Search Results
Found 9 results
510(k) Data Aggregation
(98 days)
Aerin Medical, Inc.
The Aerin Console is an electrosurgical system intended to generate radiofrequency electrical current for the use of an Aerin Medical Stylus. The Aerin Console is indicated for use in small clinic, office or hospital environments.
The Aerin Console is a radio-frequency (RF) generator designed to be used with an Aerin Medical Stylus (RhinAer Stylus, K221907 and VivAer, K200300) to deliver bipolar RF energy to tissue. The Aerin Console consists of an RF generator, a power cord, and a foot switch. It incorporates user interface software to control, monitor and regulate RF power delivery to soft tissues via a cable-connected electrosurgical hand piece and electrode.
The provided text is a 510(k) summary for a medical device called the "Aerin Console." A 510(k) submission generally focuses on demonstrating substantial equivalence to a predicate device rather than providing detailed acceptance criteria and performance studies like those typically seen for novel AI/ML devices or those requiring clinical studies.
Based on the provided text, there is no information about acceptance criteria or a study that proves the device meets specific performance criteria related to AI/ML or diagnostic accuracy. The device is an electrosurgical system (RF generator), not an AI-powered diagnostic tool. The document states:
- "Clinical testing was not necessary for this device."
- The comparison focuses on technological characteristics (design, energy type, temperature, power, stylus compatibility) to a predicate device.
- The non-clinical tests listed are for electrical safety, EMC, software validation (IEC 62304), and device security (IEC 81001-5-1). These are related to the hardware and software's safe operation, not its performance in a diagnostic or AI-assisted task.
Therefore, most of the requested information (points 1-9) which pertain to the performance evaluation of a diagnostic device, especially one involving AI/ML and human interpretative performance, cannot be extracted from this document.
However, I can extract the relevant information from the document as it applies to the device described:
-
A table of acceptance criteria and the reported device performance:
- Acceptance Criteria (Implicitly based on comparison to predicate and safety standards):
- Compliance with electrical safety, EMC, software validation, and device security standards.
- Performance characteristics (design, energy type, temperature range, output power, stylus compatibility) substantially equivalent to the predicate device.
- Reported Device Performance (from "Summary of non-clinical test"):
- Complies with:
- ISO 60601-1 (Electrical Safety)
- ISO 60601-1-2 (EMC)
- IEC 60601-4-2 (EMC)
- 47 CFR Part 15 Subpart B (FCC, unintentional radiators)
- 47 CFR Part 27 (FCC, wireless (likely related to EMC, but could imply wireless comms are part of device))
- IEC 62304 (Software validation)
- IEC 81001-5-1 (Device Security)
- Complies with:
- Acceptance Criteria (Implicitly based on comparison to predicate and safety standards):
-
Sample size used for the test set and the data provenance: Not applicable. This is not an AI/ML diagnostic device, and no test set in the context of diagnostic performance (e.g., images for classification) is mentioned. The "tests" refer to compliance with engineering standards.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for diagnostic purposes is not established for this device.
-
Adjudication method for the test set: Not applicable.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, explicitly stated "Clinical testing was not necessary for this device."
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a hardware device with controlling software, not a standalone algorithm for diagnostic interpretation.
-
The type of ground truth used: Not applicable for diagnostic performance. Ground truth for engineering tests (e.g., electrical safety) is defined by the standards themselves.
-
The sample size for the training set: Not applicable. This device does not use a training set for machine learning.
-
How the ground truth for the training set was established: Not applicable.
In summary, the provided document is a regulatory submission for an electrosurgical generator, focusing on demonstrating substantial equivalence to a predicate device and compliance with relevant safety and performance standards for hardware and control software. It does not contain the kind of information typically associated with the performance evaluation of AI/ML diagnostic devices.
Ask a specific question about this device
(51 days)
Aerin Medical Inc.
The Reach Needle is used for in injecting local anesthetics into a patient to provide regional anesthesia.
The Reach Needle is a sterile, single use, 4-inch-long needle used for administering anesthetic solutions for regional anesthesia. The Reach Needle is 27-qauge needle that is reinforced to reduce needle flex during administration of anesthesia solutions. The needle tip is a short bevel quincke style tip that extends approximately 2mm past the reinforcing shaft, providing a distal stop for consistent injection depth. The distal end of the needle has a slight bend to help access difficult to reach areas such as nasal anatomy. The needle hub is compatible with luer lock connectors. The Reach Needle is intended for use in adults (≥ 22 years).
