SYVA EMIT II PLUS ETHYL ALCOHOL ASSAY, MODELS 9K309UL/9K409UL

K993980 · Syva Co. · DML · Feb 1, 2000 · Clinical Toxicology

Device Facts

Record IDK993980
Device NameSYVA EMIT II PLUS ETHYL ALCOHOL ASSAY, MODELS 9K309UL/9K409UL
ApplicantSyva Co.
Product CodeDML · Clinical Toxicology
Decision DateFeb 1, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3040
Device ClassClass 2

Indications for Use

Emit® II Plus Ethyl Alcohol Assay is an enzyme assay. The assay is intended for use in the quantitative analysis of ethyl alcohol in human serum or plasma. The Emit® II Plus Ethyl Alcohol Assay is an enzyme assay intended for use in the quantitative rne Emit - 11 Tub any) Aconol in human urine, plasma, and serum. The Emit® II Plus Ethyl Alcohol Assay is designed for use with most clinical chemistry analyzers.

Device Story

Emit® II Plus Ethyl Alcohol Assay is an in vitro diagnostic enzyme assay for quantitative ethanol measurement in human urine, serum, or plasma. Used in clinical laboratory settings with standard clinical chemistry analyzers; operated by laboratory personnel. The assay utilizes enzymatic reactions to detect ethanol concentrations; results are provided to clinicians to assist in patient assessment. Benefits include rapid, accurate quantification of alcohol levels in biological samples to support clinical decision-making.

Clinical Evidence

Bench testing only. Performance validated through precision studies (within-run %CV 1.39-1.55%; total %CV 1.61-1.78%), linearity testing, and spike recovery (95.7-101.5%). Interference studies evaluated bilirubin, hemoglobin, triglycerides, and LDH/Lactate. Sensitivity reported as <3.0 mg/dL. Correlation to predicate method showed r=0.986 (serum) and r=0.997 (urine).

Technological Characteristics

Enzyme-based in vitro diagnostic reagent kit. Designed for use on clinical chemistry analyzers. Quantitative range 10-600 mg/dL. Sensitivity <3.0 mg/dL. Compatible with various anticoagulants (EDTA, Sodium Citrate, Sodium Heparin, Oxalate/Fluoride).

Indications for Use

Indicated for quantitative analysis of ethyl alcohol in human urine, serum, or plasma using clinical chemistry analyzers.

