040) | "The performance of the HRV and BRS algorithm was tested and found to be substantially equivalent with the performance of the algorithms of our secondary predicate device, the TASK FORCE MONITOR 3040." | This indicates that comparison testing was performed against the predicate device's algorithms. Specific quantitative performance delta or equivalence margin is not detailed. |
| Overall Finapres NOVA Software | Met all applicable requirements based on "moderate level of concern" according to FDA's Guidance for Software. | "The verification tests performed demonstrate that the new software module on the Finapres NOVA met all applicable requirements." | This relates to software validation and verification processes, not specific clinical performance metrics. |
2. Sample Size Used, Test Set, and Data Provenance
- Sample Size for Test Set: Not explicitly stated. The document mentions that "The performance of the HRV and BRS algorithm was tested," implying a test set was used, but the size of this set (number of patients/cases) is not provided.
- Data Provenance: Not explicitly stated for the algorithm performance test. The device is manufactured by Finapres Medical Systems B.V. in the Netherlands. The predicate device (TASK FORCE MONITOR 3040) is from CNSYSTEMS MEDIZINTECHNIK GMBH, Graz, Austria. Without a direct statement, it's unclear if the test data for the algorithm comparison was retrospective or prospective, or its geographic origin.
3. Number of Experts Used to Establish Ground Truth and Qualifications
- Not applicable. The document focuses on demonstrating substantial equivalence of algorithms to a predicate device's algorithms, not on establishing a physician-adjudicated ground truth for a novel diagnostic claim requiring expert consensus. The "ground truth" for the algorithm comparison would be the output of the predicate device's algorithms or a reference standard used in its validation.
4. Adjudication Method
- Not applicable. See point 3.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not done. The submission primarily focuses on the technical and algorithmic equivalence of the ANS software module to a predicate device's algorithms. There is no mention of a study involving human readers or comparing their performance with and without AI assistance.
6. Standalone Performance Study (Algorithm Only)
- Yes, a standalone performance assessment of the algorithm was conducted, implicitly. The statement "The performance of the HRV and BRS algorithm was tested and found to be substantially equivalent with the performance of the algorithms of our secondary predicate device, the TASK FORCE MONITOR 3040" indicates that the algorithm's output was directly compared to the predicate device's algorithm output. This represents an algorithm-to-algorithm comparison, which is a form of standalone evaluation in this context of substantial equivalence for a software module.
7. Type of Ground Truth Used
- Predicate device algorithm output / Reference Standard from Predicate. The "ground truth" in this context refers to the established and validated performance of the algorithms within the predicate device (TASK FORCE MONITOR 3040), or a recognized reference standard that the predicate device's algorithms were validated against for HRV and BRS. The document does not specify how the predicate device's algorithms were validated, but for the purpose of this 510(k), the predicate's performance serves as the benchmark.
8. Sample Size for Training Set
- Not explicitly stated, and likely not applicable in the traditional sense for this submission. This submission is for a new software module being added to an existing device (Finapres NOVA cleared under K141460). The focus is on the performance of the new ANS algorithms compared to a predicate, not on training a new machine learning model from scratch that would require a distinct training set. The algorithms for HRV and BRS are likely established physiological models rather than data-driven machine learning models requiring large-scale training sets in the same way.
9. How Ground Truth for Training Set Was Established
- Not applicable. See point 8.
Ask a specific question about this device
(291 days)
The Finapres NOVA is intended to be used with patients who have a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. Cardiac output derived from the flow signal requires a calibration with thermal dilution.
The Finapres NOVA has the option to include additional modules to extend its functionality with ECG and SpO2 measurements and blood pressure calibration.
When the SpO2 module is present, the Finapres NOVA can additionally monitor the functional oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate.
When the ECG module is present, the Finapes NOVA can additionally monitor the ECG parameters of a patient and their pulse rate. Alarms concerning the pulse rate will be available from the monitor.
When the blood pressure calibration module is present, the Finapres NOVA can additionally provide an upper arm non-invasive blood pressure measurement to determine the blood pressure value for calibration.
The Finapres NOVA is intended to be used for subjects above 18 years of age.
The Finapres NOVA is intended for use in a professional medical environment.
The Finapres NOVA is an instrument to noninvasively monitor blood pressure and hemodynamic parameters. The Finapres NOVA provides a characterization of the arterial circulation and its beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. The Finapres NOVA has the option to include four additional modules to extend its functionality with ECG and SpO2 measurements, blood pressure calibration and data transfer to and from the device.
The embedded software in the device provides computation of real-time and beat-tobeat blood pressure as well as hemodynamic parameters from the non-invasely measured blood pressure waveform. Hemodynamic parameters include cardiac output based on the modelflow method and total peripheral resistance.
