Search Results
Found 4 results
510(k) Data Aggregation
(202 days)
CLARITY MEDICAL SYSTEMS
- General ophthalmic imaging including retinal, corneal, and external imaging. .
- Photodocumentation of pediatric ocular diseases including retinopathy of prematurity (ROP). ♥
- Screening for Type 2 pre-threshold retinopathy of prematurity (ROP) (zone 1, stage 1 or 2, . without plus disease, or zone 2, stage 3, without plus disease) or treatment-requiring ROP, defined as Type 1 ROP (zone 1, any stage, with plus disease; zone 1, stage 3 without plus disease; or Zone 2, stage 2 or 3, with plus disease) or threshold ROP (at least 5 contiguous or 8 non-contiguous clock hours of stage 3 in zone 1 or 2, with plus disease)* in 35-37 week postmenstrual infants.
The RetCam 3 Ophthalmic Imaging System utilizes a digital camera in a handpiece with multiple field of view lenses to capture color ophthalmic digital images including retinal, corneal, and external images. An on board computer is used to store, view, retrieve, and export the digital ophthalmic images. A standard Halogen light source is used to provide illumination to the eye through the handpiece. An optional Xenon light source is also available with the RetCam 3 to facilitate imaging with fluorescein angiography. Light intensity, camera focus, and image capture are controlled by the use of a RetCam 3 footswitch and can also be controlled by a keyboard on the RetCam 3 consoles. A console monitor is provided with the RetCam 3 for viewing images. Proprietary software is installed in the computer to capture, store, view, retrieve, and export ophthalmic images.
A device modification to the RetCam 3 is being made to replace the Xenon light source module with an LED light source module. Third party testing has been performed and formal certification has been obtained verifying compliance with the applicable IEC 60601 electrical safety and electromagnetic compatibility standards. Bench testing has verified comparable safe maximum light irradiance output of the Xenon and LED light source modules. Bench testing has verified comparable fluoresce dye responses to the Xenon and LED light sources.
This document describes the RetCam 3 Ophthalmic Imaging System, and a modification to replace its Xenon light source with an LED light source. The provided text primarily focuses on clinical performance data for screening Retinopathy of Prematurity (ROP) using the RetCam system, which supports the overall indication for use.
Here's an analysis of the acceptance criteria and study data:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria in terms of numerical sensitivity and specificity targets. Instead, it presents the results of a clinical study and concludes that these results support the device's indications for use. For the purpose of this response, we'll infer the reported performance as the de-facto "met" criteria. The study aimed to measure the accuracy, reliability, and image quality of RetCam wide-field digital images to screen for ROP.
Metric | Acceptance Criteria (Implied) | Reported Device Performance (Grader A) | Reported Device Performance (Grader B) | Reported Device Performance (Grader C) |
---|---|---|---|---|
Sensitivity (for Type 2 Pre-threshold ROP or worse) | High sensitivity (to avoid missing cases) | 1.0 (26/26) (95% CI: 0.868, 1.0) | 1.0 (26/26) (95% CI: 0.868, 1.0) | 1.0 (26/26) (95% CI: 0.868, 1.0) |
Specificity (for Type 2 Pre-threshold ROP or worse) | Reasonable specificity (to limit false positives) | 0.883 (83/94) (95% CI: 0.802, 0.933) | 0.851 (80/94) (95% CI: 0.765, 0.909) | 0.851 (80/94) (95% CI: 0.765, 0.909) |
Image Quality (adequate or possibly adequate) | High percentage of adequate images | 93.3% to 100% | 93.3% to 100% | 93.3% to 100% |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Test Set): Eyes from 67 consecutive infants. The total number of eyes is not explicitly stated but implies 134 eyes if both eyes were imaged and analyzed per infant. However, the sensitivity and specificity values are based on 26 positive cases for ROP and 94 negative cases, suggesting a total of 120 cases (eyes or infants) for the ROP diagnosis evaluation. The document states "31 to 33 weeks and/or 35 to 37 weeks postmenstrual age (PMA)".
