Search Filters

Search Results

Found 11 results

510(k) Data Aggregation

    K Number
    K121330
    Device Name
    WAVE O CAGE
    Date Cleared
    2012-06-29

    (57 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The WAVE O Cage is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via an open or a minimally invasive unilateral posterior approach. Alternatively, these implants may also be implanted via a transforaminal approach. These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.

    Device Description

    The WAVE O cage is an expandable PEEK cage of various widths and heights, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft.

    AI/ML Overview

    This document is a 510(k) summary for the WAVE O Cage, an intervertebral body fusion device. It does not describe a study involving artificial intelligence or a medical imaging device. Therefore, I cannot provide the requested information.

    The document explicitly states under section "6. CLINICAL TEST SUMMARY" that "No clinical studies were performed." and under section "5. NON-CLINICAL TEST SUMMARY" that "The results of this testing indicate that the WAVE O Cage is equivalent to predicate devices." This implies that the device's performance was established through non-clinical testing and comparison to predicate devices, not through a study with acceptance criteria and reported device performance in the way described in your request for AI-enabled devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121288
    Device Name
    FUSE CAGE
    Date Cleared
    2012-06-29

    (60 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K121332
    Device Name
    LOOP CAGE
    Date Cleared
    2012-06-29

    (57 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K121333
    Device Name
    WAVE D CAGE
    Date Cleared
    2012-06-29

    (57 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K103573
    Date Cleared
    2011-12-28

    (387 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ART2 Spinal Fixation System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

    As a pedicle screw system ART2 Spinal Fixation System is indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

    Device Description

    The ART2 Spinal Fixation System has been developed with the objective of providing the surgeon with an adaptable pedicle screw system in order to carry out posterior pedicle screw fixation of the spine simply, quickly and effectively. It is a modification of the ART Pedicle Screw system (K033150).

    The system consists of a variety of color coded top loading pedicle screws. The pedicle screws are available in various lengths and diameters. The screw is connected to the rod connector. Two sizes of connectors are available, short and long. The long is used in cases of spondylolisthesis where the short connector would not be able to engage the rod. The rods are available in multiple lengths.

    The components are manufactured from titanium alloy (Ti 6Al 4V ELI) per ASTM F136.

    AI/ML Overview

    The provided document describes the ART2 Spinal Fixation System, a pedicle screw system. The acceptance criteria and the study proving it meets these criteria are outlined in the "Summary of Nonclinical Tests" section.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria StudyAcceptance Criteria (from ASTM Standards)Reported Device Performance
    Static and dynamic compressionMet criteria laid out in ASTM F1717"The acceptance criteria was/were met." (Implies the ART2 system performed comparably or better than the predicate devices and passed the ASTM F1717 requirements)
    Static torsionMet criteria laid out in ASTM F1717"The acceptance criteria was/were met." (Implies the ART2 system performed comparably or better than the predicate devices and passed the ASTM F1717 requirements)
    Static Axial SlippageMet criteria laid out in ASTM F1798"The acceptance criteria was/were met." (Implies the ART2 system performed comparably or better than the predicate devices and passed the ASTM F1798 requirements)
    Static Torsional SlippageMet criteria laid out in ASTM F1798"The acceptance criteria was/were met." (Implies the ART2 system performed comparably or better than the predicate devices and passed the ASTM F1798 requirements)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size (number of devices or tests performed) for each specific test or the data provenance (country of origin, retrospective/prospective). It only mentions what tests were conducted and that the acceptance criteria were met. For non-clinical, mechanical testing like this, the "test set" would refer to the number of devices or components subjected to each test type. The data provenance would be from laboratory testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is a submission for a spinal fixation system, which involves mechanical safety and performance testing, not diagnostic or interpretive tasks where expert ground truth would be established from patient data. The "ground truth" here is compliance with established ASTM (American Society for Testing and Materials) standards for medical devices.

    4. Adjudication Method for the Test Set

    Not applicable. As this is mechanical testing against an engineering standard, there is no human adjudication process involved. The results are quantitative measurements compared against predefined thresholds in the ASTM standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document explicitly states: "No clinical studies were performed." Therefore, no MRMC study with human readers was conducted.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No. This is a medical device (spinal fixation system) for surgical implantation, not a software algorithm that operates without human intervention. The testing described is mechanical, material, and functional testing of the physical device.

