IMMUNALYSIS COCAINE/COCAINE METABOLITE ELISA FOR ORAL FLUIDS

K051578 · Immunalysis Corporation · DIO · Nov 18, 2005 · Clinical Toxicology

Device Facts

Record IDK051578
Device NameIMMUNALYSIS COCAINE/COCAINE METABOLITE ELISA FOR ORAL FLUIDS
ApplicantImmunalysis Corporation
Product CodeDIO · Clinical Toxicology
Decision DateNov 18, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3250
Device ClassClass 2

Indications for Use

The Immunalysis Cocaine/Cocaine Metabolite ELISA test system utilizes an Enzyme Linked Immunoassay (ELISA) for the qualitative detection of Cocaine/Cocaine Metabolite at a cutoff of 20 ng/mL in ORAL FLUID SAMPLES COLLECTED WITH THE QUANTISAL™ ORAL FLUID COLLECTION DEVICE ONLY. This in vitro diagnostic device is intended for clinical laboratory use only. The Immunalysis Cocaine/Cocaine Metabolite ELISA Kit for Oral Fluids provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/ mass spectrometry (GS-MS) is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Device Story

In-vitro diagnostic ELISA test system; detects cocaine/metabolites in oral fluid collected via Quantisal device. Used in clinical laboratories by trained personnel. Provides preliminary qualitative results; requires confirmation by GC-MS. Assists healthcare providers in identifying drug presence; supports clinical decision-making regarding substance abuse. Benefits include rapid screening of oral fluid samples.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Enzyme Linked Immunoassay (ELISA) technology. Qualitative detection at 20 ng/mL cutoff. Requires Quantisal oral fluid collection device. In-vitro diagnostic format.

Indications for Use

Indicated for qualitative detection of cocaine/cocaine metabolite in oral fluid samples at 20 ng/mL cutoff. Intended for clinical laboratory use for individuals suspected of cocaine use.

Regulatory Classification

Identification

A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.

