Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K012821
    Date Cleared
    2001-10-22

    (60 days)

    Product Code
    Regulation Number
    886.4170
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The D.O.R.C. Harmony CryoStar Cryosurgical System is a surgical system for ophthalmic surgery intended for use in surgery of the posterior segment or anterior segment including cryopexy for retinal detachment, glaucoma, cataract extraction, trichiasis, and retinopathy of prematurity.

    Device Description

    The D.O.R.C. CryoStar Cryosurgical System consists of a cryogen gas bottle, a console, an ophthalmic probe, a footswitch and necessary interconnections. The console controls delivery of the gas to the probe. It also processes signals from the footswitch, enabling the probe to freeze and defrost.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device called the D.O.R.C. CryoStar Cryosurgical System. It claims substantial equivalence to predicate devices, and therefore no clinical study was performed.

    Here's an analysis of the provided text based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/Acceptance Criteria (Implied)D.O.R.C. CryoStar Cryosurgical System (Reported Performance)
    Console TechnologyElectro-mechanical
    Software ContentNon-programmable firmware
    Pressure ControlManual
    Temperature MeasurementThermocouple (for some probes in range only)
    Indicator Panel: TimerDigital display
    Indicator Panel: PressureLED array
    Indicator Panel: TemperatureLED array
    Power SourceMains
    Gas InletScrew connection
    Gas OutletBarbed fitting
    Probe ConnectionScrew fitting
    ProbesOphthalmic standard freeze and end-freeze, some with thermocouples. Re-usable when sterilized by autoclaving.
    Probe AdaptorUniversal (suits probes with or without thermocouples)
    Consumable Cryogen Gases (Nitrous Oxide)Medical grade, vapour withdrawal gas
    Consumable Cryogen Gases (Carbon Dioxide)Medical grade, vapour withdrawal gas

    Study Proving Device Meets Acceptance Criteria:

    The provided document is a 510(k) submission, not a study report. The "study" that proves the device meets the acceptance criteria is a demonstration of substantial equivalence to predicate devices. This means the manufacturer presented evidence that the D.O.R.C. CryoStar Cryosurgical System has the same intended use and similar technological characteristics to previously cleared devices (D.O.R.C. Cryo 1500 and Keeler Cryo Master) and that any differences do not raise new questions of safety or effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: Not applicable. This is not a clinical study involving a test set of data. The submission relies on a comparison of technical specifications and intended use.
    • Data Provenance: Not applicable. The data is descriptive information about the device itself and a comparison to predicate devices, not data collected from a clinical setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    Not applicable. No ground truth was established by experts for a test set in this context. The determination of substantial equivalence is made by the FDA based on the provided technical information and comparison to predicate devices.

    4. Adjudication Method for the Test Set:

    Not applicable. There was no test set or adjudication process as part of a clinical study.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a cryosurgical system, not an AI-powered diagnostic tool engaging human readers. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a surgical instrument, not an algorithm or AI system.

    7. The Type of Ground Truth Used:

    Not applicable in the typical sense of a clinical study. The "ground truth" in this 510(k) submission is the established safety and effectiveness of the predicate devices (D.O.R.C. Cryo 1500 and Keeler Cryo Master) based on their prior FDA clearance. The new device is deemed substantially equivalent if it can demonstrate it performs safely and effectively like these predicate devices.

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI/machine learning study requiring a training set.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K010082
    Date Cleared
    2001-03-12

    (61 days)

    Product Code
    Regulation Number
    886.4150
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K983797
    Date Cleared
    1999-03-03

    (126 days)

    Product Code
    Regulation Number
    886.4150
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The D.O.R.C. Harmony Total Vitrectomy System is a surgical system for ophthalmic surgery intended for use in surgery of the posterior segment (vitrectomy).