    7. Type of Ground Truth Used

    The ground truth used for performance validation is based on established engineering and material standards, specifically:

    • ASTM F1717 (Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model)
    • ASTM F1798 (Standard Test Method for Static and Dynamic Axial Compression-Shear, Torsion, and Combined Loading Fatigue Tests of Spinal Intervertebral Body Fusion Devices)

    These standards define methods and acceptance criteria for mechanical properties of spinal implants.

    8. Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI or machine learning model. Therefore, there is no "training set" in the context of data for model development. The design and manufacturing process would involve extensive engineering and material science, but not in the way of a data-driven training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI model. The "ground truth" for the device's design and functionality would be based on biomechanical principles, material science, clinical requirements for spinal stabilization, and previous generations of similar devices (the predicate devices).

    Ask a Question

    Ask a specific question about this device

    K Number
    K100945
    Device Name
    FUSE PLIF CAFE
    Date Cleared
    2011-02-09

    (309 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FUSE Cage is indicated for intervertebral body spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). FUSE implants are to be used with autogenous bone graft and implanted via an open posterior approach. The FUSE Cage is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

    Device Description

    The FUSE is rectangular in shape and it has a honeycomb structure thru the implant in the superior/inferior direction. There is an oval opening in the M/L direction. There is a roughened surface on the superior and inferior surfaces. There are neutral, 4° and 8° lordosis configurations. The posterior end has a threaded hole for attaching insertion instruments, while the other end is solid and tapered.

    AI/ML Overview

    The provided text describes a medical device, the AMT FUSE Cage, and its 510(k) summary for FDA clearance. However, it does not contain any information about acceptance criteria, specific device performance metrics, or any study design that would "prove" the device meets such criteria in the context of AI/machine learning or diagnostic accuracy.

    The information provided is typical for a 510(k) submission for a non-active implantable medical device, focusing on substantial equivalence to a predicate device based on design, materials, function, and mechanical testing.

    Therefore, I cannot populate the requested table or answer most of the questions, as the source document does not contain this type of information.

    Here's what I can extract and explain:

    Summary of Information from the Provided Text:

    The document is a 510(k) premarket notification for the AMT FUSE Cage, a device intended for intervertebral body spinal fusion. The FDA determined the device is substantially equivalent to legally marketed predicate devices.

    Missing Information:

    The provided text does not contain any information related to:

    • Acceptance criteria for performance metrics (e.g., sensitivity, specificity, accuracy).
    • Reported device performance (in terms of diagnostic accuracy or AI performance).
    • Sample size used for a test set or training set.
    • Data provenance (country of origin, retrospective/prospective).
    • Number of experts or their qualifications for ground truth.
    • Adjudication methods.
    • Multi-reader multi-case (MRMC) comparative effectiveness studies.
    • Standalone algorithm performance studies.
    • Type of ground truth used (pathology, outcomes data, expert consensus).
    • How ground truth for a training set was established.

    This is because the device described is a physical implant (a spinal cage), not an AI/ML-driven diagnostic or image analysis tool. The "study" mentioned refers to non-clinical mechanical tests, not clinical performance or diagnostic accuracy.

    What is included (and why the requested AI/ML-related questions are not applicable):

    The document states:

    • Device Name: FUSE Cages
    • Intended Use: Intervertebral body spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. To be used with autogenous bone graft and supplemental fixation via an open posterior approach, after 6 months of non-operative treatment.
    • Material: Commercially pure titanium.
    • Function: Posterior stabilization of the lumbar spinal column with the technique of Posterior Lumbar Interbody Fusion (PLIF).
    • Comparison to Predicate: "The AMT FUSE Cage has the same indications and material, and similar designs as previously cleared devices." (Predicate device: WAVE (K080401)).
    • Nonclinical Tests: "Tests performed according to ASTM F2077/F2267 indicate that the AMT FUSE Cage meets required mechanical strengths." These are mechanical strength tests relevant to the physical stability and integrity of the implant, not performance related to AI or diagnostic accuracy.
    Ask a Question

    Ask a specific question about this device

    K Number
    K080401
    Date Cleared
    2008-05-09

    (85 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SHELL Cages are indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. SHELL Cages are used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach at the C3 to C7 disc levels using autograft bone. SHELL Cages are to be used with supplemental fixation. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.

