K123891 · Spartan Bioscience, Inc. · NTI · Aug 12, 2013 · Clinical Toxicology
Device Facts
Record ID
K123891
Device Name
SPARTAN RX CYP2C19 TEST SYSTEM
Applicant
Spartan Bioscience, Inc.
Product Code
NTI · Clinical Toxicology
Decision Date
Aug 12, 2013
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 862.3360
Device Class
Class 2
Indications for Use
The Spartan RX CYP2C19 System is a qualitative in vitro diagnostic test for the identification of a patient's CYP2C19 *2, *3 and *17 genotype determined from genomic DNA obtained from a buccal swab sample. For prescription use only.
Device Story
Sample-to-result DNA testing system; uses buccal swab input. System integrates DNA extraction, PCR amplification, and fluorescent probe detection of CYP2C19 *2, *3, and *17 alleles. Operator collects buccal sample, inserts swab into reagent tube, and loads into Spartan RX Analyzer. Analyzer performs thermal cycling and automated genotype calling. Results printed via connected printer. Used in clinical settings to aid therapeutic strategy determination. System includes integrated controls for run monitoring and anomaly detection. Benefits include rapid, automated genotyping without manual DNA purification.
Clinical Evidence
Bench testing only. Reproducibility study (n=960 tests) across three sites showed 99.8% correct call rate after second-pass testing. Method comparison against bi-directional sequencing (n=325 samples) showed 100% agreement after second-pass testing. Analytical specificity evaluated against exogenous substances (mouthwash, tobacco, etc.) and endogenous substances (blood, bacteria) with acceptable performance.
Technological Characteristics
PCR-based genotyping system. Components: Spartan RX Analyzer (thermal cycler), Netbook (UI), Printer, Barcode Scanner. Consumables: Sample Collection Kits (SCK) and External Control Kits (ECK) with reagent tubes. Connectivity: Netbook interface. Sterilization: Not specified. Software: Automated genotype calling algorithm.
Indications for Use
Indicated for use as an aid to clinicians in determining therapeutic strategies for therapeutics metabolized by the Cytochrome P450 2C19 gene product, specifically affected by *2, *3, and *17 alleles. Not indicated to predict drug response or non-response.
Regulatory Classification
Identification
A drug metabolizing enzyme genotyping system is a device intended for use in testing deoxyribonucleic acid (DNA) extracted from clinical samples to identify the presence or absence of human genotypic markers encoding a drug metabolizing enzyme. This device is used as an aid in determining treatment choice and individualizing treatment dose for therapeutics that are metabolized primarily by the specific enzyme about which the system provides genotypic information.
Special Controls
The special control is FDA's guidance document entitled "Class II Special Controls Guidance Document: Drug Metabolizing Enzyme Genotyping System."
*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Drug Metabolizing Enzyme Genotyping Test System.” See § 862.1(d) for the availability of this guidance document.
K220026 — Genomadix Cube CYP2C19 System · Genomadix, Inc. · Mar 21, 2023
K120466 — VERIGENE CYP2C19 NUCLEIC ACID TES (2C19) · Nanosphere, Inc. · Nov 6, 2012
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k123891
B. Purpose for Submission:
New device
C. Measurand:
Genotype of Cytochrome P450 2C19 (CYP2C19)
D. Type of Test:
Genotyping microarray
E. Applicant:
Spartan Bioscience Inc.
F. Proprietary and Established Names:
Spartan RX CYP2C19 Test System
G. Regulatory Information:
1. Regulation section:
21 CFR 862.3360; Drug Metabolizing Enzyme Genotyping Systems
21 CFR 862.2570; Instrumentation for Clinical Multiplex Test Systems
2. Classification:
Class II
3. Product code:
NTI; Drug metabolizing enzyme genotyping systems
NSU; Instrumentation for Clinical Multiplex Test Systems
{1}
4. Panel:
Toxicology (91)
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
See indication(s) for use below.
2. Indication(s) for use:
Assay - The Spartan RX CYP2C19 System is indicated for use as an aid to clinicians in determining therapeutic strategies for therapeutics that are metabolized by the Cytochrome P450 2C19 gene product, and that are specifically affected by the *2, *3, and *17 alleles. The Spartan RX CYP2C19 Assay will be run on the Spartan RX CYP2C19 Platform from the buccal sample collected with a buccal swab. The Spartan RX CYP2C19 Assay is not indicated to be used to predict drug response or non-response.