The provided text is a U.S. FDA 510(k) summary for the "Reach Needle," an anesthesia conduction needle. It outlines the device's characteristics, intended use, and comparison to a predicate device.
Crucially, the document explicitly states under the section "Summary of clinical test": "Clinical testing was not necessary for this device."
This means that the device's acceptance was based on non-clinical testing (e.g., in vitro or bench testing) and a demonstration of substantial equivalence to a legally marketed predicate device, rather than a clinical study involving human patients or complex AI algorithms requiring extensive ground truth establishment and expert adjudication.
Therefore, I cannot provide the information requested in your prompt regarding acceptance criteria and a study proving the device meets those criteria, as the nature of the device (a medical needle, not an AI or imaging device) and the regulatory pathway chosen (510(k) based on substantial equivalence and non-clinical data) did not involve such a study.
The questions you asked are highly relevant for AI/ML-enabled medical devices or devices where performance needs to be validated against complex human assessment or outcomes, but they do not apply to this specific medical needle as per the provided FDA documentation.
Ask a specific question about this device
(29 days)
Aerin Medical Inc.
The RhinAer® Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the destruction of soft tissue in the nasal airway, including in posterior nasal nerve regions in patients with chronic rhinitis.
The RhinAer Stylus is functionally unchanged from the predicate in design and intended use to generate and deliver bipolar RF energy to treat tissue in ENT procedures. The design of the Stylus shaft has been changed to improve physician visualization and access during the procedure. The Stylus shaft diameter has been reduced and is more malleable. The treatment tip shape has been re-designed to change its position relative to the shaft to facilitate tissue apposition. The subject and predicate devices have the same material composition. The device and packaging system has been tested and validated with shelf-life studies out to 6 months. The optional custom treatment parameters (power, temperature, time) have been removed.
The provided text describes a 510(k) summary for a medical device called the RhinAer® Stylus (FG1393). This summary focuses on demonstrating substantial equivalence to a predicate device (RHIN1 Stylus, K192471) rather than presenting a study where specific acceptance criteria are defined for a performance outcome and then proven.
In the context of this 510(k) submission, "acceptance criteria" are generally related to demonstrating that the modified device maintains the safety and effectiveness of its predicate, and that changes to the device (shaft diameter, malleability, treatment tip design) do not introduce new safety or effectiveness concerns. The "study" here refers to the non-clinical tests performed to support this claim.
Given the information, here's how we can address your request:
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state quantitative acceptance criteria for a "device performance" in the sense of a clinical outcome, but rather for engineering and safety aspects to demonstrate substantial equivalence.
Acceptance Criteria Category | Specific Criteria (Implicit or Explicit in Document) | Reported Device Performance |
---|---|---|
Functional Performance | Adequate stylus strength (force load testing) | Met all performance testing requirements. |
Thermocouple accuracy | Met all performance testing requirements. | |
Thermocouple response time | Met all performance testing requirements. | |
Usability | No new use errors causing serious harm | All study objectives met; no new use errors causing serious harm identified. |
No further improvement to device/user interface necessary | All study objectives met; device safe to use. | |
Biocompatibility | Equivalent to predicate (due to similar material composition) | Predicate device's testing applicable. |
Electrical Safety | Equivalent to predicate (due to similar design and intended use) | Predicate device's testing applicable. |
Sterility Validation | Equivalent to predicate (due to similar design and intended use) | Predicate device's testing applicable. |
Shelf-life | Maintaining integrity and functionality for 6 months | Passed all aging and transit testing acceptance criteria for 6-month shelf life. |
Material Composition | Same as predicate device | Subject and predicate devices have the same material composition. |
RF Generator Compatibility | Same as predicate (Aerin Console, Model FG226) | Same |
Energy Type | Same as predicate (Bipolar radiofrequency) | Same |
Tissue Temperature | Consistent with predicate's operating range (50-70°C) | 60°C (within predicate's range) |
Power | Consistent with predicate's operating range (3-5W) | 4 W (within predicate's range) |
2. Sample Size Used for the Test Set and Data Provenance
The document only references "non-clinical tests" and "bench functional testing" and "usability testing." It does not provide specific sample sizes for these tests beyond implying a sufficient number were performed to meet test objectives.
- Sample Size (Test Set): Not specified beyond general statements about "bench functional testing," "usability testing," and "aging and transit testing."