    The WAVE/LOOP Cage is indicated for intervertebral body spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level. WAVE implants are to be used with autogenous bone graft and implanted via an open posterior approach. The WAVE Cage is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

    Device Description

    The SHELL was especially adapted to the local anatomy in order to secure the surgical result as possible. Its caudal side is flat, its cranial side is domed and the implant is formed conically from anterior to posterior. In the lateral view, the implant has a slightly lordotic form. The A/P view is trapezoidal. There are x-ray pins at the four edges, which slightly protrude the cage in order to maximize rotational fixation.

    The LOOP is a banana-shaped implant with optimized tip design and an integrated swivel joint for the instrument fixation. Both factors facilitate the implantation. The radial orientated fixation teeth define the direction of implantation and secure the implant position. LOOP is shaped with a 6° lordosis.

    The WAVE is rectangular in shape. Cross section is trapezoidal with the lateral side 1mm higher than the medial. WAVE has a neutral and a 6° lordosis. The posterior end has a threaded hole for attaching insertion instruments, while the other end is solid and tapered. The WAVE cages are implanted in pairs.

    Materials: PEEK-OPTIMA LT1 polymer (ASTM F2026 Standard Specification for Polyetheretherketone (PEEK) Polymers for Surgical Implant Applications)

    Function: The SHELL Cage was developed as an intercorporal implant for anterior cervical spondylodesis. The WAVE Cage was developed as an implant for the posterior stabilization of the lumbar spinal column with the technique of Posterior Lumbar Interbody Fusion (PLIF). The LOOP Cage was developed as an implant for the posterior stabilization of the lumbar spinal column with the technique of Transforaminal Lumbar Interbody Fusion (T-LIF).

    AI/ML Overview

    Here's an analysis of the provided information regarding the acceptance criteria and study for the AMT Interbody Fusion Devices:

    It's important to note that the provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed clinical trial report with specific acceptance criteria in the context of diagnostic performance or clinical outcomes. The "acceptance criteria" here refer to mechanical strength and similarity to predicate devices, not AI/algorithm performance.


    Description of Acceptance Criteria and Study

    The AMT Interbody Fusion Devices (SHELL, WAVE, LOOP Cages) demonstrate their suitability for market by undergoing nonclinical bench testing to demonstrate mechanical strength and by establishing substantial equivalence to previously cleared predicate devices. The "acceptance criteria" primarily revolve around meeting established mechanical standards for interbody fusion devices and demonstrating that the device's technological characteristics, indications for use, and materials are comparable to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
    Material CompliancePEEK-OPTIMA LT1 polymer (ASTM F2026 Standard Specification)The devices are made of PEEK-OPTIMA LT1 polymer, which complies with ASTM F2026.
    Mechanical StrengthASTM F2077/F2267 (Standard Test Methods for Intervertebral Body Fusion Devices)The devices "meet required mechanical strengths" as indicated by tests performed according to ASTM F2077/F2267. Specific numerical targets or results are not provided but are stated to be met.
    Indications for Use (Substantial Equivalence)Same indications as previously cleared devices.The SHELL, WAVE, and LOOP Cages have indications for use that are either identical or substantially similar to the predicate devices (e.g., DDD at specific spinal levels, use with autograft bone, supplemental fixation, non-operative treatment duration).
    Design/Technology (Substantial Equivalence)Similar designs and materials as previously cleared devices.The devices are described as having "similar designs" and the "same material" (PEEK-OPTIMA LT1) as predicate devices. Variations in geometry (e.g., SHELL, LOOP, WAVE shapes) are acknowledged but deemed acceptable in comparison to predicates.

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes a pre-market notification (510(k)) for a medical device that is a physical implant, not an AI/software device. Therefore, the concept of a "test set" and "data provenance" as typically applied to AI performance studies does not directly apply here.

    • Test Set: Not applicable in the context of AI/algorithm performance. The "test" involved ex-vivo mechanical testing of the physical implant devices. The sample size for these mechanical tests (e.g., number of devices tested for ASTM F2077/F2267) is not specified in the summary.
    • Data Provenance: Not applicable in the context of AI/algorithm performance. The "data" comes from bench testing of the physical medical devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. This device is a physical implant, and its performance is evaluated based on engineering and material standards (e.g., ASTM standards) through mechanical testing, not through expert-labeled ground truth for an algorithm.

    4. Adjudication Method for the Test Set

    Not applicable. There is no human annotation or adjudication process described for the mechanical testing of a physical implant.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. These studies are relevant for evaluating the impact of AI on human reader performance, which is not applicable to a physical interbody fusion device.

