Psychemedics Homogeneous Enzyme Immunoassay for Cocaine in Hair
Applicant
Psychemedics Corporation
Product Code
JXO · Clinical Toxicology
Decision Date
Jun 25, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3250
Device Class
Class 2
Indications for Use
The Psychemedics homogeneous enzyme immunoassay for cocaine in hair is an enzyme immunoassay for the preliminary qualitative detection of cocaine in human head and body hair using a cocaine calibrator of 5 ng/10 mg hair for the purpose of identifying cocaine use. This product is intended exclusively for in-house professional use and not for sale to anyone. The Psychemedics EIA Cocaine Assay provides only a preliminary analytical test result. To confirm positive results, a more specific alternate chemical method (e.g. LC/MS/MS) must be used. Clinical consideration and professional judgement should be applied to the interpretation of any drug-of-abuse test result.
Device Story
Device performs preliminary qualitative detection of cocaine in human head and body hair; utilizes homogeneous enzyme immunoassay (HEIA) technology. Process involves pre-analytical hair extraction into liquid matrix; competitive binding assay between drug in sample and drug-labeled recombinant G6PDH enzyme; antibody binding reduces enzyme activity; presence of drug increases enzyme activity proportionally. Result measured spectrophotometrically at 340 nm via automated clinical chemistry analyzer. Used in professional laboratory settings; results interpreted by clinicians to identify drug use; requires confirmatory testing for positive results. Benefits include rapid screening of hair samples for cocaine exposure.
Clinical Evidence
Performance validated via comparison studies using 215 anonymous workplace hair samples. HEIA results compared against washed LC/MS/MS confirmatory testing. Study 1 showed 92 positive and 123 negative samples. Discordant results (positive HEIA/negative LC/MS/MS) attributed to environmental contamination or sweat-derived drug removed during washing. Additional studies compared unwashed vs. washed LC/MS/MS, confirming that washing reduces cocaine concentration in some samples. Precision studies performed at +/- 25%, 50%, 75%, and 100% of the 5 ng/10 mg cutoff showed consistent results. Cosmetic treatments (perm, dye, shampoo, relaxer) did not affect assay results.
Technological Characteristics
Homogeneous enzyme immunoassay (HEIA). Reagents: anti-cocaine monoclonal antibody, cocaine-labeled recombinant G6PDH. Measurement: spectrophotometric at 340 nm using automated clinical chemistry analyzer (e.g., Olympus AU 640). Confirmation: LC/MS/MS (AB Sciex 3200/6500). Extraction: acidic aqueous buffer. Qualitative test.
Indications for Use
Indicated for the preliminary qualitative detection of cocaine in human head and body hair for the purpose of identifying cocaine use in professional settings. Requires confirmation by a more specific method (e.g., LC/MS/MS).
Regulatory Classification
Identification
A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.
Special Controls
*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
Psychemedics Microplate EIA for Cocaine in Hair (k111925)
Related Devices
K210212 — Psychemedics Homogeneous Enzyme Immunoassay for Amphetamines in Hair · Psychemedics Corporation · Dec 15, 2021
K201326 — Psychemedics Homogeneous Enzyme Immunoassay (HEIA) for Opiates in Hair, Psychemedics Homogeneous Enzyme Immunoassay (HEIA) for Oxycodone in Hair · Psychemedics Corporation · Jul 27, 2021
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K201228
B Applicant
Psychemedics Corporation
C Proprietary and Established Names
Psychemedics Homogeneous Enzyme Immunoassay for Cocaine in Hair
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JXO | Class II | 21 CFR 862.3250 - Cocaine and Cocaine Metabolite Test System | TX - Clinical Toxicology |
## II Submission/Device Overview:
A Purpose for Submission:
New Device
B Measurand:
Cocaine
C Type of Test:
Qualitative homogeneous enzyme immunoassay (HEIA)
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
B Indication(s) for Use:
The Psychemedics homogeneous enzyme immunoassay for cocaine in hair is an enzyme immunoassay for the preliminary qualitative detection of cocaine in human head and body hair using a cocaine calibrator of 5 ng/10 mg hair for the purpose of identifying cocaine use. This product is intended exclusively for in-house professional use and not for sale to anyone.
The Psychemedics EIA Cocaine Assay provides only a preliminary analytical test result. To confirm positive results, a more specific alternate chemical method (e.g. LC/MS/MS) must be used. Clinical consideration and professional judgement should be applied to the interpretation of any drug-of-abuse test result.
C Special Conditions for Use Statement(s):
OTC - Over The Counter
D Special Instrument Requirements:
The Psychemedics HEIA For Cocaine in Hair assay requires an Automated Clinical Chemistry Analyzer at 340 nm. The Olympus AU 640 analyzer was used for validation of the candidate assay.
