Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K140592
    Date Cleared
    2014-12-17

    (285 days)

    Product Code
    Regulation Number
    890.5880
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Ceragem International Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Ceragem® Thermal Massager Model CGM-MB-1101 Automatic Thermal Massager is to provide muscle relaxation therapy by delivering heat and soothing massage. Additionally, the product provides topical radiant infrared heat for:

    • Temporary relief of minor muscle and joint pain stiffness
    • Temporary relief of minor joint pain associated with arthritis
    • Temporary increase in local circulation where applied
    • Relaxation of muscles
    Device Description

    This device is comprised of the main body and optional components. The main body components are the main mat which includes the internal projector and the up/down movement motor, and the lower mat. The optional components are the external projectors (3-sphere). This device is an electrically powered, motorized, multi-functional physical therapy table intended to use for temporary relief of muscle and joint pain and stiffness by applying massage pressure from the vertical and horizontal movements, and heat from the internal and external projectors.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Ceragem Automatic Thermal Massager, Model CGM-MB-1101, based on the provided FDA 510(k) summary:

    This device does not involve AI/ML technology, so many of the requested fields related to AI performance, ground truth establishment, and expert involvement are not applicable and will be marked as such. The study described is primarily a comparison to a predicate device and adherence to industry standards, rather than clinical performance trials with specific disease outcomes.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are mainly demonstrated through substantial equivalence to a predicate device and adherence to recognized safety and performance standards. The "performance" in this context refers to the device's functional specifications being within acceptable limits and similar to the predicate device.

    Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance (Model CGM-MB-1101)
    Intended UseSame as predicate: Muscle relaxation therapy by delivering heat and soothing massage.Same intended use.
    Indications for UseSame as predicate: Temporary relief of minor muscle/joint pain/stiffness, arthritis pain, increased local circulation, muscle relaxation.Same indications for use.
    Technological CharacteristicsSimilar emission source (Jade Projectors), infrared emission spectrum (5-20 microns).Jade Projectors, 5-20 micron infrared emission spectrum.
    Safety - Electrical StandardsUL 60601-1, CAN/CSA-C22.2 No. 601.1-M90, IEC 60601-1:2005, EN 60601-1:2006.Tested to and meets these standards.
    Safety - EMC StandardsIEC 60601-1-2: 2007 (Electromagnetic Compatibility).Tested to and meets this standard.
    Safety - Thermal RegulationOperating Temperature Range similar to predicate, with maximal heat not exceeding predicate.Internal: 30°C – 65°C; External, main, Auxiliary: 30°C - 60°C. Predicate: 30°C – 60°C. Max temperature does not exceed predicate.
    Safety - BiocompatibilityISO 10993-10 (Tests for Irritation and Skin Sensitization).Tested to and meets this standard.
    Safety - Risk ManagementISO 14971 (Application of Risk Management).Tested to and meets this standard.
    Software Life Cycle (if applicable)IEC 62304 (Medical device Software Software Life Cycle Processes).Tested to and meets this standard.
    Performance - Mechanical ComponentsGeared DC Motor, Limit Switch, Wire-Chain, Ergonomically Molded Polyethylene Rails.Geared DC Motor, Limit Switch, Wire-Chain, Ergonomically Molded Polyethylene Rails. (Similar to predicate)
    Performance - Weight CapacityStatic Weight: 300 pounds (predicate).298 pounds (subject device declared substantially equivalent).
    Maximum Permissible Weight: 300 pounds (predicate).298 pounds (subject device declared substantially equivalent).

    Study Details

    1. Table of Acceptance Criteria and Reported Device Performance

    (See table above)

    2. Sample size used for the test set and the data provenance

    • Sample Size (Test Set): Not applicable in the context of clinical performance data for efficacy. The "test set" here refers to the device itself being tested for compliance with technical standards and compared against its predicate's specifications. No patient outcome data or imaging data is presented to require a test set sample size.
    • Data Provenance: The data primarily originates from the manufacturer's engineering specifications, technical testing reports, and a comparison against the previously cleared predicate device (K040031) by the same manufacturer. It is a retrospective analysis of the device's design and technical specifications relative to a predicate, combined with physical testing for safety standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This is not an AI/ML device that generates diagnoses or interpretations. "Ground truth" in this context would refer to the established engineering standards and the specifications of the predicate device, which are objectively measurable and defined by regulatory bodies and the manufacturer. Experts involved would be engineers and regulatory specialists, but their role is not to establish a "ground truth" for a diagnostic output.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This pertains to diagnostic outcomes and expert consensus, which does not apply to this type of device submission. The "adjudication" is essentially the FDA's review of the manufacturer's claims of substantial equivalence and compliance with standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a non-AI therapeutic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a non-AI therapeutic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" implicitly used is the established safety and performance standards for similar electromechanical thermal massage devices (e.g., UL, IEC, ISO standards) and the specifications and regulatory clearance of the predicate device (K040031). There is no diagnostic "ground truth" involved.