    Device Description

    The D.O.R.C. Harmony Total Vitrectomy System is an integrated system for posterior segment eye surgery. It provides the majority of functions required by the ophthalmic surgeon for performing a vitrectomy including vitreous cutting and aspiration, phacofragmentation, illumination, diathermy, air system and viscous fluid injection/extraction. The system is operated from one multifunction footswitch which gives the surgeon control over all of the surgical functions. The D.O.R.C. Harmony Total Vitrectomy System which includes the system, multifunction footswitch, trolley, automatic infusion pole and various sterile and non-sterile accessories provides the majority of functions needed to complete a posterior segment vitrectomy. The system is ac powered and also requires a source of medical grade compressed air to provide for some of the system functions.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for the D.O.R.C. Harmony Total Vitrectomy System. This document focuses on demonstrating substantial equivalence to existing predicate devices, rather than presenting a study with specific acceptance criteria and performance metrics in the way one might for a novel diagnostic or AI-powered medical device.

    Therefore, many of the requested elements (e.g., sample size for test/training sets, number of experts, adjudication methods, MRMC studies, standalone performance, type of ground truth) are not applicable to this type of regulatory submission.

    The "study" in this context is the comparison of technological characteristics of the D.O.R.C. Harmony Total Vitrectomy System to its predicate devices. The acceptance criteria are implicit in the FDA's "substantial equivalence" determination, meaning the new device must be as safe and effective as a legally marketed predicate device.

    Here's an breakdown based on the provided text, addressing the points where information is available:

    1. Table of Acceptance Criteria and Reported Device Performance

    For a 510(k) submission like this, there isn't a formally stated "acceptance criteria" in terms of specific performance metrics with PASS/FAIL thresholds for a novel diagnostic. Instead, the acceptance criteria are satisfied by demonstrating that the new device's technological characteristics are substantially equivalent to legally marketed predicate devices. The "reported device performance" is essentially the listed specifications of the D.O.R.C. Harmony Total Vitrectomy System, as compared to the predicate devices.

    Feature/SpecificationAcceptance Criteria (Implicit: Comparable to Predicate)D.O.R.C. Harmony Total PerformancePredicate Devices (Range/Example)
    Intended UsePosterior segment ophthalmic surgeryPosterior segment ophthalmic surgerySame
    Vitrectomy TypeGuillotine cutter (or equivalent)Guillotine cutterGuillotine or radial reciprocating
    Vitrectomy Drive MechanismPneumatic or electricPneumatic or electricPneumatic
    Maximum Cut RateComparable to predicate devices800 cuts/minute750 - 2500 cuts/minute (predicates show a range)
    Aspiration TypeVenturi (or equivalent)VenturiVenturi
    Linear ControlPresentYESYES
    Maximum VacuumComparable to predicate devices500 mm Hg500 - 600 mm Hg
    PhacofragmentationPresent (if predicate has it)YESYES
    Phaco FrequencyComparable to predicate devices40 kHz28.5 kHz - 40 kHz
    DiathermyPresent (if predicate has it)YESYES
    Diathermy FrequencyComparable to predicate devices13.56 MHz600 kHz (one predicate)
    Diathermy Max PowerComparable to predicate devices12 watts7.5 - 10 watts
    Fiber Optic Light SourcePresent (if predicate has it)YESYES
    Light OutputDual/SingleDualDual
    Lamp TypeHalogen (or equivalent)HalogenUnknown for predicates
    Lamp WattageComparable to predicate devices150 wattsUnknown for predicates
    Air InfusionPresent (if predicate has it)YESYES
    Air Infusion Max PressureComparable to predicate devices95 mm Hg99 - 120 mm Hg
    Viscous Fluid InjectionPresent (if predicate has it)YESNO (for Storz, MID Labs, Scieran)
    Viscous Fluid Injection Viscosities1000 or 5000 cs1000 or 5000 cs1000 or 5000 cs (Alcon)
    Viscous Fluid ExtractionPresent (if predicate has it)YESNO (for Storz, MID Labs, Scieran)
    Multifunction FootswitchPresentYESYES

    2. Sample size used for the test set and the data provenance

    • Not Applicable. This is a comparison of device specifications to predicate devices, not a study involving patient data or a "test set" in the context of diagnostic accuracy. The provenance of the data is the device's own specifications and publicly available information about predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. No "ground truth" derived from expert assessment of a test set is mentioned or required for this type of submission. The "ground truth" is the demonstrated functionality and specifications of the device itself and its predicates.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No adjudication process is described for a test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI or diagnostic device that involves human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. This is a surgical system, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this submission is the functional specifications and intended use of the D.O.R.C. Harmony Total Vitrectomy System and its predicate devices. Substantial equivalence is established by demonstrating that these characteristics are similar enough to ensure equivalent safety and effectiveness.