Platform - The Spartan RX CYP2C19 System is indicated for use as an aid to clinicians in determining therapeutic strategies for therapeutics that are metabolized by the Cytochrome P450 2C19 gene product, and that are specifically affected by the *2, *3, and *17 alleles. The Spartan RX CYP2C19 Platform will be used to run the Spartan RX CYP2C19 Assay.
3. Special conditions for use statement(s):
For prescription use only
The information provided from this test may supplement decision making and should only be used in conjunction with routine monitoring by a physician. Because of the variability in the knowledge of clinical utility with specific drugs that are metabolized by CYP2C19, clinicians should use professional judgment in the interpretation of results from this test. Results from this type of assay should not be used in predicting a patient's response to drugs for which the drug metabolizing enzyme activity of that allele, or drug metabolic pathway, has not been clearly established.
4. Special instrument requirements:
Spartan RX Platform
I. Device Description:
The Spartan RX CYP2C19 Test System is a sample-to-result DNA testing system with integrated DNA extraction and amplification. Genotypes are determined using PCR and fluorescent probe detection. The Spartan RX CYP2C19 Test System is comprised of the
2
{2}
Spartan RX Platform, Spartan transport system, and Spartan RX Assays (sample collection kit and external control kit). The Spartan RX CYP2C19 Assays are run on the Spartan RX Platform.
The Spartan RX CYP2C19 Test System uses a sample obtained with a buccal swab. The operator collects the buccal swab samples from a patient, inserts each sample into a reagent tube, places the reagent tube into the transport system until the reagent tubes are ready to be inserted into an Analyzer instrument. The test system then integrates and automates DNA extraction, PCR-based amplification of the target gene, detection of the *2, *3, and *17 alleles using fluorescent-probes v. fluorescent signal detection and analysis, and genotype determination. Results are presented to the end user as genotype calls. The system has integrated controls that monitor performance of a run and automatically inform the user of any anomalies in the instrument or reagents. The system detects the CYP2C19 *2, *3, and *17 genotypes in separate reagent tubes.
The Spartan RX CYP2C19 Test System uses the Spartan RX Platform which has four components:
1. The Spartan RX Analyzer is a thermal cycling instrument that automates polymerase chain reaction (PCR) amplification of the target gene, fluorescence-based detection of CYP2C19 alleles, and genotype calling.
2. The Netbook serves as the user interface for logging into the Spartan RX Platform, inputting patient information, and starting a test.
3. The Printer automatically prints the genotype results after the Spartan RX Platform finishes performing the Spartan RX CYP2C19 Assay.
4. The Barcode Scanner is used to automatically enter SCK and ECK lot numbers into the Spartan RX Platform.
The Spartan RX CYP2C19 Test System uses Spartan RX Assays (Sample collection kit and external control kit). The Spartan RX Assays are the consumable components of the Spartan RX CYP2C19 Test System.
Sample collection kits (SCKs) contain the consumables required to determine a patient's CYP2C19 *2, *3, or *17 genotype. There are three types of SCKs, specific for the *2, *3, or *17 allele. Each SCK consists of a pouch, buccal swab, and a reagent tube. The pouch contains labeling information to ensure traceability (lot number, manufacturing date, and expiry date) and two compartments (for the buccal swab and reagent tube). The pouch also includes a barcode that is compatible with a standard reader. The buccal swab is used to collect a buccal sample from the inside of a patient's cheek and transfer it into the reagent tube. The reagent tube contains chemicals for DNA extraction, PCR amplification, and fluorescent detection of the specific CYP2C19 allele. The SCKs are color-coded to indicate which CYP2C19 allele they are designed to detect.
3
{3}
The External control kits (ECKs) contain the consumables required to determine if the Spartan RX Platform and reagents are performing correctly. Each ECK consists of a pouch and reagent tube. The pouch contains labeling information to ensure traceability (lot number, manufacturing date, and expiry date). The reagent tube contains synthetic nucleic acid control and chemicals for PCR amplification and fluorescent detection of the specific CYP2C19 allele. The ECKs are color-coded to indicate which reagent they control for (blue = *2, white = *3, and black = *17). The ECKs are always produced from the same lot of reagents for which they are designed to control.