- Data Provenance: Not explicitly stated, but as these are non-clinical (bench and usability) tests conducted for a medical device submission, they would typically be performed in a controlled laboratory or simulated clinical environment. There is no indication of country of origin of patient data, as no clinical studies were performed.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- This section is not applicable (N/A) because the reported "studies" are non-clinical engineering and usability tests, not studies requiring expert-adjudicated ground truth for a clinical outcome (e.g., image interpretation or disease diagnosis).
- Usability testing likely involved healthcare professionals or trained users, but their role was to evaluate the device interface and identify potential use errors, not to establish a clinical "ground truth."
4. Adjudication Method for the Test Set
- N/A. Adjudication methods (like 2+1, 3+1) are typically used for establishing ground truth in clinical studies involving interpretation (e.g., radiology reads). The non-clinical tests described (functional, usability, shelf-life) do not involve such adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of AI vs. Without AI Assistance
- N/A. This device is an electrosurgical cutting and coagulation device, not an AI or imaging diagnostic device. Therefore, an MRMC study related to AI assistance is not relevant to this submission.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- N/A. This device is a physical medical instrument, not an algorithm, so the concept of standalone algorithm performance does not apply.
7. The Type of Ground Truth Used
- N/A for clinical ground truth. For the non-clinical tests, the "ground truth" was established by engineering specifications, safety standards, and validated test methods (e.g., force load measurements, thermocouple calibration, sterility validations, package integrity standards).
8. The Sample Size for the Training Set
- N/A. Training sets are relevant for machine learning algorithms. This device is a physical medical instrument and does not involve a training set as understood in AI/ML contexts.
9. How the Ground Truth for the Training Set Was Established
- N/A. This question is not applicable for the reasons mentioned in point 8.
Ask a specific question about this device
(67 days)
Aerin Medical Inc
The VivAer® Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue, including cartilage in the internal nasal valve area.
The VivAer® Stylus is a disposable, handheld device capable of delivering bipolar radiofrequency energy to tissue. The stylus consists of an array of bipolar electrodes positioned on a non-conductive tip which is attached to a handle via a non-conductive shaft. A temperature sensor is located on the tip to monitor tissue temperature during treatment. The product's intended users are physicians including otolaryngologists, maxillofacial surgeons and other physicians specialized in nasal procedures.
The VivAer® Stylus improves nasal breathing by modifying the soft tissues of the nasal airway through the use of low doses of radiofrequency energy. The low-power radiofrequency generates heat within the submucosal tissue, creating a coagulation lesion. As the lesion heals, the tissue retracts and stiffens. This decreases the nasal airflow resistance thereby improving inflow of air through the nose.
Here's a breakdown of the acceptance criteria and study information for the VivAer Stylus based on the provided FDA 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
Test/Characteristic | Acceptance Criteria | Reported Device Performance |
---|---|---|
Packaging Validation | ||
Sterilization (Ethylene Oxide) | Sterility Assurance Level (SAL) of 10-6 | Demonstrated ability to achieve SAL of 10-6 |
Shelf-life (Accelerated Aging) | Pass all testing acceptance criteria | Passed all accelerated aging and transit testing |
Usability | No use-related errors that may result in serious harms | No use-related errors identified |
Biocompatibility | Not applicable (no new patient-contacting materials) | Not repeated |
Device Performance Testing | ||
Thermocouple Accuracy & Response Time | Met all requirements | Met all requirements |
Treatment Parameter Verification | Confirmed correct programming of parameters | Confirmed correct programming of parameters |
Reuse Prevention Verification | Met all requirements | Met all requirements |
Force Load Testing (Shaft Strength) | Met all requirements | Met all requirements |
Tissue Heating & Conductive Gel Compatibility | Confirmed energy delivery to tissue samples at variety of treatment parameters | Confirmed energy delivery to tissue samples at variety of treatment parameters |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state a specific sample size for a "test set" in the context of an AI/algorithm-based device. The testing described is primarily non-clinical performance testing of a physical medical device (VivAer Stylus) and its packaging.
- Sample Size: Not specified in terms of numerical units for each test, but implied to be sufficient for demonstrating compliance with acceptance criteria.
- Data Provenance: The studies were conducted internally by Aerin Medical Inc. in Sunnyvale, California, as part of their 510(k) submission. These are prospective tests performed on the manufactured device.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not applicable as the described testing is for a surgical device focused on its physical and functional performance, not for an AI/algorithm that requires expert-established ground truth for image interpretation or diagnosis. The ground truth for these tests is based on engineering specifications, regulatory standards, and objective measurements (e.g., temperature readings, force measurements, sterility verification).