Regulatory Classification

Identification

An alcohol test system is a device intented to measure alcohol (e.g., ethanol, methanol, isopropanol, etc.) in human body fluids (e.g., serum, whole blood, and urine). Measurements obtained by this device are used in the diagnosis and treatment of alcohol intoxication and poisoning.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS For Emit® Ethyl Alcohol Assay #### 1. Manufacturer and Contact Information: | Manufacturer: | Syva Company | |---------------|-------------------------| | | 3403 Yerba Buena Rd. | | | P.O. Box 49013 | | | San Jose, CA 95161-9013 | Paul Rogers Contact Information: Syva Company - Dade Behring Inc. 3403 Yerba Buena Road San Jose, CA 95161-9013 Tel: 408-239-2000 ### 2. Device Classification Name: The Clinical Chemistry and Clinical Toxicology Devices Panel have classified "Ethyl Alcohol System" as Class II. #### 3. Intended Use: Emit® II Plus Ethyl Alcohol Assay is an enzyme assay. The assay is intended for use in the quantitative analysis of ethyl alcohol in human serum or plasma. #### 4. Device Description and Characteristics: This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of SMDA 1990. Syva Company - Dade Behring Inc. is submitting the Premarket Notification, 510(k) with the intention to commercially market an in vitro diagnostic reagent test kit for the analysis of ethyl alcohol in human urine, serum, or plasma. The Syva Emit® II Plus Ethyl Alcohol Assay is an enzyme assay intended for use in quantitative analysis of ethyl alcohol in human urine, serum, or plasma. The Syva Emit® II Plus Ethyl Alcohol Assay has been found to be equivalent to the predicate device: Emit® Ethyl Alcohol Assay (K903153) with regard to intended use, assay sample, and overall performance characteristics. Comparative Analysis: The Syva Emit® II Plus Ethyl Alcohol Assay showed excellent correlation to the predicate method. The comparative analysis to the predicate method resulted in a correlation of 0.986 with serum and 0.997 with urine, with a slope value of 1.000 with both urine and serum. {1}------------------------------------------------ ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS For Emit® Ethyl Alcohol Assay (cont.) Precision: A Precision study was performed and the Syva Emit® II Plus Ethyl Alcohol Assay demonstrated acceptable within-run precision with coefficients of variation (%CV) ranging from 1.39% to 1.55% and acceptable total precision with coefficients of variation (%CV) ranging from 1.61% to 1.78%. Specificity: A study was performed on potentially cross-reacting compounds and the Syva Emit® II Plus Ethyl Alcohol Assay demonstrated acceptable levels of cross-reactivity, ranging from <0.1 -- 14.2%. Sensitivity: The sensitivity level of the Emit® II Plus Ethyl Alcohol Assay is less than 3.0 mg/dL. This level represents the lowest concentration of amphetamines that can be distinquished from 0 ng/mL with a confidence level of 95%. Endogenous Interference: Endogenous interference in serum, due to either bilirubin, hemoglobin, or trigvcerides, in the Syva Emit® II Plus Ethyl Alcohol Assay are 99, 100, and 97% recovery compared to controls. The presence of LDH/Lactate does not cause a significant interference in the Svva Emit® II Plus Ethyl Alcohol Assay (0.4 mg/dL), A series of potentially interfering endogenous substances in urine were tested by the Syva Emit® II Plus Ethyl Alcohol Assav. The recoveries for the interfering substances range from 99 - 104%. Spiked Sample Recovery: Spike recovery of ethanol in human serum and urine by the Syva Emit® II Plus Ethyl Alcohol Assay was performed across the quantitative range of between 10 and 600 mg/dL. Recovery of ethanol spikes ranged between 95.7 - 101.5%. High Sample Dilution: Dilution of high serum and urine samples by water the Negative Calibrator were performed in the Syva Emit® II Plus Ethyl Alcohol Assav. Recovery from the normal concentrations in both types of samples diluted with either water or Negative Calibrator ranged from 97.1 - 100.9%. Anticoagulants: The performance of the anticoagulants EDTA (K3), Sodium Citrate, Sodium Heparin, and Oxalate/Fluoride, as compared to serum, was assayed by the Syva Emit® II Plus Ethyl Alcohol Assay. Average recovery versus the serum control ranged from95.5 – 102.3%. Linearity: The linearity of diluted serum and urine samples were tested in the Syva Emit® II Plus Ethyl Alcohol Assay. Both serum and urine samples are considered linear according to the acceptance criteria of the CAP linearity survey. Urine Preservatives: the effect of urine preservatives was tested by the Syva Emit® II Plus Ethyl Alcohol Assay. Recoveries of urine preservatives compared to controls ranged from 97.5 -- 99.1%. {2}------------------------------------------------ # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS For Emit® Ethyl Alcohol Assay (cont.) Calibration Stability: The calibration stability of the Syva Emit® II Plus Ethyl Alcohol Assay on the SYVA- 30R Biochemical System was determined to be 25 days. ## 5. Substantial Equivalence: In conclusion, Syva Company considers the Emit® II Plus Ethyl Alcohol Assay to be substantially equivalent to the Ethyl Alcohol Assay (K903153) with regard to intended use, assay sample, and overall performance characteristics. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird in flight, composed of three curved lines. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 FEB 1 2000 Mr. Paul L. Rogers Jr. Senior Manager, Regulatory Affairs Syva Company - Dade Behring Inc. P.O. Box 49013 3403 Yerba Buena Road San Jose, California 95161-9013 Re: K993980 > Trade Name: Emit® II Plus Ethyl Alcohol Assay Regulatory Class: II Product Code: DML Dated: November 23, 1999 Received: November 24, 1999 Dear Mr. Rogers: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Autman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (If known): Emit® II Plus Ethyl Alcohol Assay Device Name: ## Indications for Use: The Emit® II Plus Ethyl Alcohol Assay is an enzyme assay intended for use in the quantitative rne Emit - 11 Tub any) Aconol in human urine, plasma, and serum. The Emit® II Plus Ethyl Alcohol Assay is designed for use with most clinical chemistry analyzers. # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use_ (Optional Format 1-2-96) Device Name: Emit® II Plus Ethyl Alcohol Assay Patricia Bernhardt for S.D.C. Division of Clinical Laboratory Devices 510(k) Number. く.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...