The confirmation assay uses an AB Sciex 3200 LC/MS/MS linked to two LC-20AD Prominence Liquid Chromatograph Pump and a LEAP PAL HTC-xt autosampler or a AB Sciex 6500 LC/MS/MS linked to two Shimadzu LC-30AD pumps and a LEAP PAL-HTC-xt autosampler with a DLW wash system.
IV Device/System Characteristics:
A Device Description:
The Psychemedics homogeneous enzyme immunoassay (HEIA) test for cocaine in hair consists of two parts; a pre-analytical hair treatment procedure (to extract cocaine from the solid hair matrix to a measurable liquid matrix) and the screening assay, the Psychemedics Cocaine HEIA.
Performance of the HEIA requires a sample extract solution and cocaine-glucose-6-phosphate dehydrogenase (G6PDH) conjugate. The Psychemedics Cocaine HEIA consists of reagents R1 (anti-cocaine monoclonal antibody with substrate) and R2 (cocaine labeled recombinant G6PDH).
The confirmation assay consists of an AB Sciex 3200 LC/MS/MS linked to two LC-20AD Prominence Liquid Chromatograph Pump and a LEAP PAL HTC-xt autosampler or a AB Sciex 6500 LC/MS/MS linked to two Shimadzu LC-30AD pumps and a LEAP PAL-HTC-xt autosampler with a DLW wash system.
B Principle of Operation:
Hair sample extracts are combined with antibody and glucose-6-phosphate dehydrogenase (G6PDH)-cocaine immunoconjugate in a cuvette in a clinical chemistry analyzer. The assay is
K201228 - Page 2 of 13
{2}
based on competition between the cocaine-labeled G6PDH and free drug from the extracted hair sample for a fixed amount of specific antibody binding sites. Int the absence of free drug from the samples, the specific antibody binds the cocaine-G6PDH and causes a decrease in enzyme activity. This phenomenon creates a direct relationship between the drug concentration in hair and enzyme activity. The enzyme activity is determined spectrophotometrically at 340 nm by measuring the conversion of nicotinamide adenine dinucleotide (NAD) to NADH.
For samples that are presumptive positive by the screening assay, a new aliquot of the hair sample is weighed, washed extensively to remove externally derived cocaine, extracted by a different procedure and confirmed by LC/MS/MS for the presence of cocaine.
V Substantial Equivalence Information:
A Predicate Device Name(s):
Psychemedics Cocaine EIA
B Predicate 510(k) Number(s):
K111925
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K201228 | K111925 |
| --- | --- | --- |
| Device Trade Name | Psychemedics Homogeneous Enzyme Immunoassay for Cocaine in Hair | Psychemedics Cocaine Assay |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Same | For the detection of Cocaine in human hair. The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. |
| Product Code | Same | JXO |
| Measurand | Same | Cocaine |
| Confirmation Method | Same | LC/MS/MS |
| Sample Matrix | Same | Human Hair |
| Cutoff | Same | 5 ng/10 mg hair |
| General Device Characteristic Differences | | |
| Method of Measurement | Automated Clinical Chemistry Analyzer at 340 nm | Microplate reader, read at 450 nm |
K201228 - Page 3 of 13
{3}
K201228 - Page 4 of 13
| Device & Predicate Device(s): | K201228 | K111925 |
| --- | --- | --- |
| Type of Test | Homogeneous Enzyme Immunoassay | Microplate Enzyme Immunoassay |
| Extraction Method | Acidic aqueous buffer | Patented Digestion Method |
VI Standards/Guidance Documents Referenced:
None referenced.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Precision of the Psychemedics HEIA for Cocaine in Hair assay
Precision studies were performed by spiking negative hair before digestion to achieve concentrations of negative, the cutoff concentration of 5 ng cocaine/10 mg hair, and +/- 75%, +/-50% and +/-25% of the cutoff concentration.
Intra-Assay Precision - Four samples at each concentration were tested in replicates of 2 on a single day. Results are summarized in the table below.
Inter-Assay Precision – Testing was performed for 10 days, with 2 runs per day in replicates of 4. Results are summarized in the table below.
| Summary Intra-Assay Precision | | |
| --- | --- | --- |
| Level | NEG | POS |
| -100% | 8 | 0 |
| -75% | 8 | 0 |
| -50% | 8 | 0 |
| -25% | 8 | 0 |
| +25% | 0 | 8 |
| +50 | 0 | 8 |
| +75% | 0 | 8 |
| +100% | 0 | 8 |
| Summary Inter-Assay Precision | | |
| --- | --- | --- |
| Level | NEG | POS |
| -100% | 80 | 0 |
| -75% | 80 | 0 |
| -50% | 80 | 0 |
| -25% | 80 | 0 |
| +25% | 0 | 80 |
| +50 | 0 | 80 |
| +75% | 0 | 80 |
| +100% | 0 | 80 |
Precision of the LC-MS/MS method
Intra-assay precision was determined around the cutoff of 5 ng cocaine per 10 mg hair, 0.5 ng of benzoylecgonine, cocaethylene and norcocaine per 10 mg of hair and 1 pg/mg of para-hydroxy cocaine, meta-hydroxy cocaine and ortho-hydroxy cocaine with 5 replicates for each concentration per drug metabolite.