    6. Standalone (Algorithm Only Without Human-in-the Loop Performance) Study

    No, a standalone performance study (in the context of an algorithm) was not done. This document pertains to a physical medical device.

    7. Type of Ground Truth Used

    The "ground truth" for the device's technical performance is based on established engineering standards and material specifications. Specifically:

    • Mechanical Integrity: Compliance with ASTM F2077/F2267 as objective, quantitative benchmarks for intervertebral body fusion devices.
    • Material Composition: Conformance to ASTM F2026 for PEEK-OPTIMA LT1 polymer.
    • Substantial Equivalence: Comparison to the known characteristics (indications, materials, design principles) of existing, legally marketed predicate devices.

    8. Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. Since there is no training set for an AI algorithm, there is no ground truth established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K053473
    Date Cleared
    2006-02-03

    (51 days)

    Product Code
    Regulation Number
    890.5500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Maestro MDTL Laser System is intended to emit energy in the visible red & infrared spectrums to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, the temporary increase in local blood circulation and/or promoting relaxation of muscle.

    Device Description

    The Maestro MDTL Laser System is a non-invasive, portable therapeutic medical laser designed to deliver light energy to the target tissue. The System operates by AC power and can be used with a variety of laser probes.

    The Maestro MDTL Laser System is comprised of a Base Unit and select Laser Probes. The Control Unit houses the electronics, circuits and controls to power the Laser Probes. The Laser Probes are connected to the Control Unit by a cable, which plugs into the rear of the Control Unit. The Laser Probes house the laser diode and circuitry to deliver the light energy to the designated treatment areas.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Maestro MDTL Laser System, which is an infrared lamp. This type of submission relies on demonstrating substantial equivalence to predicate devices rather than conducting extensive clinical studies with specific acceptance criteria and detailed performance reporting as might be expected for novel or high-risk devices.

    Therefore, the information you've requested regarding specific acceptance criteria, a study proving the device meets them, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, or standalone algorithm performance does not apply to this 510(k) submission.

    Here's an breakdown of the available information based on your request, highlighting what is not present in this document:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Not explicitly stated as such, but inferred from 510(k) process)Reported Device Performance (as stated in the document)
    Substantial Equivalence to Predicate Devices in intended use, technical, and performance characteristics"The Maestro System and the named predicate devices have the same intended uses and similar technical and performance characteristics."
    Compliance with International Standards"ISO 9000:2000 EN46001 Directive 89/336 regarding electromagnetic compatibility"
    Compliance with Applicable Performance Standards (e.g., 21 CFR 1010 and 1040)"The System is designed to comply with applicable performance standards promulgated by Federal Food and Drug Administration, such as 21 CFR 1010 and 1040."
    No new safety or efficacy issues raised"The Maestro MDTL Laser System performs as intended and do not raise any new safety or efficacy issues."
    Functional Performance Testing"Testing of the System includes functional performance testing..."
    Electrical Safety Testing"...and electrical safety testing."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not applicable. This 510(k) submission describes functional performance testing and electrical safety testing of the device itself, not a clinical study involving a "test set" of patients or data in the sense of a diagnostic or prognostic device. The testing described is against engineering and safety standards, and comparison to predicate devices, not clinical efficacy trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable. As above, there is no "test set" requiring expert ground truth in the context of this 510(k) notification. The "ground truth" for this device's performance is compliance with electrical and functional standards and similarity to already cleared predicate devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. No clinical test set or adjudication process is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. The Maestro MDTL Laser System is a therapeutic laser for pain relief, not an AI-powered diagnostic or assistive device. Therefore, no MRMC study, AI assistance, or effect size on human readers is relevant or performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This device is hardware, a therapeutic laser, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The "ground truth" for this 510(k) relies on:
      • Engineering and Safety Standards: Compliance with ISO 9000:2000, EN46001, Directive 89/336, 21 CFR 1010, and 1040.
      • Functional Testing: The device performing as per its specifications (e.g., emitting light in the specified spectrum, maintaining power output).
      • Comparison to Predicate Devices: Demonstrating the new device has "the same intended uses and similar technical and performance characteristics" to previously cleared predicate devices.

    8. The sample size for the training set:

    • Not applicable. As this is not an AI/ML device, there is no "training set."

    9. How the ground truth for the training set was established:

    • Not applicable. As this is not an AI/ML device, there is no "training set" ground truth to establish.