    8. The sample size for the training set

    • Not applicable. This is a non-AI device; there is no training set.

    9. How the ground truth for the training set was established

    • Not applicable. This is a non-AI device; there is no training set or associated ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K111128
    Device Name
    LABONACHECK A1C
    Date Cleared
    2012-09-13

    (511 days)

    Product Code
    Regulation Number
    864.7470
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    CERAGEM INTERNATIONAL, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LabonaCheck™ A1c is intended for the quantitative measurement of glycated hemoglobin in venous whole blood and capillary fingerstick samples. This device is intended for multiple patient, professional use. Measurement of percent glycated hemoglobin (HbA 1c) is effective in monitoring long-term glucose control in individuals with diabetes mellitus by using the LabonaCheck™ A1c. Only auto-disabling, single use lancing devices should be used with this system.

    Device Description

    The system consists of HbA1c Analyzer (MH 200) and HbA1c Test kit. The LabonaCheck™ AIc HbAic Analyzer measures the coloured responses of the LabonaCheck™ A1c HbA1c Test Kit by spectral reflectance.

    AI/ML Overview

    The LabonaCheck™ A1c device is intended for the quantitative measurement of glycated hemoglobin (HbA1c) in venous whole blood and capillary fingerstick samples for professional use. The provided document details several performance characteristic studies.

    Here's an analysis of the acceptance criteria and study data:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal acceptance criteria in a dedicated section with pass/fail thresholds for all studies. However, the manufacturer's conclusions and noted percentage differences are used to infer an implied acceptance criterion where applicable.

    Performance CharacteristicImplied Acceptance CriteriaReported Device Performance
    Precision/ReproducibilityNot explicitly stated, but typical expectations for HbA1c CV% are generally below 3-5% for clinical utility.Internal Study (Within-run, Day-to-day, and Total Precision):
    • Reference A1c 4.3% (Mean 4.4%): Total Precision CV 3.0%
    • Reference A1c 6.2% (Mean 6.1%): Total Precision CV 2.6%
    • Reference A1c 8.9% (Mean 9.1%): Total Precision CV 1.9%
    • Reference A1c 11.7% (Mean 11.6%): Total Precision CV 2.2%
    • Reference A1c 14.3% (Mean 14.2%): Total Precision CV 1.6%
      External Study (Point-of-Care Sites):
    • Interval 1 (approx 4.8% HbA1c): Overall CV% between 2.9% - 3.5% (across operators and sites)
    • Interval 2 (approx 8.8% HbA1c): Overall CV% between 2.7% - 3.6% (across operators and sites)
    • Interval 3 (approx 13.2% HbA1c): Overall CV% between 2.7% - 3.5% (across operators and sites) |
      | Linearity/Assay Measuring Range | The study aims to support linearity from 4% to 15%. % Recovery between real and expected values should be close to 100%. (Implied successful if sponsor claims support for the range). | Supported linearity from 4% to 15%.
    • Expected 3.2% (Real 3.5%): 108% Recovery
    • Expected 5.6% (Real 5.8%): 103% Recovery
    • Expected 8.0% (Real 8.1%): 101% Recovery
    • Expected 10.4% (Real 10.5%): 101% Recovery
    • Expected 12.7% (Real 12.9%): 102% Recovery
    • Expected 15.1% (Real 15.4%): 102% Recovery
    • Expected 17.5% (Real 17.5%): 100% Recovery |
      | Detection Limit | A reportable range of 4% to 15% A1c. | LoB Mean: 3.10%, LoD Mean: 4.12%. This suggests the lower end of the reportable range (4%) is supported. |
      | Analytical Specificity (Interference) | Recovery within ±10% of control results for interfering substances. (For Hemoglobin F: average bias within ±10% up to 22.7% HbF. For variants C, D, E, S, Carbamylated Hb, and Rheumatoid factor: within ±10% of reference difference). For Hemoglobin concentration: within 10% difference. | Known Substances: All tested substances (Ascorbic acid, Bilirubin, Glucose, Hemoglobin (up to 20g/dL), Lipid, Albumin, K3EDTA, Heparin, Sodium fluoride, Sodium citrate, Acetaminophen, Metformin, Acetylsalicylic acid, Glybenclamide, Ibuprofen) showed no significant interference (implied within ±10%).
      Hemoglobin Variants:
    • HbF: Average bias within ±10% up to 22.7%. The device is affected by HbF > 20%.
    • HbC, D, E, S: Satisfied acceptance criteria (within ±10% of reference difference).
      Other interfering factors:
    • Carbamylated Hemoglobin & Rheumatoid Factor: Satisfied acceptance criteria (within ±10% of reference difference).
      Hemoglobin Concentration: Satisfied acceptance criteria (within 10% difference) for concentrations ranging from 8.7g/dL to 21.0g/dL. The claimed functional range is 10.0g/dL - 20.0g/dL. |
      | Incubation Time Study | Reference difference within 5% when R1 reagent and whole blood sample incubated for 2 to 3 minutes. (Implied acceptance met if results within 5%). | When R1 reagent and whole blood sample incubated for 2 to 3 minutes, reference difference was within 5%. (Acceptance criteria satisfied). |
      | Method Comparison (Accuracy) w/ Predicate Device | Not explicitly stated but typical metrics for point-of-care devices include a certain percentage of results within ±X% or correlation coefficient/slope/intercept criteria. For this submission, comparison to predicate and reference devices is key. For the difference % to predicate, 90% or more within 10% is often an acceptance. | LabonaCheck™A1c vs. Tosoh G7: 100% (50/50) within ±10% difference. Slope 1.0552, Intercept -0.4360, Correlation 0.9708.
      NycoCard vs. Tosoh G7: 96% (48/50) within ±10% difference. Slope 1.0613, Intercept -0.3304, Correlation 0.9695.
      LabonaCheck™A1c vs. NycoCard: 96% (48/50) within ±10% difference. Slope 0.9873, Intercept -0.0516, Correlation 0.9873. |
      | Point-of-Care Study (Accuracy) | Slope (0.9-1.1), Intercept (less than 1%), Correlation coefficient results of three POC sites satisfied the acceptance criteria. | Venous Blood Samples (All sites combined): Slope 1.0005 (95% CI 0.97961.0215), Intercept -0.0292 (95% CI -0.2036-0.1451), Correlation 0.9870. (All within acceptance criteria).
      Capillary Blood Samples (All sites combined): Slope 0.9692 (95% CI 0.94810.9903), Intercept 0.4230 (95% CI 0.25230.5937), Correlation 0.9859. (All within acceptance criteria). |
      | Matrix Comparison | Not explicitly stated, but the sponsor concluded that the stated anticoagulants can be used, implying satisfactory performance based on the linear regression analysis. | K3EDTA vs. Sodium Heparin (combined results): Slope 0.9682 (95% CI 0.92791.0085), Intercept 0.3190 (95% CI -0.05700.6951), Correlation 0.9670.
      K3EDTA vs. NaF (combined results): Slope 0.9487 (95% CI 0.90330.9941), Intercept 0.4364 (95% CI 0.01270.8602), Correlation 0.9569. (Concluded that these anticoagulants can be used). |

