Search Results
Found 5 results
510(k) Data Aggregation
(102 days)
DePuy Mitek, Inc.
The FMS VUE™ Fluid Management and Tissue Debridement System is intended to provide controlled fluid distention and suction, controlled cutting, shaving and abrading of bone and tissue for use in arthroscopic surgery of the shoulder, knee, ankle, elbow, wrist and hip joints.
The FMS VUE Fluid Management and Tissue Debridement System is an arthroscopic pump system designed to provide optimal visibility of the surgical field by computer-controlled fluid regulation of intra-articular pressure, and flow during arthroscopic procedures. The system integrates a tissue debridement for controlled cutting, burring, shaving and abrading of bone and soft tissue.
The FMS VUE Fluid Management and Tissue Debridement System consists of irrigation pump, which controls joint pressure, and the suction pump, which controls the flow of saline and waste from the joint.
The FMS VUE Fluid Management and Tissue Debridement System includes an integrated shaver console, intended to provide controlled cutting, burring, shaving and abrading of bone and soft tissue.
The FMS VUE Fluid Management and Tissue Debridement System is designed to work with the existing FMS VUE platform accessories including: foot pedal, remote hand control, shaver handpieces, FMS connect cable, tubing sets, blades, and burrs.
The FMS VUE Fluid Management and Tissue Debridement System contains updates in the hardware, which include replaced near obsolescence components and new real panel connectors.
The FMS VUE Fluid Management and Tissue Debridement System software modification includes an additional shaver oscillation profile algorithm.
The provided document is a 510(k) summary for the DePuy Mitek, Inc. FMS VUE™ Fluid Management and Tissue Debridement System, seeking substantial equivalence to a previously cleared predicate device (K130169). The submission highlights hardware updates and a software modification (an additional shaver oscillation profile algorithm).
Crucially, this document is for a medical device (arthroscopic pump and debridement system) and not an AI/ML powered device that would typically have the acceptance criteria and study information you're asking for related to AI model performance (e.g., sensitivity, specificity, MRMC studies, ground truth establishment by experts).
The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context are related to the hardware and software changes for a physical medical device, demonstrating that the modified device remains as safe and effective as its predicate.
Therefore, many of the requested points in your prompt are not applicable to the information contained in this 510(k) summary. I will address the applicable points based on the document's content.
Acceptance Criteria and Device Performance (Non-AI/ML Context)
The acceptance criteria here are focused on demonstrating that the updated FMS VUE system maintains its functional performance, electrical safety, electromagnetic compatibility, and software integrity, proving it is substantially equivalent to the predicate device despite hardware and minor software changes.
Table of Acceptance Criteria and Reported Device Performance (as inferred from the document):
Acceptance Criteria (General Area) | Reported Device Performance (Summary from Document) |
---|---|
Functional Performance | - Pressure Transducer Test: Performed. Results not detailed but implied to be acceptable for intended use. |
- Pressure Security Test (Low & High Pressure Fault Checks): Performed. Results not detailed but implied to be acceptable.
- DUO (Inflow/Outflow) Performance: Performed. Results not detailed but implied to be acceptable.
- SOLO (Inflow) Performance: Performed. Results not detailed but implied to be acceptable.
- RFID Capability Functional Testing with existing electrical accessories (Foot Pedal, Remote Hand Control, Shaver Handpieces, FMS Connect Cable): Performed. Results not detailed but implied to be acceptable.
- Backflow-Prevention Valve Testing (Dye Leak Testing): Performed. Results not detailed but implied to be acceptable.
- Microbial and Viral Ingress Testing: Performed. Results not detailed but implied to be acceptable.
Overall: "Results of functional performance testing were compared to the predicate device and have demonstrated that the proposed device is suitable for its intended use." |
| Software Verification & Validation | - Software verification and validation (V&V) testing was conducted. Documentation provided as per FDA guidance "Content of Premarket Submission for Device Software Functions" (Basic Documentation level). - "The software development process for the subject FMS VUE Fluid Management and Tissue Debridement System software conforms to all clauses of IEC 62304 standard." (IEC 62304 is a standard for medical device software life cycle processes). |
| Electrical Safety & EMC | - "The subject FMS VUE Fluid Management and Tissue Debridement System complies with the IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC." (These are international standards for medical electrical equipment safety and electromagnetic compatibility). |
Study Information (as applicable to this non-AI/ML medical device submission)
-
Sample size used for the test set and the data provenance:
- The document describes "functional performance testing" of the physical device. This doesn't involve "test sets" of patient data in the way an AI/ML model would. Instead, it refers to testing the physical device and its components (e.g., pressure transducers, pumps, software algorithms for fluid control, electrical safety tests).