{4}
Intra-assay Precision Around the Cutoff for Cocaine, Benzoylecgonine, Cocaethylene and Norcocaine.
| Intra-assay Precision Around the Cutoff | | | | | |
| --- | --- | --- | --- | --- | --- |
| ng analyte/10 mg hair | | | | | |
| Cocaine | 2.5 | 3.75 | 5 | 6.25 | 7.5 |
| Average | 2.588 | 3.884 | 5.305 | 6.748 | 8.172 |
| S.D | 0.086 | 0.056 | 0.212 | 0.200 | 0.217 |
| %C.V. | 3.31 | 1.44 | 3.99 | 2.96 | 2.65 |
| 95% C.I. | 2.512 – 2.663 | 3.835 – 3.933 | 5.120 – 5.491 | 6.572 – 6.923 | 7.983 – 8.361 |
| % of Target | 103.5 | 103.6 | 106.1 | 108.0 | 109.0 |
| | | | | | |
| Benzoylecgonine | 0.25 | 0.375 | 0.50 | 0.625 | 0.75 |
| Average | 0.2516 | 0.3672 | 0.4962 | 0.6340 | 0.7992 |
| S.D | 0.019 | 0.027 | 0.022 | 0.032 | 0.029 |
| %C.V. | 7.40 | 7.25 | 4.34 | 5.06 | 3.66 |
| 95% C.I. | 0.235-0.267 | 0.343-0.390 | 0.477-0.515 | 0.605-0.662 | 0.773-0.824 |
| % of Target | 100.6 | 97.9 | 99.2 | 101.4 | 106.6 |
| | | | | | |
| Cocaethylene | 0.25 | 0.375 | 0.50 | 0.625 | 0.75 |
| Average | 0.258 | 0.389 | 0.514 | 0.686 | 0.823 |
| S.D | 0.010 | 0.007 | 0.016 | 0.013 | 0.020 |
| %C.V. | 3.99 | 1.85 | 3.18 | 1.96 | 2.43 |
| 95% C.I. | 0.249-0.267 | 0.382-0.395 | 0.500-0.528 | 0.674-0.698 | 0.805-0.840 |
| % of Target | 103.4 | 103.8 | 102.9 | 109.8 | 109.7 |
| | | | | | |
| Norcocaine | 0.25 | 0.375 | 0.50 | 0.625 | 0.75 |
| Average | 0.256 | 0.399 | 0.532 | 0.680 | 0.803 |
| S.D. | 0.007 | 0.018 | 0.019 | 0.019 | 0.021 |
| %C.V. | 2.93 | 4.60 | 3.68 | 2.91 | 2.65 |
| 95% C.I. | 0.249-0.262 | 0.382-0.415 | 0.515-0.549 | 0.663-0.698 | 0.785-0.822 |
| % of Target | 102.4 | 106.4 | 106.5 | 108.9 | 107.2 |
Intra-assay Precision Around the Cutoff for Para-Hydroxycocaine, Meta-Hydroxycocaine and Ortho-Hydroxycocaine.