Special Controls

*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k051578 B. Purpose for Submission: New device C. Measurand: Cocaine/Cocaine Metabolite D. Type of Test: Qualitative immunoassay E. Applicant: Immunalysis Corporation F. Proprietary and Established Names: Immunalysis Cocaine/Cocaine Metabolite ELISA for Oral Fluids G. Regulatory Information: 1. Regulation section: 21 CFR §862.3250, Cocaine/Cocaine Metabolite Test System 2. Classification: Class II 3. Product code: DIO 4. Panel: Toxicology (91) H. Intended Use: 1. Intended use(s): See indications for use below. 2. Indication(s) for use: "The Immunalysis Cocaine/Cocaine Metabolite ELISA test system utilizes an Enzyme Linked Immunoassay (ELISA) for the qualitative detection of Cocaine/Cocaine Metabolite at a cutoff of 20 ng/mL in ORAL FLUID SAMPLES COLLECTED WITH THE QUANTISAL™ ORAL FLUID COLLECTION DEVICE ONLY. This in vitro diagnostic device is intended for clinical laboratory use only. The Immunalysis Cocaine/Cocaine Metabolite ELISA Kit for Oral Fluids provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/ mass spectrometry (GS-MS) is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used." 3. Special conditions for use statement(s): See above. {1} 4. Special instrument requirements: This device must be read on a spectrophotometer that reads wavelengths of 450 nm and 620 nm. I. Device Description: The device consists of a saliva collection device and a Cocaine/Cocaine Metabolite ELISA kit. An oral fluid specimen is collected by placing the collection device, a cellulose pad affixed to a propylene stem, under the tongue until approximately one milliliter saliva has saturated the pad. A blue indicator on the stem indicates when enough sample has been collected. The collector is transferred to a provided polypropylene tube containing preservative buffer (3 ml) and closed, ready for transport or storage. The ELISA assay is consists of 8-well microstrips coated with a sheep anti-Benzoylecgonine and a monoclonal anti-Cocaine antibody, a plate frame, conjugated Cocaine, negative and positive controls, a cut-off calibrator, TMB substrate, and stop reagent. J. Substantial Equivalence Information: 1. Predicate device name(s): DRI Cocaine EIA Assay 2. Predicate 510(k) number(s): K960187 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Analyte | Cocaine and Benzoylecgonine | Cocaine and Benzoylecgonine | | Methodology | Immunoassay (EIA) | Immunoassay (ELISA) | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Test Matrix | Oral Fluid | Urine | | Cutoff | 20 ng/mL | 300 ng/mL | K. Standard/Guidance Document Referenced (if applicable): None referenced. L. Test Principle: Enzyme-labeled drug and drug present in the sample compete for limited anti-cocaine antibody binding sites. Binding of the enzyme-labeled drug inhibits its reaction with the substrate, thereby influencing the rate of absorbance change measured by the instrument. The rate of absorbance change is proportional to the concentration of drug in the sample. Concentrations of controls and unknowns are calculated from the standard curve. Results are read at 450nm and 620 nm. {2} M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Precision was tested by spiking negative oral fluid with 0, 10, 20, 30, and 40 ng/mL of Benzoylecgonine; this corresponded to 0, 50%, 100%, 150%, and 200% of the cutoff. One milliliter of the spiked fluids were pipetted onto the collection pad of the oral fluid collector then processed as per instructions. Intra-assay precision was assessed with sixteen replicates of each concentration analyzed in one run: Intra-assay Precision: Immunalysis Cocaine/Cocaine Metabolite for Oral Fluid Assay | Benzoylecgonine concentration (ng/mL) | Mean O.D. | S.D. | C.V.% | B/B0 % | | --- | --- | --- | --- | --- | | 0 | 2.4351 | 0.150 | 6.17 | 100 | | 10 (50% of cutoff) | 1.3943 | 0.108 | 7.74 | 57.26 | | 20 (100% of cutoff) | 1.2644 | 0.117 | 9.28 | 51.93 | | 30 (150% of cutoff) | 1.1456 | 0.036 | 3.16 | 47.04 | | 40 (200% of cutoff) | 1.0475 | 0.059 | 5.63 | 43.02 | B= Absorbance of the mean concentration B0= Absorbance of the mean zero dose Inter-Assay Precision Negative oral fluid was spiked with 0, 10, 20, 30 and 40 ng/mL of Benzoylecgonine. This corresponded to 0, 50%, 100%, 150% and 200% of the recommended screening cutoff for Benzoylecgonine in oral fluids. One mL of these different levels of Benzoylecgonine spiked oral fluids were pipetted on to the collection pad of the Quantisal Oral Fluid collector and the pad then placed in the transport tube with the preservative buffer. Eight replicates of the five different Benzoylecgonine concentrations were analyzed in each run. This was repeated in 10 different assay