- The provenance of data is from laboratory testing and verification conducted by the manufacturer, DePuy Mitek, Inc. (a Johnson & Johnson company). It is implicitly prospective testing for the design verification and validation of the modified device. No specific patient data or geographic origin is relevant here.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This question is not applicable. "Ground truth" in the context of an AI/ML model's diagnostic or predictive performance is not relevant for the functional performance testing of a fluid management and tissue debridement system. The "ground truth" here is the engineering specifications and performance requirements established by the manufacturer, verified through standard test methods and adherence to regulatory standards (e.g., IEC 60601-1, IEC 62304).
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies, particularly for establishing ground truth from multiple expert readers in imaging studies. This is a functional and safety verification of a physical device.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This question is not applicable. The device is a physical arthroscopic fluid management and tissue debridement system, not an AI/ML-powered diagnostic or assistive tool for human readers. No MRMC study was performed or required. The document explicitly states: "No human clinical testing was conducted to determine the safety and effectiveness of the FMS VUE Fluid Management and Tissue Debridement System."
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The device has an integrated shaver console with an "additional shaver oscillation profile algorithm." The software verification and validation would have tested this algorithm's performance within the system, which can be seen as a form of standalone testing for the algorithm's functional correctness. The document states: "Software verification and validation testing was conducted..." and "The software development process... conforms to all clauses of IEC 62304 standard." This implicitly covers testing of the algorithm without a human-in-the-loop, to ensure it performs as programmed.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- This question is not applicable in the typical sense for AI/ML. The "ground truth" for the device's functional performance are the engineering specifications, validated test methods, and compliance with recognized industry standards (e.g., pressure output accuracy, flow rates, software logic, electrical safety limits).
-
The sample size for the training set:
- This question is not applicable. This is not an AI/ML system that requires a "training set" of data in the manner of deep learning models. The software algorithms are likely rule-based or control-loop algorithms, not learned from large datasets.
-
How the ground truth for the training set was established:
- This question is not applicable for the same reason as point 7.
Ask a specific question about this device
(17 days)
DEPUY MITEK INC., A JOHNSON AND JOHNSON COMPANY
The Healix Advance Knotless Anchors are indicated for use in the following procedures for reattachment of soft tissue to bone:
Shoulder
Rotator Cuff .
. Biceps Tenodesis
The proposed Healix Advance Knotless Anchor is a one piece implantable cannulated, threaded anchor designed to secure soft tissue to bone. The anchor is provided loaded on a disposable inserter driver device. The proposed anchors will be offered in a 6.5 mm size. The proposed 6.5 mm Healix Advance Knotless PEEK Anchor is manufactured from PEEK (Polyetheretherketone) material.
The DePuy Mitek HEALIX ADVANCE KNOTLESS PEEK ANCHOR (6.5mm) is a medical device designed for reattachment of soft tissue to bone. This submission (K131683) is for a larger version (6.5mm) of previously cleared predicate devices (4.75mm and 5.5mm Healix Advance Knotless PEEK Anchors).
Acceptance Criteria and Device Performance Study:
Since this FDA submission is a Special 510(k) based on substantial equivalence to predicate devices, formal acceptance criteria and a single study proving the device meets them in the traditional sense (e.g., a clinical trial with specific performance endpoints) are not presented in the provided documents. Instead, the submission focuses on demonstrating that the new device has the same design, technological characteristics, and safety/performance profile as the predicate devices, despite the size change.
The "Substantial Equivalence" framework in 510(k) submissions relies on showing that the new device is as safe and effective as a legally marketed predicate device. The "acceptance criteria" here can be inferred as showing that the new device performs equivalently or comparably to the predicate in non-clinical tests relevant to its function.
1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria (Inferred fromPredicate Equivalence) | Reported Device Performance (Non-clinical Testing) |
---|---|
Similar Mechanical Properties: | |
- Insertion Torque (comparable to predicate) | Performed; demonstrated substantial equivalence |
- Torque to Failure (comparable to predicate) | Performed; demonstrated substantial equivalence |
- Anchor Pullout Strength (comparable to predicate) | Performed; demonstrated substantial equivalence |
Biocompatibility: | Manufactured from PEEK material, same as predicate; inherently considered biocompatible based on predicate. |
Sterility: | Not explicitly mentioned but assumed to meet standards for implantable devices. |
Explanation: The "acceptance criteria" are implied by the need to demonstrate substantial equivalence to the predicate devices (4.75mm and 5.5mm Healix Advance Knotless PEEK Anchors). The non-clinical testing performed (Insertion Torque, Torque to Failure, Anchor Pullout) aimed to show that the 6.5mm anchor's mechanical performance is comparable to the predicate devices, thus meeting the "acceptance criteria" for safety and effectiveness in the context of a Special 510(k).
