IMMAGE IMMUNOCHEMISTRY SYSTEM GENTAMICIN(GEN)/TOBRAMYCIN(TOB) REAGENT / BECKMAN DRUG CLAIBRATOR 3 PLUS
K964009 · Beckman Instruments, Inc. · DNO · Feb 11, 1997 · Clinical Toxicology
Device Facts
Record ID
K964009
Device Name
IMMAGE IMMUNOCHEMISTRY SYSTEM GENTAMICIN(GEN)/TOBRAMYCIN(TOB) REAGENT / BECKMAN DRUG CLAIBRATOR 3 PLUS
Applicant
Beckman Instruments, Inc.
Product Code
DNO · Clinical Toxicology
Decision Date
Feb 11, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3450
Device Class
Class 2
Indications for Use
The IMMAGE Immunochemistry System Gentamicin (GEN) reagent, when used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 3 Plus, is intended for the quantitative determination of cgentamicin in human serum or plasma by rate nephelometric inhibition immunoassay. The IMMAGE Immunochemistry System Tobramycin (TOB) reagent, when used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 3 Plus, is intended for the quantitative determination of tobramycin in human serum or plasma by rate nephelometric inhibition immunoassay. Beckman Drug Calibrator 3 Plus, used in conjunction with IMMAGE reagents, is intended for use on Beckman's IMMAGE™ Immunochemistry Systems for the calibration of Gentamicin and Tobramycin test systems.
Device Story
The IMMAGE Immunochemistry System uses liquid-stable reagents and Beckman Drug Calibrator 3 Plus to measure gentamicin and tobramycin concentrations in human serum or plasma. The system employs a rate nephelometric inhibition immunoassay principle. Samples are processed on the IMMAGE Immunochemistry System, which detects light scattering changes to quantify drug levels. The device is intended for clinical laboratory use. Results assist healthcare providers in therapeutic drug monitoring to optimize dosing and minimize toxicity. The system provides quantitative outputs used for clinical decision-making regarding patient drug therapy.
Clinical Evidence
Bench testing only. Method comparison studies against Abbott TDx predicates showed high correlation for Gentamicin (n=135, r=0.998, slope=0.994) and Tobramycin (n=94, r=0.996, slope=1.009). Within-run imprecision studies (n=80 per level) demonstrated %C.V. values ranging from 1.7% to 6.2% across therapeutic ranges. Stability testing confirmed 24-month shelf-life and 14-day open-container/calibration stability.
Technological Characteristics
Liquid-stable reagents; rate nephelometric inhibition immunoassay principle; mouse-derived antibodies; automated system-based analysis; intended for use with Beckman IMMAGE Immunochemistry System.
Indications for Use
Indicated for the quantitative determination of gentamicin and tobramycin concentrations in human serum or plasma for therapeutic drug monitoring.
Regulatory Classification
Identification
A gentamicin test system is a device intended to measure gentamicin, an antibiotic drug, in human specimens. Measurements obtained by this device are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to ensure appropriate therapy.
K963673 — IMMAGE IMMUNOCHEMISTRY SYSTEM TDM (CAR,PHE,PHY, & THE) REAGENTS · Beckman Instruments, Inc. · Feb 4, 1997
Submission Summary (Full Text)
{0}
200 South Kraemer Boulevard, Box 8000. Brea. CA 92822-8000 • (714) 993-5321
# BECKMAN
## Summary of Safety & Effectiveness
### IMMAGE™ Immunochemistry System Therapeutic Drug Monitoring Reagents
#### [Gentamicin (GEN) and Tobramycin (TOB)]
FEB 11 1997
### 1.0 Submitted By:
Annette Hellie
Regulatory Specialist, Product Submissions
Beckman Instruments, Inc.
200 S. Kraemer Blvd., W-337
Brea, California 92822-8000
Telephone: (714) 993-8767
FAX: (714) 961-4457
### 2.0 Date Submitted:
03 October 1996
### 3.0 Device Name(s):
#### 3.1 Proprietary Names
IMMAGE™ Immunochemistry System Gentamicin (GEN) Reagent
IMMAGE™ Immunochemistry System Tobramycin (TOB) Reagent
Beckman Drug Calibrator 3 Plus
#### 3.2 Classification Names
Gentamicin Test System (21 CFR §862.3450)
Tobramycin Test System (21 CFR §862.3900)
Calibrator (21 CFR §862.1150)
### 4.0 Predicate Device(s):
| IMMAGE System Reagent | Predicate | Manufacturer | Docket Number |
| --- | --- | --- | --- |
| Gentamicin (GEN) Reagent | TDx®** Gentamicin Reagent | Abbott® Laboratories, Inc. | K904226 |
| Tobramycin (TOB) Reagent | TDx Tobramycin Reagent | Abbott Laboratories, Inc. | K904226 |
*Abbott Laboratories, Abbott Park, IL 60064
**Trademark of Abbott Laboratories
Beckman Instruments, Inc.