The Transport System is designed to keep the sample collection kits and collected samples cold as they are moved from the freezer to the patient and from the patient to the Spartan RX Analyzer. The sample transport system is an insulated transport bag that contains a polystyrene box and a Cold Block.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Infiniti CYP2C19 Assay
2. Predicate 510(k) number(s):
K101683
3. Comparison with predicate:
Assay:
| Similarities | | |
| --- | --- | --- |
| Item | Spartan RX CYP2C19 Test System | Infiniti CYP2C19 Assay k101683 (Predicate) |
| Indication(s) for Use | The Assay is indicated for use as an aid to clinicians in determining therapeutic strategies for therapeutics that are metabolized by the Cytochrome P450 2C19 gene product, and that are specifically affected by the *2, *3, and *17 alleles.. | Same |
| Target Gene | CYP4502C19 *2, *3, and *17. | Same |
| Limitations | Not intended to be used to predict drug response or non-response. | Same |
| DNA Sequencing | Detects specific DNA sequences through | Same |
| | samples of DNA sequences | |
| Assay | Not intended to be used to predict drug response or non-response. | Same |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Spartan RX CYP2C19 Test System | Infiniti CYP2C19 Assay k101683 (Predicate) |
| | recognition of DNA targets. | |
| Assay Results | Assay signal results are interpreted by a software program. Assay results are provided as genotype calls reported to the end user in a report format. | Same |
| Differences | | |
| --- | --- | --- |
| Item | Spartan RX CYP2C19 Test System | Infiniti CYP2C19 Assay k101683(Predicate) |
| Specimen Type | Buccal swab sample. | Purified DNA from EDTA-anti-coagulated whole blood sample. |
Platform:
| Similarities | | |
| --- | --- | --- |
| Item | Spartan RX CYP2C19 Test Platform | AutoGenomics INFINITI Analyzer (Predicate) |
| Indication(s) for Use | The Platform runs the assay for Cytochrome P450 2C19 gene product. | Same |
| Technology | Utilizes thermal cycling and target DNA amplification. | Same |
| Differences | | |
| --- | --- | --- |
| Item | Spartan RX CYP2C19 Test Platform | AutoGenomics INFINITI Analyzer (Predicate) |
| Description | Fluorescent probe PCR-based genotyping test for multiplex analysis of DNA sequences. | Microarray-based genotyping test for simultaneous (multiplex system) of DNA sequences. |
| Sample preparation | Automated | Off-line |
# K. Standard/Guidance Document Referenced (if applicable):
None
{5}
L. Test Principle:
The Spartan RX CYP2C19 Test System utilizes PCR amplification of target DNA and detection of the presence or absence of specific CYP2C19 alleles (*2, *3, *17) using molecular beacon probes. The Spartan RX CYP2C19 Test System makes use of a two-temperature PCR cycling program which combines the annealing and extension steps.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Site-to-Site Reproducibility
The reproducibility study was conducted at three sites (two external and one internal). The three sites are designated, site 1, site 3, and site 4. Samples used for the study were buccal scrapes collected from *1/*17, *2/*3, *1/*1, *17/*17, *1/*3, *2/*17, *2/*2, and *1/*2 individuals (one individual per genotype); samples were collected from eight individuals at each site. Genotypes of the individuals included in the study were confirmed by bi-directional DNA sequencing prior to initiation of the study. Reproducibility was evaluated at each site in two sessions each day. During each session the following was performed twice: Eight individuals were tested by two operators. If the result of the first test performed using the Spartan RX CYP2C19 System for a particular individual was inconclusive, the test was repeated (second pass), as per the RX System's Instructions For Use. External control tests were performed on each system, as per the RX System's Instructions For Use, and testing proceeded if the results passed.
A total of 960 tests were performed; 10 second pass tests were performed due to inconclusive calls on the first pass test. After second pass testing of inconclusive genotype results, the Spartan RX CYP2C19 Test System produced a correct call rate of 99.8%, with a 95% one-sided lower confidence limit of 99%. There was one incorrect call. The root cause of the incorrect call could not conclusively be determined.
