EMIT 2000 GENTAMICIN ASSAY AND CALIBRATORS
Device Facts
| Record ID | K962519 |
|---|---|
| Device Name | EMIT 2000 GENTAMICIN ASSAY AND CALIBRATORS |
| Applicant | Behring Diagnostics, Inc. |
| Product Code | LCD · Clinical Toxicology |
| Decision Date | Sep 4, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3450 |
| Device Class | Class 2 |
Indications for Use
The Emit® 2000 Gentamicin Assay is an in vitro diagnostic reagent test kit for the quantitative analysis of gentamicin in human serum or plasma.
Device Story
Emit 2000 Gentamicin Assay; in vitro diagnostic reagent kit. Input: human serum or plasma samples. Principle: homogeneous enzyme immunoassay (Emit technology). Output: quantitative gentamicin concentration (ng/mL). Used in clinical laboratory settings; operated by laboratory technicians/pathologists. Results used by clinicians for therapeutic drug monitoring of gentamicin to optimize dosing and minimize toxicity. Benefits: rapid, accurate quantification of drug levels to guide patient treatment.
Clinical Evidence
Bench testing only. Precision evaluated using three control levels (subtherapeutic, therapeutic, upper therapeutic) per NCCLS EP5-T2; within-run CVs 1.6-5.6%, total CVs 2.7-6.7%. Accuracy assessed via 104 patient samples compared to Abbott TDx FPIA method; correlation coefficient 0.995, slope 0.94, intercept 0.10 ng/mL.
Technological Characteristics
Homogeneous enzyme immunoassay reagent kit. Quantitative analysis performed on automated clinical chemistry analyzers (e.g., Syva 30R).
Indications for Use
Indicated for the quantitative measurement of gentamicin levels in human serum or plasma to assist in therapeutic drug monitoring.
Regulatory Classification
Identification
A gentamicin test system is a device intended to measure gentamicin, an antibiotic drug, in human specimens. Measurements obtained by this device are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to ensure appropriate therapy.
Predicate Devices
- Abbott TDx Gentamicin Assay (K904226)
Related Devices
- K962819 — GENTAMICIN FLEX REAGENT CARTRIDGE · Dade Intl., Inc. · Sep 13, 1996
- K020704 — ROCHE ONLINE GENTAMICIN · Roche Diagnostics Corp. · Jul 12, 2002
- K060709 — MULTIGENT GENTAMICIN · Seradyn, Inc. · Jun 15, 2006
- K042479 — VITROS CHEMISTRY PRODUCTS GENT, TOBRA & VANC REAGENTS; VITROS CHEMISTYR PRODUCTS CALIBRATOR KITS 11, 13 & 14; VITROS CHE · Ortho-Clinical Diagnostics, Inc. · Jan 6, 2005