Search Results
Found 1 results
510(k) Data Aggregation
K Number
K231969Device Name
BASEN Wearable Breast Pump (Model: YM-801, YM-803, YM-805, YM-806, YM-808)
Manufacturer
Date Cleared
2024-08-30
(424 days)
Product Code
Regulation Number
884.5160Why did this record match?
Applicant Name (Manufacturer) :
Dongguanshi Yiyingmei Technology Co., Ltd.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The BASEN™ Wearable Breast Pump (Model: YM-801, YM-805, YM-805, YM-806, YM-808) is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.
Device Description
The BASEN™ Wearable Breast Pump is an electrically powered wearable breast pump consisting of the following key components: pump, pump diaphragm, container valve, 5-ounce (150 milliliter) container bowl, flange, seal ring, joint pipe, and USB charging cable. It is designed to work in the user's bra and has a rechargeable battery so it can be used hands-free without any external power cords. The motor unit includes a press-button user interface, pump body, and LED display. Pumping can be performed on one breast (single pumping). The user interface allows the user to switch from stimulation, massage mode, and expression mode and control the vacuum levels within those modes.
Both stimulation and expression mode consist of 9 vacuum levels. The pump can provide vacuum levels from -40 to -105 mmHg with cycling rates from 70 to 114 cycles per minute in stimulation and massage mode. Expression mode had vacuum levels from -40 to -245 mmHg with cycling rates from 23 to 90 cycles per minute.
The motor unit operates on embedded software. Software updates by end-users are not supported. The subject device is for repeated use by a single user in a home environment. The breast pump expresses by creating a seal around the nipple using the flange and applying and releasing suction to the nipple. The milk is collection container bowl. To prevent milk from flowing into the vacuum system, a backflow protection membrane physically separates the milk-contacting pathway from the vacuum system. The device is provided not sterile.
The motor unit operates on a rechargeable battery and does not function when charging. The rechargeable battery can be charged from the external USB adapter if the motor unit is not in operation.
The subject device components are made of the following materials:
- . Motor unit: Acrylonitrile Butadiene Styrene (ABS) plastic
- . Flange, tube, valve, diaphragm: Silicone
- Linker, milk collection container: Polypropylene
All milk contacting components are compliant with 21 CFR 174-179.