    8. The sample size for the training set

    • Not Applicable. This is a hardware surgical system, not a machine learning model requiring a training set.

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K980636
    Date Cleared
    1998-05-20

    (90 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fiber Optic Endoillumination Probe is indicated for intraocular illumination in vitreoretinal surgery.

    Device Description

    The Fiber Optic Endoillumination Probe provides intraocular illumination. The light is transmitted from the source to the inside of the eye through a flexible optical fiber, which has been threaded through an illumination probe.

    AI/ML Overview

    This 510(k) summary (K980636) for the Fiber Optic Endoillumination Probe does not describe a study with acceptance criteria and reported device performance. Instead, it demonstrates substantial equivalence to a predicate device (Infinitech, Inc. Fiber Optic Probe, K870942) based on operational and technological characteristics.

    Therefore, most of the requested information regarding acceptance criteria, study details, and performance metrics is not available in the provided text.

    Here's a breakdown of what can be extracted and what is not present:

    1. Table of Acceptance Criteria and Reported Device Performance

    Not explicitly provided.

    The submission relies on a comparison table to establish substantial equivalence, not to report performance against specific acceptance criteria. The "Comparison Criteria" act as attributes that are matched for equivalence rather than performance metrics with thresholds.

    Comparison CriteriaAcceptance Criteria (Implied for Equivalence)Reported Device Performance (Dutch Ophthalmic, USA Fiber Optic Endoillumination Probe)
    Device Type: Fiber Optic ProbeMust be "Yes" (a Fiber Optic Probe)Yes
    Indication: Intraocular illumination in vitreoretinal surgeryMust be "Yes" (for intraocular illumination in vitreoretinal surgery)Yes
    Patient Contact Materials: PMMA and fluoropolymer fiber and medical grade Surgical stainless steelMust be "Yes" (matching materials)Yes
    Sterilization Method: Validated ETOMust be "Validated ETO" (or equivalent, typically requiring validation data)Validated ETO
    Packaging: Validated Heat Sealed TyvekMust be "Validated Heat Sealed Tyvek" (or equivalent, typically requiring validation data)Validated Heat Sealed Tyvek

    Note: The predicate device details for "Sterilization Method" and "Packaging" are "Not Known," indicating that Dutch Ophthalmic's methods were assessed for safety and effectiveness on their own merits and found acceptable, rather than directly comparing to a known predicate value.

    2. Sample size used for the test set and the data provenance

    Not applicable/Not provided. This submission does not detail an experimental test set with sample sizes. It's a regulatory submission for substantial equivalence based on descriptive and technological characteristics.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable/Not provided. No test set or ground truth established by experts is described.

    4. Adjudication method for the test set

    Not applicable/Not provided. No test set or adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is not an AI-enabled device. It's a fiber optic endoillumination probe. Therefore, no MRMC study or AI-related effectiveness is discussed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This is not an algorithm-only device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable/Not provided. No ground truth is described as the device is for illumination, not diagnosis or measurement requiring ground truthful comparison.

    8. The sample size for the training set

    Not applicable/Not provided. This is not an AI/machine learning device, so there is no training set mentioned.

    9. How the ground truth for the training set was established

    Not applicable/Not provided. As there's no training set, there's no ground truth established for one.

    Ask a Question

    Ask a specific question about this device

    K Number
    K973229
    Date Cleared
    1997-11-25

    (90 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The D.O.R.C. HEXON Illumination System is indicated for intraocular illumination in vitreoretinal surgery.

