BAYER ADVIA DIGOXIN ASSAY FOR THE INTEGRATED MODULE SYSTEM

K033007 · Bayer Diagnostics Corp. · KXT · Nov 14, 2003 · Clinical Toxicology

Device Facts

Record IDK033007
Device NameBAYER ADVIA DIGOXIN ASSAY FOR THE INTEGRATED MODULE SYSTEM
ApplicantBayer Diagnostics Corp.
Product CodeKXT · Clinical Toxicology
Decision DateNov 14, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3320
Device ClassClass 2

Indications for Use

The Bayer ADVIA® IMS™ Digoxin assay is for in vitro diagnostic use to quantitatively measure digoxin, a cardioactive drug, in human serum. Measurements obtained are used as an aid in the diagnosis of digoxin overdose and in monitoring therapeutic levels of digoxin to ensure appropriate therapy.

Device Story

In vitro diagnostic assay for human serum; measures digoxin concentration. Uses heterogeneous competitive magnetic separation principle. Reagents (antibody conjugate, enzyme conjugate, magnetic particles) react with patient sample; alkaline phosphatase-labeled digoxin competes with patient digoxin for antibody binding sites. System incubates at 37°C; magnetic particles capture antibody-hapten complexes; unbound components washed away. pNPP substrate added; alkaline phosphatase activity produces para-nitrophenoxide, measured via absorbance at 405 nm. Dose-response curve inversely proportional to digoxin concentration. Operated on ADVIA IMS Analyzer by laboratory personnel. Output provides quantitative digoxin levels (ng/mL) to assist clinicians in therapeutic drug monitoring and overdose assessment.

Clinical Evidence

Bench testing only. Precision studies (n=76-100 per level) showed total CVs ranging from 2.0% to 9.1%. Analytical sensitivity is 0.04 ng/mL. Method comparison against predicate (n=72) yielded correlation: ADVIA IMS = 1.017 x Immuno 1 + 0.055 (r=0.985). Analytical specificity tested against nine potential cross-reactants; significant cross-reactivity observed for digoxigenin, digoxigenin bis-digitoxoside, digoxigenin mono-digitoxoside, and lanatoside C.

Technological Characteristics

Heterogeneous competitive magnetic separation assay. Two liquid reagents. Storage 2-8 °C. Analytical range 0.04-6 ng/mL. System-specific packaging for ADVIA IMS analyzer.

Indications for Use

Indicated for quantitative measurement of digoxin in human serum for patients requiring monitoring of therapeutic drug levels or diagnosis of digoxin overdose.