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size for Test Set: Not explicitly stated. The document mentions "Product Design Verification and Design Validation activities," but does not detail the number of samples used for each test (Insertion Torque, Torque to Failure, Anchor Pullout).
- Data Provenance: The testing was "Non-clinical Testing" performed by the manufacturer, DePuy Mitek. The country of origin of the data is not specified, but it would have been generated in a controlled laboratory setting by the company. It is retrospective in the sense that the testing was conducted prior to submission to demonstrate equivalency.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Not applicable in this context. The "ground truth" for non-clinical mechanical testing is derived from the physical properties and performance characteristics of the devices themselves, measured through established engineering and biomechanical methods. No human experts are used to interpret the "ground truth" of these test results; rather, engineers and scientists conduct and analyze the tests against predetermined specifications or predicate performance.
4. Adjudication Method for the Test Set:
- Not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or image-based diagnostic evaluations where human interpretation of outcomes is involved. For non-clinical mechanical testing, results are quantitative and objective, and thus do not require an adjudication process.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of images or data (e.g., radiologists reading mammograms with and without AI assistance). The Healix Advance Knotless PEEK Anchor is an implantable surgical device, and its performance is evaluated through biomechanical testing and clinical outcomes, not through human reader interpretation of data.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:
- No, a standalone (algorithm only) study was not done. This type of study is specifically for evaluating the performance of AI algorithms. The device described is a physical surgical implant, not an algorithm.
7. Type of Ground Truth Used:
- The "ground truth" for the non-clinical testing was based on objective, quantitative mechanical measurements (e.g., torque values, force measurements for pullout strength). These measurements establish the intrinsic performance characteristics of the device. The comparison of these measurements to the predicate device's known performance serves as the basis for the substantial equivalence claim.
8. Sample Size for the Training Set:
- Not applicable. This device is a physical product, not an AI/ML algorithm that requires a training set. The concept of a "training set" is relevant to machine learning models.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable, as there is no training set for this device.
Ask a specific question about this device
(89 days)
DEPUY MITEK INC., A JOHNSON AND JOHNSON COMPANY
ORTHOCORD Suture is indicated for use in general soft tissue approximation, and/or ligation, including orthopaedic procedures.
Orthocord® Suture is a synthetic, sterile, braided, composite suture composed of dyed (D&C Violet #2 or D&C Blue #6) absorbable polydioxanone (PDS) and un-dyed non-absorbable polyethylene. The partially absorbable suture is coated with a copolymer of 90% caprolactone and 10% glycolide.
The provided text does not contain a study or detailed performance data that would allow for a complete answer to your request. The document is a 510(k) summary for a medical device (Small Size Orthocord® Suture) seeking substantial equivalence to previously cleared devices.
Instead of a detailed study with acceptance criteria and performance metrics, it states that:
- "The determination of substantial equivalence for this device was based on a detailed device description, and conformance to consensus and voluntary standards."
- "Non-clinical laboratory testing was performed demonstrating that the device conformed to the USP monograph for absorbable sutures, except for oversized diameter."
This indicates that the device's equivalence and safety/performance were established primarily through comparison to existing predicate devices and adherence to general standards, rather than a novel clinical or non-clinical study with specific acceptance criteria and reported device performance as you've requested for AI/software-based devices.
Therefore, I cannot populate the table or answer most of your detailed questions because the information is not present in the provided text.
However, I can extract the following limited information:
1. A table of acceptance criteria and the reported device performance:
Acceptance Criteria Category | Reported Device Performance |
---|---|
Conformance to Standards | Conformed to consensus and voluntary standards. |
USP Monograph for Absorbable Sutures | Conformed to the USP monograph for absorbable sutures, except for oversized diameter. (The specific parameters within the USP monograph are not detailed, nor is the extent of the oversized diameter mentioned). |
Material Composition | Comprised of the same materials as the predicate Size 2 Orthocord suture (K040004/K043298). |
Design Similarity | Similar design to the predicate Size 2 Orthocord suture (K040004/K043298), with differences in size and braid construction. |
2. Sample sized used for the test set and the data provenance: Not applicable. No specific "test set" or data provenance (country of origin, retrospective/prospective) is described for a study in the document.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth establishment by experts is not mentioned for this type of submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a surgical suture, not an AI/imaging diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable. The "ground truth" for this device's performance would be its physical and mechanical properties conforming to established standards and predicate device performance, not clinical ground truth in the AI sense.