    Device Description

    The D.O.R.C. HEXON Illumination System consists of the Illumination Unit and accessories. The Illumination Unit uses a metal halide discharge lamp and utilizes an internal focusing system to focus the light into the end of the optical fiber. Accessories to the Illumination Unit include adapters and color filters. Accessories to the system include single use and reusable fiberoptic probes, fibers, and microinstruments which require storilization prior to use.

    AI/ML Overview

    The provided text is a 510(k) premarket notification summary for the D.O.R.C. HEXON Illumination System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting a full clinical study to prove novel performance or efficacy. Therefore, much of the requested information regarding acceptance criteria, specific study design, expert involvement, and detailed performance metrics as would be found in a clinical trial for a new therapeutic or diagnostic device is not applicable or not provided in this document.

    Here's a breakdown based on the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in the traditional sense of a performance study with specific thresholds. Instead, it demonstrates performance by comparing the device's characteristics to those of predicate devices to establish substantial equivalence.

    CharacteristicAcceptance Criteria (Implied by Predicate - "YES")Reported Device Performance (HEXON Illum. System)
    Indication: Endoillumination for vitreoretinal surgeryYES (Matches Predicate)YES
    Lamp TypeAcceptable (Differences noted vs. Grieshaber)Metal Halide
    Lamp RatingNot specified by Predicate24 watts
    Light Output (lumens)Not specified by Predicate1850
    Color Temperature (degrees K)Acceptable (Similar to VitLite I)4700
    Variable IntensityYES (Matches Predicate)Yes
    Wavelength Range (nm)Acceptable (Similar to VitLite I)400-800
    UV filtrationYES (Matches Predicate)Yes
    Infrared filtrationYES (Matches Predicate)Yes
    Additional color filtration (Green, Yellow, Red, Blue)YES (Matches Grieshaber)Yes (for all listed)
    Additional color filtration (Daylight)Not specified by Predicate (HEXON offers it)Yes
    Multiple probe diametersYES (Matches Predicate)Yes
    Panoramic light probeYES (Matches VitLite I)Yes
    Straight probeYES (Matches Predicate)Yes
    Illuminated accessoriesYES (Matches Predicate)Yes H

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not applicable. This is not a study assessing device performance on patient samples. The "test set" here refers to the characteristics of the device itself and its comparison to predicate devices, not data from patients.
    • Data Provenance: Not applicable in the context of clinical data. The data provenance relates to the specifications and characteristics of the D.O.R.C. HEXON Illumination System and the predicate devices (Grieshaber Light Source and Escalon VitLite I), derived from engineering specifications and existing market information.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. Ground truth in a clinical or diagnostic sense is not being established. The "truth" here is the stated technical specifications of the devices being compared.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. There is no expert adjudication of clinical findings in this type of submission. The comparison is based on documented device specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is an ophthalmic illumination system, not an AI or diagnostic tool that would involve "human readers" or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a medical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this 510(k) submission is the technical specifications and intended use of the legally marketed predicate devices. The D.O.R.C. HEXON Illumination System demonstrates substantial equivalence by showing that it has similar technological characteristics and the same intended use as these predicates.

    8. The sample size for the training set

    • Not applicable. There is no "training set" in the context of device development for a 510(k) involving an illumination system.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K973757
    Date Cleared
    1997-11-07

    (36 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The D.O.R.C. SOLUX Light Source is indicated for intraocular illumination in vitreoretinal surgery.

    Device Description

    The D.O.R.C. SOLUX Light Source uses a halogen lamp and an internal focusing system to focus the light into the end of the optical fiber. Accessories to the Illumination Unit include adapters and color filters.

    AI/ML Overview

    This 510(k) summary (K973757) describes the D.O.R.C. SOLUX Light Source, an ophthalmic light source intended for intraocular illumination in vitreoretinal surgery. The submission establishes substantial equivalence to existing predicate devices (Grieshaber Light Source and Escalon VitLite I) primarily based on operational and technological characteristics, rather than clinical performance studies with acceptance criteria in the typical sense of AI/software devices.