| Intra-assay Precision Around the Cutoff | | | | | |
| --- | --- | --- | --- | --- | --- |
| ng analyte/10 mg hair | | | | | |
| Para-HO-Cocaine | 0.5 | 0.75 | 1.0 | 1.25 | 1.5 |
| Average | 0.467 | 0.716 | 0.934 | 1.243 | 1.418 |
| S.D | 0.008 | 0.031 | 0.027 | 0.029 | 0.040 |
| %C.V. | 1.66 | 4.33 | 2.84 | 2.33 | 2.85 |
| 95% C.I. | 0.460-0.474 | 0.689-0.743 | 0.911-0.957 | 1.218-1.268 | 1.383-1.453 |
| % of Target | 93.5 | 95.6 | 93.5 | 99.5 | 94.6 |
| | | | | | |
| Meta-HO-Cocaine | 0.5 | 0.75 | 1.0 | 1.25 | 1.5 |
| Average | 0.467 | 0.721 | 0.913 | 1.186 | 1.369 |
K201228 - Page 5 of 13
{5}
| Intra-assay Precision Around the Cutoff | | | | | |
| --- | --- | --- | --- | --- | --- |
| ng analyte/10 mg hair | | | | | |
| S.D | 0.023 | 0.032 | 0.027 | 0.049 | 0.058 |
| %C.V. | 4.97 | 4.39 | 2.90 | 4.12 | 4.22 |
| 95% C.I. | 0.446-0.487 | 0.693-0.749 | 0.890-0.937 | 1.143-1.229 | 1.318-1.419 |
| % of Target | 93.4 | 96.2 | 91.4 | 94.9 | 91.3 |
| | | | | | |
| Ortho-HO-Cocaine | 0.5 | 0.75 | 1.0 | 1.25 | 1.5 |
| Average | 0.471 | 0.734 | 0.959 | 1.176 | 1.392 |
| S.D | 0.014 | 0.040 | 0.018 | 0.048 | 0.053 |
| %C.V. | 2.96 | 5.43 | 1.89 | 4.07 | 3.82 |
| 95% C.I. | 0.459-0.483 | 0.699-0.769 | 0.943-0.975 | 1.134-1.218 | 1.345-1.438 |
| % of Target | 94.3 | 97.9 | 95.9 | 94.1 | 92.8 |
## Inter-assay Precision for Cocaine, Benzoylecgonine, Cocaethylene and Norcocaine
Inter-assay precision for cocaine, benzoylecgonine, and cocaethylene, was determined using 50, 100 and 150 µL of the cocaine in hair assay 5 ng standard; inter-assay precision for norcocaine was determined using 50, 100 and 150 µL of the cocaine in hair assay 0.5 ng norcocaine standard and the cocaine in hair assay 0.01 ng hydroxy cocaine standard. Inter-assay precision was performed over 5 separate days, testing each respective sample concentration in quintuplicate with results summarized below.
| Inter-assay Precision | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| ng Analyte/10 mg hair | | | | | | | | | | | | |
| | Cocaine | | | Benzoylecgonine | | | Cocaethylene | | | Norcocaine | | |
| | 2.5 | 5 | 7.5 | 0.25 | 0.5 | 0.75 | 0.25 | 0.5 | 0.75 | 0.25 | 0.5 | 0.75 |
| Average | 2.56 | 5.02 | 7.57 | 0.257 | 0.527 | 0.827 | 0.251 | 0.501 | 0.767 | 0.247 | 0.500 | 0.762 |
| S.D. | 0.1180 | 0.2687 | 0.4092 | 0.0216 | 0.0322 | 0.0463 | 0.0152 | 0.0244 | 0.0434 | 0.0136 | 0.0329 | 0.0501 |
| %CV | 4.61 | 5.35 | 5.41 | 8.42 | 6.11 | 5.60 | 6.04 | 4.88 | 5.65 | 5.50 | 6.57 | 6.58 |
| 95% Confidence Interval | 2.45 - 2.66 | 4.78 - 5.25 | 7.21 - 7.92 | 0.238 - 0.276 | 0.499 - 0.555 | 0.786 - 0.868 | 0.238 - 0.264 | 0.480 - 0.522 | 0.729 - 0.805 | 0.235 - 0.259 | 0.471 - 0.529 | 0.718 - 0.806 |
| % of Target | 102.27 | 100.39 | 100.88 | 102.67 | 105.42 | 110.21 | 100.51 | 100.15 | 102.28 | 98.78 | 100.06 | 101.61 |
K201228 - Page 6 of 13
{6}
Inter-assay Precision for para-hydroxycocaine, meta-hydroxycocaine and ortho-hydroxycocaine.
| Inter-assay Precision | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | para-HO-Cocaine | | | meta-HO-Cocaine | | | ortho-HO-Cocaine | | |
| pg analyte/mg hair | | | | | | | | | |
| | 0.5 | 1 | 1.5 | 0.5 | 1 | 1.5 | 0.5 | 1 | 1.5 |
| Average | 0.525 | 1.015 | 1.502 | 0.502 | 0.992 | 1.491 | 0.509 | 1.038 | 1.520 |
| S.D. | 0.0463 | 0.0739 | 0.0815 | 0.0484 | 0.0803 | 0.1118 | 0.0377 | 0.0826 | 0.0994 |
| %CV | 8.81 | 7.28 | 5.43 | 9.65 | 8.10 | 7.50 | 7.40 | 7.96 | 6.54 |
| % Confidence Interval | 0.489 - 0.571 | 0.950 - 1.080 | 1.431 - 1.573 | 0.460 - 0.544 | 0.922 - 1.062 | 1.393 - 1.589 | 0.476 - 0.542 | 0.966 - 1.110 | 1.433 - 1.607 |
| % of Target | 105.10 | 101.50 | 100.14 | 100.44 | 99.16 | 99.39 | 101.84 | 103.80 | 101.31 |
2. Linearity:
The screening immunoassay is a qualitative test only; therefore, a linearity evaluation is not applicable.
