Search Results
Found 2 results
510(k) Data Aggregation
K Number
K090720Device Name
COCHLEAR BAHA BP100
Manufacturer
COCHLEAR BONE ANCHORED SYSTEMS AB
Date Cleared
2009-06-17
(91 days)
Product Code
LXB
Regulation Number
874.3302Why did this record match?
Applicant Name (Manufacturer) :
COCHLEAR BONE ANCHORED SYSTEMS AB
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Cochlear Baha BP100 sound processor is intended for use with the Baha auditory osseointegrated implant (for children aged 5 and older, or adults), or with the Baha Headband or Baha Softband (no age limitations), for the following patients and indications:
- Patients who have a conductive or mixed hearing loss and can still benefit from sound amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 45 dB HL.
- Bilateral fitting of the BP100 is intended for patients who meet the above criterion in both ears, with bilaterally symmetric moderate to severe conductive or mixed hearing loss. Symmetrical bone-conduction thresholds are defined as less than a 10 dB average difference between ears (measured at 0.5, 1, 2, and 3 kHz), or less than a 15 dB difference at individual frequencies.
- Patients who suffer from unilateral sensorineural deafness in one ear with normal hearing in the other ear (i.e. single-sided deafness or "SSD"). Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL.
- Baha for SSD is also indicated for any patient who is indicated for an airconduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
Device Description
The Baha implant system works by combining a sound processor with an abutment and a small titanium implant placed in the skull behind the ear. The system is based on the process of "osseointegration" through which living tissue integrates with titanium in the implant. Thus, the titanium implant becomes one with the bone, allowing high-quality amplified and processed sound to be conducted via the skull bone directly to a cochlea with residual functionality. Baha sound processors can also be fitted as external (non-implanted) bone-conduction vibrators with either the Baha Headband or Baha Softband as an alternative to a conventional bone-conduction hearing aid.
The Baha BP100 will be the newest addition to a current family of three marketed sound processors for use with the Baha auditory osseointegrated implant, or with the Baha Headband/Softband setup. In its initial release, it will provide a moderate gain ear-level sound processor that will serve as a premium-feature option to the currently marketed Baha Divino. At a later date, the BP100 platform may be used to provide a range of programmable products to meet end-user needs.
Ask a Question
K Number
K080363Device Name
BAHA CORDELLE II
Manufacturer
COCHLEAR BONE ANCHORED SYSTEMS AB
Date Cleared
2008-04-10
(59 days)
Product Code
MAH, LXB
Regulation Number
874.3302Why did this record match?
Applicant Name (Manufacturer) :
COCHLEAR BONE ANCHORED SYSTEMS AB
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Baha Cordelle II sound processor is intended for use with the Baha auditory osseointegrated implant for the following patients and indications;
- . Patients who have a conductive or mixed hearing loss and can still benefit from sound amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 65 dB HL.
- Bilateral fitting of the Cordelle II is intended for patients who meet the above e criterion in both ears, with bilaterally symmetric moderate to severe conductive or mixed hearing loss. Symmetrical bone-conduction thresholds are defined as less than a 10 dB average difference between ears (measured at 0.5, 1, 2, and 3 kHz), or less than a 15 dB difference at individual frequencies.
- Patients who suffer from unilateral sensorineural deafness in one ear with normal # hearing in the other ear (i.e. single-sided deafness or "SSD"). Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL.
- . Baha for SSD is also indicated for any patient who is indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
Device Description
The Baha system works by combining a sound processor with an abutment and a small titanium implant placed in the skull behind the ear. The system is based on the process of "osseointegration" through which living tissue integrates with titanium in the implant. Thus, the titanium implant becomes one with the bone, allowing high-quality amplified and processed sound to be conducted via the skull bone directly to a cochlea with residual functionality.
The Cordelle II is one of three currently marketed sound processors for use with the Baha auditory osseointegrated implant. It is the only analog signal processing and body-worn sound processor, and it offers the highest power output of all the devices.