twx: 910-592-1260 • telex: 06-78413
{1}
Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System TDM (GEN & TOB) Reagents
Summary of Safety & Effectiveness
## 5.0 Description:
The IMMAGE Immunochemistry System (TDM) GEN and TOB Reagents in conjunction with Beckman Drug Calibrator 3 Plus, are intended for use in the quantitative determination of gentamicin and tobramycin concentrations respectively in human serum and plasma samples on Beckman's IMMAGE Immunochemistry System.
## 6.0 Intended Use:
The IMMAGE Immunochemistry System Gentamicin (GEN) reagent, when used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 3 Plus, is intended for the quantitative determination of cgentamicin in human serum or plasma by rate nephelometric inhibition immunoassay.
The IMMAGE Immunochemistry System Tobramycin (TOB) reagent, when used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 3 Plus, is intended for the quantitative determination of tobramycin in human serum or plasma by rate nephelometric inhibition immunoassay.
Beckman Drug Calibrator 3 Plus, used in conjunction with IMMAGE reagents, is intended for use on Beckman's IMMAGE™ Immunochemistry Systems for the calibration of Gentamicin and Tobramycin test systems.
## 7.0 Comparison to Predicate(s):
The following table shows similarities and differences between the predicates identified in Section 4.0 of this summary.
| Reagent | Aspect/Characteristic | Comments |
| --- | --- | --- |
| GEN | | |
| IMMAGE System (GEN & TOB) Reagents | Intended use. | Same as Abbott TDx Reagents |
| | Liquid stable reagents. | |
| TOB | | |
| IMMAGE System (GEN & TOB) Reagents | IMMAGE utilizes nephelometric inhibition immunoassay | Abbott TDx reagents utilize fluorescence polarization immunoassay |
| IMMAGE System (GEN & TOB) Reagents | Antibody source for IMMAGE GEN & TOB reagents is mouse. | Antisera source for TDx Gentamicin and Tobramycin is sheep. |
file: dc3510k.sae
{2}
Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System TDM (GEN & TOB) Reagents
Summary of Safety & Effectiveness
## 8.0 Summary of Performance Data:
The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method comparison, stability, and imprecision experiments that relate results obtained from the Abbott TDx Reagents to the IMMAGE System Reagents.
### Method Comparison* Study Results
IMMAGE System GEN, TOB, PHY, and THE Reagents
| Analyte | Slope | Intercept | r | n | Predicate Method |
| --- | --- | --- | --- | --- | --- |
| Gentamicin | 0.994 | -0.05 | 0.998 | 135 | Abbott TDx
Gentamicin |
| Tobramycin | 1.009 | -0.12 | 0.996 | 94 | Abbott TDx
Tobramycin |
* Deming Regression statistics used
### Stability Study Results
| Reagent | Product Claim |
| --- | --- |
| IMMAGE GEN & TOB | 24 month shelf-life
14 day open container stability
14 day calibration stability |
### Estimated Within-Run Imprecision
| Sample | Mean (μg/mL) | S.D. (μg/mL) | %C.V. | N |
| --- | --- | --- | --- | --- |
| Gentamicin (GEN) | | | | |
| Level 1 | 1.52 | 0.095 | 6.2 | 80 |
| Level 2 | 5.96 | 0.107 | 1.8 | 80 |
| Level 3 | 11.0 | 0.26 | 2.4 | 80 |
| Sample | Mean (μg/mL) | S.D. (μg/mL) | %C.V. | N |
| --- | --- | --- | --- | --- |
| Tobramycin (TOB) Reagent | | | | |
| Level 1 | 1.5 | 0.073 | 4.9 | 80 |
| Level 2 | 5.93 | 0.098 | 1.7 | 80 |
| Level 3 | 10.9 | 0.23 | 2.1 | 80 |
This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.
file: dc3510k.sse
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.