Found 2 results

510(k) Data Aggregation

    K Number
    K122978
    Date Cleared
    2013-05-23

    (239 days)

    Product Code
    Regulation Number
    880.6880
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Revolutionary Science Saniclave 102 is designed to be used in medical and dental clinics, hospitals and other facilities where reusable sterile equipment is used. It is intended to sterilize unwrapped heat and moisture stable solid instruments, mated surfaces, knurled and hinged devices (excluding lumened devices and dental hand pieces) that are compatible with saturated steam sterilization at 121 degrees Celsius for 30 minutes. Unwrapped instruments that were sterilized should be used immediately after sterilization is completed (immediate use sterilization). The chamber must be manually drained and wiped dry after each cycle.

    Device Description

    The Saniclave 102 by Revolutionary Science (model number RS-SC-102) is a 120 volt autoclave. The Saniclave works like most other table top steam sterilizers by boiling water in a pressurized vessel (or chamber). As the water boils, the chamber pressurizes and the steam sterilizes the instruments placed inside. The Saniclave technology is based on the scientific concept that prolonged saturated pressurized steam at or over a temperature of 121 degrees Celsius kills bacteria.

    AI/ML Overview

    The Revolutionary Science Saniclave 102 is a steam sterilizer (autoclave) intended for use in medical and dental clinics, hospitals, and other facilities for sterilizing unwrapped heat and moisture-stable solid instruments.

    Here's an analysis of its acceptance criteria and the study performed:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" for performance in the typical sense of quantitative measures like sensitivity, specificity, or accuracy for a diagnostic device. Instead, the "acceptance criteria" are implied through substantial equivalence to a predicate device and adherence to industry standards for sterilizers.

    The core performance requirement for a sterilizer is to achieve sterilization. This is demonstrated by meeting specific thermal and time parameters.

    FeatureAcceptance Criteria (Implied / Predicate Device)Reported Device Performance (Saniclave 102)
    Sterilization Temperature121°C121°C
    Sterilization Time30 minutes30 minutes
    Pressure15 psi15 psi (controlled mechanically via ZPD valve and pressure relief valve)
    Process Equivalent Time (F₀)Greater than or equal to 30 minutesGreater than or equal to 30 minutes
    Sterilization EfficacyAchieve sterilizationDemonstrated by physical and biological testing in accordance with ANSI/AAMI ST55:2010
    Intended Use CompatibilitySteam sterilizer for unwrapped instrumentsSteam sterilizer for unwrapped instruments

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "Physical and biological testing were performed in accordance with ANSI/AAMI ST55:2010." However, it does not specify the sample size used for these tests (e.g., number of cycles, number of biological indicators).

    The data provenance is not explicitly stated in terms of country of origin, but the testing was conducted by Revolutionary Science, the manufacturer, in support of their 510(k) submission to the FDA (USA). The nature of the testing is prospective, as it's an evaluation of the device's performance against a standard.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This is not applicable in the context of device sterilization testing. The "ground truth" for a sterilizer's performance isn't established by human experts in the same way as, for example, image interpretation.

    Instead, the ground truth for sterilization efficacy is established by:

    • Physical measurements: Verifying that temperature, time, and pressure parameters are met throughout the cycle.
    • Biological indicators: Using standardized biological indicators (e.g., spore strips) containing highly resistant bacterial spores (e.g., Geobacillus stearothermophilus) to confirm spore inactivation. A successful test (ground truth) is no growth of microorganisms after incubation.

    The study refers to using "Class 5 integrators" which "confirm that pressure, temperature and time were achieved," and implies their use for each cycle. These integrators serve as a chemical indication of whether the critical sterilization parameters were met, further supporting the "ground truth" that sterilization conditions were achieved.

    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1, 3+1 consensus) are typically used for subjective assessments, such as expert review of medical images or clinical outcomes. This is not applicable to the objective physical and biological testing of a sterilizer. The results of biological indicator tests (growth/no growth) and physical parameter measurements are objective.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC study is not applicable to this device. MRMC studies are used to evaluate the diagnostic accuracy of AI systems in combination with human readers, typically for image interpretation tasks. The Saniclave 102 is a physical sterilization device, not a diagnostic tool that involves human interpretation of output.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop)

    While the device incorporates a microprocessor-based circuit board (software/firmware controlled), a "standalone performance" study in the context of AI algorithms is not applicable. The device's function is automated sterilization. The non-clinical testing performed (physical and biological testing per ANSI/AAMI ST55:2010) inherently assesses the "standalone" performance of the device and its control system in achieving sterilization parameters. There is no human "in the loop" required beyond operating the device.

    7. Type of Ground Truth Used

    The ground truth used for proving the device meets acceptance criteria is based on:

    1. Objective Physical Parameters: Confirmation that the specified sterilization temperature (121°C), time (30 minutes), and pressure (15 psi) are consistently achieved and maintained within the chamber during the sterilization cycle.
    2. Biological Inactivation (Biological Indicators): Absence of growth from highly resistant bacterial spores after exposure to the sterilization cycle and subsequent incubation.
    3. Chemical Integration (Class 5 Integrators): Confirmation by Class 5 chemical integrators that all critical parameters (time, temperature, and steam) for sterilization have been met.

    8. Sample Size for the Training Set

    The concept of a "training set" is not applicable here. The Saniclave 102 is a physical device, not an AI model that learns from data. Its operational parameters (time, temperature, pressure) are pre-programmed and fixed, not learned from a training dataset.

