Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K043512
    Date Cleared
    2005-04-08

    (109 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EASYGLUCO Blood Glucose Monitoring System is used for the quantitative measurement of glucose level in whole blood as an aid in monitoring the effectiveness of diabetes management in the home and in clinical settings. The EASYGLUCO Blood Glucose Monitoring System is for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip testing along with alternate site testing on the arm, palm, thigh, and calf.

    Device Description

    The EASYGLUCO Blood Glucose Monitoring System is an in vitro diagnostic device designed for measuring the concentration of glucose in whole blood, which is used with the EASYGLUCO Test Strips. The test principle is: This device is an in vitro diagnostic product intended for the measurement of glucose concentration in human blood. The principle of the test relies upon a specific type of glucose in the blood sample, the dehydrogenase glucose that reacts to electrodes in the test strip. The test strip employs an electrochemical signal generating an electrical current that will stimulate a chemical reaction. This reaction is measured by the Meter and displayed as your blood glucose result.

    AI/ML Overview

    The provided text is a 510(k) summary for the EASYGLUCO Blood Glucose Monitoring System. It does not contain the detailed study information required to answer all the questions. The document focuses on establishing substantial equivalence to predicate devices and provides basic device description and intended use.

    Therefore, I can only address the questions for which information is available in the provided text.

    Here's a breakdown of the information that cannot be provided from the given document:

    • A table of acceptance criteria and the reported device performance: This detailed information is typically found in the full submission, not the summary.
    • Sample sized used for the test set and the data provenance: Not mentioned in the summary.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned in the summary.
    • Adjudication method for the test set: Not mentioned in the summary.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This device is a blood glucose monitor, not an AI-assisted diagnostic imaging device, so an MRMC study is not applicable.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: As a blood glucose monitor, its primary function is standalone measurement, but the specific performance study details are not provided.
    • The type of ground truth used: While the principle of "in vitro diagnostic product intended for the measurement of glucose concentration in human blood" implies comparison to a reference method, the specific ground truth used in a validation study is not detailed.
    • The sample size for the training set: Not mentioned in the summary. The device's operation doesn't suggest a traditional "training set" in the context of machine learning, but rather calibration/validation data.
    • How the ground truth for the training set was established: Not mentioned in the summary.

    Based on the provided text, here's what can be extracted:

    1. A table of acceptance criteria and the reported device performance

    The provided text does not contain a table of acceptance criteria or reported device performance data. The 510(k) summary states that the submission provides "sufficient data to understand the basis for a determination of substantial equivalence," implying that such data would be in the full submission.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not available in the provided 510(k) summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not available in the provided 510(k) summary. Given that this is an in vitro diagnostic device for glucose measurement, "experts" in this context would likely be reference laboratory personnel performing comparative measurements, rather than medical specialists interpreting images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/not available in the context of an in vitro diagnostic blood glucose monitor. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective interpretation of medical images or other diagnostic tests where there's potential for inter-observer variability. For a quantitative device like a blood glucose meter, validation involves comparing its output against a reference method, not expert adjudication of its readings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done, as this device is a blood glucose monitoring system, not an AI-assisted diagnostic imaging system that involves human readers interpreting results.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The EASYGLUCO Blood Glucose Monitoring System is designed as a standalone device for measuring glucose concentration. The device itself performs the measurement and displays the result without direct human intervention in the measurement process beyond sample application. The 510(k) summary implies that its performance was validated in this standalone capacity, but the details of such a study are not provided.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document doesn't explicitly state the specific "ground truth" used. However, for a blood glucose monitoring system, the ground truth would typically be established by a highly accurate laboratory reference method for glucose measurement (e.g., hexokinase method or isotope dilution-mass spectrometry) on the same blood samples used for device testing.

    8. The sample size for the training set

    This information is not available in the provided 510(k) summary.