4. Adjudication Method for the Test Set
This is not applicable as there is no mention of a human-in-the-loop diagnostic or interpretive task that would require adjudication. The testing involves engineering and performance validations.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for AI-powered diagnostic or screening devices to assess the impact of AI on human reader performance. The VivAer Stylus is a surgical device, not a diagnostic AI.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No, a standalone algorithm performance study was not done. This device is a physical electrosurgical tool and does not employ an AI algorithm in the manner described. The "algorithm" here refers to the pre-programmed treatment parameters of the device (power, temperature, time), which were verified for correct programming.
7. The Type of Ground Truth Used
The ground truth used for the described non-clinical tests is based on:
- Engineering Specifications: E.g., for thermocouple accuracy, shaft strength, power output.
- Regulatory Standards: E.g., for sterilization (SAL of 10-6), biocompatibility.
- Objective Measurements: E.g., confirmed energy delivery to tissue samples, verification of programmed parameters.
8. The Sample Size for the Training Set
This is not applicable. The device is a surgical instrument, not an AI/machine learning model that requires a training set of data.
9. How the Ground Truth for the Training Set Was Established
This is not applicable for the same reason as #8.
Ask a specific question about this device
(101 days)
Aerin Medical Inc.
The RHIN I Stylus is indicated for use in otothing of the destruction of soft tissue in the nasal airway, including in posterior nasal nerve regions in patients with chronic rhinitis.
The RHIN1 Stylus is a handheld bipolar radiofrequency (RF) probe designed for use in otorhinolaryngology (ENT) surgery. The Stylus comprises a handle, shaft, and treatment tip. The treatment tip consists of an array of bipolar electric es and a temperature sensor that allows for monitoring of tissue temperature during RF energy delivery. The shaft is malleable to allow the user to bend the shaft in order to more easily access the desired treatment area(s). The Stylus is designed for use with the Aerin Console. It includes features to allow compatibility with and authentication by the Aerin Console. The Stylus connects to the Aerin Console via a flexible cable.
The RHIN1 Stylus is used to treat patients experiencing chronic rhinitis. During a treatment procedure, the clinician inserts the shaft of the RHIN1 Stylus into a patient's nostril to position the treatment tip and deliver low power RF energy to the target tissue of the nasal airway. The low-power radiofrequency energy generates heat within the tissue, causing destruction of the unwanted soft tissue. The shaft of the RHIN1 Stylus is then removed from the nostril.
The provided text describes the 510(k) premarket notification for the RHIN1 Stylus, a radiofrequency probe for treating chronic rhinitis. While the document details the device's technical characteristics, regulatory classification, and a summary of non-clinical and clinical tests, it does not include information about an AI/algorithm-based device. Instead, the study outlined is a clinical trial assessing the effectiveness and safety of the RHIN1 Stylus itself in human subjects for a medical procedure.
Therefore, the requested information regarding acceptance criteria and a study proving the device meets those criteria for an AI/algorithm-based device cannot be extracted from this document. The document pertains to a physical medical device and its performance in a clinical setting.
To answer your request, I would need a document describing an AI/algorithm-based medical device.
Ask a specific question about this device
(106 days)
Aerin Medical, Inc.
The Vivaer ARC Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue, in the internal nasal valve area.
The Vivaer ARC Stylus is a handheld bipolar radiofrequency (RF) probe designed for use in otorhinolaryngology (ENT) surgery. The Stylus comprises a handle, shaft and treatment tip. The treatment tip consists of an array of bipolar electrodes and a temperature sensor that allows for monitoring of tissue temperature during RF energy delivery. The Stylus is designed for use with the Aerin Console. It includes features to allow compatibility with and authentication by the Aerin Console. The Stylus connects to the Aerin Console via a flexible cable.