Reproducibility – First set of testing:
| Sample Type | Site | Samples tested | No calls | Incorrect calls | Correct calls | Correct call rate | 95% Confidence Limit |
| --- | --- | --- | --- | --- | --- | --- | --- |
| *1/*1 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 3 | 40 | 1 | 0 | 39 | 97.5% | 90% |
| | 4 | 40 | 0 | 1 | 39 | 97.5% | 90% |
| | Total | 120 | 1 | 1 | 118 | 98.3% | 95% |
{6}
| *1/*2 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | 3 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 4 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | Total | 120 | 0 | 0 | 120 | 100.0% | 98% |
| Sample Type | Site | Samples tested | No calls | Incorrect calls | Correct calls | Correct call rate | 95% Confidence Limit |
| *2/*2 | 1 | 40 | 2 | 0 | 38 | 95.0% | 86% |
| | 3 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 4 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | Total | 120 | 2 | 0 | 118 | 98.3% | 95% |
| *1/*3 | 1 | 40 | 1 | 0 | 39 | 97.5% | 90% |
| | 3 | 40 | 1 | 0 | 39 | 97.5% | 90% |
| | 4 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | Total | 120 | 2 | 0 | 118 | 98.3% | 95% |
| *2/*3 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 3 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 4 | 40 | 3 | 0 | 37 | 92.5% | 83% |
| | Total | 120 | 3 | 0 | 117 | 97.5% | 94% |
| *1/*17 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 3 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 4 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | Total | 120 | 0 | 0 | 120 | 100.0% | 98% |
| *17/*17 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 3 | 40 | 1 | 0 | 39 | 97.5% | 90% |
| | 4 | 40 | 1 | 0 | 39 | 97.5% | 90% |
| | Total | 120 | 2 | 0 | 118 | 98.3% | 95% |
| *2/*17 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 3 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 4 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | Total | 120 | 0 | 0 | 120 | 100.0% | 98% |
| TOTAL | 1 | 320 | 3 | 0 | 317 | 99.1% | 98% |
| | 3 | 320 | 3 | 0 | 317 | 99.1% | 98% |
| | 4 | 320 | 4 | 1 | 315 | 98.4% | 97% |
| | Total | 960 | 10 | 1 | 949 | 98.9% | 98% |
{7}
After re-testing those samples that gave inconclusive results in the first set of testing:
| Sample Type | Site | Samples tested | No calls | Incorrect calls | Correct calls | Correct call rate | 95% Confidence Limit |
| --- | --- | --- | --- | --- | --- | --- | --- |
| *1/*1 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 3 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 4 | 40 | 0 | 1 | 39 | 97.5% | 90% |
| | Total | 120 | 0 | 1 | 119 | 99.2% | 96% |
| *1/*2 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 3 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 4 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | Total | 120 | 0 | 0 | 120 | 100.0% | 98% |
| *2/*2 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 3 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 4 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | Total | 120 | 0 | 0 | 120 | 100.0% | 98% |
| *1/*3 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 3 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 4 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | Total | 120 | 0 | 0 | 120 | 100.0% | 98% |
| *2/*3 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 3 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 4 | 40 | 1 | 0 | 39 | 97.5% | 90% |
| | Total | 120 | 1 | 0 | 119 | 99.2% | 96% |
| *1/*17 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 3 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 4 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | Total | 120 | 0 | 0 | 120 | 100.0% | 98% |
| *2/*17 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 3 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 4 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | Total | 120 | 0 | 0 | 120 | 100.0% | 98% |
| *2/*17 | 1 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 3 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | 4 | 40 | 0 | 0 | 40 | 100.0% | 94% |
| | Total | 120 | 0 | 0 | 120 | 100.0% | 98% |
| TOTAL | 1 | 320 | 0 | 0 | 320 | 100.0% | 99% |
| | 3 | 320 | 0 | 0 | 320 | 100.0% | 99% |
| | 4 | 320 | 1 | 1 | 318 | 99.4% | 99% |
| | Total | 960 | 1 | 1 | 958 | 99.8% | 99% |
{8}
Lot-to-lot Reproducibility
The Spartan FRX CYP2C19 Test System uses independent reagents for detection of each of the three CYP2C19 mutations (*2, *3, *17). Lot-to-lot reproducibility was assessed using three independent lots of each reagent type. Samples used for the study were buccal scrapes collected from *1/*1, *1/*17, *2/*17, *17/*17, *2/*3, and *2/*2 individuals. After second pass testing of inconclusive genotype results, the Spartan RX CYP2C19 Test System produced a correct call rate of 99.0%, with a 95% one-sided lower confidence limit of 98%.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Controls:
External Controls: The system contains an External Control kit. The External Control reagent has two synthetic target nucleic acids that contain binding sites for the primers and wild-type or mutant probe are added. The external control kit contains a reagent for each of the three CYP2C19 loci (*2, *3 and *17) tested in the assay. At each locus under investigation, there is control material that is specific for the recessive (positive control) and dominant (negative control) genotype. PCR amplification and bi-directional DNA sequencing of each of the 6 control materials is used to confirm the sequence of the 6 oligonucleotides. External Controls result in a pass or fail and do not provide a genotype result.
