ACON TCA ONE STEP TRICYCLIC ANTIDEPRESSANT TEST STRIP AND DEVICE

K021526 · ACON Laboratories, Inc. · LFG · Jul 17, 2002 · Clinical Toxicology

Device Facts

Record IDK021526
Device NameACON TCA ONE STEP TRICYCLIC ANTIDEPRESSANT TEST STRIP AND DEVICE
ApplicantACON Laboratories, Inc.
Product CodeLFG · Clinical Toxicology
Decision DateJul 17, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3910
Device ClassClass 2

Indications for Use

The ACON® TCA One Step Tricyclic Antidepressant Test Strip and ACON TCA One Step Tricyclic Antidepressant Test Device are rapid chromatographic immunoassays for the qualitative detection of Tricyclic Antidepressant in urine at a cut-off concentration of 1,000 ng/mL in reference to Nortriptyline. They are intended for healthcare professionals at point-of-care sites.

Device Story

Rapid chromatographic immunoassay for qualitative screening of tricyclic antidepressants in human urine; competitive binding lateral flow principle; utilizes mouse monoclonal antibody. Input: urine sample; Output: visual colored line in test region (negative) or absence of line (positive); control line confirms proper wicking. Used at point-of-care sites by healthcare professionals. No instrumentation required. Provides preliminary analytical result; assists in clinical decision-making regarding drug screening.

Clinical Evidence

Clinical evaluation of 226 urine specimens compared subject devices against predicate and HPLC. Results: 99% overall agreement with predicate; 95% positive agreement; 100% negative agreement. Comparison to HPLC at 1,000 ng/mL cut-off showed >99% positive agreement and 89% negative agreement for samples near cut-off.

Technological Characteristics

Lateral flow immunochromatographic assay; competitive binding; mouse monoclonal antibody; visual qualitative readout; no instrumentation; point-of-care use.

Indications for Use

Indicated for qualitative detection of Tricyclic Antidepressants in human urine at 1,000 ng/mL cut-off (Nortriptyline reference). Intended for use by healthcare professionals at point-of-care sites.