The measurement of blood pressure in a finger is based on the arterial volume-clamp method of the Czech physiologist J. Peñáz, and the Physiocal - physiological calibration criteria for the proper unloading of the finger arteries of K.H. Wesseling. With this method, finger arterial pressure is measured using a finger cuff and an inflatable bladder in combination with an infrared plethysmograph, which consists of an infrared light source and detector.
The SpO2, upper arm calibration and ECG modules used in the Finapres NOVA are commercially available OEM modules that are used in FDA approved systems. The finger blood pressure measurement module used in the system is similar to other Finapres Medical Systems B.V. devices available on the market. The analog input/output module has been developed by Finapres during the Finapres NOVA development.
The provided text describes the Finapres NOVA device and its substantial equivalence determination. However, it does not contain the specific information required to complete all sections of your request, particularly regarding detailed acceptance criteria testing and clinical performance studies as you've outlined.
The document explicitly states: "Clinical performance data was not required to demonstrate substantial equivalence." This means that a specific clinical study with acceptance criteria for the Finapres NOVA's performance (e.g., accuracy against a gold standard in a human population) was not conducted or presented in this 510(k) summary.
Instead, the submission relies on non-clinical performance data by testing the device against various medical device safety and performance standards (listed in Table 3 in the document) and by demonstrating substantial equivalence to predicate devices that have already been cleared.
Here's a breakdown of what can and cannot be answered from the provided text:
1. A table of acceptance criteria and the reported device performance
The document lists numerous standards that the Finapres NOVA was tested against (Table 3), indicating that compliance with these standards serves as the "acceptance criteria" for non-clinical performance. However, it does not provide specific numerical values for acceptance criteria or reported device performance for blood pressure accuracy (e.g., Mean Difference, Standard Deviation as per AAMI SP10 or ISO 81060-2:2009 for the Finapres NOVA itself). It only states that the device "met all applicable requirements."
The document mentions compliance with "AAMI SP10:2002 and its successor IEC 81060-2:2009" for blood pressure calibration, but it does not present the direct results of such a clinical validation for the Finapres NOVA. It implies that the ability to be calibrated allows for accurate data, rather than directly presenting the Finapres NOVA's accuracy results against those standards.
Therefore, a table with specific numerical acceptance criteria and reported performance cannot be fully constructed directly from this document for the Finapres NOVA's clinical accuracy.
2. Sample size used for the test set and the data provenance
Since clinical performance data was explicitly "not required to demonstrate substantial equivalence" for the Finapres NOVA device itself, there is no test set sample size or data provenance provided for a clinical study of the Finapres NOVA.
The non-clinical performance testing against standards doesn't typically involve a "test set" in the sense of clinical data from patients.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable, as no clinical test set for the Finapres NOVA's performance is described.
4. Adjudication method for the test set
Not applicable, as no clinical test set for the Finapres NOVA's performance is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The Finapres NOVA is a non-invasive blood pressure and hemodynamic monitor, not an AI-assisted diagnostic tool that would involve "human readers" or "AI assistance" in the context of interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The Finapres NOVA is described as a standalone device with embedded software that computes real-time blood pressure and hemodynamic parameters. Therefore, its "standalone" performance is implicitly what the non-clinical testing against standards evaluated. However, specific performance metrics for the algorithm itself (e.g., accuracy of cardiac output calculation against a gold standard) are not provided in numerical detail.
7. The type of ground truth used
For the non-clinical performance testing, the "ground truth" would be established by the specifications and measurement methods defined within the various international standards the device was tested against (e.g., IEC, ISO, AAMI standards for electrical safety, EMC, basic safety, essential performance, and specific requirements for sphygmomanometers, oximeters, and ECG equipment).
The document mentions: "Cardiac output derived from the flow signal requires a calibration with thermal dilution." This implies that for cardiac output measurements, thermal dilution is the clinical "gold standard" or "ground truth" for calibration. However, it's not stated that the Finapres NOVA's cardiac output was validated against this in a clinical study presented here.
8. The sample size for the training set
Not applicable. This device is not described as involving a machine learning or AI algorithm that requires a "training set" in the conventional sense for image analysis or classification. Its embedded software is based on established physiological models (e.g., arterial volume-clamp method, Modelflow method).
9. How the ground truth for the training set was established
Not applicable, as there's no machine learning "training set" described.
In summary, the provided 510(k) summary focuses on demonstrating substantial equivalence primarily through non-clinical performance data (compliance with device standards) and comparison to predicate devices, rather than through a new clinical study with specific performance acceptance criteria for the Finapres NOVA itself.
Ask a specific question about this device
(113 days)
Ask a specific question about this device
Page 1 of 1