K Number
K182988Device Name
Afinion HbA1c Dx on Afinion 2
Manufacturer
Alere Technologies AS
Date Cleared
2018-11-29
(31 days)
Product Code
PDJ, JQT
Regulation Number
862.1373Why did this record match?
Applicant Name (Manufacturer) :
Alere Technologies AS
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Afinion™ HbA1c Dx is an in vitro diagnostic test for quantitative determination of glycated hemoglobin (% homoglobin Alc, HbAlc) in human venous and capillary whole blood.
This test is to be used as an aid in the diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes.
The measurement of % HbA1c is recommended as a marker of long-term metabolic control in persons with diabetes mellitus.
The Afinion™ 2 System consisting of the Afinion™ 2 Analyser and the Afinion™ Test Cartridges is for in vitro diagnostic use only. The Afinion™ 2 Analyzer is a compact multi-of-care testing and is designed to analyze the Afinion™ Test Cartridges.
Device Description
The Afinion™ HbA1c Dx test system is a CLIA moderate complexity test for diagnosing diabetes and identifying patients who may be at risk for developing diabetes, as a marker of long-term metabolic control in persons with diabetes mellitus.
The Afinion 2 is a multi-assay analyzer for point-of-care testing, designed for use with Afinion assay test cartridges and Afinion controls. It has the same functionality as the Afinion AS100 analyzer performing identical assay processing.
Ask a Question
K Number
K180296Device Name
Afinion HbA1c Dx
Manufacturer
Alere Technologies AS
Date Cleared
2018-05-07
(94 days)
Product Code
PDJ
Regulation Number
862.1373Why did this record match?
Applicant Name (Manufacturer) :
Alere Technologies AS
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Afinion™ HbA1c Dx is an in vitro diagnostic test for quantitative determination of glycated hemoglobin (% hemoglobin A1c, HbA1c) in human venous and capillary whole blood.
This test is to be used as an aid in the diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes.
The measurement of % HbA1c is recommended as a marker of long-term metabolic control in persons with diabetes mellitus.
Device Description
The Afinion™ HbA1c Dx is a fully automated boronate affinity assay for the determination of the percentage of hemoglobin A1c in human whole blood. The Afinion™ HbA1c Dx is a modification of the existing device, Alere Afinion™ HbA1c for use with the Alere Afinion™ AS100 Analyzer, with the addition of a diagnostic intended use.
The test begins with a blood sample collected with the integrated sampling device before the test cartridge is placed in the cartridge chamber of the Alere Afinion™ AS100 Analyzer. The sample is then automatically diluted and mixed with a solution that releases hemoglobin from the erythrocytes. After the hemoglobin is precipitated, the sample mixture is transferred to a blue boronic acid conjugate which binds to the cis-diols of glycated hemoglobin. This reaction mixture is soaked through a filter membrane and all precipitated hemoglobin, conjugate-bound and unbound (i.e. glycated and non-glycated hemoglobin) remains on the membrane. Excess conjugate is removed with a washing reagent. The analyzer measures the reflectance of the precipitate on the membrane as blue (glycated hemoglobin) and red (total hemoglobin) color intensities. The analyzer calculates a ratio proportional to the percentage of HbA1c in the sample and displays as the % HbA1c (NGSP).
Ask a Question
K Number
K151809Device Name
Alere Afinion HbA1c, Alere Afinion AS100 Analyzer
Manufacturer
Alere Technologies AS
Date Cleared
2015-09-25
(85 days)
Product Code
LCP, JOT
Regulation Number
864.7470Why did this record match?
Applicant Name (Manufacturer) :
Alere Technologies AS
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Alere Afinion™ AS100 Analyzer with Alere Afinion™ Data Connectivity Converter (ADCC) is a compact multi-assay analyzer for point-of-care testing, designed to analyze the Afinion™ Test Cartridges. The ADCC is a small device for automatic transfer of data, including patient and control assay results, from the Alere Afinion™ Analyzer to a laboratory information system or another electronic journal system.
Alere Afinion™AS100 Analyzer System, consisting of Alere Afinion™ AS100 Analyzer with Alere Afinion™Data Connectivity Converter (ADCC), Alere Afinion™ Test Cartridges and Alere Afinion™ Controls is for in vitro diagnostic use only.