Here's a summary of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
---|---|---|
Device Performance (Non-Clinical) | System responsiveness | Met all performance testing requirements. |
Effectiveness for clinical use | Met all performance testing requirements. | |
Adequate shaft strength (Force load testing) | Verified. | |
Thermocouple accuracy and response time (water bath immersion) | Efficacy supported. | |
Tissue heating time (using representative soft tissues of the nasal airway) | Efficacy supported. | |
Comparable thermal lesion volumes to predicate device | Showed that the Vivaer ARC Stylus creates thermal lesion volumes comparable to the predicate device. | |
Device Safety (Non-Clinical) | Biocompatibility (AAMI/ANSI/ISO 10993-1 (2009/(R) 2013)) | Demonstrated through testing. |
Sterilization validation (ANSI/AAMI/ISO 11135:2014 and ISO 10993-7:2008) | Performed in accordance with standards. | |
Electrical safety (IEC 60601-1:2005/A1:2012 and IEC 60601-2-2:2009) | Demonstrated through testing. | |
Electromagnetic compatibility (IEC 60601-1-2:2007/AC:2010) | Performed according to standard. | |
Packaging validation (Transit testing, visual inspection, accelerated aging, gross leak detection, seal strength testing in accordance with ASTM D4169-14, D4332-14, F1980-07(2011), F2096-11, F88/F88M-15) | Met all acceptance criteria. | |
Long-term safety and healing in nasal tissue (in vivo canine study) | Treatment resulted in durable change of submucosal gland population, no significant fibrosis, tissue contraction, or cartilage degeneration after 6 months. Healing process complete or near-complete at 6 months. Supports long-term safety. | |
Clinical Efficacy | Primary Efficacy Endpoint: Statistically significantly greater than 15-point improvement in mean NOSE score from baseline to 26 weeks post-procedure. | Met. Mean NOSE score decreased 55 points (SD 21) from baseline to 26 weeks (paired t-test, p |
Ask a specific question about this device
(97 days)
Aerin Medical, Inc.
InSeca ARC Stylus: The InSeca ARC Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue.
Aerin Console: The Aerin Console is an electrosurgical system intended to generate radiofrequency electrical current for the use of an ARC Stylus (e.g., InSeca ARC Stylus). The Aerin Console is indicated for use in small clinic, office or hospital environments.
InSeca ARC Stylus: The InSeca ARC Stylus is a handheld bipolar radiofrequency (RF) probe designed for use in otorhinolaryngology (ENT) surgery. The Stylus comprises a handle, shaft and treatment tip. The treatment tip consists of an array of bipolar electrodes and a temperature sensor that allows for monitoring of tissue temperature during RF energy delivery. The Stylus is designed for use with the Aerin Console. It includes features to allow compatibility with and authentication by the Aerin Console. The Stylus connects to the Aerin Console via a flexible cable.
The InSeca ARC Stylus is used to treat patients experiencing chronic nasal airway obstruction. During a treatment procedure, the clinician inserts the tip of the InSeca ARC Stylus into a patient's nostril to deliver low power RF energy to the target tissue of the nasal airway. Radiofrequency treatment of tissue creates a coagulative lesion which fibroses and retracts as it heals, thereby shrinking the tissue to lessen the degree of obstruction.
Aerin Console: The Aerin Console is an RF generator designed to be used with the InSeca ARC Stylus to deliver bipolar RF energy to tissue. The Aerin Console consists of an RF generator, a power cord, and a foot switch. It incorporates user interface software to control, monitor and regulate RF power delivery to soft tissues via a cableconnected electrosurgical hand piece and electrode.
The provided document discusses a 510(k) premarket notification for the InSeca ARC Stylus and Aerin Console. This submission focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and a study to prove meeting those criteria in the way a clinical trial or a performance study for a novel device would.
Based on the provided text, here's a breakdown of the information requested, with "N/A" where the information is not provided because the submission is for substantial equivalence rather than a full performance study with specific acceptance criteria in the traditional sense:
1. Table of Acceptance Criteria and Reported Device Performance
The document describes non-clinical performance testing. The "acceptance criteria" are implied by the statement "The subject devices met all the performance testing requirements." Specific quantitative acceptance criteria for each test are not explicitly listed in a table, but the areas of testing are detailed.
