Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K251245
    Date Cleared
    2025-08-11

    (111 days)

    Product Code
    Regulation Number
    868.1720
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Bio-Med Device, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OxyMinder Pro is an oxygen monitor with integrated pressure monitoring intended for continuous monitoring of the concentration of oxygen and pressure being delivered to patients ranging from newborns to adults. This device can be used in the hospital and subacute settings. The monitor is not intended as a life supporting device or for diagnostics.

    Device Description

    The OxyMinder Pro is an oxygen and pressure monitor capable of measuring the oxygen concentration from 18 to 100% (cleared under K213948) and pressure from 0 to 60 cmH2O (subject of this submission) that for convenience can be mechanically mounted on to the cleared blender.

    The pressure is measured via a disposable pressure tubing that connects from the monitor to an adapter placed in the patient circuit. This sampling line is identical to that cleared in predicate Maxtec K221734. We have only updated the labeling to reflect the name of the sponsor.

    As indicated the oxygen monitoring portion has been previously cleared, K213948. It utilizes a cleared oxygen sensor which outputs a voltage to determine the concentration of oxygen. The OxyMinder Pro calibrates at ambient air (21%) and 100% oxygen. The OxyMinder Pro is software controlled. Again, the oxygen monitoring feature and functions are unchanged and previously cleared under reference K213948.

    The new pressure monitoring feature utilizes a pressure sensor which measures the pressure within a patient circuit. There is a disposable pressure tubing that connects between the patient circuit and the pressure sensor.

    AI/ML Overview

    This document, K251245, describes the 510(k) clearance for the OxyMinder Pro, an oxygen and pressure monitor. It uses the Maxtec MaxO2ME+p (K221734) as its predicate device for the combined oxygen and pressure monitoring features, and the Bio-Med Devices OxyMinder (K213948) as a reference for the oxygen monitoring aspects, which were previously cleared.

    The core of the submission focuses on the new pressure monitoring feature, as the oxygen monitoring component is largely based on a previously cleared device (OxyMinder, K213948). The acceptance criteria and testing detailed largely pertain to the performance of the device's measurement capabilities and adherence to relevant safety and performance standards.


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes the performance specifications of the OxyMinder Pro and compares them to the predicate and reference devices. These specifications serve as the acceptance criteria for the device's functionality. The "Reported Device Performance" is implied by the similarity claims and the statement that "The test results met the applicable standards".

    Acceptance Criteria (Performance Specification)OxyMinder Pro Reported PerformancePredicate (Maxtec MaxO2ME+p K221734)Reference (Bio-Med Devices OxyMinder K213948)
    Oxygen Measurement Range18% – 100% O20.0 to 100% O218% – 100% O2
    Pressure Measurement Range0 to 60 cmH2O-15.0 to 60.0 cmH2ON/A (Oxygen only)
    Oxygen ResolutionO2 % displayed to nearest whole integer0.1% O2O2 % displayed to nearest whole integer
    Pressure Resolution0.1 cmH2O0.5 cmH2ON/A
    Oxygen Accuracy and Linearity±1% of full scale+1% of full scale±1% of full scale
    Pressure Accuracy+0.5 cmH2O+1.0 cmH2ON/A
    Total Oxygen Accuracy±2.5% Actual oxygen level over full operating temperature range+3% Actual oxygen level over full operating temperature range±2.5% Actual oxygen level over full operating temperature range
    Oxygen Response Time (90% final value)approx. 6 secondsapprox. 15 seconds at 23oCapprox. 6 seconds
    Warm-up TimeNone requiredNone requiredNone required
    Operating Temperature0° - 50° C [32° - 122° F]15oC – 40oC (59oF – 104oF)0° - 50° C [32° - 122° F]
    Storage Temperature0° - 40° C [32° - 104° F]-15oC – 50oC (5oF – 122oF)0° - 40° C [32° - 104° F]
    Atmospheric Pressure700 – 1010 mBars800 – 1012 mBars700 – 1010 mBars
    Humidity5 - 95%0-95% (non-condensing)5 - 95%
    Battery Life16 hours at 100% brightnessApprox. 5000 hours, typical use16 hours at 100% brightness
    Low Oxygen Alarm Range18% - 100% (>1% lower than high alarm)15% - 99% (>1% lower than high alarm)18% - 100% (>1% lower than high alarm)
    Low Pressure Alarm RangeOff – 55 cmH2O (> 1cmH2O lower than high pressure alarm)Off, 1-30 cmH2ON/A
    High Oxygen Alarm Range19% - 105% (>1% lower than low alarm)16% - 100% (>1% higher than low alarm)N/A
    High Pressure Alarm Range5 – 60 cmH2O (>1 cmH2O higher than low pressure alarm)1-60 cmH2O, OffN/A
    Alarm AccuracyExact to display alarm valueExact to display alarm valueExact to display alarm value
    Pressure Alarm Resolution1 cmH2O1 cmH2ON/A

