Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K230987
    Device Name
    FlowStar Analog Mixer Flowmeter
    Manufacturer
    Baldus Sedation GmbH & Co. KG
    Date Cleared
    2023-09-14

    (161 days)

    Product Code
    BZR
    Regulation Number
    868.5330
    Why did this record match?
    Applicant Name (Manufacturer) :

    **Baldus Sedation GmbH **& Co

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    Indicated for administering an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen to a conscious, spontaneously breathing patient
    Device Description
    This is a device which can precisely dose a mixture of medical oxygen and medical nitrous oxide, depending on the adjusted mixture, for the sedation of patients in hospitals and at the dentist. The operator, a dentist is able to adjust the total flow and the concentration of nitrous oxide of this flow. Depending on these settings the level of sedation of the patient can be controlled by the operator. The FlowStar ® Analog Mixer device has been created with the following safety features. The total flow emitted can consist of a maximum of 70 % nitrous oxide, the mix up of gas types is prevented by mechanical encodings and the flow of pure nitrous oxide in the absence of a missing oxygen supply is prevented by a fail-safe device. Two rotary knobs are used to select the desired total flow and N20/O2 gas mixture. The total flow can consist of between 0 (100% O₂) and 70 (30% O₂) percent N₂O. The minimum output of 30 % O2 prevents from a hypoxic mixture for the patient. The FlowStar® Analog Mixer device has an integrated O2 Flush knob which delivers at least over 40 l/min of pure oxygen if the user presses the knob to flush the N20 out of the patient circuit after the treatment or in the situation the patient is over sedated. The FlowStar Analog Mixer device has an alarm system which generates an alarm in the case of an insufficient O2 supply. The device will shut off the sedation when the O2 supply reaches a critical level. The FlowStar ® Analog Mixer has an integrated Bag T. The mixed gas to it. The Bag T is a device with one additional input and two outputs. The additional input is a valve which opens when the adjusted total flow is too low for the patient. The gas mixture is then enriched with additional ambient air. The first output of the Bag T is for the connected breathing bag which is filled by the mixing device when the patient exhales and is emptied when the patient inhales. The second output leads the gas mixture to the connected scavenger system and finally to the patient
    Ask a Question
    K Number
    K222794
    Device Name
    FlowStar Touch Digital Mixer Flowmeter
    Manufacturer
    Baldus Sedation GmbH & Co. KG
    Date Cleared
    2023-07-28

    (315 days)

    Product Code
    BZR
    Regulation Number
    868.5330
    Why did this record match?
    Applicant Name (Manufacturer) :

    **Baldus Sedation GmbH **& Co

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    Indicated for administering an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen to a conscious, spontaneously breathing patient.
    Device Description
    The device will be available in two models: (Details below) FLOWSTAR TOUCH DIGITAL MIXER FLOWMETER 50 %, and FLOWSTAR TOUCH DIGITAL MIXER FLOWMETER 70 %. 50% and 70% represent the maximum nitrous oxide concentration. The FlowStar Touch O2- N₂O Mixer for Analgesia is a device which can precisely dose a mixture of medical oxygen and medical nitrous oxide, depending on the adjusted mixture, for the sedation of patients in hospitals and at the dentist. The operator, a doctor or a dentist is able to adjust the total flow and the concentration of nitrous oxide of this flow. Depending on these settings the level of sedation of the patient can be controlled by the operator. The Baldus® Touch O2- N2O Mixer for Analgesia device has been created with the following safety features. The total flow emitted can consist of a maximum of 70 % nitrous oxide, the mix up of gas types is prevented by mechanical encodings and the flow of pure nitrous oxide in the absence of a missing oxygen supply is prevented by a fail-safe device. An integrated Touch pad is used to select the desired N₂O /O2 gas mixture. The total flow can consist of between 0 (100% O2) and 70 (30% O2) percent N₂O. The minimum output of 30 % O2 prevents from a hypoxic mixture for the patient. The FlowStar TouchO2- N2O Mixer for Analgesia device has an integrated O2 Flush function which delivers at least 30 l/min of pure oxygen to flush the N20 out of the patient circuit after the treatment. The FlowStar TouchO2- N₂O Mixer for Analgesia device has an alarm system which generates an alarm in the case of an insufficient O2 supply. The device will shut off the sedation when the O2 supply reaches a critical level. The gas mixture output of the FlowStar TouchO2- N2O Mixer for Analgesia device is connected to the Bag T. The Bag T is a device with two inputs and two outputs. One input is the gas mixture from the mixing device. The other input is a valve which opens when the adjusted total flow is too low for the patient. The gas mixture is then enriched with additional ambient air. The first output of the Bag T is for the connected breathing bag which is filled when the patient exhales and is emptied when the patient inhales. The second output leads the gas mixture to the scavenger system and finally to the patient. The main accessories for the proposed device are a Breathing Bag and a Double Mask Scavenger System. The device includes a microprocessor-controlled valve system that precisely meters Oxygen and Nitrous Oxide gases for analgesic purposes. The inputs of the touchpad influence the gas flow inside of the device. The firmware is considered to be of a moderate level of concern.
    Ask a Question

    Page 1 of 1