Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K243821
    Device Name
    i-Cut
    Date Cleared
    2025-04-25

    (134 days)

    Product Code
    Regulation Number
    884.1720
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    A.M.I. Agency for Medical Innovations GmbH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The i-Cut system is indicated for morcellating and extracting tissue in laparoscopic, gynecologic surgical procedures.

    Device Description

    The i-Cut is a single use laparoscopic power morcellator. It is provided sterile.

    The i-Cut consists of a rotating cutting tube with a trocar that provides protection against the cutting blade when locked in the "CLOSED" position. The device comes with an obturator for placement into the patient's body and a locking clip that keeps the trocar in the "closed" position during insertion. The activation button must be pressed to cut tissue.

    The lumen of the device is designed for use with a standard grasper or Tenaculum forceps with a diameter between 10 to 14 mm. The i-Cut is designed to be used with surgical instruments of diameters between 10 to 14 mm. The lumen is fitted with a silicone valve to prevent gas loss during use of the device.

    The i-Cut is electrically operated by a DC Motor which is powered by a 24V AC/DC mains adapter. The non-sterile i-Cut Power Supply with power cord is supplied separately to power the device.

    AI/ML Overview

    This 510(k) clearance letter pertains to the i-Cut, a single-use laparoscopic power morcellator. The provided text details regulatory information, device description, indications for use, and non-clinical performance testing. However, it does not include information about clinical studies with human participants, acceptance criteria tables, sample sizes for test sets (in the context of clinical performance), expert ground truth establishment, or multi-reader multi-case (MRMC) comparative effectiveness studies.

    The document primarily focuses on non-clinical bench testing to demonstrate substantial equivalence to a predicate device (LiNA eXcise, K101458). Therefore, I cannot address most of the requested points related to clinical study acceptance criteria and performance as this information is not present in the provided text.

    Here's an analysis of what can be extracted from the provided text, and where gaps exist:


    Preamble:

    The provided 510(k) clearance letter for the i-Cut device focuses on demonstrating substantial equivalence through non-clinical performance testing (bench testing) rather than clinical trials with human participants. Therefore, many of the requested details pertaining to clinical study design, human reader performance, and expert consensus for ground truth are not present in this document. The "acceptance criteria" and "device performance" discussed here apply to the bench testing conducted, not clinical outcomes in human subjects.


    1. A table of acceptance criteria and the reported device performance

    Based on the provided non-clinical performance testing section, we can infer some "acceptance criteria" as implied by the successful completion of each test. The "reported device performance" is generally stated as "passed" or "met requirements." No numerical data for specific criteria are provided beyond the stated specifications (e.g., cutting rate, speed).

    Acceptance Criteria (Inferred from Test Purpose)Reported Device Performance
    Sterilization Validation (ANSI/AAMI/ISO 11135:2014)Passed sterilization validation
    Shelf-life Validation (ASTM F1980-21:2023, etc.)Supports 5-year shelf life (package integrity and device performance)
    Simulated Shipping (ASTM D4332-22:2022, etc.)Passed simulated shipping tests
    Biocompatibility (ISO 10993-5, -10, -11, -23)Passed biocompatibility tests (Cytotoxicity, Sensitization, Intracutaneous irritation, Acute systemic toxicity)
    Electrical Safety (IEC 60601-1:2020)Passed electrical safety
    Electromagnetic Compatibility (EMC) (IEC 60601-1-2:2020)Passed EMC
    Load Test (Withstand expected load without failure)Passed load test
    Gas Tightness Test (Maintain pneumoperitoneum)Passed gas tightness test
    Obturator Abrasion Test (No visible wear marks)Passed obturator abrasion test (no abrasion occurs)
    Liquid Tightness Test (Meet requirements after load test)Passed liquid tightness test
    Grip of the Housing Surface Test (Surgeon's ability to properly grip)Passed grip of the housing surface test
    Surface Reflection Test (Cutting blade not cause reflections impeding visibility)Passed surface reflection test
    Trocar Sleeve – Pull-off Force Test (Connection suitable for application)Passed pull-off force test
    Functional Test (Morcellate and extract tissue in model specimen)Passed functional test
    Insertion Test (Force needed to expose cutting blade in "Closed" position)Passed insertion test
    Usability testing (Compatibility with containment system, integrity assessment)Passed usability testing
    Cutting Rate (Stated Specification)>21.5 g/min
    Speed (Stated Specification)1100 rpm

    2. Sample sizes used for the test set and the data provenance

    • Sample Size: The document does not specify the sample sizes (e.g., number of devices, number of tests) for each of the non-clinical performance tests. It only states that the tests were performed and passed.
    • Data Provenance: The data provenance is from non-clinical bench testing conducted by the manufacturer, A.M.I. Agency for Medical Innovations GmbH. The country of origin of the manufacturer is Austria. This data is inherently prospective as it involves new testing for the device's clearance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This was a non-clinical bench testing study. "Ground truth" in this context pertains to engineering specifications and performance metrics rather than clinical diagnoses established by medical experts. Usability testing involved "surgeons of different experience with laparoscopic morcellators," but it was to assess device compatibility and integrity, not to establish a clinical ground truth for a diagnostic AI.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This was a non-clinical bench testing study. Adjudication methods are typically used in clinical studies where multiple human readers interpret data to establish a consensus ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. No MRMC comparative effectiveness study was performed or described. The clearance is based on substantial equivalence demonstrated through non-clinical bench testing. The device is a physical surgical tool, not an AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical surgical tool and does not employ a standalone algorithm for diagnostic or interpretative purposes.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For non-clinical bench testing, the "ground truth" is defined by engineering specifications, regulatory standards (e.g., ISO, ASTM, IEC), and predefined pass/fail criteria for device performance (e.g., measuring force, checking for wear, verifying electrical safety compliance). There is no "expert consensus" or "pathology" in the medical sense for these tests.

    8. The sample size for the training set

    Not applicable. This device is a mechanical/electrical surgical tool, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1