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "Bench testing was performed" and lists various types of tests (Shelf-life / Aging, Software Verification and Validation, Safety and ElectroMagnetic Compatibility, etc.). It claims that "The test results met the applicable standards". However, the sample size for the test set is not explicitly stated.

    The data provenance is not explicitly mentioned as country of origin, nor is it specified if the testing was retrospective or prospective. Given it's bench testing for a device clearance, it is implicitly prospective testing within a laboratory/manufacturing environment.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This type of information (number of experts, their qualifications, and their role in establishing ground truth) is not applicable or not provided for this specific 510(k) submission. Medical device performance testing, especially for devices like oxygen and pressure monitors, relies on calibrated instruments and established physical and electrical standards to determine "ground truth" (e.g., a calibrated gas mixture for oxygen concentration, a pressure calibrator for pressure measurements), not human expert consensus.

    4. Adjudication Method for the Test Set

    The concept of an adjudication method (like 2+1 or 3+1 used in clinical trials or image interpretation studies) is not applicable here. The "ground truth" for the device's performance (e.g., accuracy of oxygen or pressure readings) is established through comparison to validated and calibrated measurement standards, not through human interpretation that would require adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC study is not applicable and was not done for the OxyMinder Pro. MRMC studies are typically used to assess human reader performance, often in diagnostic imaging, with and without AI assistance. The OxyMinder Pro is a direct measurement device; its performance is based on its physical and electrical accuracy, not on human interpretation or an AI algorithm assisting human interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The OxyMinder Pro itself is a standalone device. Its performance, as described by its measurement accuracy, resolution, response time, etc., is its standalone performance without a human-in-the-loop influencing the measurement itself. The device is intended to present data to a human for monitoring, but its core function (measurement) is standalone. The testing conducted, as per the "Non-Clinical Testing Summary," assesses the device's inherent performance.

    7. Type of Ground Truth Used

    The ground truth for the performance testing is established using calibrated instruments and reference standards. For oxygen measurements, this would involve calibrated gas mixtures with known oxygen concentrations. For pressure measurements, this would involve calibrated pressure sources or transducers with known pressure values. It is based on physical and engineering measurements against established standards, not expert consensus, pathology, or outcomes data.

    8. Sample Size for the Training Set

    The OxyMinder Pro uses an oxygen sensor (galvanic cell) and a pressure transducer. While it is "software controlled," it does not appear to employ machine learning or AI algorithms that would require a "training set" in the conventional sense (i.e., a dataset used to train a model). The software is likely for control, data processing, display, and alarm functions, using fixed algorithms based on physical principles, not learning from data. Therefore, a training set sample size is not applicable.

    9. How the Ground Truth for the Training Set Was Established

    As there is no indication of a training set in the context of machine learning, this question is not applicable. The device's operation is based on pre-programmed algorithms and calibrated sensor outputs.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1