    9. How the ground truth for the training set was established

    This information is not available in the provided 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K043543
    Date Cleared
    2005-04-08

    (109 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FREEDOM Blood Glucose Monitoring System is used for the quantitative measurement of glucose level in whole blood as an aid in monitoring the effectiveness of diabetes management in the home and in clinical settings. The FREEDOM Blood Glucose Monitoring System is for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip testing along with alternate site testing on the arm, palm, thigh and calf.

    Device Description

    The FREEDOM Blood Glucose Monitoring System is an in vitro diagnostic device designed for measuring the concentration of glucose in whole blood, which is used with the FREEDOM Test Strips.

    The test principle is:

    This device is an in vitro diagnostic product intended for the measurement of glucose concentration in human blood. The principle of the test relies upon a specific type of glucose in the blood sample, the dehydrogenase glucose that reacts to electrodes in the test strip. The test strip employs an electrochemical signal generating an electrical current that will stimulate a chemical reaction. This reaction is measured by the Meter and displayed as your blood glucose result.

    AI/ML Overview

    This is an in vitro diagnostic device, not an AI/ML device, so many of the requested categories (e.g., number of experts, adjudication method, MRMC study, sample size for training set) are not applicable. The information provided focuses on the device's accuracy against established references, which is typical for this type of device.

    Acceptance Criteria and Device Performance

    The provided 510(k) summary does not explicitly state acceptance criteria in a quantitative table format nor reported device performance in a similar table. However, based on the nature of blood glucose monitoring systems and the reference to substantial equivalence to predicate devices, the acceptance criteria would implicitly involve the device demonstrating accuracy and precision comparable to or better than the predicate devices and meeting relevant ISO standards for blood glucose monitoring.

    From the information provided, it states that the device measures "the concentration of glucose in whole blood" and is "an aid in monitoring the effectiveness of diabetes management." The device claims "substantial equivalence to the LifeScan, Inc., OneTouch® Ultra®" and others. This implies that the performance (accuracy, precision) of the FREEDOM Blood Glucose Monitoring System must be within acceptable limits relative to these established predicate devices.

    Implicit Acceptance Criteria (based on common standards for IVD blood glucose meters) & Reported Device Performance:

    Acceptance Criteria (Inferred from industry standards for IVD Blood Glucose Meters)Reported Device Performance (Inferred from "substantial equivalence")
    Accuracy: Results within a certain percentage/absolute difference of reference (e.g., 75 mg/dL,
    Ask a Question

    Ask a specific question about this device

    K Number
    K031501
    Date Cleared
    2003-10-16

    (156 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EASYGLUCO™ Monitoring System is used by individuals with diabetes. It is for the quantitative measurement of glucose levels in whole blood, as an aid in monitoring the effectiveness of diabetes management in the home and in clinical settings.

    Device Description

    The EASYGLUCO™ Monitor is an in vitro diagnostic device designed for measuring the concentration of glucose in whole blood, which is used with the EASYGLUCO™ Test Strips. The test principle is: This device is an in vitro diagnostic product intended for the measurement of glucose concentration in human blood. The principle of the test relies upon a specific type of glucose in the blood sample, the dehydrogenase glucose that reacts to electrodes in the test strip. The test strip employs an electrochemical signal generating an electrical current that will stimulate a chemical reaction. This reaction is measured by the Meter and displayed as your blood glucose result.

    AI/ML Overview

    The provided text does not contain detailed information about acceptance criteria or specific study results for the EASYGLUCO™ device. The document is a 510(k) summary for an in vitro diagnostic device, primarily focusing on establishing substantial equivalence to a predicate device.

    Therefore, I cannot extract the requested information regarding:

    • A table of acceptance criteria and reported device performance
    • Sample size used for the test set or data provenance
    • Number of experts or their qualifications for ground truth establishment
    • Adjudication method
    • MRMC comparative effectiveness study
    • Standalone performance
    • Type of ground truth used
    • Sample size for the training set
    • How ground truth for the training set was established

    The available information only describes the device, its intended use, and its substantial equivalence to other blood glucose monitoring systems. It does not include performance data or study design details.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1