AXSYM VANCOMYCIN II
Device Facts
| Record ID | K955851 |
|---|---|
| Device Name | AXSYM VANCOMYCIN II |
| Applicant | Abbott Laboratories |
| Product Code | LEH · Clinical Toxicology |
| Decision Date | May 9, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3950 |
| Device Class | Class 2 |
Indications for Use
The intended use of both assays is for the quantitative determination of vancomycin in human serum or plasma (sodium heparin, tripotassium EDTA, potassium oxalate, and sodium citrate).
Device Story
Automated fluorescence polarization immunoassay (FPIA) for quantitative vancomycin measurement in human serum or plasma; utilizes Abbott AxSYM system; reagents, calibrators, and controls used for clinical monitoring; provides quantitative results to healthcare providers for therapeutic drug management.
Clinical Evidence
Bench testing only; correlation study comparing AxSYM Vancomycin II to predicate (n=217) yielded slope 1.03, y-intercept -1.08, correlation coefficient 0.99, standard error 2.74.
Technological Characteristics
Automated fluorescence polarization immunoassay (FPIA); reagents, calibrators, and controls prepared gravimetrically; real-time stability testing performed.
Indications for Use
Indicated for quantitative determination of vancomycin in human serum or plasma for clinical monitoring.
Regulatory Classification
Identification
A vancomycin test system is a device intended to measure vancomycin, an antibiotic drug, in serum. Measurements obtained by this device are used in the diagnosis and treatment of vancomycin overdose and in monitoring the level of vancomycin to ensure appropriate therapy.
Predicate Devices
- AxSYM Vancomycin assay
Related Devices
- K964587 — VANCOMYCIN ASSAY OR THE TECHNICON IMMUNO 1 SYSTEM · Bayer Corp. · Jan 13, 1997
- K983280 — VANCOMYCIN EIA ASSAY · Diagnostic Reagents, Inc. · Nov 6, 1998
- K963267 — VANCOMYCIN FLEX REAGENT CARTRIDGE · Dade Intl., Inc. · Sep 13, 1996
- K043386 — VITROS CHEMISTRY PRODUCTS VANC REAGENT, CALIBRATOR KIT 11 AND TDM PERFORMANCE VERIFIERS I, II AND III · Ortho-Clinical Diagnostics, Inc. · Feb 7, 2005
- K013076 — SYNCHRON SYSTEMS VANCOMYCIN REAGENT; SYNCHRON SYSTEMS VANCOMYCIN CALIBRATOR · Beckman Coulter, Inc. · Dec 12, 2001