GENTAMICIN FLEX REAGENT CARTRIDGE
Device Facts
| Record ID | K962819 |
|---|---|
| Device Name | GENTAMICIN FLEX REAGENT CARTRIDGE |
| Applicant | Dade Intl., Inc. |
| Product Code | LCD · Clinical Toxicology |
| Decision Date | Sep 13, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3450 |
| Device Class | Class 2 |
Indications for Use
The GENT FLEX™ reagent cartridge used on the Dimension® clinical chemistry system is an in vitro diagnostic test intended to measure gentamicin, an aminoglycoside antibiotic drug, in human specimens. Measurements obtained by this assay are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to ensure appropriate therapy.
Device Story
GENT FLEX™ reagent cartridge; used on Dimension® clinical chemistry system. Input: human serum or plasma samples. Principle: Particle Enhanced Turbidimetric Inhibition Immunoassay (PETINIA) using latex particle-gentamicin conjugate and gentamicin-specific monoclonal antibody. Output: quantitative gentamicin concentration (µg/mL). Used in clinical laboratory settings by trained personnel. Results assist clinicians in managing gentamicin therapy and identifying potential overdose. Benefits: automated, rapid monitoring of antibiotic levels to ensure therapeutic efficacy and patient safety.
Clinical Evidence
Method comparison study; 197 human serum/plasma samples tested on Dimension® system vs. Abbott AxSYM®. Results: slope 1.02, intercept -0.15, correlation coefficient 0.987, range 0.1-10.6 µg/mL.
Technological Characteristics
PETINIA (Particle Enhanced Turbidimetric Inhibition Immunoassay) technology. Reagents: latex particle-gentamicin conjugate, gentamicin-specific monoclonal antibody. Automated liquid reagent system. Designed for use on Dimension® clinical chemistry analyzers.
Indications for Use
Indicated for the measurement of gentamicin in human serum or plasma to diagnose gentamicin overdose and monitor therapeutic levels in patients undergoing aminoglycoside antibiotic treatment.
Regulatory Classification
Identification
A gentamicin test system is a device intended to measure gentamicin, an antibiotic drug, in human specimens. Measurements obtained by this device are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to ensure appropriate therapy.
Predicate Devices
- Abbott AxSYM® Gentamicin assay
Related Devices
- K060709 — MULTIGENT GENTAMICIN · Seradyn, Inc. · Jun 15, 2006
- K962519 — EMIT 2000 GENTAMICIN ASSAY AND CALIBRATORS · Behring Diagnostics, Inc. · Sep 4, 1996
- K012978 — RANDOX GENTAMICIN · Randox Laboratories, Ltd. · Sep 18, 2001
- K020704 — ROCHE ONLINE GENTAMICIN · Roche Diagnostics Corp. · Jul 12, 2002
- K963267 — VANCOMYCIN FLEX REAGENT CARTRIDGE · Dade Intl., Inc. · Sep 13, 1996