    In summary, this 510(k) notification demonstrates substantial equivalence by comparing the Maestro MDTL Laser System to existing predicate devices based on intended use, performance (functional and electrical safety testing), and compliance with relevant standards. It does not involve clinical studies with patient test sets, AI algorithms, or expert ground truth adjudication in the way your questions imply for diagnostic AI systems.

    Ask a Question

    Ask a specific question about this device

    K Number
    K043353
    Date Cleared
    2005-07-01

    (207 days)

    Product Code
    Regulation Number
    890.5500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Omega Xp Laser System is indicated for use as an adjunctive treatment modality to provide temporary relief of pain associated with rotator cuff tendonitis.

    Device Description

    The Omega Xp Laser System is a non-invasive, portable therapeutic medical laser designed to deliver light energy to the target tissue. The System operates by either AC or battery power and can be used with a variety of laser probes. The Omega Xp Laser System is intended to emit energy in the infrared spectrum to provide temporary relief of pain associated with rotator cuff tendonitis.

    The Omega Xp Laser System is comprised of a Base Unit and select Laser Probes. The Base Unit houses the electronics, circuits and controls to power the Laser Probes. The Laser Probes are connected to the Base Unit by an umbilical cord. The Laser Probes house the laser diode and circuitry to deliver the light energy to the designated treatment areas.

    AI/ML Overview

    The provided text describes the Omega Xp Therapeutic Laser System, a Class II device for temporary relief of pain associated with rotator cuff tendonitis. However, it does not provide explicit acceptance criteria in a quantitative or tabular format for device performance. Instead, it relies on a determination of substantial equivalence to predicate devices and summarizes the safety and efficacy studies with general statements.

    Below is an attempt to structure the information based on your request, highlighting where specific details are missing or implied:

    1. Table of Acceptance Criteria and Reported Device Performance

    As mentioned, explicit, quantified acceptance criteria are not provided in the document. The "reported device performance" is primarily the conclusion of substantial equivalence and a statement that the device performs as intended without new safety or efficacy issues.

    Acceptance Criteria (Explicitly Stated)Reported Device Performance
    Intended Use:
    "The Omega Xp Laser System is indicated for use as an adjunctive treatment modality to provide temporary relief of pain associated with rotator cuff tendonitis."The clinical trial assessed the device's ability to provide temporary pain relief associated with rotator cuff tendonitis. Outcome was deemed sufficient to support the claim.
    Safety:
    No new safety issues raised.The device performs as intended and does not raise any new safety issues.
    Efficacy:
    No new efficacy issues raised.The device performs as intended and does not raise any new efficacy issues.
    Technical Characteristics:
    Similar to predicate devices (Photo Thera, Inc. Acculaser Pro 4 (K023060) and Stargate International, Inc. Excalibur Light Therapy System (K041530)).The Omega System and the named predicate devices have the same intended uses and similar technical and performance characteristics.
    Compliance with Standards:
    • ISO 9000:2000
    • EN46001
    • Directive 89/336 (electromagnetic compatibility)
    • 21 CFR 1010 and 1040 (applicable performance standards) | The device is manufactured to comply with these international standards and applicable FDA performance standards. |

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated in the provided text. The document only mentions "the randomized, blinded clinical trial conducted."
    • Data Provenance: Not explicitly stated. The documentation is from a US FDA submission, but the origin of the clinical trial data (e.g., country) is not specified. It is a retrospective summary for a 510(k) submission, confirming the clinical trial was completed prior to submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not provided in the given text. For a device like a therapeutic laser, "ground truth" often refers to patient-reported pain scores or functional outcomes, not necessarily expert consensus on images.

    4. Adjudication Method for the Test Set

    This information is not provided in the given text.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? This type of study (MRMC) is typically relevant for diagnostic imaging devices where different readers interpret cases. For a therapeutic device like the Omega Xp Laser System, a "multi-reader multi-case" study in the traditional sense is not applicable or mentioned. The study was a "randomized, blinded clinical trial," which compares the device's effect to a control (e.g., sham treatment or standard care).
    • Effect size of human readers with/without AI assistance: This is not applicable as the device is a therapeutic laser, not an AI-assisted diagnostic tool for human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Was a standalone study done? This concept is also not applicable as the Omega Xp Laser System is a physical therapeutic device operated by a human, not an AI algorithm. Its "performance" is its ability to deliver light energy and the resulting clinical outcome in patients, which is always with human involvement for operation and patient interaction.