Previously characterized genomic DNA can be tested by 1 μl of purified genomic DNA at 0.1 ng/μl using a pipette. The acceptable range of genomic DNA is 0.03 ng to 0.5 ng. The recommended amount is 0.1 ng because this has been determined to be just above the limit of detection of the assay (see 1d below). The acceptable A₂₆₀/A₂₈₀ ratio is 1.5 to 1.9. As a control, a buccal sample from a person with a known genotype may also be run.
Stability
The Sample Collection Kits and External Control kits are shipped frozen on dry ice. The kits contain a Timestip Plus disposable temperature indicator to inform the user if the package has remained frozen during shipping. Sample Collection Kits and External Control kits must be stored in a manual defrost freezer (one that does not say "frost-free") at a temperature between -20°C to -80°C. When stored under these conditions, Sample Collection Kits and External Control kits are stable until the expiry date marked on the product label.
9
{9}
# Sample stability
Sample stability was tested using 52 samples on 9 systems. Samples were stored in the transport system for up to 2 hours and run on the CYP2C19 Test System.
Protocols and acceptance criteria were reviewed and deemed acceptable. Information submitted supports the sponsor's claims that 1) the Sample Collection Kits must be used within 45 minutes of being removed from the freezer and 2) the buccal swab sample inserted into the reagent tube is stable for 1 hour if contained in the sample transport system.
# d. Detection limit:
The input material for the Spartan RX CYP2C19 Test System is a buccal swab collected from an individual, which is inserted directly into the reagent tube without any requirement for the user to extract or purify DNA. The end user can only use one buccal swab per test. 40 individuals had buccal swabs taken in the following range: 5 pooled swabs, 2 pooled swabs, 1 normal swab, one stroke swab, and buccal touch. The detection limit was assessed by analysis of buccal swab samples collected from $*1/*1$ , $*2/*17$ , $*17/*17$ , and $*2/*3$ individuals. Genotypes of all individuals were confirmed by bi-directional sequencing prior to initiation of the study. The concentration of genomic DNA isolated from each individual was determined using absorbance of DNA at $260~\mathrm{nm}$ . For each individual and each genomic DNA sample, PCR performance was assessed using 4 replicates. The DNA concentration was determined using a genomic standard curve. The quantity and quality was determined using absorbance at $260\mathrm{nm}$ and $280\mathrm{nm}$ , respectively. The following is the determined amount of DNA on a swab:
| Swabbing condition | Average DNA Amount (ng) |
| --- | --- |
| 5 pooled swabs | 3.3 |
| 2 pooled swabs | 1.6 |
| Normal swab | 1.3 |
| One stroke | 0.6 |
| Inside mouth touch | 0.1 |
{10}
Results are summarized below:
| First set of testing | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Test Condition | Genotype | # Samples tested | # Correct calls | # Incorrect calls | # Inconclusive calls | % Correct calls | 95% One-sided confidence lower limit) |
| 5 Pooled swabs | *1/*1 | 13 | 11 | 0 | 2 | 84.6% | 63% |
| | *17/*17 | 13 | 12 | 0 | 1 | 92.3% | 72% |
| | *2/*3 | 13 | 8 | 0 | 5 | 61.5% | 40% |
| | *2/*17 | 13 | 13 | 0 | 0 | 100.0% | 83% |
| | Total | 52 | 44 | 0 | 8 | 84.6% | 75% |
| 2 Pooled swabs | *1/*1 | 13 | 13 | 0 | 0 | 100.0% | 83% |
| | *17/*17 | 13 | 13 | 0 | 0 | 100.0% | 83% |
| | *2/*3 | 13 | 13 | 0 | 0 | 100.0% | 83% |
| | *2/*17 | 13 | 13 | 0 | 0 | 100.0% | 83% |
| | Total | 52 | 52 | 0 | 0 | 100.0% | 95% |
| Normal Swab | *1/*1 | 13 | 13 | 0 | 0 | 100.0% | 83% |
| | *17/*17 | 13 | 13 | 0 | 0 | 100.0% | 83% |
| | *2/*3 | 13 | 12 | 0 | 1 | 92.3% | 72% |
| | *2/*17 | 13 | 13 | 0 | 0 | 100.0% | 83% |
| | Total | 52 | 51 | 0 | 1 | 98.1% | 92% |
| 1 Half Stroke | *1/*1 | 13 | 12 | 0 | 1 | 92.3% | 72% |
| | *17/*17 | 13 | 13 | 0 | 0 | 100.0% | 83% |
| | *2/*3 | 13 | 13 | 0 | 0 | 100.0% | 83% |
| | *2/*17 | 13 | 13 | 0 | 0 | 100.0% | 83% |
| | Total | 52 | 51 | 0 | 1 | 98.1% | 92% |
| Inside Mouth Touch | *1/*1 | 13 | 10 | 0 | 3 | 76.9% | 54% |
| | *17/*17 | 13 | 12 | 0 | 1 | 92.3% | 72% |
| | *2/*3 | 13 | 9 | 0 | 4 | 69.2% | 46% |
| | *2/*17 | 13 | 13 | 0 | 0 | 100.0% | 83% |
| | Total | 52 | 44 | 0 | 8 | 84.6% | 75% |
e. Analytical specificity:
Effect of Oral Rinse
Buccal swab samples were collected from $*1/*1, *2/*17, *17/*17$ , and $*2/*3$ individuals. Genotypes of all individuals were confirmed by bi-directional sequencing prior to initiation of the study.
