510(k) Data Aggregation

    K Number
    K103747
    Device Name
    ELITECH CLINCAL SYSTEMS CHOLESTEROL HDL SL 2G, HDL 2G CALIBRATOR, LDL SL 2G, LDL 2G CALIBRATOR, ELITROL II & ELITECH SYS
    Manufacturer
    ELITECHGROUP SEPPIM S.A.S.
    Date Cleared
    2011-08-31

    (251 days)

    Product Code
    LBS, JIT, JJY, MRR
    Regulation Number
    862.1475
    Why did this record match?
    Applicant Name (Manufacturer) :

    ELITECHGROUP SEPPIM S.A.S.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    ELITech Clinical Systems CHOLESTEROL HDL SL 2G is intended for use with ELITech Clinical Systems CHOLESTEROL HDL 2G CALIBRATOR and ELITech Clinical Systems ELITROL I and ELITROL II on ELITech Clinical Systems Selectra analyzers for the quantitative in vitro diagnostic determination of High Density Lipoprotein (HDL) Cholesterol in human serum and plasma. It is not intended for use in Point of Care settings. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. ELITech Clinical Systems CHOLESTEROL HDL 2G CALIBRATOR is a calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems CHOLESTEROL HDL SL 2G on the ELITech Clinical Systems Selectra analyzers. ELITech Clinical Systems CHOLESTEROL LDL SL 2G is intended for use with ELITech Clinical Systems CHOLESTEROL LDL 2G CALIBRATOR and ELITech Clinical Systems ELITROL I and ELITROL II for the quantitative in vitro diagnostic determination of Low Density Lipoprotein (LDL) Cholesterol in human serum and plasma on ELITech Clinical Systems Selectra analyzers. It is not intended for use in Point of Care settings. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. ELITech Clinical Systems CHOLESTEROL LDL 2G CALIBRATOR is a calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems CHOLESTEROL LDL SL 2G on ELITech Clinical Systems Selectra Analyzers. ELITech Clinical Systems ELITROL I & ELITROL II are multi-parametric control sera for in vitro diagnostic use in quality control of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers.
    Device Description
    The device for this submission is available as kit only. It consists of 2 reagents, "R1" and "R2". Reagent R1 contains: Good's buffer, Cholesterol oxidase (CO bacterial), Peroxidase (horseradish), Ascorbate oxidase (bacterial), N,N-bis(4-sulphobutyl)-m-toluidine-disodium (DSBmT), Accelerator. Reagent R2 contains: Good's buffer, Cholesterol esterase (CHE bacterial), 4-Amino-Antipyrine (4-AA), detergent. ELITech Clinical Systems CHOLESTEROL HDL 2G CALIBRATOR is a lyophilized calibrator based on human serum containing lipoprotein from the various lipoprotein classes including high density lipoproteins and sodium azide as preservative. CHOLESTEROL HDL 2G CALIBRATOR is prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods. The device for this submission is available as kit only. It consists of 2 reagents, "R1" and "R2". Reagent R1 contains: MES buffer, Detergent 1, Cholesterol esterase (CHE bacterial), Cholesterol oxidase (CO bacterial), Peroxidase (horseradish), 4-Amino-Antipyrine (4-AA), Ascorbate oxidase (vegetal). Reagent R2 contains: MES buffer, Detergent 2, N,N-bis(4-sulphobutyl)-m-toluidine-disodium (DSBmT). ELITech Clinical Systems CHOLESTEROL LDL 2GCALIBRATOR is a lyophilized calibrator based on human serum containing lipoprotein from the various lipoprotein classes including low density lipoproteins and sodium azide as preservative. CHOLESTEROL LDL 2G CALIBRATOR is prepared from plasma donor units tested individually by FDA - approved methods and found to be negative for HbsAg, anti-HCV antibody and anti-HIV1&2 antibodies. ELITech Clinical Systems ELITROL I and ELITROL II are two level quality control products consisting of lyophilized human serum containing constituents at desired levels. Elitrol I and Elitrol II are prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods or methods in compliance with the European Directive 98/79/EC, Annex II, List A.
    Ask a Question

    Page 1 of 1