    7. Type of Ground Truth Used

    For a therapeutic device, the "ground truth" for efficacy is typically clinical outcomes data, such as:

    • Patient-reported pain reduction (e.g., using a Visual Analog Scale - VAS or Numeric Rating Scale - NRS).
    • Improvement in functional ability.
    • Other subjective or objective measures of disease activity related to rotator cuff tendonitis.

    The text states "Further evidence of the safety and efficacy of the Omega System was assessed by the randomized, blinded clinical trial conducted," implying that clinical outcomes were the "ground truth" for efficacy.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: This information is not provided in the text. For a traditional medical device (non-AI), there isn't typically a "training set" in the machine learning sense. The clinical trial data would be the primary evidence for efficacy.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: This is not applicable in the context of this device. The efficacy of the device was established through a "randomized, blinded clinical trial," where the "ground truth" for participant outcomes would have been gathered during the trial (e.g., pain scores, functional assessments, etc.).
    Ask a Question

    Ask a specific question about this device

    K Number
    K033148
    Date Cleared
    2004-04-12

    (195 days)

    Product Code
    Regulation Number
    888.3060
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ART Anterior Spinal System is intended for anterolateral screw fixation of the T6-L5 spine.

    Indications for use include:

    • degenerative disc disease (ddd) defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
    • spondylolisthesis
    • trauma (i.e., fracture or dislocation) .
    • spinal stenosis .
    • deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis) .
    • tumor .
    • pscudoarthrosis .
    • failed previous fusion .
    Device Description

    The ART Anterior Spinal System consists of a Plate that attaches to the anterior or antero-lateral vertebral body of the thoraco-lumbar and lumbar spine. One place on each vertebral body. The single size plate is attached to the vertebral body with two diverging Ø6.5mm Screws. The superior portion of the plate is U-shaned to receive a Ø6mm Rod. The inner surface of the U-shaped opening of the plate is threaded to receive a Locking Nut. When tightened the Locking Nut firmly secures the Rod to the Plate. In this manner the Rods connect two or more Plates. Sce schematic in Exhibit III.

    A rod connector component, called a Domino, attaches two rods for use in multi-level fusions. Rods can be connected in series or parallel.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the "ART Anterior Spinal System." It is a spinal intervertebral body fixation orthosis. However, the document does not contain any acceptance criteria, reported device performance data, details of a study on acceptance criteria, AI components, or human reader effectiveness studies.

    The information provided focuses on:

    • Device Description: The physical components of the ART Anterior Spinal System.
    • Intended Use/Indications for Use: The medical conditions for which the device is designed (e.g., degenerative disc disease, spondylolisthesis, trauma).
    • Equivalency: Stating that the device is similar to previously cleared devices (Kaneda SR Anterior Spinal System and M-2 Anterior Plate).
    • Nonclinical Tests: A brief mention that "Testing was performed according to ASTM F1717." This refers to a standard for spinal implant testing, but no specific results or acceptance criteria from these tests are detailed.
    • Regulatory Information: Communication from the FDA confirming substantial equivalence to predicate devices, classification, and general regulatory requirements.

    Therefore, requested information about acceptance criteria, device performance, study details (sample sizes, provenance, ground truth, experts, adjudication, MRMC, standalone), and AI components cannot be extracted from this document, as it is not present.

    To summarize, for the given input:

    1. A table of acceptance criteria and the reported device performance:

    • Not Available. The document states "Testing was performed according to ASTM F1717" but does not provide specific acceptance criteria or the results of these tests.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not Available. No information on specific test sets, sample sizes, or data provenance is provided. The testing mentioned (ASTM F1717) is for mechanical properties, not clinical data sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not Available. This information is for clinical studies, which are not described in this document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Available. This information is for clinical studies, which are not described in this document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This document describes a physical medical device (spinal fixation system), not an AI-enabled diagnostic or therapeutic tool for which MRMC studies are typically performed. There is no mention of AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not Applicable. This document describes a physical medical device, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    • Not Available. This information is for clinical studies, which are not described in this document. The "ground truth" for a mechanical device would be its physical properties meeting engineering standards.

    8. The sample size for the training set:

    • Not Applicable. No training set is mentioned as this is a physical medical device clearance, not an AI model.

    9. How the ground truth for the training set was established:

    • Not Applicable. No training set is mentioned.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2