The following three different rinse procedures were evaluated:
- No rinse before swab collection
- 10 seconds rinse prior to collection
- 60 seconds rinse prior to collection
{11}
A total of 58 samples were tested under a no rinse condition. 56 samples were tested by rinsing with 15 ml water for at least 10 seconds immediately prior to swab collection. 56 samples were tested by rinsing with 30 ml water for at least 60 seconds immediately prior to swab collection. If the result of the first test performed using the Spartan RX CYP2C19 Test System for a particular individual was inconclusive; the test was repeated (second pass). Five samples produced inconclusive results; after second pass repeats of these samples, the system produced a 100% correct call rate for all rinse conditions tested. A water rinse is recommended to mitigate the risk that substances in the mouth that may interfere with performance of the Spartan RX CYP2C19 Test System.
## Interference (Exogenous Substances):
Interference from potential exogenous interfering substances was evaluated using buccal swab samples collected from *1/*1, *2/*17, *17/*17, and *2/*3 individuals. Genotypes of all individuals were confirmed by bi-directional sequencing prior to initiation of the study. To test the impact of exogenous substances on performance of the Spartan RX CYP2C19 Test System, buccal samples were collected from individuals immediately after exposure to the substance followed by a water rinse, which was performed as per the RX CYP2C19 Test System's instructions for use. A total of 16 samples were tested for each potential exogenous interfering substance (n=4 per genotype tested). If the result of the first test performed using the Spartan RX CYP2C19 Test System for a particular individual was inconclusive, the test was repeated (second pass). The following exogenous substances and amounts were tested:
- Antiseptic mouthwash - 20 ml - Rinse around mouth for 30 seconds
- Toothpaste - 3/4" strip - Brush teeth for 2 min, spit
- Baking soda solution - 30 ml (0.1 g/ml) - Rinse around mouth for 10 seconds
- Cough syrup - 30 ml - Rinse around mouth for 10 seconds
- Cranberry juice - 30 ml - Rinse around mouth for 10 seconds
- Salt water - 30 ml (0.01 g/ml) - Rinse around mouth for 10 seconds
- Sugar water - 30 ml (0.01 g/ml) - Rinse around mouth for 10 seconds
- Meat - 15 g - Chew for 10 seconds
- Chewing gum - 1 standard piece - Chew for 1 min
- Hard candy - 1 standard piece - Suck until fully dissolved
- Tobacco smoking - 1 cigar - 5 puffs
- Denture paste - Apply 3 strips to roof of mouth, leave for 5 min, remove
Mouthwash and tobacco smoke produced one inconclusive call each; these samples all produced correct calls for the second pass test. Baking soda, and chewing gum, produced two inconclusive calls; these samples also produced correct calls for the second pass test. Chewing gum produced 11 inconclusive calls, with 1 inconclusive call remaining in the second pass. After first and second pass analysis, the overall correct call rates were 91.5% and 99.55%, respectively. All substances tested met the acceptance criteria.
12
{12}
Interference (Endogenous Substances):
Interference from potential endogenous interfering substances was evaluated using buccal swab samples collected from *1/*1, *2/*17, *17/*17, and *2/*3 individuals. Genotypes of all individuals were confirmed by bi-directional sequencing prior to initiation of the study. A total of 16 samples were tested for each potential endogenous interfering substance (n=4 per genotype tested). Each of the following substances were added directly to the reagent tube immediately prior to insertion of the buccal swab sample.