Regulatory Classification

Identification

A tricyclic antidepressant drugs test system is a device intended to measure any of the tricyclic antidepressant drugs in serum. The tricyclic antidepressant drugs include imipramine, desipramine, amitriptyline, nortriptyline, protriptyline, and doxepin. Measurements obtained by this device are used in the diagnosis and treatment of chronic depression to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). A tricyclic antidepressant drugs test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### SUMMARY OF 510(k) 8. This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The Assigned 510(k) number is ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ # Submitter: ACON Laboratories, Inc. 4108 Sorrento Valley Boulevard San Diego, California 92121 Tel .: 858-535-2030 Fax: 858-535-2038 # Date: May 8, 2002 # Contact Person: Edward Tung, Ph.D. # Product Names: ACON® TCA One Step Tricyclic Antidepressant Test Strip ACON® TCA One Step Tricyclic Antidepressant Test Device # Common Name: Immunochromatographic test for the qualitative detection of Tricyclic Antidepressant in urine # Device Classification: The ACON TCA One Step Tricyclic Antidepressant Test Strip and ACON TCA One Step Tricyclic Antidepressant Test Device are similar to other FDA-cleared devices for the qualitative detection of Tricyclic Antidepressant in urine specimens. These tests are used to provide a preliminary analytical result. Tricyclic Antidepressant test systems have been classified as Class II devices with moderate complexity. The product code for these devices is LFG and the regulation number is 862.3910. {1}------------------------------------------------ # Classification Name: Tricyclic Antidepressant test system # Intended Use: The ACON® TCA One Step Tricyclic Antidepressant Test Strip and ACON TCA One Step Tricyclic Antidepressant Test Device are rapid chromatographic immunoassays for the qualitative detection of Tricyclic Antidepressant in urine at a cut-off concentration of 1,000 ng/mL in reference to Nortriptyline. They are intended for healthcare professionals at point-of-care sites. 23-1-1-1-1 \$\pi_i\$ # Description: The ACON TCA One Step Tricyclic Antidepressant Test Strip and ACON TCA One Step Tricyclic Antidepressant Test Device are competitive binding, lateral flow immunochromatographic assays for the qualitative screening of Tricyclic Antidepressant in a urine sample. The test is based on the principle of antigen-antibody immuncchemistry. It utilizes mouse monoclonal antibody to selectively detect elevated levels of Tricyclic Antidepressant in urine at a cut-off concentration of 1000 ng/mL for Nortriptyline. These tests can be performed without the use of an instrument. A drug-positive urine specimen will not generate a colored-line in the designated test region, while a negative urine specimen or a urine specimen containing Tricyclic Antidepressant at the concentration below the cut-off level will generate a colored-line in the test region. To serve as a procedural control. a colored-line will always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred. # Predicate Device: Status DSTM TCA One step Tricyclic Antidepressants Test 510(k) Number: K980249 # Comparison to a Predicate Device: A comparison of the features of the ACON TCA One Step Tricyclic Antidepressant Test Strip and ACON TCA One Step Tricyclic Antidepressant Test Device versus the Status™ One step Antidepressants Test is shown below: - Both tests are assays intended for the qualitative detection of Tricyclic Antidepressant in urine . samples. - . Both tests are intended as a screening method that provides a preliminary analytical test result. - . Both tests are immunochromatographic, lateral flow assays for the rapid detection of Tricyclic Antidepressant with a visual, qualitative end result. {2}------------------------------------------------ - Both tests utilize the same basic immunoassay principles that rely on antigen/ antibody . interactions to indicate a positive or negative result. - Both tests have a cut-off for Nortriptyline concentration of 1,000 ng/mL. . ### Safety and Effectiveness Data: ### Accuracy A clinical evaluation was conducted using 226 clinical urine specimens including over 10% of the samples with Antidepressants concentrations at -25% cut-off to +25% cut-off range. This evauation compared the test results between ACON® TCA One Step Tricyclic Antidepressants Test Strip and Test Device with Status DS™ One Step Tricyclic Antidepressants Test; as well as against data obtained from the customary HPLC analysis. The comparisons of data obtained from this study yielded the following results: ACON TCA One Step Tricyclic Antidepressant Test Strip versus the Status DSTM One Step Tricyclic Antidepressants Test: > Positive Agreement: 55 / 58 = 95% (86% - 99%*) Negative Agreement: 164 / 164 = 100% (98% - 99%*) Overall Agreement: 219 / 222 = 99% (96% - 99 %*) * 95% Confidence Intervals ACON TCA One Step Tricyclic Antidepressant Test Device versus the Status DSTM One Step Tricyclic Antidepressant Test: > Positive Agreement: 55 / 58 = 95% (86% - 99%*) Negative Agreement: 164 / 164 = 100% (98% - 99%*) Overall Agreement: 219 / 222 = 99% (96% - 99 %*) * 95% Confidence Intervals ACON TCA One Step Tricyclic Antidepressant Test Strip versus HPLC at the cutoff of 1,000 ng/ml: | Acon<br>TCA test<br>strip | Drug -<br>free urine | <-25%<br>cutoff | -25% cutoff<br>to cutoff | Cutoff to<br>+25% cutoff | >+25%<br>cutoff | % agreement | |---------------------------|----------------------|-----------------|--------------------------|--------------------------|-----------------|----------------------| | Negative | 150 | 17 | 0 | 0 | 0 | 89%<br>(84% - 93%)* | | Positive | 0 | 12 | 8 | 15 | 20 | >99%<br>(90% - 99%)* | * 95% Confidence Intervals {3}------------------------------------------------ | Acon<br>TCA test<br>Device | Drug - free<br>urine | <-25%<br>cutoff | -25% cutoff<br>to cutoff | Cutoff to<br>+25% cutoff | >+25%<br>cutoff | % agreement | |----------------------------|----------------------|-----------------|--------------------------|--------------------------|-----------------|---------------------| | Negative | 150 | 17 | 0 | 0 | 0 | 89%<br>(84% - 93%)* | | Positive | 0 | 12 | 8 | 15 | 20 | >99%<br>(90% 99%)* | ACON TCA One-Step Tricyclic Antidepressant Test Device versus HPLC at the cutoff of 1,000 ng/ml: · 95% Confidence Intervals #### Conclusion: These clinical studies demonstrate the substantial equivalency between the ACON TCA One Step Tricyclic Antidepressant Test Strip, ACON TCA One Step Tricyclic Antidepressant Test Device and the Status DSTM One Step Tricyclic Antidepressant Test, which has already being marketed in the United States. It is also demonstrated that these tests are safe and effective in detecting Tricyclic Antidepressant at a concentration of 1,000 ng/mL. The POL study demonstrated that these tests are suitable for healthcare professionals including professionals at point-of-care site. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings and a wavy line representing its body. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUL 1 7 2002 Edward Tung, Ph.D. Director of Regulatory Affairs ACON Laboratories, Inc. 4108 Sorrento Valley Blvd. San Diego, CA 92121 Re: k021526 > Trade/Device Name: ACON® TCA One Step Tricyclic Antidepressants Test Strip ACON® TCA One Step Tricyclic Antidepressants Test Device Regulation Number: 21 CFR 862.3910 Regulation Name: Tricyclic antidepressant drugs test system Regulatory Class: Class II Product Code: LFG Dated: May 8, 2002 Received: May 10, 2002 Dear Dr. Tung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html". Sincerely yours, Steven Putman Steven I. Gutman, M.D., M.B.A. - -Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ K021526 #### INDICATIONS FOR USE 10. 510(k) Number: Koals Ble Device Name: ACON® TCA One Step Tricyclic Antidepressants Test Strip ACON® TCA One Step Tricyclic Antidepressants Test Device 德 12/200 Indications for Use: The ACON TCA One Step Tricyclic Antidepressants Test Strip and ACON TCA One Step Tricyclic Antidepressant Test Device are rapid chromatographic immunoassays for the qualitative detection of Tricyclic Antidepressants in human urine at a cut-off concentration of 1,000 ng/mL in reference to Nortriptyline. They are intended for Healthcare professionals including professionals at the point-of-care sites. Sean Cooper (Division Sign-Off) Division of Clinical Laboratory Devices Division of Clinical Laboratory Devices 510(k) Number K021526 (Please do not write below this point) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use_ V (Per 21 CFR 801.109) Over-The-Counter Use Or
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...