- Data Provenance: The study was a prospective trial. The geographical origin of the data is not explicitly stated, but the study was conducted by "Chiang, et. al." and published in "Archives of Ophthalmology," which suggests it's likely from a US-based medical institution, common for such publications.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Number of Experts for Device Interpretation: Three (3) expert retinal specialist graders interpreted the images. Their specific qualifications (e.g., years of experience) are not detailed beyond "expert retinal specialist graders."
- Number of Experts for Ground Truth: The reference standard (ground truth) was established by experienced pediatric ophthalmologists using binocular indirect ophthalmoscopy (BIO). The exact number of these ophthalmologists is not specified.
4. Adjudication Method for the Test Set
The document states, "Images were interpreted by three expert retinal specialist graders who provided a diagnosis and evaluation of image quality. Findings were compared with a reference standard of binocular indirect ophthalmoscopy (BIO) by experienced pediatric ophthalmologists."
This implies that each of the three graders interpreted the images independently. There is no mention of a consensus or adjudication process among the image graders for their reported performance. Their individual performances are presented. The "ground truth" was established separately by BIO.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study with AI: The study described is not an MRMC study comparing human readers with and without AI assistance. It focuses on the diagnostic accuracy of human readers interpreting images acquired by the RetCam system compared to a clinical gold standard (BIO). The device is an imaging system, not an AI diagnostic tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- No Standalone Algorithm Performance: The device being evaluated is an imaging system, not an AI algorithm. The performance reported is that of human experts interpreting images generated by the device. Therefore, no standalone algorithm performance was conducted.
7. The Type of Ground Truth Used
- The ground truth used was expert clinical diagnosis by experienced pediatric ophthalmologists using binocular indirect ophthalmoscopy (BIO). This is a recognized gold standard for ROP diagnosis.
8. The Sample Size for the Training Set
- The document does not describe a training set for an AI algorithm. The study is a prospective trial evaluating the diagnostic performance of human interpreters using the RetCam imaging system.
9. How the Ground Truth for the Training Set Was Established
- As there is no mention of an AI algorithm or a training set, the establishment of ground truth for a training set is not applicable to this document.
Ask a specific question about this device
(289 days)
CLARITY MEDICAL SYSTEMS
Each and all above listed RetCam Systems are indicated for:
- General ophthalmic imaging including retinal, corneal, and external imaging; .
- Photodocumentation of pediatric ocular diseases including retinopathy of prematurity (ROP). .
- Screening for Type 2 pre-threshold retinopathy of prematurity (ROP) (zone 1, stage 1 or 2, . without plus disease, or zone 2, stage 3, without plus disease) or treatment-requiring ROP, defined as Type 1 ROP (zone 1, any stage, with plus disease; zone 1, stage 3 without plus disease; or zone 2, stage 2 or 3, with plus disease) or threshold ROP (at least 5 contiguous or 8 non-contiguous clock hours of stage 3 in zone 1 or 2, with plus disease)* in 35-37 week postmenstrual infants.
RetCam Ophthalmic Imaging Systems utilize a digital camera in a handpiece with multiple field of view lenses to capture color ophthalmic digital images including retinal, corneal, and external images. An on board computer (RetCam II and RetCam 3) or laptop computer (RetCam Shuttle and RetCam Portable) is used to store, view, retrieve, and export the digital ophthalmic images. A standard Halogen light source is used in all RetCam configurations to provide illumination to the eye through the handpiece. An optional Fluorescein light source is also available with the RetCam II and RetCam 3 configurations. Light intensity, camera focus, and image capture are controlled by the use of a footswitch on all configurations and can also be controlled by a keyboard on the RetCam II and RetCam 3 consoles. Console monitors are provided with the RetCam II and RetCam 3 configurations for viewing images. The laptop monitor is used for viewing images with the RetCam Shuttle and RetCam Portable. Proprietary software is installed on the computers to capture, store, view, retrieve, and export ophthalmic images.