    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" in the context of this physical device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112811
    Device Name
    SANICLAVE 200
    Date Cleared
    2012-01-11

    (106 days)

    Product Code
    Regulation Number
    880.6880
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Revolutionary Science Saniclave 200 is designed to be used in medical and dental clinics, hospitals and other facilities where reusable sterile equipment is used. It is intended to sterilize wrapped and unwrapped heat and moisture stable solid instruments, mated surfaces, knurled and hinged devices (excluding lumened devices and dental hand pieces) that are compatible with saturated steam sterilization at 121 degrees Celsius for 30 minutes.

    Device Description

    The Saniclave 200 by Revolutionary Science (model number RS-SC-200) is a 120 volt front loading autoclave. The Saniclave works like most other table top steam sterilizers by boiling water in a pressurized vessel (or chamber). As the water boils, the chamber pressurizes and the steam sterilizes the instruments placed inside. The single heating element (permantently affixed to the bottom inside the chamber with bulkhead fittings) generates all heat for the autoclave, including preheat, sterilization and dry modes. It is immersed in the water of the chamber. When the cycle is initiated the heater turns on and boils the water.

    AI/ML Overview

    Here's an analysis of the Revolutionary Science Saniclave 200 (RS-SC-200) based on the provided 510(k) summary, structured to address your specific points:

    Since the device is a steam sterilizer (autoclave), the "acceptance criteria" and "device performance" are typically defined by demonstrating that the device effectively sterilizes according to established standards. The study mentioned is the validation testing against these standards.

    Revolutionary Science Saniclave 200 (RS-SC-200) Performance Study

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from ANSI/AAMI ST55:2010)Reported Device Performance (Saniclave 200)
    Meets all requirements of ANSI/AAMI ST55:2010 for steam sterilization.Testing showed that Saniclave 200 meets all requirements of this standard.
    Specific Performance Parameters (for Recommended Use):
    Sterilization Temperature: 121°COperates at 121°C for sterilization (for both wrapped and unwrapped).
    Sterilization Time: 30 minutesOperates for 30 minutes for sterilization (for both wrapped and unwrapped).
    Ability to sterilize wrapped instruments (up to 4.5 lbs)Validated for wrapped instruments (single pouching) up to 4.5 lbs.
    Ability to sterilize unwrapped instruments (up to 4.5 lbs)Validated for unwrapped instruments up to 4.5 lbs.
    Effective sterilization of heat and moisture stable solid instruments, mated surfaces, knurled and hinged devices (excluding lumened devices and dental hand pieces).Intended to sterilize these types of instruments. (Implicitly met by standard compliance.)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify the exact sample size (e.g., number of cycles, number of biological indicators, number of devices) used for the physical and biological testing. It only states that "physical and biological testing were performed."
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. Given that it's a 510(k) submission for a new device, the testing would be prospective and conducted by Revolutionary Science or a contracted testing facility.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For a steam sterilizer, "ground truth" related to sterilization efficacy is objectively established through physical measurements (temperature, pressure, time) and biological indicator (BI) kill-rates, which demonstrate the elimination of highly resistant microorganisms (e.g., Geobacillus stearothermophilus spores). This doesn't typically involve human expert consensus in the same way an AI diagnostic tool would.

    • Number of Experts: Not applicable in the context of expert consensus for ground truth. Testing is performed under controlled laboratory conditions, often overseen by engineers and microbiologists.
    • Qualifications of Experts: Not specified, but implied to be qualified personnel experienced in medical device sterilization validation according to ANSI/AAMI ST55.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. Sterilization efficacy is assessed against pre-defined, objective chemical and biological indicators, not through human consensus or adjudication of subjective interpretations.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size (AI vs. Without AI Assistance)

    • MRMC Comparative Effectiveness Study: No. This type of study is relevant for AI-powered diagnostic or interpretive tools that assist human readers (e.g., radiologists interpreting images). A steam sterilizer does not involve human "readers" or "AI assistance" in its direct function of sterilizing instruments.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone Performance: Not applicable in the context of an algorithm. The device itself (the Saniclave 200) is a standalone physical product that performs sterilization. Its performance is evaluated based on its ability to meet the established physical and biological parameters for sterilization, independent of human intervention during the cycle.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The ground truth for sterilization efficacy is established through:
      • Physical Monitoring: Verification that physical parameters (temperature, pressure, time) achieve the specified sterilization conditions within the chamber.
      • Biological Indicators (BIs): Use of biological indicators containing highly resistant bacterial spores (e.g., Geobacillus stearothermophilus) to confirm that all test microorganisms are destroyed, indicating a high probability of sterility. This represents a robust, objective measure of sterilization effectiveness.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. The Saniclave 200 is a physical device, not an AI algorithm that undergoes "training" on a dataset. Its design and operational parameters are based on established scientific principles of steam sterilization, not machine learning from a training set.

    9. How the Ground Truth for the Training Set Was Established

    • How Ground Truth for Training Set was Established: Not applicable, as there is no training set in the context of this device. The scientific and engineering principles of saturated steam sterilization (killing bacteria at 121°C for 30 minutes) are the foundational "ground truth" for the device's design, derived from decades of scientific research and validated standards like ANSI/AAMI ST55. The device is designed to execute these established principles to achieve sterility.