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision Studies (Internal): The internal study details whole blood samples, tested per day using two meters and two lots of reagents. The table shows 5 reference A1c levels. The specific number of distinct patient samples is not explicitly stated, but rather "whole blood samples" were used, with replicates for statistical analysis, implying a controlled, prospective study design. Provenance is internal (CERAGEM Medisys Inc.).
    • Precision Studies (External): 3 intervals of HbA1c EDTA blood were analyzed. The specific number of distinct patient samples is not stated directly, but it specifies "three intervals of HbA1c EDTA blood." This was a multi-site prospective study conducted at three Point-of-Care (POC) sites. Provenance is external (three POC sites).
    • Linearity/Assay Measuring Range: 7 mixed control solutions were used. This is likely a prospective, controlled study using prepared samples. The provenance is internal to the manufacturer.
    • Detection Limit: A zero sample (blank) and five low HbA1c samples (n=60, Ns=60 for each, assayed twice a day in replicates of ten for three days). This is prospective, controlled study. Provenance is internal to the manufacturer.
    • Analytical Specificity (Interference):
      • Common Substances: Whole blood sample pools (3 intervals) were spiked with interfering substances. This is a controlled, prospective study. Provenance is internal to the manufacturer.
      • Hemoglobin Variants (F, C, D, E, S): Samples provided by the National Glycohemoglobin Standardization Program (NGSP). This implies retrospective use of well-characterized samples from the NGSP program. Provenance: NGSP (likely international, given their role in standardization).
      • Carbamylated Hemoglobin & Rheumatoid Factor: 10 samples (whole blood and spiked samples). This is a controlled, prospective study. Provenance is internal to the manufacturer.
      • Hemoglobin Study: 50 whole blood samples. This is a prospective study using patient samples. Provenance is internal to the manufacturer.
    • Incubation Time Study: Whole blood samples of 3 HbA1c concentrations (Low, Medium, High). This is a controlled, prospective study. Provenance is internal to the manufacturer.
    • Method Comparison w/ Predicate Device: 50 EDTA venous whole blood patient samples (25 at each of two POC sites). Samples spanned 4.2-14.5% A1c. This is a prospective study using patient samples. Provenance is two Point-of-Care (POC) sites.
    • Point-of-Care Study (Accuracy):
      • Venous Blood: 120 EDTA venous whole blood samples (40 at each of three POC sites). Samples spanned 4.5-14.6% A1c. This is a prospective study using patient samples. Provenance is three Point-of-Care (POC) sites.
      • Capillary Blood: 120 capillary blood samples (40 at each of three POC sites). Samples spanned 4.2-14.8% A1c. This is a prospective study using patient samples. Provenance is three Point-of-Care (POC) sites.
    • Matrix Comparison: Venous whole blood samples (K3EDTA, Sodium heparin, NaF) collected from 40 donors. Samples spanned 4.5-14.1% A1c. This is a prospective study using patient samples. Provenance is internal to the manufacturer.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The studies primarily involve quantitative measurements for HbA1c.

