    K Number
    K243842
    Device Name
    Intra-Operative Positioning System (IOPS®)
    Manufacturer
    Centerline Biomedical, Inc.
    Date Cleared
    2025-03-06

    (83 days)

    Product Code
    DQK, DQX, DQY
    Regulation Number
    870.1425
    Why did this record match?
    Applicant Name (Manufacturer) :

    Centerline Biomedical, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.
    Device Description
    The Intra-Operative Positioning System (IOPS) consists of a surgical navigation technology and a number of associated accessories. The navigation technology is a non-contact reusable multi-patient use device. The associated accessories are single use devices provided sterile (EtO). The IOPS displays the position and orientation of sensor equipped catheters, guidewires, and tracking pad utilizing electromagnetic tracking technology. The system enables mapping of the patient's vascular system utilizing previously acquired CT scan data. IOPS registers the location and orientation of the sensors in real time, superimposing navigation of the catheters and guidewires to the patient's vascular map displayed on a monitor. The system is for use as an adjunct to fluoroscopy and does not make a diagnosis. The associated accessories include: - Guidewire - Catheters - Fiducial Tracking Pad - Guidewire Handle
    Ask a Question
    K Number
    K242133
    Device Name
    Intra-Operative Positioning System (IOPS®) (Fiducial Tracking Pad); Intra-Operative Positioning System (IOPS®) (Guidewire Handle)
    Manufacturer
    Centerline Biomedical, Inc.
    Date Cleared
    2024-10-11

    (81 days)

    Product Code
    DQK
    Regulation Number
    870.1425
    Why did this record match?
    Applicant Name (Manufacturer) :

    Centerline Biomedical, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and quidewires used in endovascular interventions in the descending aorta. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.
    Device Description
    IOPS Fiducial Tracking Pad is a sterile, single use tracking pad intended for use with the Intra-Operative Positioning System (IOPS), manufactured by Centerline Biomedical. It is equipped with a single tracking sensor allowing IOPS to track gross patient motion to allow maintenance of patient registration during a procedure. It will not track minor patient motion such as breathing or cardiac movement. The tracking pad is equipped with radiopaque beads which allow registration of a conebeam CT scan of the patient in their current position to a previously acquired CT scan. IOPS Guidewire Handle is a sterile, single use, non-sensorized device intended for use with the Intra-Operative Positioning System (OPS) and its sensorized 0.035″ guidewire, manufactured by Centerline Biomedical. It is meant to be connected to an IOPS sensorized guidewire to allow detection and visualization of the guidewire tip position, in real time, on a 3D rendering of the patient's vascular map.
    Ask a Question
    K Number
    K241243
    Device Name
    Intra-Operative Positioning System (IOPS®) Viewpoint (Simple Curve Catheters); Intra-Operative Positioning System (IOPS®) Viewpoint (Double Curve Catheters)
    Manufacturer
    Centerline Biomedical, Inc.
    Date Cleared
    2024-07-24

    (82 days)

    Product Code
    DQK, DOK, DOY
    Regulation Number
    870.1425
    Why did this record match?
    Applicant Name (Manufacturer) :

    Centerline Biomedical, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the descending aorta. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.
    Device Description
    The IOPS Viewpoint Catheters are sterile, single use 6 Fr catheters intended for use with the Intra-Operative Positioning System (IOPS), manufactured by Centerline Biomedical. They are designed to be inserted into the femoral artery, brachial artery, or axillary artery and navigate through vasculature in the descending aorta to access branch vessels near to, or involved in, the lesion during endovascular procedures. The IOPS Viewpoint Catheters are available in two tip shape configurations (Simple and Double). Each tip shape is available in two lengths (75cm and 125cm). All models are equipped with multiple tracking sensors allowing the loPS to detect and visualize the catheter tip position and shape, in real time, on a 3D rendent's vascular map. The catheters have a radiopaque marker near the distal tip and a luer locking hub on the proximal end.
    Ask a Question
    K Number
    K230309
    Device Name
    Intra-Operative Positioning System (IOPS®) (MC-1); IOPS Simple Curve Catheter (SCC-1) and IOPS Reverse Curve Catheter (RCC-1); IOPS Guidewire (ATW-2); IOPS Tracking Pad (TP-1); IOPS Guidewire Handle (SSH-1)
    Manufacturer
    Centerline Biomedical, Inc.
    Date Cleared
    2023-06-21

    (138 days)

    Product Code
    DQK, DOK
    Regulation Number
    870.1425
    Why did this record match?
    Applicant Name (Manufacturer) :

    Centerline Biomedical, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The IOPS® (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor equipped compatible catheters and guidewires used in endovascular interventions in the descending aorta. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.
    Device Description
    The IOPS system displays the position and orientation of sensor equipped guidewires and catheters utilizing electromagnetic tracking technology. The system enables mapping of the patient's vascular system utilizing previously acquired CT scan data. IOPS registers the location and orientation of the sensors in real time superimposing navigation of the catheters and guidewires to the patient's vascular map.
    Ask a Question
    K Number
    K190106
    Device Name
    Intra-Operative Positioning System; Simple Curve Catheter, Reverse Curve Catheter ; Angled Tip Guidewire ; Tracking Pad ; Guidewire Handle
    Manufacturer
    Centerline Biomedical, Inc.
    Date Cleared
    2019-06-24

    (153 days)

    Product Code
    DQK
    Regulation Number
    870.1425
    Why did this record match?
    Applicant Name (Manufacturer) :

    Centerline Biomedical, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor equipped compatible catheters and guidewires used in endovascular interventions in the descending aorta. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.
    Device Description
    The IOPS system displays the position and orientation of sensor equipped guidewires and catheters utilizing electromagnetic tracking technology. The system enables mapping of the patient's vascular system utilizing previously acquired CT scan data. IOPS registers the location and orientation of the sensors in real time superimposing navigation of the catheters and guidewires to the patient's vascular map. The patient's vascular map is generated using a contrast enhanced, high resolution CT scan. The IOPS creates a 3D rendering of that structure. A bone segmented 3D rendering may optionally be created to provide anatomical, skeletal points visible in relation to the vascular rendering. The main principles of action for the IOPS are similar to those used in Global Positioning System (GPS) tracking. The navigation components generate a time-varying magnetic field in which the position and orientation of sensor embedded catheters and guidewires are read. The computing unit visually displays the location of the sensor on the patient's vascular map. The system is intended for use by trained clinicians for patients undergoing endovascular interventional procedures of the descending aorta, such as stent grafting. The system promotes more efficient use of operating room time and minimizes the need for fluoroscopy. The clinician uses the IOPS catheters and guidewires to navigate through the aorta to access branch vessels near to, or involved in, the lesion. The catheters and guidewires are not for angiographic or diagnostic use. IOPS is composed of a mobile cart which houses a monitor, computer, keyboard, pointing device, uninterruptable power supply (UPS), and cables. IOPS includes a tracking system composed of a system control unit (SCU), system interface unit (SIU), field generator, mounting brackets, and cables. These components are reusable and not patient contacting. The IOPS is integrated with software to generate the mapping and overlay of the live sensors. The IOPS works with a sensor embedded catheter, guidewire, and tracking pad which are provided sterile and not intended for re-use.