Linearity of the LC-MS/MS confirmation method
The linearity of the confirmation method was evaluated by spiking cocaine and cocaine metabolites into negative hair. The following concentrations were analyzed in replicates of 5 for each analyte: 0.0, 0.25, 0.50, 1.0, 3.0, 5.0, 50, 100 and 150 ng cocaine per 10 mg hair; 0.0, 0.25, 0.50, 1.0, 3.0, 5.0, 50 and 100 ng benzoylecgonine, cocaethylene and norcocaine per 10 mg hair; 0.0, 0.25, 0.3, 0.5, 0.75, 1.0, 10, 20, 100, 200 and 300 pg of para-, meta-, and ortho-hydroxycocaine per mg hair. over the range of 0 to 1000 pg/mg hair.
All results were within 20% of the target concentrations. Agreement of the 5 determinations at each concentration was ≤10% CV. The regression coefficient was >0.995.
3. Analytical Specificity/Interference:
Specificity of the immunoassay
Immunoassay: Interference from structurally unrelated compounds
The following potentially interfering compounds were tested using the screening immunoassay on 3 negative hair samples confirmed negative for cocaine, opioids, PCP, amphetamines and marijuana. The negative hair samples were spiked with cocaine at ±50% of the cutoff to assess possible positive or negative interference. All potential interferents listed below were tested at a concentration of 100 ng/10 mg hair. No positive or negative interference was observed in this study.
Anhydroecgonine methyl ester, Atropine, Bupropion, Cotinine, Cannabinol, Chlorpheniramine maleate, O-Desmethylvenlafaxine, Desipramine, Doxylaminesuccinate, 1S, 2R (+)-Ephedrine, Naproxen, Nicotine, Nortriptyline, H-Propoxyphene, R,R (-)Pseudoephedrine, Thioridazine, cis-Tramadol, Venlafaxine, (±)-11-nor-9-Carboxy-Δ9-THC,
K201228 - Page 7 of 13
{7}
Pentazocine, Amoxicillin, Propanolol, Promethazine, Phenmetrazine, Phendimetrazine, Benzocaine, Ecgonine, Dextromethorphan, Amitriptyline, R-(-)Phenylephrine, Glutethimide, Meprobamate, Lidocaine, Carbamazepine, Diazepam, Nordiazepam, AM-2201, JWH-019, JWH-081, JWH-122, Acetaminophen, Caffeine, Dyphylline, Methaqualone, Theophylline, CP47.497, CP47.497 C8 Homologue, HU-211, JWH-200, JWH-250, Ibuprofen, Naproxen, R,R-(-)-Pseudoephedrine, Ethosuximide, $(\pm)$ Epinephrine, Norepinephrine, Barbital, Metanephrine, Normetanephrine, Methocarbamol, Alprazolam, Cimetidine, Citalopram, Clopidogrel, bisulfate, Fluconazole, Hydrochlorothiazide, Lamotrigine, L-Thyroxine, Methyl Phenidate, Omeprazole, Amlopidine Besylate, Atorvastatin, Azithromycin, Bupivacaine, Certrizine, Dimenhydrinate, Lisinopril, Methsuximide, Phensuximide, N-Normethyl Suximide, Butabarbital, Amobarbital, Secobarbital, Hexobarbital, Phenobarbital, Mephytoin, Ethotoin, Mephobarbital, PEMA, 10,11-Dihydro-carbamazepine, Medazepam, Chlorpramazine, Flurazepam, Lorazepam, Temazepam, Bromazepam, Primodone, 5,5-Diphenyl Hydantoin, Triamterene, Nordoxepin, Oxazepam, Levitriacetam, Metformin, Phenytoin, R-Phenylephrine, Sertraline, Topiramate, Zolpidem Tartrate, Vanilmandelic acid, 5-Hydroxy Indole 3-Acetic acid
## Immunoassay: Cross-reactivity with structurally related compounds
Hair samples (8 mg each) were spiked with the compounds listed in the tables below. Compounds that are potentially unlikely to cross-react were tested in duplicate at concentrations of 100 ng/10 mg hair. Potentially cross-reacting compounds were tested individually in duplicate up to concentrations of 1000 ng/10 mg hair. The samples were then extracted and analyzed by the assay device, followed by comparison to results from the cutoff calibrator. For structurally related compounds that were expected to show cross-reactivity, additional concentrations were tested to determine the concentration that is equivalent to the cutoff shown in the 1st table below. Compounds that did not show cross-reactivity are shown in the 2nd table below.
