510(k) Data Aggregation

    K Number
    K243343
    Device Name
    BD CTGCTV2
    Date Cleared
    2025-04-22

    (179 days)

    Product Code
    Regulation Number
    866.3393
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    BD Integrated Diagnostic Solutions/Becton,

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD CTGCTV2 assay incorporates automated DNA extraction and real-time polymerase chain reaction (PCR) for the direct, qualitative detection of DNA from:

    • Chlamydia trachomatis (CT)
    • Neisseria gonorrhoeae (GC)
    • Trichomonas vaginalis (TV)

    The assay may be used for detection of CT, GC and/or TV DNA in patient- or clinician-collected vaginal swab specimens (in a clinical setting) and male and female urine specimens. The assay may also be used for the detection of CT and GC DNA in endocervical swab and Liquid-Based Cytology (LBC) specimens in ThinPrep® PreservCyt® Solution using an aliquot that is removed prior to processing for the ThinPrep® Pap test. The assay may also be used for the detection of CT and GC DNA in clinician-collected rectal and oropharyngeal swab specimens.

    The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial, gonococcal, and/or trichomoniasis urogenital disease and chlamydial and gonococcal extragenital infection.

    The BD CTGCTV2 assay is available for use with the BD MAX™ System (urogenital specimens) or the BD COR™ System (urogenital and extragenital specimens), as described above.

    Device Description

    The BD CTGCTV2 assay, performed on the BD COR™ System (hereafter referred to as BD CTGCTV2), is designed for use with the applicable BD Molecular specimen collection and transport devices for male and female urine, rectal swabs, oropharyngeal swabs, vaginal swabs, endocervical swabs, and LBC specimens (PreservCyt®). Specimens are collected and transported to the testing laboratory using their respective transport devices under conditions of time and temperature that have been determined to maintain the integrity of the target nucleic acids.

    The BD COR™ MX Instrument, when combined with the BD COR™ PX Instrument, is to be used for automated sample preparation, extraction, and purification of nucleic acids from multiple specimen types, as well as the automated amplification and detection of target nucleic acid sequences by fluorescence-based real-time PCR for simultaneous and differential detection of Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis.

    The BD CTGCTV2 assay extraction reagents are dried in 96-well microtiter plates that contain binding magnetic affinity beads and Sample Processing Control (SPC). Each tube is capable of binding and eluting sample nucleic acids. The SPC monitors the integrity of the reagents and the process steps involved in DNA extraction, amplification and detection, as well as for the presence of potential assay inhibitors.

    The BD CTGCTV2 assay liquid reagent plate includes Wash, Elution and Neutralization buffers. The beads (described above), together with the bound nucleic acids, are washed and the nucleic acids are eluted by a combination of heat and pH. When performed on BD COR™ System, there is an additional buffer to rehydrate the dried extraction mix. Eluted DNA is neutralized and transferred to the Amplification reagent (described below) to rehydrate the PCR reagents. After reconstitution, the BD COR™ System dispenses a fixed volume of PCR-ready solution containing extracted nucleic acids into the BD PCR Cartridge. Microvalves in the BD PCR Cartridge are sealed by the system prior to initiating PCR in order to contain the amplification mixture and thus prevent evaporation and contamination.

    The BD CTGCTV2 assay is comprised of two targets for Chlamydia trachomatis (detected on the same optical channel), two targets for Neisseria gonorrhoeae (detected on two different optical channels) and one target for Trichomonas vaginalis (detected on one optical channel). Only one Chlamydia trachomatis target is required to be positive in order to report a positive result. Both Neisseria gonorrhoeae targets are required to be positive in order to report a positive result.

    The amplified DNA targets are detected using hydrolysis (TaqMan®) probes, labeled at one end with a fluorescent reporter dye (fluorophore), and at the other end, with a quencher moiety. Probes labeled with different fluorophores are used to detect the target analytes in different optical channels of the BD COR™ System. When the probes are in their native state, the fluorescence of the fluorophore is quenched due to its proximity to the quencher. However, in the presence of target DNA, the probes hybridize to their complementary sequences and are hydrolyzed by the 5'-3' exonuclease activity of the DNA polymerase as it synthesizes the nascent strand along the DNA template. As a result, the fluorophores are separated from the quencher molecules and fluorescence is emitted. The BD COR™ System monitors these signals at each cycle of the PCR and interprets the data at the end of the reaction to provide qualitative test results for each analyte (i.e., positive or negative).

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the BD CTGCTV2 assay detail its performance in detecting Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) in extragenital specimens (rectal and oropharyngeal swabs).

    Here's an analysis based on your request:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the BD CTGCTV2 assay are implicitly demonstrated through its clinical performance results, where the assay's sensitivity (Positive Percent Agreement - PPA) and specificity (Negative Percent Agreement - NPA) for extragenital specimens are compared against a Composite Comparator Algorithm (CCA). The FDA's clearance indicates that these performance metrics met the necessary standards for substantial equivalence.

    Table of Acceptance Criteria and Reported Device Performance:

    While explicit numerical acceptance criteria (e.g., "PPA must be >= X%") are not directly stated in the provided text, the reported performance measures are the ones that met the FDA's requirements for clearance.

    MetricTarget/Condition (Implicit Acceptance Criteria)Reported Device Performance (BD CTGCTV2)
    Chlamydia trachomatis (CT) - Oropharyngeal
    Sensitivity (PPA)Sufficiently high for diagnostic use100% (86.2–100% CI)
    Specificity (NPA)Sufficiently high for diagnostic use99.8% (99.5–99.9% CI)
    Neisseria gonorrhoeae (GC) - Oropharyngeal
    Sensitivity (PPA)Sufficiently high for diagnostic use92.8% (85.8–96.5% CI)
    Specificity (NPA)Sufficiently high for diagnostic use99.5% (99.1–99.7% CI)
    Chlamydia trachomatis (CT) - Rectal
    Sensitivity (PPA)Sufficiently high for diagnostic use97.7% (93.5–99.2% CI)
    Specificity (NPA)Sufficiently high for diagnostic use99.4% (99.0–99.7% CI)
    Neisseria gonorrhoeae (GC) - Rectal
    Sensitivity (PPA)Sufficiently high for diagnostic use95.8% (89.7–98.4% CI)
    Specificity (NPA)Sufficiently high for diagnostic use99.8% (99.5–99.9% CI)
    Non-Reportable Rate (Total CT and GC)Reasonably low for clinical utility (e.g.,