Test Category | Implied Acceptance Criteria | Reported Device Performance |
---|---|---|
Mechanical Performance | Adequate shaft strength | Met all performance testing requirements |
Force Load Testing | ||
Functional Performance (Efficacy) | Accurate thermocouple readings | Met all performance testing requirements |
Thermocouple Accuracy & Response Time | Appropriate tissue heating time | Met all performance testing requirements |
Tissue Heating Time Testing | Thermal lesion volumes comparable to predicate device | Created thermal lesion volumes comparable to predicate |
Thermal Effects Comparison | ||
Software Validation | Software functions as intended for control, monitoring, and regulation of RF power | Met all performance testing requirements |
Usability Testing | N/A (General usability, no specific criteria given) | Met all performance testing requirements |
Biocompatibility | Biocompatibility in accordance with ISO 10993-1 | Demonstrated biocompatibility |
Sterilization & Packaging | Sterilization per ISO 11135:2014 & ISO 10993-7:2008; packaging per ASTM standards | Sterilization and packaging validated and met all criteria |
Electrical Safety | Compliance with IEC 60601-1 & IEC 60601-2-2 | Demonstrated electrical safety |
Electromagnetic Compatibility (EMC) | Compliance with IEC 60601-1-2 | Demonstrated EMC |
2. Sample Size Used for the Test Set and Data Provenance
The document describes non-clinical bench testing. Specific sample sizes for each test (e.g., number of styluses tested for force load, number of tissue samples for heating time) are not specified. The data provenance is generally "bench testing" and "tissue heating time testing performed using tissues representative of the submucosal tissues of the nasal airway." There is no mention of country of origin for bench testing, and it is entirely retrospective in the sense that it's laboratory testing, not a prospective human clinical trial.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
Not applicable. The ground truth for bench and functional testing is based on engineering specifications, regulatory standards, and objective measurements (e.g., temperature, force, voltage, thermal lesion volume). There is no mention of human experts establishing "ground truth" for the non-clinical tests described.
4. Adjudication Method for the Test Set
Not applicable. As the testing described is non-clinical bench and laboratory testing against established specifications and regulatory standards, an adjudication method for a "test set" in the context of expert review or consensus is not relevant or described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study typically involves human readers (e.g., radiologists, clinicians) evaluating cases (e.g., images) to determine the effectiveness of an intervention (like an AI algorithm). This type of study is not mentioned in the document as it pertains to a physical electrosurgical device rather than an AI or image-based diagnostic system. Therefore, no effect size of human readers improving with AI assistance is provided.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Not applicable. The InSeca ARC Stylus and Aerin Console are electrosurgical devices used by a clinician during a procedure; they are not an AI algorithm designed to operate in a standalone diagnostic or predictive capacity. While the Aerin Console incorporates "user interface software to control, monitor and regulate RF power delivery," its performance is intrinsically linked to its use by a human operator with the stylus. No "algorithm only" performance is discussed.
7. The Type of Ground Truth Used
The "ground truth" for the non-clinical tests described in the document is based on:
- Engineering specifications and design requirements: For mechanical integrity, electrical safety, and software functionality.
- Regulatory standards: Such as AAMI/ANSI/ISO 10993-1 for biocompatibility, ANSI/AAMI/ISO 11135 and ISO 10993-7 for sterilization, IEC 60601-1/60601-2-2 for electrical safety, and IEC 60601-1-2 for EMC.
- Physical measurements and objective comparisons: For thermocouple accuracy, tissue heating time, and thermal lesion volumes (compared to the predicate device).
8. The Sample Size for the Training Set
Not applicable. This document describes a 510(k) submission for a medical device based on substantial equivalence, with non-clinical performance and safety testing. It does not involve machine learning or AI models with "training sets."
9. How the Ground Truth for the Training Set Was Established
Not applicable. As there is no "training set" referenced for a machine learning model, the establishment of its ground truth is not relevant to this document.
Ask a specific question about this device
(65 days)
Aerin Medical, Inc.
The Aerin Medical Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue.
The Aerin Medical Stylus is a handheld bipolar radiofrequency (RF) probe designed for use in otorhinolaryngology (ENT) surgery. The Stylus comprises a handle, shaft and treatment tip. The treatment tip consists of an array of bipolar electrodes and a temperature sensor that allows for monitoring of tissue temperature during radiofrequency energy delivery. The stylus is connected, via a flexible cable, to a commercially-available RF generator (the ORA-50 S, cleared under K993854) meeting the following requirements: RF operating frequency of 460 kHz (± 5 kHz); bipolar low power (3 watts to 5 watts) RF energy delivery; and sensitive temperature control with low overshoot (50°C to 70°C).
This document describes a 510(k) premarket notification for the Aerin Medical Stylus, a radiofrequency wand intended for use in otorhinolaryngology (ENT) surgery to coagulate soft tissue in the nasal airway to treat nasal airway obstruction. The submission asserts substantial equivalence to the predicate device, Aerin Medical Wand (K150637).
