510(k) Data Aggregation

    K Number
    K182683
    Date Cleared
    2019-06-27

    (274 days)

    Product Code
    Regulation Number
    892.5770
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Chinan Biomedical Technology, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Z-Robot™ Patient Positioning System is an electro-mechanical robotic arm for patient positioning in radiotherapy and medical imaging. It is designed for positioning a patient with a high degree of accuracy and repeatability.

    Device Description

    The Z-Robot™ Patient Positioning System (hereinafter referred to as Z-Robot™) is an electro-mechanical robotic arm with the motion capability in six degrees of freedom. These axes allow the couch to be precisely positioned translationally and rotationally within the specified movement limits. Therefore, Z-Robot™ could be used to accurately position patient in radiotherapy and medical imaging.

    AI/ML Overview

    The requested information is contained within the provided text. The device in question is the "Z-Robot™ Patient Positioning System", which is an electro-mechanical robotic arm for patient positioning in radiotherapy and medical imaging. The acceptance criteria and the study that proves the device meets these criteria are primarily based on non-clinical testing.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly provide a "table of acceptance criteria and reported device performance" for all tests. However, it lists the performance parameters tested and indicates that "All test results were acceptable per the acceptance criteria detailed in the corresponding protocol and test report." It also provides a direct comparison to the predicate device for several key performance metrics, implying these are the relevant acceptance criteria for substantial equivalence.

    Performance CharacteristicAcceptance Criteria (Implied by Predicate & "Same")Reported Device Performance (Z-Robot™ Patient Positioning System)
    Accuracy± 0.5 mm and ± 0.2°± 0.5 mm and ± 0.2° (Same as predicate)
    Linear XYZ maximal medical speed10 cm/s10 cm/s (Same as predicate)
    Linear XYZ maximum maintenance speed20 cm/s20 cm/s (Same as predicate)
    Angular Yaw Pitch Roll maximum medical speed5 °/s5 °/s (Same as predicate)
    Angular Yaw Pitch Roll maximum maintenance speed12 °/s12 °/s (Same as predicate)
    Collision detectionDetection of a 150 N forceDetection of a 150 N force (Same as predicate)
    "Overtravel" in case of emergency stop