8. The sample size for the training set: Not applicable.
9. How the ground truth for the training set was established: Not applicable.
Ask a specific question about this device
(76 days)
DEPUY MITEK INC., A JOHNSON AND JOHNSON COMPANY
The DePuy Mitek Tibial Tapered Screw and Tibial Sheath are indicated for fixation of soft tissue grafts during cruciate ligament reconstruction.
The proposed INTRAFIX System consists of two components, a non-absorbable INTRAFIX Expansion Sheath and a non-absorbable INTRAFIX Expansion Screw. Also included within the system is the instrumentation to place devices and establish the tunnel. The device functions by establishing the tunnel and placing the expansion Sheath into the tunnel by the use of Sheath Inserter. This is followed by screwing an Expansion Screw into the Sheath expanding the Sheath which compresses the graft against the tunnel and creating fixation.
The provided text describes a 510(k) summary for the INTRAFIX® PEEK Tapered Screw, which is a medical device. This type of submission focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing rather than a separate clinical study with predefined acceptance criteria for the device itself. Therefore, the questions related to clinical study design, sample sizes, expert involvement, and ground truth for a device performance study are not directly applicable in the context of this 510(k) summary.
However, I can extract the information relevant to non-clinical testing which served as the "study" to demonstrate substantial equivalence, and interpret "acceptance criteria" as meeting the performance of the predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Performance of Predicate Device) | Reported Device Performance (INTRAFIX® PEEK Tapered Screw) |
---|---|
Torque Capability (of predicate INTRAFIX® Tapered Screws) | Demonstrated equivalent Torque Capability |
Pullout Strength (of predicate INTRAFIX® Tapered Screws) | Demonstrated equivalent Pullout Strength |
Note: The document states "Results of performance and safety testing have demonstrated that the modified device is substantially equivalent to the predicate devices." This implies that the PEEK Tapered Screw's performance in these non-clinical tests was comparable to or met the performance standards of the predicate devices. The exact numerical values for these metrics are not provided in this summary, but the conclusion is that equivalence was achieved.
2. Sample size used for the test set and the data provenance
- Sample size for test set: Not specified in the provided summary.
- Data provenance: Non-clinical testing. The country of origin and whether it was retrospective or prospective is not applicable or specified for this type of testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. This was non-clinical engineering and mechanical testing, not a study requiring expert clinical ground truth.
4. Adjudication method for the test set
- Not applicable. This was non-clinical engineering and mechanical testing.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is a medical device for orthopedic fixation, not an AI diagnostic or image analysis device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a medical device for orthopedic fixation.
7. The type of ground truth used
- The "ground truth" for demonstrating substantial equivalence in this context was the established performance (e.g., torque capability, pullout strength) of the predicate devices.
8. The sample size for the training set
- Not applicable. This is a medical device; there is no "training set" in the context of machine learning.
9. How the ground truth for the training set was established
- Not applicable. This is a medical device; there is no "training set."
Ask a specific question about this device
(50 days)
DEPUY MITEK INC., JOHNSON AND JOHNSON COMPANY
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction;
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair;
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis;
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction:
Hip: Capsular repair, Acetabular Labral Repair.
The HEALIX ADVANCE BR Anchor is absorbable threaded suture anchor preloaded on a disposable inserter assembly intended for fixation soft tissue to bone. The suture options may include needles to facilitate suture passage through tissue. HEALIX ADVANCE BR Anchor is provided sterile and is for single patient use only.
The provided document describes a medical device, the HEALIX ADVANCE™ BR Anchor, and its substantial equivalence to previously marketed predicate devices. However, the document does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of a diagnostic or AI-based device.
This document is a 510(k) summary for a bone anchor, which is a physical medical device (implant) used for fixation of soft tissue to bone. The "Safety and Performance" section states: "Design verification activities, such as Anchor Torque, Anchor Pull Out (at T=0 and in vitro testing throughout the healing period) were performed against pre-defined acceptance criteria according to the intended use. Results of performance testing have demonstrated that the proposed devices are suitable for their intended use."
This indicates that mechanical performance testing was done for the physical implant, but the details of the acceptance criteria, specific study results, sample sizes for these tests, or any form of "ground truth" or expert adjudication are not included in this regulatory summary. These details would typically be found in the full 510(k) submission or design verification reports, which are not provided here.
Therefore, I cannot fulfill the request as it pertains to acceptance criteria and a study proving a diagnostic device's performance, an AI-based device's performance, or a multi-reader multi-case (MRMC) comparative effectiveness study. The request's questions are geared towards the evaluation of diagnostic algorithms or software, which is not the subject of this document.
Ask a specific question about this device
Page 1 of 1