Based on the provided text, the following information can be extracted regarding the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document does not provide a formal table of specific quantitative acceptance criteria for device performance. Instead, it describes general performance testing conducted and states that the device "met all performance testing requirements." The relevant performance aspects mentioned are:
Acceptance Criteria (General Description) | Reported Device Performance |
---|---|
Biocompatibility | Demonstrated through testing per AAMI / ANSI / ISO 10993-1. |
Sterilization and Packaging Validation | Validated per AAMI / ANSI / ISO 11137-1 and 11137-2. Packaging met acceptance criteria per ASTM standards. |
Electrical Safety | Demonstrated through testing to IEC 60601-1 and 60601-2-2. |
Electromagnetic Compatibility | Performed according to IEC 60601-1-2. |
Usability (ergonomics, visualization, orientation, force application) | Evaluated as part of a cadaver study, implying met acceptance. |
Shaft Strength (force load testing) | Conducted, implying adequate strength and met acceptance. |
Thermocouple Accuracy and Response Time | Tested via water bath immersion, implying met accuracy and response criteria. |
Tissue Heating Time | Tested using representative tissues, implying met heating time criteria. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size:
- Cadaver Study (for usability): The document mentions a "cadaver study" but does not specify the number of cadavers or individual tests performed.
- Tissue Heating Time: The number of tissue samples or tests is not specified.
- Other tests (Biocompatibility, Sterilization, Electrical Safety, EMC, Packaging): These tests typically involve a defined number of device units or samples according to the relevant standards, but specific numbers are not provided in this summary.
- Data Provenance: Not specified in the provided text. The submission is from Aerin Medical, Inc. based in Sunnyvale, CA, USA, so it is likely that the testing was conducted in the USA or by facilities adhering to relevant international standards. The document does not indicate whether data was retrospective or prospective for any of the non-clinical tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable for this device. The testing described is primarily non-clinical engineering and performance testing. "Ground truth" in the traditional sense of expert consensus on diagnostic images or clinical outcomes is not relevant here. Decisions on usability would likely have been made by engineers or clinical specialists involved in the cadaver study, but their qualifications and numbers are not specified.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. As noted above, this summary focuses on non-clinical performance and safety testing, not clinical studies requiring adjudication of outcomes or diagnoses.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The Aerin Medical Stylus is a surgical device, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted or is relevant here.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. For the non-clinical studies described:
- Biocompatibility and Sterilization were validated against established ISO/AAMI standards.
- Electrical Safety and EMC were validated against IEC standards.
- Usability was evaluated in a cadaver study (presumably by observing ease of use, visualizing the treatment, etc.).
- Thermocouple accuracy and response time were measured against known temperatures in a water bath.
- Tissue heating time was measured directly in representative tissues.
The "ground truth" for these tests are the objective measurements against predefined engineering specifications and regulatory standards.
8. The sample size for the training set
Not applicable. This is a physical device, not an AI algorithm requiring a training set.
9. How the ground truth for the training set was established
Not applicable.
Ask a specific question about this device
(253 days)
Aerin Medical, Inc.
The Aerin Medical Wand is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue.
The Aerin Medical Wand is a handheld bipolar radiofrequency (RF) probe designed for use in otorhinolaryngology (ENT) surgery. The Wand comprises a handle, shaft and treatment tip. The treatment tip consists of an array of bipolar electrodes and a temperature sensor that allows for monitoring of tissue temperature during radiofrequency energy delivery. The Wand is connected, via a flexible cable, to a commercially-available RF generator (the ORA-50 S, cleared under K993854) meeting the following requirements: RF operating frequency of 460 kHz (± 5 kHz); bipolar low power (3 watts to 5 watts) RF energy delivery; and sensitive temperature control with low overshoot (50°C to 70°C).
All patient-contacting materials of the Aerin Medical Wand are polymers or stainless steel that have been shown to be biocompatible for short-term contact (less than 24 hours) with breached or compromised surfaces. The Wand is individually packaged and supplied radiation-sterilized for single use.
The Aerin Medical Wand is used in the treatment of patients experiencing chronic nasal airway obstruction. During a treatment procedure, the clinician inserts the tip of the Aerin Medical Wand into a patient's nostril to deliver low power RF energy to the target tissue of the nasal airway. Radiofrequency treatment of tissue creates a coagulative lesion which fibroses and retracts as it heals, thereby shrinking the tissue to lessen the degree of obstruction.