| Cross Reactivities of Structurally Related Compounds | | |
| --- | --- | --- |
| Compound | % Cross Reactivity | Concentration Equivalent to 5.0 ng Cocaine/10 mg Hair |
| Benzoylecgonine | 2.0 | 250 |
| Norcocaine | 20 | 25 |
| Cocaethylene | 55.5 | 9 |
| Ecgonine | <1 | >500 |
| Ecgonine Methyl Ester | <1 | >500 |
| Anhydroecgonine Methyl Ester | <1 | >500 |
| Compounds That were not found to Cross React in the Cocaine HEIA (1000 ng/10 mg hair) | | | | |
| --- | --- | --- | --- | --- |
| Lisinopril Dihydrate | Atropine | Bupropion | Cotinine | Cannabinol |
| Chlorpheniramine maleate | O-Desmethylvenlafaxine | Desipiramine | Doxylamine succinate | 1S, 2R Ephedrine |
| Amitriptyline | Dextromethorphan | Lidocaine | Methocarbamol | Nordoxepin |
| Pentazocine | Phenylephrine | Triamterene | Naproxen | Nicotine |
K201228 - Page 8 of 13
{8}
| Compounds That were not found to Cross React in the Cocaine HEIA (1000 ng/10 mg hair) | | | | |
| --- | --- | --- | --- | --- |
| Nortriptyline | Propoxyphene | R,R-pseudoephedrine | Thioridazine | Cis-Tramadol |
| Venlafaxine hydrochloride | 8- (-)-11-nor-9-carboxy-delta 9 THC | 11-nor-9-carboxy-delta 9 THC | Amoxicillin | Propanolol |
| Promethazine | Phenmetrazine | Phendimetrazine | Benzocaine | Dimenhydrinate |
| Metanephrin | Carbamazepine | Diazepine | Nordiazepam | Oxazepam |
| Acetominophen | Caffeine | Dyphylline | Methaqualone | Theophylline |
| Amphetamine | Phenmetrazine | Imipramine | Methamphetamine | Phencyclidine |
| Phenylpropanolamine | Amitryptiline | Dextromethorphan | Lidocaine | Methocarbamol |
| Nordoxepin | Pentazocine | Phenylephrine | Triamterene | Ethosuximide |
| Alpha-methyl-alpha-propylsuccinimide | Metharbital | Barbital | Methsuximide | Phensuximide |
| N-Normethsuximide | Mephyton | Ethotoin | Mephobarbital | PEMA |
| Phenobarbital | Methyl PEMA | 10, 11-Dihydrocarbamazepine | Primidone | 5,5-Diphenylhydantoin |
| 4-Methylprimidone | Butabarbital | Amobarbital | Secobarbital | Hexobarbital |
| Phenobarbital | Glutethimide | Methaqualone | Diazepam | Chlorpromazine |
| Flurazepam | AM2201 | JWH-019 | JWH-081 | JWH-122 |
| CP47,497 (±) | CP47, 497 (±), C8 Homologue | HU-211 | JWH-200 | JWH-250 |
| Ibuprofen | Ephinephrine (±) | Norephenephrine (±) | Metanephrine (±) | Normetanephrine (±) |
| Vanilmandelic acid (±) | 5-Hydroxyindole-3-acetic acid | Homovanillic acid | Alprazolam | Cimetidine |
| Citalopram HBr | Clonazepam | Clopidogrel bisulfate | Dextromethorphan | Fluconazole |
| Hydrochlorothiazide | Lamotrigine | L-Thyroxine | Methylphenidate HCl | Omeprazole |
| Levetiracetam | Sertaline HCl | Topiramate | Zolpidem Tartrate | Zonisamide |
| Amlopidine Besylate | Atorvastatin Calcium Salt | Azithromycin Dihydrate | Bupivacaine HCl | Cetrizine Di-HCl |
## Immunoassay: Effect of cosmetic treatments
Ten cocaine-negative head hair samples were treated with perm, dye, shampoo and relaxer and the results compared to the same samples without treatments. Seven cocaine-positive head hair samples were treated with perm, dye, shampoo and relaxer and the results compared to the same samples without the treatments. All of the positive samples were previously analyzed by the confirmation procedure for cocaine. After the treatments these samples and an aliquot of the same hair samples untreated were extracted for cocaine analysis and then assayed by the HEIA cocaine assay. The EIA results with and without treatment are compared
K201228 - Page 9 of 13
{9}
to detect interference in the assay or loss of drug due to the treatments. In each case of the 10 negative samples treated with a type of cosmetic product, all samples remained negative before and after the treatments. In each case of the 7 cocaine positive samples treated with a type of cosmetic product, all samples remained positive before and after the treatments.