- Whole blood - 3.5 -μl of 0.5% blood
- Human oral bacteria - 9 x 10⁴ cells
| Substance | # Samples tested | # Correct Calls | # Incorrect calls (all three mutations) | Inconclusive calls (all three mutations) | % Correct calls |
| --- | --- | --- | --- | --- | --- |
| Whole blood | 16 | 15 | 0 | 1 | 93.8% |
| Oral bacteria | 16 | 15 | 0 | 1 | 93.8% |
| Total | 32 | 30 | 0 | 2 | 93.8% |
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Method comparison testing was evaluated by technicians at three sites, Ottawa Hospital Research Institute, Ottawa, ON, Children's Hospital of Eastern Ontario, Ottawa, ON, and Mount Sinai Hospital, Toronto, ON. 325 samples of genomic DNA were extracted from saliva samples for bi-directional sequencing. These samples were compared to results from the Spartan RX CYP2C19 System for precision and reproducibility testing. 5 samples gave an inconclusive result on the first pass test and were re-tested; after the second pass, all samples were called correctly. No samples were called incorrectly. The overall correct call rate for the first pass and second pass tests was 98.5% and 100%, respectively. The agreement between the RX CYP2C19 system and bi-directional sequencing is summarized below.
{13}
First set of testing:
| Genotype | # Samples tested | # Correct Calls | # Incorrect calls | # Inconclusive calls | % Correct calls | 95% One-sided confidence |
| --- | --- | --- | --- | --- | --- | --- |
| *1/*1 | 130 | 128 | 0 | 2 | 98.5% | 95% |
| *1/*2 | 95 | 94 | 0 | 1 | 98.9% | 95% |
| *2/*2 | 19 | 19 | 0 | 0 | 100.0% | 88% |
| *1/*3 | 7 | 7 | 0 | 0 | 100.0% | 72% |
| *3/*3 | 1 | 1 | 0 | 0 | 100.0% | 27% |
| *1/*17 | 40 | 39 | 0 | 1 | 97.5% | 90% |
| *17/*17 | 11 | 11 | 0 | 0 | 100.0% | 80% |
| *2/*3 | 6 | 6 | 0 | 0 | 100.0% | 69% |
| *2/*17 | 15 | 15 | 0 | 0 | 100.0% | 85% |
| *3/*17 | 1 | 1 | 0 | 0 | 100.0% | 27% |
| Total | 325 | 321 | 0 | 4 | 98.8% | 97% |
After re-testing those samples that gave inconclusive results in the first set of testing:
| Genotype | # Samples tested | # Correct Calls | # Incorrect calls | # Inconclusive calls | % Correct calls | 95% One-sided confidence |
| --- | --- | --- | --- | --- | --- | --- |
| *1/*1 | 130 | 130 | 0 | 0 | 100.0% | 98% |
| *1/*2 | 95 | 95 | 0 | 0 | 100.0% | 97% |
| *2/*2 | 19 | 19 | 0 | 0 | 100.0% | 88% |
| *1/*3 | 7 | 7 | 0 | 0 | 100.0% | 72% |
| *3/*3 | 1 | 1 | 0 | 0 | 100.0% | 27% |
| *1/*17 | 40 | 30 | 0 | 0 | 100.0% | 94% |
| *17/*17 | 11 | 11 | 0 | 0 | 100.0% | 80% |
| *2/*3 | 6 | 6 | 0 | 0 | 100.0% | 69% |
| *2/*17 | 15 | 15 | 0 | 0 | 100.0% | 85% |
| *3/*17 | 1 | 1 | 0 | 0 | 100.0% | 27% |
| Total | 325 | 325 | 0 | 0 | 100.0% | 99% |
b. Matrix comparison:
Not applicable
{14}
15
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The product insert lists the following information:
The Spartan RX CYP2C19 identifies the CYP2C19*2, CYP2C19*3, and CYP2C19*17 alleles. The following table lists the alleles recognized by the Spartan RX CYP2C19 Assay; the polymorphism for each allele; enzyme activity; and references.
| CYP2C19 Allele | Polymorphism Recognized by Spartan RX CYP2C19 Assay | Enzyme Activity | References |
| --- | --- | --- | --- |
| *1 | None
*1 genotype for the Spartan RX CYP2C19 Assay indicates only the absence of *2, *3, and *17 alleles. | Normal | Romkes (1991), Richardson (1995), Blaisdell (2002) |
| *2 | 19154G>A | None | de Morais (1994), Ibeanu (1998), Fukushima-Uesaka (2005), Lee (2009) |
| *3 | 17948G>A | None | de Morais (1994), Fukushima-Uesaka (2005) |
| *17 | -806C>T | Increased | Sim (2006), Rudberg (2008) |
{15}
CYP2C19*4, *5, *6, *7, *8, and other alleles may be associated with absent or reduced enzyme activity, but are significantly less frequent than the CYP2C19*2 and *3 alleles. The Spartan RX CYP2C19 Assay provides no information for these alleles.