Here's an analysis of the provided text regarding the Aerin Medical Wand (Model FG011) and its acceptance criteria and study data:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of numerical acceptance criteria for device performance. However, it implicitly defines successful performance by meeting the following:
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Safety: | |
Freedom from unanticipated serious adverse events (SAEs) | Met: No SAEs reported in the study. |
Minimal complications (e.g., temporary bleeding, edema, crusting, mucosal tears) | Very minimal complications observed, similar to those reported with the predicate device. All adverse events were anticipated observations from thermal treatment and healing, except for one common cold. |
No changes in outward appearance of the nose. | Met: External photographs of each treated nostril revealed no changes in outward appearance. |
Biocompatibility | Demonstrated through testing per AAMI / ANSI / ISO 10993 series. |
Radiation sterilization | Validated per AAMI / ANSI / ISO 11137 series. |
Electrical safety | Demonstrated through testing to IEC 60601-1 and 60601-2-2. |
Electromagnetic compatibility | Performed according to IEC 60601-1-2. |
Packaging integrity | Met all acceptance criteria of ASTM D4169-09, ASTM F1980-07, ASTM F2096-11, and ASTM F88/F88M-09. |
No significant variation or hot spots in thermal imaging | Met: Thermal imaging showed uniform heating without significant variation or hot spots. |
Effectiveness: | |
Technical feasibility (ability to deliver RF energy) | Evaluated and met. |
Clinically significant improvement in NOSE score | At 90 days, a clinically significant improvement was reported in 88% of subjects, as assessed with the NOSE scale. This is comparable to the 82-88% reported for the predicate device. One subject lost to follow-up reported a 25-point improvement at 30 days. |
Progression in healing and nasal obstruction improvement | Demonstrated during the 90-day follow-up period via endoscopic intranasal assessments, similar to literature for predicate device. |
Create thermal lesion volumes comparable to predicate device | Bench testing showed the Aerin Medical Wand creates thermal lesion volumes comparable to volumes created with the predicate device. Histological changes were evaluated as absent, minimal, or mild for tissue treated within specified parameters during in vivo preclinical testing. |
Meet all design requirements (usability, thermocouple response, tissue heating times) | Met all design requirements. |
2. Sample Sizes Used for the Test Set and Data Provenance
-
Clinical Study (Test Set):
- Sample Size: 33 subjects were enrolled.
- Treatments: 47 nostrils were treated in the index procedures. 10 subjects had repeat treatments (12 nostrils), for a total of 59 nostril treatments.
- Data Provenance: Prospective, multi-center, non-randomized study conducted in the United States.
-
Pre-Clinical Testing (Animal Study - also a form of test set):
- Sample Size: Not explicitly stated how many animals were used, but "Several combinations of energy delivery durations and target temperatures were used" in in vivo testing.
- Data Provenance: This was an in vivo animal study. Location not specified, but typically conducted in a controlled lab/research environment.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Clinical Study:
- The document implies that physicians performed "physician evaluation of the tissue healing progression" and assessed "endoscopic intranasal assessments." Specific number of physicians or detailed qualifications are not provided, but they would be medical professionals (likely ENT specialists) involved in the study.
- Pre-Clinical Testing:
- A veterinary pathologist evaluated the treated tissue histologically after 8 weeks of healing. The number of pathologists and their specific experience level are not stated.
4. Adjudication Method (for the test set)
- The document does not describe a formal adjudication method (e.g., 2+1, 3+1) for the clinical study's outcomes. Clinical outcomes like SAE reporting and NOSE score improvements are generally direct measures or subject self-assessments complemented by physician review.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study involving AI assistance was not done. This device is an electrosurgical wand, not an AI-powered diagnostic or assistive tool. The study focused on the device's safety and effectiveness in treating nasal obstruction, comparing its outcomes to medical literature on a predicate device.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not applicable. This report is for a physical medical device (electrosurgical wand), not an algorithm or software. Therefore, the concept of "standalone algorithm performance" does not apply.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)
-
For the Clinical Study (human subjects):
- Outcomes Data: Primarily based on the patient's subjective "Nasal Obstruction Symptom Evaluation (NOSE) score" (a validated patient-reported outcome measure).
- Physician Evaluation: Physician assessment of tissue healing progression via endoscopic intranasal assessments.
- Safety Data: Freedom from SAEs, reporting of adverse events, and external photographic assessments.
-
For the Pre-Clinical Testing (animal study):
- Pathology: Histological evaluation of treated tissue by a veterinary pathologist.
8. The Sample Size for the Training Set
- Not applicable. This document describes a medical device, not a machine learning algorithm. Therefore, there is no "training set" in the context of AI/ML. All samples discussed (clinical subjects, animal models) are generally considered test sets for assessing the final device's performance.
9. How the Ground Truth for the Training Set Was Established
- Not applicable, as there is no training set mentioned in the context of an AI/ML algorithm.
Ask a specific question about this device
Page 1 of 1