PARAMAX SALICYLATE REAGENT
Device Facts
| Record ID | K964715 |
|---|---|
| Device Name | PARAMAX SALICYLATE REAGENT |
| Applicant | Dade Intl., Inc. |
| Product Code | DKJ · Clinical Toxicology |
| Decision Date | Apr 25, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3830 |
| Device Class | Class 2 |
Indications for Use
The proposed and the predicate device are intended for use as quantitative determination of salicylate in plasma and serum.
Device Story
Paramax® Salicylate Reagent; in vitro diagnostic reagent for quantitative salicylate measurement in plasma and serum; utilizes colorimetric methodology; used on Paramax® analytical systems; provides salicylate concentration values to clinicians; aids in monitoring salicylate levels; supports clinical decision-making regarding toxicity or therapeutic management.
Clinical Evidence
Bench testing only. Correlation study performed using 49 clinical samples compared against DuPont aca Salicylate Reagent. Results: correlation coefficient 0.999; regression equation: Paramax® value = (1.16 x aca value) - 1.8 mg/dL.
Technological Characteristics
Colorimetric reagent system; quantitative assay for salicylate; designed for use on automated clinical chemistry analyzers (Paramax® system).
Indications for Use
Indicated for the quantitative determination of salicylate in human plasma and serum for clinical diagnostic purposes.
Regulatory Classification
Identification
A salicylate test system is a device intended to measure salicylates, a class of analgesic, antipyretic and anti-inflammatory drugs that includes aspirin, in human specimens. Measurements obtained by this device are used in diagnosis and treatment of salicylate overdose and in monitoring salicylate levels to ensure appropriate therapy.
Predicate Devices
- DuPont aca Salicylate Reagent (K771805)
Related Devices
- K014173 — SYNCHRON SYSTEMS SALICYLATE REAGENT AND CALIBRATOR · Beckman Coulter, Inc. · Feb 15, 2002
- K042329 — SALICYLATE-SL ASSAY · Diagnostic Chemicals , Ltd. · Dec 27, 2004
- K093939 — ADVIA CHEMISTRY SALICYLATE REAGENT · Siemens Healthcare Diagnostics · May 19, 2010
- K981872 — SALICYLATE-SL ASSAY CATALOGUE NUMBER 511-20,511-40 · Diagnostic Chemicals , Ltd. · Jun 29, 1998