Samples with the CYP2C19*10 genotype may be miscalled by the Spartan RX CYP2C19 Assay. Both CYP2C19*2 and CYP2C19*10 are poor metabolizers.
## References
1. Desta Z et al. (2002). Clinical significance of the cytochrome P450 2C19 genetic polymorphism. *Clin Pharmacokinet.* 41(12): 913–958.
2. Sim SC et al. (2006). A common novel CYP2C19 gene variant causes ultrarapid drug metabolism relevant for the drug response to proton pump inhibitors and antidepressants. *Clin Pharmacol Ther.* 79(1): 103–113.
3. Rudberg I et al. (2008). Impact of the ultrarapid CYP2C19*17 allele on serum concentration of escitalopram in psychiatric patients. *Clin Pharmacol Ther.* 83(2): 322–327.
4. http://www.cypalleles.ki.se/cyp2c19.htm
5. Romkes M et al. (1991). Cloning and expression of complementary DNAs for multiple members of the human cytochrome P450IIC subfamily. *Biochemistry.* 30(13): 3247–3255.
6. Richardson TH et al. (1995). A universal approach to the expression of human and rabbit cytochrome P450s of the 2C subfamily in *Escherichia coli*. *Arch Biochem Biophys.* 323(1): 87–96.
7. Blaisdell J et al. (2002). Identification and functional characterization of new potentially defective alleles of human CYP2C19. *Pharmacogenetics.* 12(9): 703–711.
8. de Morais SM et al. (1994). The major genetic defect responsible for the polymorphism of S- mephenytoin metabolism in humans. *J Biol Chem.* 269(22): 15419–15422.
9. Ibeanu GC et al. (1998). Identification of new human CYP2C19 alleles (CYP2C19*6 and CYP2C19*2B) in a Caucasian poor metabolizer of mephenytoin. *J Pharmacol Exp Ther.* 286(3): 1490–1495.
10. Fukushima-Uesaka H et al. (2005). Genetic variations and haplotypes of CYP2C19 in a Japanese population. *Drug Metab Pharmacokinet.* 20(4): 300–307.
16
{16}
11. Lee S et al. (2009). Identification of new CYP2C19 variants exhibiting decreased enzyme activity in the metabolism of S-mephenytoin and omeprazole. *Drug Metab Dispos.* 37(11): 2262–2269.
N. Instrument Name:
Spartan RX CYP2C19 Platform
O. System Descriptions:
1. Modes of Operation:
The Spartan RX CYP2C19 Platform is a multi-well thermal cycler with optical detection capability. The unit has two optical detection channels (labeled as green and red).
The system is comprised of:
- Spartan RX Analyzer
- Power adapter for Spartan RX Analyzer
- Netbook computer
- Power adapter for netbook
- Printer
- Power adapter for printer
- USB cable for connecting netbook and printer
- Ethernet cable for connecting netbook and Spartan RX Analyzer
- Barcode scanner and stand
2. Software:
The Spartan RX user interface has been pre-loaded on the netbook. It will automatically run when the user logs into the notebook. The Hazard Analysis and Software Documentation was reviewed and deemed acceptable.
3. Specimen Identification:
The specimens are identified by a barcode, and the tubes are color coded.
4. Specimen Sampling and Handling:
The system is automated. After a buccal swab is taken, it is directly placed into the assay color coded tube (closing the tube), and loaded into the platform.
5. Calibration:
Not applicable.
17
{17}
6. Quality Control (QC):
The device has an External Control Kit which consists of a reagent tube containing synthetic nucleic acid control material, PCR amplification chemicals, and fluorescent oligonucleotide probes. The reagent tube is removed from the packaging and placed directly into the Spartan RX Analyzer according to the External Control procedure. The reagent tube is never opened.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
None
Q. Proposed Labeling
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
18
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.