    Here's an analysis based on the provided document, addressing your specific points:

    #Information RequestedDetails from K973757
    1Table of acceptance criteria and reported device performanceAcceptance Criteria (Implied by Predicate Comparison): The device (SOLUX Light Source) must have technological and operational characteristics similar to legally marketed predicate devices, particularly in areas like intended use (intraocular illumination for vitreoretinal surgery), lamp type (Halogen), variable intensity, UV filtration, and ability to adapt to multiple accessories. While specific numerical "acceptance criteria" are not explicitly stated for each characteristic in the way one would for a diagnostic accuracy study, the implicitly accepted performance is that which is "substantially equivalent" to the predicate devices. Reported Device Performance (SOLUX Light Source):- Indication: Endoillumination for vitreoretinal surgery (YES)- Lamp Type: Halogen- Lamp Rating: 100 watts- Light Output (lumens): High: 3800, Low: 3200- Color Temperature (degrees K): High: 3220, Low: 3160- Variable Intensity: Yes- Wavelength Range (nm): 400-800- UV filtration: Yes- Additional color filtration: Green, Yellow, Red, Blue, Daylight (all Yes)- Adapts to multiple accessories: Yes
    2Sample size for the test set and data provenanceThis is a conventional medical device (light source), not a software/AI device. Therefore, there is no "test set" or "data provenance" in the context of diagnostic performance evaluation. The substantial equivalence is based on engineering specifications and comparison to predicate devices, not on data from patient cases.
    3Number of experts to establish ground truth for test set and qualificationsNot applicable. No test set for ground truth establishment.
    4Adjudication method for the test setNot applicable. No test set.
    5MRMC comparative effectiveness studyNot applicable. This is a light source, not a diagnostic AI system where human readers would improve with AI assistance.
    6Standalone (algorithm only) performance studyNot applicable. This is a hardware device, not an algorithm. Standalone performance for such a device would refer to its electrical, optical, and mechanical specifications as an engineering product, which were implicitly demonstrated or compared to predicates to establish substantial equivalence.
    7Type of ground truth usedNot applicable. The "ground truth" for this type of device is its physical specifications and functionality, which are compared against known capabilities of predicate devices. There is no diagnostic "ground truth" from pathology, expert consensus, or outcomes data in this context.
    8Sample size for the training setNot applicable. This is not an AI/ML device. There is no "training set."
    9How the ground truth for the training set was establishedNot applicable. No training set.

    Summary of the Study and Acceptance Criteria:

    The "study" in this context is a 510(k) premarket notification which aims to demonstrate substantial equivalence to legally marketed predicate devices. The primary method is a comparison of technological and operational characteristics.

    • Acceptance Criteria (Implied): The D.O.R.C. SOLUX Light Source is accepted if its technical specifications and intended use are similar enough to the predicate devices (Grieshaber Light Source and Escalon VitLite I) that it can be considered equally safe and effective. The comparison table (page 1) serves as the core of this demonstration. For instance, having "YES" for "Indication: Endoillumination for vitreoretinal surgery" for all devices shows alignment in intended use. The fact that its lamp type is Halogen, like the Grieshaber, and its wavelength range (400-800 nm) encompasses that of the Escalon VitLite (450-700 nm), along with features like UV and color filtration, contribute to this claim of equivalence.

    • Proof of Meeting Criteria: The provided table titled "Comparison of SOLUX Light Source to Predicate Devices" directly shows how the D.O.R.C. SOLUX Light Source's characteristics align with or are similar to the predicate devices across various attributes like lamp type, variable intensity, filtration, and intended use. The FDA's letter (page 2-3) confirming "substantial equivalence" is the official recognition that the device meets these implied acceptance criteria based on the information provided in the 510(k) submission.

    This 510(k) is characteristic of traditional medical device clearances based on engineering comparisons rather than clinical performance data typical of AI/ML-powered diagnostic devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K962135
    Date Cleared
    1996-08-29

    (87 days)

    Product Code
    Regulation Number
    886.4115
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K954842
    Date Cleared
    1996-04-15

    (175 days)

    Product Code
    Regulation Number
    886.4360
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1