Search Results
Found 3 results
510(k) Data Aggregation
(152 days)
COEUR LABORATORIES, INC.
For use with Medrad Mark II, III,and IV Angiographic Injectors.
The Coeur syringe is a six piece device (plunger jacket, barrel, retainer, dust cap, fill tube) which is used in conjunction with power injectors to inject diagnostic fluids associated with angiographic procedures. In particular, the plunger jacket acts as the seal when the plunger is moved to aspirate or inject the contrast media solution. This premarket notification describes the modification to the Coeur syringe, which is a material change to the plunger jacket. The new material Coeur intends to use for the plunger jacket is chlorobutyl.
This document is a 510(k) summary for a medical device submitted to the FDA, which generally focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed clinical study with acceptance criteria in the same way a new drug or novel medical device might.
Based on the provided text, the acceptance criteria and study information are quite limited, as the submission focuses on a material modification to an existing device (changing the plunger jacket material from natural rubber to chlorobutyl). The performance evaluation is primarily comparative to the original device.
Here's an attempt to extract and interpret the requested information based on the available text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criteria (Inferred) | Reported Device Performance |
---|---|---|
Functional Performance | Perform as well as natural rubber (current material for plunger jacket). | "Pre-sterilization test results passed all test criteria." |
Maintain intended function with Medrad Mark II, III, and IV Angiographic Injectors. | "functional performance of the chlorobutyl plunger jacket within the Coeur syringe" was evaluated. "At this time the chlorobutyl material performs as well as natural rubber". | |
Biocompatibility | Acceptable biocompatibility for chlorobutyl material. | "Biocompatability testing has been conducted on the chlorobutyl material and all test results are acceptable." |
Sterilization | Withstand Ethylene Oxide gas sterilization with acceptable post-sterilization functional performance. | "Currently the test product is being sterilized so that post sterilization functional performance can be evaluated." (This indicates it was in progress at the time of the summary, but the final FDA clearance implies it was found acceptable.) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated. The text mentions "Pre-production testing" and "the test product," but does not specify the number of syringes or tests conducted.
- Data Provenance: Not explicitly stated, but implied to be internal testing conducted by Coeur Laboratories, Inc. (the manufacturer). This would be considered retrospective testing of internal manufacturing and material changes. There is no mention of country of origin for the data, but the company is based in Raleigh, NC, USA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not applicable/Not mentioned. This type of submission (material change to a syringe) does not typically involve expert consensus to establish "ground truth" in the way clinical diagnostic devices do. The "truth" is established by adherence to functional and biocompatibility standards.
4. Adjudication Method for the Test Set
- Not applicable/Not mentioned. Adjudication methods like 2+1 or 3+1 are used in clinical studies for diagnostic accuracy, which is not the focus here. The performance was assessed against predefined functional and biocompatibility criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No. This type of study is not relevant for a material change to a syringe. MRMC studies are used for diagnostic accuracy involving human readers interpreting images or data.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- No. This is a physical medical device (syringe), not an algorithm or AI system.
7. The Type of Ground Truth Used
- The "ground truth" for this device's performance is based on engineering and biocompatibility standards and benchmarks, specifically comparing the chlorobutyl material's performance to the previously used natural rubber, and ensuring it meets established functional requirements for angiographic syringes, along with acceptable biocompatibility profiles according to recognized tests. This is a conformance to specifications approach rather than a diagnostic "ground truth."
8. The Sample Size for the Training Set
- Not applicable/Not mentioned. There is no "training set" in the context of this device. Training sets are relevant for machine learning algorithms.
9. How the Ground Truth for the Training Set Was Established
- Not applicable/Not mentioned. See point 8.
Ask a specific question about this device
(88 days)
COEUR LABORATORIES, INC.
The Coeur Front Load Injector Retrofit Kit is intended to be used with Medrad MCT and MCT Plus Injectors to hold the Coeur 200 mL Front Load Syringe in place in front of the Iniector Ram during the injection of contrast media into the vascular system for diagnostic purposes.
The Coeur Front Load Retrofit Kit consists of two accessory pieces for use with Medrad MCT and MCT Plus Injector Systems and older models of Medrad injectors upgraded by Medrad to resemble the MCT and MCT Plus. The kit contains a face plate and ram tip extension that allow easy use of the Coeur 200 mL Front Load Syringe on the modified versions of Medrad injectors. This premarket notification describes the Coeur Face Plate Retrofit Kit, which is a modification of the Coeur Front Load Injector Turret, which allows use of the Coeur 200 mL Front Load Syringe on older models of Medrad injectors.
The Face Plate uses a standard bayonet mount to attach directly onto the Medrad injectorbase on the MCT and MCT Plus injectors. The ram tip extension is then attached to the existing ram tip and tightened into place to prevent slippage. Once in place, the Coeur 200 mL Front Load Syringe may be used easily on the injectors.
This is a submission for a medical device (injector retrofit kit), not an AI/ML device. Therefore, many of the requested categories (e.g., ground truth, training set, multi-reader multi-case studies) are not applicable.
Here's the information derived from the provided text, focusing on what is relevant for a hardware modification:
Acceptance Criteria and Device Performance for Coeur Front Load Injector Retrofit Kit
Acceptance Criteria Category | Acceptance Criteria | Reported Device Performance |
---|---|---|
Component Specifications | Components manufactured to specifications. | "Verification data was summarized to demonstrate that the Coeur Front Load Injector Retrofit Kit components could be produced to specification." |
Simulated Clinical Usage | Performance with Retrofit Kit and Coeur 200 mL Front Load Syringe is equivalent to predicate device (Medrad injector with Coeur Front Load Injector Turret and Coeur 200 mL Front Load Syringe). | "Simulated clinical usage tests showed that a Medrad injector fitted with the Retrofit Kit and the Coeur 200 mL Front Load Syringe performed as well as the Medrad injector fitted with the previously cleared Coeur Front Load Injector Turret and Coeur 200 mL Front Load Syringe." |
Additional Information:
- Sample Size for Test Set and Data Provenance: Not explicitly stated. The text mentions "verification data" and "simulated clinical usage tests" but does not specify the number of units or tests conducted.
- Number of Experts and Qualifications: Not applicable. This is a hardware modification, not a diagnostic or AI device requiring expert interpretation for ground truth.
- Adjudication Method: Not applicable.
- Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not applicable as this is a hardware device. No human readers or AI assistance are involved in performance evaluation in this context.
- Standalone Performance (Algorithm only without human-in-the-loop): Not applicable as this is a hardware device. No algorithm is involved.
- Type of Ground Truth Used: The "ground truth" here is the adherence to manufacturing specifications and functional equivalence to a predicate device in simulated use. This is based on physical measurements and functional testing against established benchmarks (predicate device performance).
- Sample Size for Training Set: Not applicable as this is a hardware device and does not involve machine learning.
- How Ground Truth for Training Set was Established: Not applicable.
Ask a specific question about this device
(185 days)
COEUR LABORATORIES, INC.
Ask a specific question about this device
Page 1 of 1