Alere Afinion™HbA1c is an in-vitro diagnostic test for quantitative determination of glycated hemoglobin (% hemoglobin A1c, % HbA1c) in human whole blood. The measurement of % HbA1c is recommended as a marker of longterm metabolic control in persons with diabetes mellitus.
Alere Afinion™HbA1c Controls have been designed for use with the Alere Afmion™AS100 Analyzer System. Quality control using the Alere Afinion™HbA1c Control should be done to confirm that the Alere Afinion™AS100 Analyzer System is working properly and provides reliable result.
The Alere Afinion™ ACR assay is an in vitro diagnostic test for quantitative determination of albumin, creatinine and albumin/creatinine ratio (ACR) in human urine using the Alere Afinion™ AS100 Analyzer. The measurement of urine albumin, creatinine and albumin/creatinine ratio, aids in the early diagnosis of nephropathy.
The Alere Afinion™ ACR Control kit contains liquid preparations of albumin and creatinine in citrate buffer. The controls should be used to confirm that the Alere Afinion™AS100 Analyzer System is working properly and provides reliable results.
The Afinion™ Lipid Panel is an in vitro diagnostic test for quantitative determination of total cholesterol (Chol), high-density lipoprotein (HDL) cholesterol and triglycerides (Trig) in serum. Values for low-density lipoprotein (LDL) cholesterol and Chol/HDL ratio are calculated by the Afinion™ AS100 Analyzer. Chol, HDL cholesterol, Trig, and calculated LDL cholesterol, non-HDL cholesterol and Chol/HDL ratio) are used in the diagnosis and treatment of disorders involving excess or low cholesterol in the blood and lipid and lipoprotein metabolism disorders. Afinion™ Lipid Panel Control has been designed for use with the Afinion™ AS100 Analyzer and Afinion™ Lipid Panel. Afinion™ Lipid Panel Control is intended for use as assayed control material for total cholesterol (Chol), high-density lipoprotein (HDL) cholesterol and triglycerides (Trig). The controls should be used to confirm that the Afinion™ AS100 Analyzer System is working properly and provides reliable results.
For use in clinical laboratories and point of care laboratory settings. For prescription use only.
Device Description
The Alere Afinion HbA1c test cartridge contains all of the reagents necessary for the determination of % HbA1c. The sample material is collected from a fingerstick or a sample tube with the integrated sampling device before the test cartridge is placed in the cartridge chamber of the Alere Afinion AS100 Analyzer. The blood sample is then automatically diluted and mixed with a solution that releases hemoglobin from the erythrocytes. The hemoglobin precipitates. This sample mixture is transferred to a blue boronic acid conjugate, which binds to the cis-diols of glycated hemoglobin. This reaction mixture is soaked through a filter membrane and all precipitated hemoglobin, conjugate-bound and unbound (i.e. glycated and non-glycated hemoglobin) remains on the membrane. Any excess of conjugate is removed with a washing reagent. The analyzer evaluates the precipitate on the membrane. By measuring the reflectance, the blue (glycated hemoglobin) and the red (total hemoglobin) color intensities are evaluated, the ratio between them of which is proportional to the percentage of HbA1c in the sample. The % HbA1c is displayed on the Alere Afinion™ AS100 Analyzer.
The analyzer uses different chemical and mechanical assay methods combined with advanced, computerized processing and measuring technology to perform a variety of diagnostic assays.
A test cartridge with patient sample or control is placed in the cartridge chamber of the analyzer. By manually closing the lid, the test cartridge is transported into the analysis compartment of the analyzer. There is no further user intervention or activity required to perform the assay. Test and lot specific information are obtained from the test cartridge barcode label, which then initiates the processing of the test. The sample and reagents are automatically transferred between the wells of the test cartridge.
A monochrome solid-state camera monitors the entire process. When the assay is completed, lightemitting diodes (LEDs) illuminate the final reaction area, which can be either a colored membrane or a reaction well. The camera detects the reflected or transmitted light, which is converted to a test result and displayed on the touch screen. When the user accepts the result, the lid covering the cartridge chamber opens automatically and the used test cartridge can be removed and discarded. The analyzer is then ready for the next run.
The current sampling device integrated in the test cartridge is made of plastic with a glass capillary tube glued into the sampling device. The new, modified plastic sampling device will be manufactured in one piece of plastic of the same base material (polystyrene) as the plastic part of the current sampling device.