    • For accuracy studies (Method Comparison, POC Study, Matrix Comparison), the ground truth was established by a reference method, Tosoh G7, and the predicate device, NycoCard® HbA1c, not by human experts interpreting results. The Tosoh G7 is likely an NGSP-certified laboratory method, considered a gold standard for HbA1c. The NycoCard® HbA1c is a legally marketed predicate device.
    • For Precision studies, there isn't a "ground truth" in the sense of expert interpretation; rather, the "reference A1c (%)" values are used as targets for reproducibility.
    • While operators (6 intended users across 3 sites for external precision and POC studies) performed the tests, they were not establishing ground truth, but rather generating test results to be compared against established reference values. The specific qualifications of these operators are not detailed beyond "intended users."

    4. Adjudication Method for the Test Set

    Not applicable. This device provides a quantitative measurement of a biomarker (HbA1c). The "ground truth" is established by laboratory reference methods (Tosoh G7, NGSP certified methods) or comparison to a predicate device, not by expert consensus or adjudication of qualitative interpretations.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not explicitly done. The studies mentioned involve multiple operators or sites (e.g., 6 operators at 3 POC sites), but this is primarily to assess reproducibility and accuracy in a real-world setting, not to measure the human diagnostic improvement with vs. without AI assistance. The device is a standalone quantitative analyzer, not an AI-assisted diagnostic aid for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, the core of the submission provides standalone performance of the LabonaCheck™ A1c device. The studies evaluate the device's accuracy, precision, linearity, detection limits, and interference characteristics directly, comparing its results against reference methods or a predicate device. Human involvement is as operators performing the test, not as interpreters whose performance is being augmented by an algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for these studies is primarily reference method data and, in some cases, predicate device data.

    • For accuracy and method comparison studies, the Tosoh G7 (a laboratory reference method, likely NGSP-certified) served as the primary reference for establishing "true" HbA1c values.
    • For interference studies of hemoglobin variants, NGSP-provided samples were used, with comparisons made against NGSP data (results of Primus - the predicate manufacturer).
    • For precision and linearity, controlled samples with known or targeted HbA1c concentrations were used.

    8. The Sample Size for the Training Set

    This submission describes a medical device (HbA1c analyzer and test kit) that performs a chemical assay for quantitative measurement. It is not an AI/ML algorithm that requires a "training set" in the conventional sense of machine learning. Therefore, a training set sample size is not applicable to this type of device.

    9. How the Ground Truth for the Training Set Was Established

    As stated above, this is not applicable as the device is not an AI/ML algorithm requiring a training set. The device operates based on a boronate affinity assay, a chemical method.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062476
    Date Cleared
    2006-10-31

    (68 days)

    Product Code
    Regulation Number
    890.5880
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    CERAGEM INTERNATIONAL, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the CERAGEM – RH1 Automatic Thermal Massager is to provide the user with muscle relaxation therapy by delivering heat and soothing massage. Additionally, the product provides topical radiant infrared heat for:

    • ◊ temporary relief of minor muscle and joint pain, and stiffness
    • ُ the temporary relief of minor joint pain associated with arthritis
    • the temporary increase in local circulation where applied
    • 0 relaxation of muscles
    Device Description

    The CERAGEM - RH1 Automatic Thermal Massager is an electrically powered motorized multi-functional physical therapy table. Its intended use is to provide muscle relaxation therapy by delivering heat and soothing massage. The massage function is delivered by massage rollers mounted on an independent carriage underneath a pad on the table's torso section. The heat function is delivered by two components: 1) heated jade massage rollers mounted together on the moving carriage and 2) heated Epoxy Carbon Panels. In both cases, radiant infrared heat is emitted. Together, the massage action and heated pressure points apply light pressure as well as heat to the user.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study proving the device meets acceptance criteria. The document is a 510(k) premarket notification summary for the CERAGEM - RH1 Automatic Thermal Massager, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting performance data against specific acceptance criteria.

    Therefore, I cannot fulfill your request for the tables and details about a study to prove acceptance criteria, as that information is not present in the given text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1