Regulatory Classification

Identification

A digoxin test system is a device intended to measure digoxin, a cardiovascular drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of digoxin overdose and in monitoring levels of digoxin to ensure appropriate therapy.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k033007 B. Analyte: Digoxin C. Type of Test: Quantitative, Heterogeneous Competitive Magnetic Separation Assay D. Applicant: Bayer Diagnostics E. Proprietary and Established Names: ADVIA IMS® Digoxin Reagent F. Regulatory Information: 1. Regulation section: 21 CFR 862.3320, Digoxin test system 2. Classification: Class II 3. Product Code: KXT 4. Panel: Clinical Toxicology (91) G. Intended Use: 1. Indication(s) for use: The Bayer ADVIA® IMS™ Digoxin assay is for in vitro diagnostic use to quantitatively measure digoxin, a cardioactive drug, in human serum. Measurements obtained are used as an aid in the diagnosis of digoxin overdose and in monitoring therapeutic levels of digoxin to ensure appropriate therapy. 2. Special condition for use statement(s): Prescription Use Only 3. Special instrument Requirements: ADVIA IMS® Analyzer Only H. Device Description: The reagent consists of Digoxin Antibody Conjugate (R1), containing Mouse monoclonal anti-digoxin conjugate, 76 µg/L (nominal quantity); Buffer; BSA; Mouse G Globulins; Surfactant; Preservative, and Digoxin Enzyme Conjugate (R2), {1} Page 2 of 6 containing Digoxin ALP conjugate, 250 µg/L (nominal quantity); Buffer; BSA; BGG; and surfactant. In this procedure, Digoxin Antibody Conjugate (R1) is reacted with patient sample. The labeled Digoxin Enzyme Conjugate (R2) is then added and the antibody/hapten complex binds during a second incubation on the system at 37° C. The mIMP™ (monoclonal ImmunoMagnetic Particle) Reagent is added and another incubation occurs during which the antibody/hapten complex is bound. The mIMP/antibody complex is then washed and the pNPP (para-nitrophenyl phosphate) substrate is added. The alkaline phosphatase in the conjugate reacts with pNPP to form paranitrophenoxide and phosphate. Increasing absorbance, due to the formation of para-nitrophenoxide is monitored at 405 nm. The indicator reaction occurs as follows: ![img-0.jpeg](img-0.jpeg) A sample having no digoxin will have the minimum label bound, while samples having high digoxin concentrations will have less label bound. Thus, the dose/response curve is inversely proportional to the digoxin concentration in the sample. I. Substantial Equivalence Information: 1. Predicate device name(s): Immuno 1 Digoxin Assay 2. Predicate K number(s): K912616 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Same | Quantitative measurement of digoxin | | Principle | Same | Heterogeneous Competitive Magnetic Separation Assay | | Storage | Same | 2-8°C | | Sensitivity | Same | 0.04 ng/mL | | Analytical Range | Same | 0.04 – 6.0 ng/mL | | Reagents | Same | Two liquid reagents contained in system specific packaging | {2} Page 3 of 6 J. Standard/Guidance Document Referenced (if applicable): None Referenced K. Test Principle: Heterogeneous Competitive Magnetic Separation Assay. In this assay, digoxin in serum competes with ALP-labeled digoxin for binding sites on an anti-digoxin fluorescein-labeled antibody in the reagent. The fluoresceinated antibody bound to either the free digoxin in the sample or the digoxin-ALP now binds to the magnetic particles. The ALP in the digoxin-ALP then dephosphorylates dioxetane phosphate, resulting in luminescence. The amount of luminescence is inversely proportional to the concentration of digoxin in the original sample. L. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Four precision studies were performed with the following results: | n | Mean (ng/mL) | Within Run SD | Within Run CV (%) | Total SD | Total CV (%) | | --- | --- | --- | --- | --- | --- | | 80 | 0.45 | 0.03 | 6.1 | 0.04 | 9.1 | | 80 | 1.40 | 0.04 | 2.8 | 0.07 | 5.0 | | 80 | 2.74 | 0.05 | 2.0 | 0.10 | 3.8 | | 76 | 0.75 | 0.02 | 3.3 | 0.10 | 3.8 | | 80 | 1.92 | 0.05 | 2.5 | 0.09 | 4.6 | | 80 | 3.29 | 0.07 | 2.1 | 0.08 | 2.6 | | n | Mean (ng/mL) | Within Run SD | Within Run CV (%) | Total SD | Total CV (%) | | --- | --- | --- | --- | --- | --- | | 99 | 0.43 | 0.02 | 5.4 | 0.03 | 7.5 | | 100 | 1.39 | 0.03 | 1.9 | 0.05 | 3.6 | | 100 | 2.72 | 0.06 | 2.1 | 0.10 | 3.7 | | 99 | 0.74 | 0.03 | 3.7 | 0.04 | 6.0 | | 100 | 1.87 | 0.05 | 2.4 | 0.11 | 5.9 | | 100 | 3.26 | 0.07 | 2.1 | 0.12 | 3.8 | | n | Mean (ng/mL) | Within Run SD | Within Run CV (%) | Total SD | Total CV (%) | | --- | --- | --- | --- | --- | --- | | 80 | 0.45 | 0.02 | 5.1 | 0.03 | 5.6 | | 80 | 1.42 | 0.03 | 2.2 | 0.04 | 2.8 | | 80 | 2.78 | 0.05 | 1.7 | 0.06 | 2.0 | | 78 | 0.75 | 0.03 | 3.7 | 0.04 | 4.9 | | 80 | 1.91 | 0.03 | 1.6 | 0.08 | 4.1 | | 80 | 3.37 | 0.06 | 1.7 | 0.10 | 2.8 | {3} | n | Mean (ng/mL) | Within Run SD | Within Run CV (%) | Total SD | Total CV (%) | | --- | --- | --- | --- | --- | --- | | 100 | 0.38 | 0.02 | 5.4 | 0.03 | 8.5 | | 100 | 1.33 | 0.04 | 3.1 | 0.05 | 3.6 | | 100 | 2.64 | 0.05 | 2.0 | 0.10 | 3.7 | | 100 | 0.72 | 0.026 | 3.6 | 0.034 | 4.8 | | 100 | 1.87 | 0.04 | 2.3 | 0.05 | 2.7 | | 100 | 3.25 | 0.05 | 1.6 | 0.08 | 2.3 | # b. Linearity/assay reportable range: The analytical range of the assay is $0.04 - 6.0\mathrm{ng / mL}$ The recovery/linearity studies were performed using the Trial lot 2 of the ADVIA IMS Digoxin reagents. The following is the data collected: ![img-1.jpeg](img-1.jpeg) # c. Traceability (controls, calibrators, or method): Calibrators and controls were previously cleared and are traceable to a United States Pharmacopeia (USP) standard # d. Detection limit: Analytical sensitivity is $0.04\mathrm{ng / mL}$ , calculated as 2 Standard Deviations from the zero calibrator. # e. Analytical specificity: Nine (9) potential cross reactants to Digoxin were spiked into serum pools and tested for cross-reactivity. {4} | Cross Reactants | Spiked Conc. (ng/mL) | Endogenous Conc. (ng/mL) | Observed Conc. (ng/mL) | ADVIA IMS % Cross-Reactivity | | --- | --- | --- | --- | --- | | Digitoxin | 120 | 0.77 | 0.80 | 0.03 | | | 120 | 1.94 | 1.95 | 0.01 | | | 120 | 3.58 | 3.67 | 0.08 | | Digoxigenin | 40 | 0.75 | 3.33 | 6.4 | | | 40 | 1.90 | 4.18 | 5.7 | | | 40 | 3.42 | 5.52 | 5.2 | | Digoxigenin bis-digitoxoside | 1.5 | 0.42 | 1.12 | 46.6 | | | 1.5 | 1.67 | 2.21 | 36.0 | | | 1.5 | 2.73 | 3.42 | 46.0 | | Digoxigenin mono-digitoxoside | 0.8 | 0.42 | 0.94 | 65.4 | | | 0.8 | 1.67 | 2.09 | 52.5 | | | 0.8 | 2.73 | 3.27 | 67.5 | | Dihydrodigoxin | 120 | 0.77 | 0.86 | 0.07 | | | 120 | 1.90 | 1.99 | 0.08 | | | 120 | 3.36 | 3.37 | 0.01 | | Furosemide | 120 | 0.61 | 0.68 | 0.05 | | | 120 | 1.61 | 1.66 | 0.04 | | | 120 | 2.95 | 3.19 | 0.20 | | Lanatoside C | 2.5 | 0.42 | 1.83 | 56.5 | | | 2.5 | 1.67 | 2.88 | 48.4 | | | 2.5 | 2.73 | 4.14 | 56.4 | | Ouabaine | 120 | 0.77 | 0.76 | -0.01 | | | 120 | 1.94 | 1.89 | -0.04 | | | 120 | 3.58 | 3.51 | -0.06 | | Spironolactone | 120 | 0.75 | 0.77 | 0.02 | | | 120 | 1.99 | 1.90 | -0.07 | | | 120 | 3.46 | 3.34 | -0.10 | f. Assay cut-off: N/A # 2. Comparison studies: a. Method comparison with predicate device: Seventy-two (72) serum samples (0.04 to $4.73\mathrm{ng / mL}$ ) were tested on both the Immuno 1 and the ADVIA IMS. The correlation results are summarized as follows: ADVIA IMS = 1.017 x Immuno 1 + 0.055 n = 72, r = 0.985, Syx = 0.265 {5} Page 6 of 6 ![img-2.jpeg](img-2.jpeg) ADVIA IMS vs. IMMUNO 1 IMMUNO 1 (ng/mL) b. Matrix comparison: N/A 3. Clinical studies: a. Clinical sensitivity N/A b. Clinical specificity: N/A c. Other clinical supportive data (when a and b are not applicable): N/A 4. Clinical cut-off: N/A 5. Expected values/Reference range: Therapeutic Range 0.8 - 2.0 ng/mL. The manufacturer recommends that each laboratory should determine its own reference ranges for the diagnostic evaluation of patient results. M. Conclusion: Based upon the information provided for the file, I recommend that the ADVIA IMS® Digoxin Reagent is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...