Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Powder Free Latex Patient Examination Gloves, with Protein Content labeling Claim (Contains 50 Micrograms per dm- of Glove or Less of Water Extractable Protein), meets all of the requirements of ASTM D3578:05-2010

    AI/ML Overview

    The provided text describes the acceptance criteria and the performance of an A1 Brand Powder Free Patient Examination Glove. This document is a 510(k) summary for a medical device (patient examination glove), which is often supported by non-clinical performance data rather than extensive clinical studies as seen with AI/software as a medical device (SaMD). Therefore, many of the requested points, particularly those related to AI algorithm performance, ground truth establishment, expert adjudication, and multi-reader multi-case studies, are not applicable to this type of device.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and the Reported Device Performance

    CHARACTERISTICSSTANDARDSSPECIFICATION (Acceptance Criteria)DEVICE PERFORMANCE/ACTUAL SPECIFICATION (Reported Performance)
    DimensionsASTM D3578:05-2010Thickness Palm: min 0.08mm Finger: min 0.08mmMeet. Palm Thickness: 0.09 mm Finger Thickness: 0.11 mm
    Length (by size) S: min 220 M: min 220 L: min 230Length (by size) S: 240 mm M: 240 mm L: 240 mm
    Physical PropertiesASTM D3578:05-2010
    ASTM D412:06
    ASTM D573:04-2010Tensile Strength (MPa) Before aging: Min 18 MPa After aging: Min 14 MPaMeet.
    Tensile Strength (MPa) Before aging: 22.91 MPa After Aging: 20.98 MPa
    Elongation at Break (%) Before aging: Min 650mm After aging: Min 500mmElongation at Break (%) Before aging: 850 mm After aging: 820 mm
    Freedom from pin-holesASTM D5151:06-2011AQL 1.5Meet
    Powder Free ResidueASTM D6124:06-2011Below 2mg/gloveMeet
    Protein ContentASTM D5712:10-2010Below 50ug/dm²Meet
    BiocompatibilityAnimal Irritation Test (ISO 10993-10:2010)Under the conditions tested, the subject glove was not an irritant.Under the conditions tested, the subject glove was not an irritant.
    Dermal Sensitization (ISO 10993-10:2010)Under the condition tested, the subject glove was not a sensitizer.Under the condition tested, the subject glove was not a sensitizer.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes (e.g., number of gloves tested for each characteristic) used for the non-clinical tests. It refers to compliance with ASTM and ISO standards, which typically include sampling plans. The data provenance is Malaysia, where A1 Glove Sdn. Bhd. is located. These are non-clinical (laboratory/bench) tests, not retrospective or prospective human subject data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is not an AI/SaMD and does not use "ground truth" derived from expert consensus on medical images or diagnoses. The "ground truth" for glove performance is defined by the objective measurements and chemical analyses specified by the ASTM and ISO standards.

    4. Adjudication method for the test set

    Not applicable. There is no adjudication method described as this is not a study involving human interpretation or subjective assessment of medical data. The tests are objective measurements against defined standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI device.

    7. The type of ground truth used

    The "ground truth" for this device's performance is based on established international standards and objective laboratory measurements (e.g., precise measurements of thickness, tensile strength, chemical residue, and biological reactivity according to ASTM and ISO protocols). There is no pathology, outcomes data, or expert consensus used in the typical sense for medical imaging or diagnostic algorithms.

    8. The sample size for the training set

    Not applicable. This is not an AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1