## Specificity of the LC-MS/MS method
An interference study was conducted for the LC-MS/MS method by spiking potential interferents into hair specimens containing cocaine and cocaine metabolites, and the results were compared to hair specimens containing cocaine and cocaine metabolites without potential interferent. All of the samples spiked with the potential interfering compounds and cocaine (2 ng/10 mg hair) and cocaine metabolites (Benzoylecgonine, Cocaethylene, Norcocaine each at 0.2 ng/10 mg hair and Para-hydroxycocaine, Meta-hydroxycocaine, Ortho-hydroxycocaine each at 0.4pg/mg hair) produced a concentration within ± 20% of the target value. The sponsor concluded that the potentially interfering compounds listed below did not cross-react or cause interference with the LC-MS/MS assay.
Interferents tested at 20 ng/10mg hair: Clonazepam, Estazolam, Flurazepam, Flunitrazepam, Midazolam, Nitrazepam, Prazepam, Triazolam, Zolpidem, Phentermine, R,R-(-)-Pseudoephedrine; Interferents tested at 50 ng/10 mg hair: Ibuprofen, Meprobamate, Naproxen, (+)-Propoxyphene; Interferents at 200 ng/10 mg hair: Morphine, Oxycodone, Codeine, Phencyclidine, S-(+)-Methamphetamine, S-(+)-Amphetamine, (±)-Methadone, Phenobarbital, Phenytoin, R-(-)-Phenylephrine HCl, Carbamazepine, Salicylic Acid, Valproic Acid; Interferents tested at 500 ng/10 mg hair: Oxcarbazepine, Gabapentin, Acetaminophen, (-)-Cotinine, S-(-)-Nicotine, Caffeine; Interference Mix 1 (200 ng/10 mg hair): (±)-Methadone, Buprenorphine (SYN), cis-Tramadol HCl, Codeine, Fentanyl, Hydrocodone, Hydromorphone, Meperidine, Morphine, Naloxone, Naltrexone, Oxycodone, Oxymorphone; Interference Mix 2 (500 ng/10 mg hair): Gabapentin, Pregabalin, Salicylic acid, Valproic acid, Vigabatrin; Interference Mix 3 (100-2000 ng/10 mg hair): (+)-Propoxyphene, Aripiprazole, Lacosamide, Oxcarbazepine, Rufinamide, Warfarin; Interference Mix 4 (100 ng/ 10 mg hair): Carbamazepine, Levetiracetam, Metformin HCl, Phenobarbital, Phenytoin, R(-)-Phenylephrine HCl, Sertraline hydrochloride, Topiramate, Zolpidem tartrate, Zonisamide; Interference Mix 5 (2.5-50 ng/ 10 mg hair): Amlodipine besylate, Atorvastatin calcium salt, Azithromycin dihydrate, Bupivacaine HCl.1H2O, Cetirizine dihydrochloride, Dimenhydrinate, Lisinopril dihydrate, Loratadine; Interference Mix 6 (25-50 ng/ 10 mg hair): Montelukast sodium salt, Pioglitazone hydrochloride, Prednisolone, Prednisone, Procainamide HCl, Simvastatin.
## 4. Assay Reportable Range:
Not applicable. The screening immunoassay is a qualitative test only.
## 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
### Traceability:
The cocaine calibrator is traceable to a commercial standard with a certified reference material Certificate of Analysis.
K201228 - Page 10 of 13
{10}
K201228 - Page 11 of 13
## Sample Storage and Shipping Stability:
Five hair samples containing cocaine were stored in the same collection foil and card-envelope provided with the hair collection kit. After a period of 5 months, the samples were shipped overnight to another location and returned to the original location by overnight shipping. Samples were tested by the routine LC-MS/MS procedure both before being shipped and after their return. The sponsor stated that all samples that tested positive via the HEIA device and confirmed positive via LC-MS/MS before shipping and storage remained positive after shipping and storage.
## 6. Detection Limit:
The screening immunoassay is a qualitative test only; therefore, a detection limit evaluation is not applicable. See section VII.A.1. above for sensitivity around the immunoassay device cutoff.
## Detection Limit of the LC-MS/MS method
The sponsor performed a study to determine the Lower limit of Quantitation (LLOQ) of the LC-MS/MS assay for cocaine and cocaine metabolites.
The LLOQ is the lowest claimed concentration that can be quantitated within 20% of the expected value for COC, BE, CE and NCOC, and within ± 25% for p-OHCOC, m-OHCOC, o-OHCOC. This was found to be 0.25 ng/10 mg hair for COC, BE, CE and NCOC and 0.4 pg/mg hair for p-OHCOC, m-OHCOC, and o-OHCOC.
## 7. Assay Cut-Off:
Analytical performance of the device around the claimed cutoff is described in precision section VII.A.1 above.