runs, twice per day over a period of 5 working days. | Benzoylecgonine | 10 ng/mL B/B0 % | 20 ng/mL B/B0 % | 30 ng/mL B/B0 % | 40 ng/mL B/B0 % | | --- | --- | --- | --- | --- | | MEAN | 58.93 | 48.47 | 43.18 | 39.74 | | S.D. | 3.253 | 3.190 | 2.98 | 3.168 | | C.V. % | 5.52 | 6.58 | 6.90 | 7.97 | {3} B= Absorbance of the micro-plate well for the particular concentration B0= Absorbance of the zero calibrator The B/B0 % is a dimensionless number that is independent of the Raw absorbance Reproducibility of the oral fluid collection device was assessed by collecting oral fluid from 50 subjects with a pre-weighed collector and tube as per the package instructions. After the volume indicator turned blue, the collector and tube were weighed and the net weight of the saliva was determined and converted to volume (mLs). Quantisal Oral Fluid Collection Device: Volume Adequacy Study | Avg. Vol. (mL) | Std. Dev. | C.V. | Mean + 3 SD (mL) | Mean – 3 SD (mL) | | --- | --- | --- | --- | --- | | 0.993 | 0.029 | 2.88% | 1.079 | 0.907 | These results support the sponsor's claim that the device collects 1 mL ± 10% saliva. b. Linearity/assay reportable range: Not applicable. This assay is intended for qualitative use. c. Traceability, Stability, Expected values (controls, calibrators, or methods): All calibrator and control stock solutions are prepared from commercially available DEA-exempt solutions; Benzoylecgonine concentration is confirmed by GC/MS analysis. Claimed shelf-life of the oral fluid collection device and the ELISA kit and components is 24 months and 12 months respectively. Real-time studies are ongoing. Stability of benzoylecgonine in the collection device was determined by spiking a pool of negative saliva with benzoylecgonine at a concentration around the cutoff and a pool at a concentration two-fold higher (2X). Samples were stored at 4°C or at room temperature. The specimens kept at room temperature were assayed in duplicate by GC-MS after 7 days, 14 days and 30 days; samples kept at 4°C were assayed were in duplicate by GC-MS after 14 days and 30 days. The sponsor's acceptance criterion was recovery of ± 20% of the initial value. {4} 5 # Stability of Benzoylecgonine in Quantisal Buffer ## STABILITY STUDY AT ROOM TEMPERATURE | Data point | BE Spike ng/mL | % of initial value | BE Spike 2X ng/mL | % of initial value | | --- | --- | --- | --- | --- | | 0 | 13.659 | 100 | 30.213 | 100 | | 7 | 15.555 | 113.88 | 35.089 | 116.14 | | 14 | 15.332 | 112.25 | 30.102 | 99.63 | | 21 | 15.105 | 110.59 | 30.271 | 100.19 | | 30 | 14.221 | 104.11 | 28.785 | 95.27 | ## STABILITY STUDY AT 4°C | Data point | BE Spike ng/mL | % of initial value | BE Spike 2X ng/mL | % of initial value | | --- | --- | --- | --- | --- | | 0 | 13.659 | | 30.213 | | | 14 | 13.078 | 95.75 | 33.162 | 109.76 | | 30 | 17.367 | 127.15 | 28.088 | 92.97 | A shipping study showed that benzoylecgonine spiked into saliva had acceptable recovery (± 15%) after transport. ### d. Detection limit: See the Precision/Reproducibility section above for performance around the stated cutoff concentration. ### e. Analytical specificity: Cross-reactivity of structurally similar compounds was determined by spiking concentrations of different drugs into synthetic oral fluid. | Compound | Concentration ng/mL | B/B0 % | BE equivalents ng/mL | Percent (%) Cross-reactivity | | --- | --- | --- | --- | --- | | Benzoylecgonine | 10 | 64.6 | 10 | 100 | | Benzoylecgonine | 20 | 47.06 | 20 | 100 | | Benzoylecgonine | 40 | 35.95 | 40 | 100 | | Cocaine | 20 | 50.88 | 17.26 | 86.3 | | Cocaine | 40 | 43.38 | 25.17 | 62.92 | | Cocaethylene | 20 | 48.2 | 19.62 | 98.1. | | Ecgonine | 2000 | 32.54 | 49.68 | 2.5 | | Ecgonine Methyl Ester | 2000 | 93.88 | 4.05 | 0.2 | | Norcocaine | 1000 | 35.11 | 41.51 | 4.1 | {5} Structurally unrelated compounds were spiked into synthetic oral fluid at a concentration of 10000 ng/mL; none of the compounds in the table below had an immunoassay response greater than the 50% control of the assay (i.e. Benzoylecgonine at 10 ng/mL). ## Compounds tested for Cross-reactivity: Immunalysis Cocaine/Cocaine metabolite Oral Fluid Assay | Acetaminophen | Ethylmorphine | Mereridine | | --- | --- | --- | | Amitriptyline | Flurazepam | Nalorphine | | Amobarbital | Glutethimide | Nicotine | | Barbital | Hexobarbital | Nordoxepin | | Benzoylecgonine | Hydromorphone | n-Normethsuximide | | Butabarbital | Imipramine | Nortriptyline | | Bromazepam | Lidocaine | Oxazepam | | Caffeine | Lorazepam | Oxycodone | | Carbamazepine | Medazepam | Phenobarbital | | Cocaine | Methadone | Phensuximide | | Codeine | EDDP | Phenytoin | | Chlorpromazine | Methaqualone | Primidone | | Desipramine | Metharbital | Protriptyline | | Diacetylmorphine | Mephenytoin | Quinine | | Dihydrocarbamazepine | Methyl-propylsuccinimide | Secobarbital | | Diazepam | Mephobarbital | Temazepam | | Doxepin | Methyl PEMA | Theophylline | | Dyphylline | Methsuximide | Trimipramine | | Ethosuximide | 4-Methylprimidone | | | Ethotoin | Morphine | | Commonly ingested substances were tested for interference. The effects of sugar, toothpaste, cranberry juice, baking soda, cola carbonated beverages, cough syrup, mouthwash and distilled water were studied. Each of these diluted commonly ingested substances was spiked in separate aliquots with 10 ng/mL of Benzoylecgonine (50% of the cutoff) and 30 ng/mL Benzoylecgonine (150% of the cutoff). Quantisal oral fluid collection devices were used with each of the spiked commonly ingested substances and the immunoassay results obtained utilizing the spikes with the above commonly ingested substances were compared to spikes of Benzoylecgonine in synthetic negative oral fluid. If a spike at 10 ng/mL of Benzoylecgonine (50% of the cutoff calibrator) remained negative with respect to the 20 ng/mL Benzoylecgonine calibrator in synthetic negative oral fluid it was deemed that there was no interference. If a spike at 30 ng/mL of Benzoylecgonine (150% of the cutoff calibrator) remained positive with respect to the 20 ng/mL of Benzoylecgonine calibrator in synthetic negative oral fluid, it was deemed that there was no interference. 6 {6} | Specimen with Benzoylecgonine concentration | Mean Absorbance | B/B0% | POS/NEG | Effect | | --- | --- | --- | --- | --- | | Zero | 2.486 | 100 | | | | 10 ng/mL in Synthetic Oral Fluid | 1.709 | 68.76 | NEG | | | 20 ng/mL in Synthetic Oral Fluid | 1.348 | 54.22 | Cutoff | | | 40 ng/mL in Synthetic Oral Fluid | 1.062 | 42.724 | POS | | | 10ng/mL in sugar water | 1.474 | 59.3 | NEG | | | 10 ng/mL in toothpaste slurry | 1.41 | 56.72 | NEG | | | 10 ng/mL in cranberry juice | 1.639 | 65.96 | NEG | | | 10 ng/mL in baking soda solution | 1.842 | 74.10 | NEG | | | 10 ng/mL in Orange juice | 1.539 | 61.91 | NEG | | | 10 ng/mL in Cola beverage | 1.594 | 64.13 | NEG | | | 10 ng/mL in Cough syrup | 1.735 | 69.79 | NEG | | | 10 ng/mL in mouthwash | 1.725 | 69.39 | NEG | | | 10 ng/mL in Distilled Water | 1.743 | 70.12 | NEG | | | 30 ng/mL in sugar water | 1.075 | 43.23 | POS | | | 30 ng/mL in toothpaste slurry | 1.249 | 50.27 | POS | | | 30 ng/mL in cranberry juice | 1.189 | 47.81 | POS | | | 30 ng/mL in baking soda solution | 1.355 | 54.49 | NEG | Negative Effect | | 30 ng/mL in Orange juice | 1.121 | 45.08 | POS | | | 30 ng/mL in Cola beverage | 1.163 | 46.79 | POS | | | 30 ng/mL in Cough syrup | 1.107 | 44.51 | POS | | | 30 ng/mL in mouthwash | 1.185 | 47.67 | POS | | | 30 ng/mL in Distilled Water | 1.153 | 46.36 | POS | | From the above data, only the 30 ng/mL Benzoylecgonine spike in the 25mg/mL baking soda dissolved in Distilled Water had an effect on the immunoassay. Under these conditions, a sample spiked at 150% of the cutoff (i.e. 30 ng/mL of Benzoylecgonine) read as a borderline negative, roughly equivalent to a 20 ng/mL Benzoylecgonine cutoff. None of the other substances tested under these conditions had an effect on the immunoassay. f. Assay cut-off: Performance around the assay cut-off of 20 ng/mL is demonstrated in the intra-assay precision section above. The Substance Abuse and Mental Health Services Administration (SAMHSA) has recommended 20 ng/mL as a cutoff level for cocaine and metabolites oral fluid tests. 2. Comparison studies: a. Method comparison with predicate device: Oral fluid and urine samples were collected in the same visit from 100 admitted cocaine users in a clinical setting. Urine samples were tested by the predicate assay using a cutoff of 150 ng/mL. Oral fluid samples were tested in duplicate using a screening cutoff of 20 ng/mL; all samples were tested by GC/MS at an independent facility. {7} 8 # Comparison of Immunalysis Cocaine/Cocaine metabolite Oral Fluid Assay and the Predicate Urine Assay | | Predicate Urine Assay | | | | --- | --- | --- | --- | | | | Pos | Neg | | Cocaine/Cocaine metabolite Oral Fluid Assay | Pos | 53 | 4 | | | Neg | 4 | 39 | Positive agreement: 93 % Negative agreement: 91 % Overall agreement: 92 % # Comparison of Immunalysis Cocaine/cocaine metabolite Oral Fluid Assay and GC/MS | | GC/MS | | | | --- | --- | --- | --- | | | | Pos | Neg | | Cocaine/Cocaine metabolite Oral Fluid Assay | Pos | 50 | 5 | | | Neg | 2 | 43 | Positive agreement: 91% Negative agreement: 96 % Overall agreement: 93 % b. Matrix comparison: Not applicable; this device is intended for use with oral fluid only. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: Not applicable. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...