Here's a breakdown of the acceptance criteria and the study that supports the RetCam Ophthalmic Imaging System, based on the provided text:
Acceptance Criteria and Device Performance
The acceptance criteria are not explicitly stated in a quantitative manner as "acceptance criteria" but are implied by the reported performance metrics of the device in screening for ROP. The study establishes the device's accuracy by demonstrating high sensitivity and acceptable specificity for identifying specific stages of ROP.
Acceptance Criteria (Implied) | Reported Device Performance (Sensitivity and Specificity for Type 2 Threshold ROP Or Worse) |
---|---|
Sensitivity for Type 2 pre-threshold or worse ROP in 35-37 week postmenstrual infants. | |
Defined as: |
- Type 2 pre-threshold ROP (Zone 1, Stage 1 or 2, without plus disease, or Zone 2, Stage 3, without plus disease)
- Treatment-requiring ROP (Type 1 ROP: Zone 1, any stage, with plus disease; Zone 1, stage 3 without plus disease; or Zone 2, stage 2 or 3, with plus disease)
- Threshold ROP (at least 5 contiguous or 8 non-contiguous clock hours of stage 3 in zone 1 or 2, with plus disease) | Grader A: 1.0 (26/26)
Grader B: 1.0 (26/26)
Grader C: 1.0 (26/26) |
| Specificity for Type 2 pre-threshold or worse ROP in 35-37 week postmenstrual infants.
Same definition as above. | Grader A: 0.883 (83/94)
Grader B: 0.851 (80/94)
Grader C: 0.851 (80/94) |
| Image Quality: Technical quality of images captured at 35-37 weeks PMA by trained NICU nurses. | 93.3% to 100% of images found to be "adequate" or "possibly adequate" |
Study Details
The study referenced is:
Chiang, et. al. (Telemedical Retinopathy of Prematurity Diagnosis Accuracy and, Reliability, and Image Quality: Archives of Ophthalmology; 2007;125(11):1531-1538)
-
Sample size used for the test set and the data provenance:
- Test set sample size: Images from 67 consecutive infants were used. While the text refers to "eyes," the performance data is given for "26 cases of Type 2 Threshold ROP Or Worse" and "94 cases of no Type 2 Threshold ROP Or Worse," suggesting a total of 120 eyes/instances for the performance evaluation against the reference standard.
- Data provenance: Prospective trial. Country of origin not explicitly stated, but the journal (Archives of Ophthalmology) is a US-based publication, suggesting US data or an internationally recognized standard of practice.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of experts for image interpretation: Three expert retinal specialist graders interpreted the images. Their specific qualifications (e.g., years of experience) are not detailed beyond "expert retinal specialist graders."
- Number of experts for ground truth (reference standard): "Experienced pediatric ophthalmologists" performed binocular indirect ophthalmoscopy (BIO) to establish the reference standard. The number of these ophthalmologists is not specified.
-
Adjudication method for the test set:
- The text describes three individual expert retinal specialist graders (Grader A, B, C) interpreting the images, and their findings were compared to a reference standard. It does not indicate an adjudication method (like 2+1 or 3+1 consensus) among these three graders to form a single "AI" diagnosis. Instead, the performance is reported per grader. The reference standard itself was established by experienced pediatric ophthalmologists using BIO.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, this was not an MRMC comparative effectiveness study involving human readers with vs. without AI assistance. The study evaluates the standalone performance of reading RetCam images by human experts against a clinical gold standard (BIO). The RetCam system itself produces the images, which are then interpreted by human experts; it does not feature an AI component that assists human interpretation in the sense of a diagnostic algorithm providing results.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. The study describes human interpretation of images captured by the RetCam device. The RetCam is an imaging system, not an AI algorithm that provides diagnostic output. The "performance" refers to the accuracy of human graders interpreting the images acquired by the device.
-
The type of ground truth used:
- Clinical Gold Standard: Binocular Indirect Ophthalmoscopy (BIO) performed by experienced pediatric ophthalmologists. This is a direct clinical assessment by experts, considered the gold standard for ROP diagnosis.