## B Comparison Studies:
### 1. Method Comparison with Predicate Device:
#### Accuracy of the immunoassay screening method
A total of 92 de-identified hair samples were used to compare the results of the Psychemedics HEIA for Cocaine in Hair with the LC-MS/MS method. The studies were comprised of the following hair samples: 79 samples from males, 13 samples from females; 39 black hair samples, 47 brown hair samples (from light brown to dark brown), 6 salt/pepper hair samples, 65 head hair samples, and 27 body hair samples. The hair samples were then tested using the Psychemedics Microplate EIA for Cocaine in Hair assay and the Psychemedics LC-MS/MS confirmatory assay, results from both methods were compared. The comparison of Psychemedics Microplate EIA for Cocaine in Hair assay with LC-MS/MS confirmatory assay is shown in the following table.
{11}
K201228 - Page 12 of 13
| Comparison of Cocaine HEIA Screening Results with Unwashed Cocaine LC/MS/MS | | | | |
| --- | --- | --- | --- | --- |
| HEIA Result | LC/MS/MS Result, ng Cocaine/10 mg hair (% of cutoff calibrator) | | | |
| | < 2.50 (< 50% below cutoff) | 2.50 – 4.99 (≥ 50% below cutoff to cutoff) | 5.00 – 7.50 (cutoff to ≤50% above cutoff) | > 7.50 (> 50% above cutoff) |
| Positive | 0 | 2 | 4 | 35 |
| Negative | 44 | 7 | 0 | 0 |
| Comparison of Cocaine HEIA Screening Results with Washed Cocaine LC/MS/MS | | | | |
| --- | --- | --- | --- | --- |
| HEIA Result | LC/MS/MS Result, ng Cocaine/10 mg hair (% of cutoff calibrator) | | | |
| | < 2.50 (< 50% below cutoff) | 2.50 – 4.99 (≥ 50% below cutoff to cutoff) | 5.00 – 7.50 (cutoff to ≤ 50% above cutoff) | > 7.50 (> 50% above cutoff) |
| Positive | 0 | 2 | 8 | 31 |
| Negative | 48 | 3 | 0 | 0 |
Discordant Results Between HEIA and Unwashed LC/MS/MS
| Sample | HEIA Result | LC/MS/MS Result (ng Cocaine/10 mg hair) |
| --- | --- | --- |
| 1 | POS | 4.43 |
| 2 | POS | 4.96 |
Discordant Results Between HEIA and Washed LC/MS/MS
| Sample | HEIA Result | LC/MS/MS Result (ng Cocaine/10 mg hair) |
| --- | --- | --- |
| 1 | POS | 3.75 |
| 2 | POS | 4.27 |
A separate set of 30 de-identified samples were prepared by weighing approximately 12 mg per new sample aliquot. The samples were then confirmed unwashed by LC/MS/MS. The studies were comprised of the following hair samples: 29 samples from males, 1 sample from a female; 21 black hair samples, 7 brown hair samples (from light brown to dark brown), 2 salt/pepper hair samples; 14 head hair samples and 16 body hair samples. The hair samples were then tested using the Psychemedics Microplate EIA for Cotinine in Hair assay and the Psychemedics LC-MS/MS confirmatory assay, results from both methods were compared. The comparison of Psychemedics Microplate EIA for Cotinine in Hair assay with LC-MS/MS confirmatory assay is shown in the following table.
| Comparison of Cocaine HEIA Screening Results with Unwashed Cocaine LC/MS/MS | | | | |
| --- | --- | --- | --- | --- |
| HEIA Result | LC/MS/MS Result, ng Cocaine/10 mg hair (% of cutoff calibrator) | | | |
| | < 2.50 (< 50% below cutoff) | 2.50 – 5.0 (≥ 50% below cutoff to cutoff) | 5.01 – 7.50 (cutoff to ≤50% above cutoff) | > 7.50 (> 50% above cutoff) |
| Positive | 0 | 0 | 3 | 13 |
| Negative | 11 | 1 | 2 | 0 |
{12}
| Discordant Results Between HEIA and Unwashed LC/MS/MS | | |
| --- | --- | --- |
| Sample | HEIA Result | LC/MS/MS Result (ng Cocaine/10 mg hair) |
| 1 | POS | 4.43 |
| 2 | POS | 4.96 |
2. Recovery Study
Recovery from LC-MS/MS Analysis
The sponsor conducted a study to evaluate recovery for cocaine, benzoylecgonine, cocaethylene and norcocaine at a concentration of 5 ng/10 mg hair. Ten tubes were prepared and analyzed by LC/MS/MS. The recovery of the analytes ranged from 97% to 100%.
3. Matrix Comparison:
Not Applicable.
C Clinical Studies:
1. Clinical Sensitivity:
Not Applicable.
2. Clinical Specificity:
Not Applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not Applicable.
D Clinical Cut-Off:
Not Applicable.
E Expected Values/Reference Range:
Not Applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and substantially equivalent to the predicate.
K201228 - Page 13 of 13
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.