-
The sample size for the training set:
- The text does not mention a training set. This study describes a clinical validation of an imaging device where human experts interpret the images, not an AI model that requires a training set.
-
How the ground truth for the training set was established:
- Not applicable, as no training set for an AI model is described or implied.
Ask a specific question about this device
(43 days)
CLARITY MEDICAL SYSTEMS
The RetCam Portable Ophthalmic Imaging System is indicated for general ophthalmic imaging including retinal, corneal and external imaging.
The RetCam Portable Imaging System is designed to quickly and easily capture wide field, digital images and videos of the eye. The optional hard side or soft side packaging allows for easy transport from one location to another within or outside the hospital or clinic. The reduced packaging dimensions enhance the ease of use in confined space situations.
This 510(k) summary for the RetCam Portable Ophthalmic Imaging System (K083771) indicates that its substantial equivalence to a predicate device (RetCam Shuttle Ophthalmic Imaging System, K081858) was established without specific performance criteria in terms of accuracy metrics (e.g., sensitivity, specificity, AUC). Instead, the review focused on device description, intended use, and comparison to the predicate.
Therefore, the provided text does not contain the information needed to fill out the table and answer all the questions about acceptance criteria and a study proving device performance using quantitative metrics.
However, based on the information provided, here's what can be inferred or stated as "not applicable" given the context of a 510(k) for an ophthalmic imaging system that relies on substantial equivalence rather than novel performance claims:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Quantitative Metric) | Reported Device Performance (Quantitative Result) |
---|---|
Not specified in 510(k) summary | Not specified in 510(k) summary |
Explanation: The 510(k) summary does not define specific quantitative acceptance criteria for image quality, diagnostic accuracy, or other performance metrics. The basis for clearance is substantial equivalence to a predicate device (RetCam Shuttle Ophthalmic Imaging System, K081858) for "general ophthalmic imaging including retinal, corneal, and external imaging." This typically means the device performs similarly in its core function (capturing images) as the predicate, but numerical performance targets are not commonly included in these summaries.
2. Sample size used for the test set and the data provenance
- Sample Size: Not specified.
- Data Provenance: Not specified. (The 510(k) summary focuses on regulatory approval based on equivalence rather than clinical trial data demonstrating specific performance against a test set).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of Experts: Not specified.
- Qualifications of Experts: Not specified.
4. Adjudication method for the test set
- Adjudication Method: Not applicable/Not specified. (A dedicated test set with expert adjudication for performance metrics is not mentioned or implied by this 510(k) summary.)
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No. This device is an imaging system, not an AI-powered diagnostic tool. The 510(k) summary does not mention any AI components or MRMC studies.
- Effect Size: Not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone Performance Study: No. This is an imaging device, not an algorithm, and the 510(k) summary does not describe any standalone algorithm performance testing.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Type of Ground Truth: Not applicable/Not specified. The 510(k) summary doesn't describe a study requiring ground truth for performance evaluation; rather, it establishes equivalence based on device specifications and intended use.
8. The sample size for the training set
- Sample Size: Not applicable/Not specified. The device is an ophthalmic imaging system, not a machine learning algorithm that requires a training set in this context.
9. How the ground truth for the training set was established
- Ground Truth for Training Set: Not applicable/Not specified. (See point 8).
Conclusion based on the provided document:
The provided 510(k) summary for the RetCam Portable Ophthalmic Imaging System (K083771) details the device's regulatory classification, indications for use, and its substantial equivalence to a predicate device (RetCam Shuttle Ophthalmic Imaging System, K081858). It does not contain information regarding specific performance acceptance criteria (e.g., sensitivity, specificity, image quality metrics) or a study that quantitatively evaluates such performance with a test set, ground truth, or expert readers. The clearance is based on the device being a comparable imaging tool to an already legally marketed device.
Ask a specific question about this device
(127 days)
CLARITY MEDICAL SYSTEMS
Ask a specific question about this device
Page 1 of 1