THERMO SCIENTIFIC CEDIA MULTI-DRUG OFT CALIBRATORS
Applicant
Microgenics Corp.
Product Code
DKB · Clinical Toxicology
Decision Date
Apr 8, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3200
Device Class
Class 2
Indications for Use
The CEDIA® Multi-Drug OFT Calibrators are intended for use in the calibration of d-Amphetamine, Benzoylecgonine, Morphine, and Phencyclidine (PCP) in human Oral Fluid when used with the CEDIA Amphetamine, Cocaine, Opiate, and Phencyclidine (PCP) OFT Assays on the MGC 240 analyzer. This in vitro diagnostic device is intended for clinical laboratory use only.
Device Story
CEDIA® Multi-Drug OFT Calibrators are liquid, ready-to-use solutions containing known concentrations of d-Amphetamine, Benzoylecgonine, Morphine, and Phencyclidine (PCP) spiked into a buffer matrix. The device consists of three levels: Negative, Cutoff, and High. It is used in clinical laboratories by technicians to calibrate the CEDIA® Multi-Drug OFT Assays on the MGC 240 analyzer. The Cutoff Calibrator serves as a qualitative reference to distinguish positive from negative patient samples. Value assignment for the calibrators is performed using LC-MS/MS. The device ensures the accuracy and reliability of the associated drug assays, facilitating clinical decision-making regarding drug testing results.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated through bench testing to verify device performance and design specifications.
Technological Characteristics
Liquid, ready-to-use calibrators; buffer matrix; stored at 2-8°C. Used for calibration of automated clinical chemistry analyzers (MGC 240).
Indications for Use
Indicated for clinical laboratory use to calibrate d-Amphetamine, Benzoylecgonine, Morphine, and Phencyclidine (PCP) assays in human oral fluid samples.
Regulatory Classification
Identification
A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K051088 — MULTIPLE ANALYTE URINE DRUGS OF ABUSE CALIBRATORS AND CONTROLS · Lin-Zhi International, Inc. · Jun 24, 2005
K021393 — PRECISET DAT PLUS · Quantimetrix Corp. · Jun 13, 2002
K031775 — PRECISET DAT PLUS I CALIBRATORS · Roche Diagnostics Corp. · Aug 5, 2003
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k101752
B. Purpose for Submission:
New device
C. Measurand:
Amphetamine, Benzoylecgonine, Morphine, Phencyclidine
D. Type of Test:
Calibrator materials
E. Applicant:
Microgenics Corporation
F. Proprietary and Established Names:
Thermo Scientific CEDIA® Multi-Drug OFT Calibrators
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| DKB | Class II | 21 CFR§ 862.3200 | Toxicology (91) |
H. Intended Use:
1. Intended use(s):
See Indications for Use below.
2. Indication(s) for use:
The CEDIA® Multi-Drug OFT Calibrators are intended for use in the calibration of d-Amphetamine, Benzoylecgonine, Morphine, and Phencyclidine (PCP) in
{1}
human Oral Fluid when used with the CEDIA Amphetamine, Cocaine, Opiate, and Phencyclidine (PCP) OFT Assays on the MGC 240 analyzer. This in vitro diagnostic device is intended for clinical laboratory use only.
3. Special conditions for use statement(s):
For in vitro diagnostic use
For prescription use
4. Special instrument requirements:
For use with CEDIA® Amphetamine, Cocaine, Opiate, and Phencyclidine (PCP) OFT Assays on the MGC240 analyzer
I. Device Description:
The CEDIA® Multi-Drug OFT Calibrators are liquid ready-to-use solutions prepared by spiking known quantities of Amphetamine, Benzoylecgonine, Morphine, and PCP into buffer matrix. The CEDIA® Multi-Drug OFT Calibrators consists of 3 calibrator levels: Negative Calibrator, Cutoff Calibrator, and High Calibrator. The Cutoff Calibrator is used as a qualitative cutoff reference for distinguishing "positive" from "negative" samples. The volume per vial is 20.0 mL for the Negative Calibrator and 10.0 mL for the Cutoff and High Calibrator. Amphetamine, Benzoylecgonine, Morphine, and PCP are value assigned for use with CEDIA® OFT Assays. The Negative Calibrator is free of analyte.
J. Substantial Equivalence Information:
1. Predicate device name(s):
CEDIA® DAU Multi-Drug Calibrators
2. Predicate 510(k) number(s):
k980853
3. Comparison with predicate:
Similarities
| Item | Candidate Device | Predicate Device |
| --- | --- | --- |
| Intended Use | The CEDIA® Multi-Drug OFT Calibrators are intended for use in the calibration of CEDIA® OFT Assays. | Same |
{2}
3
| Form | Liquid, ready to use | Same |
| --- | --- | --- |
| Storage temperature | 2 to 8° C | Same |
## Differences
| Item | Candidate Device | Predicate Device |
| --- | --- | --- |
| Analytes | Amphetamine, Benzolyecgonine, Morphine, Phencyclidine | Benzolecgonine, EDDP, d-Methamphetamine, Morphine, Nitrazepam, Phencyclidine, Secobarbital |
| Matrix | Buffer | Urine |
| Calibrator Levels | 3 Levels – Negative, Cutoff, High | 4 Levels – Primary, Secondary, Intermediate, High |
| Stability | Shelf Life – 16 months
Open Vial – 60 days | Until indicated expiration date |
## K. Standard/Guidance Document Referenced (if applicable):
None were referenced
## L. Test Principle:
Not applicable
## M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable
b. Linearity/assay reportable range:
Not applicable
{3}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
## Traceability
The sponsor claims traceability to all measurands in the calibrator (amphetamine, benzoylecgonine, morphine, phencyclidine) to a commercially available methanolic standard.
## Stability
Stability testing protocols and acceptance criteria were reviewed and found to be acceptable. Stability characteristics of the CEDIA® Multi-Drug OFT Calibrators were determined using accelerated and real time closed (unopened) vial, as well as real-time opened vial studies. A closed vial shelf-life of 16 months at the recommended storage temperature (2°C to 8°C) was demonstrated based on real-time studies. Open vial stability of 60 days was also demonstrated at the recommended storage temperature (2°C to 8°C). The closed vial shelf-life and open vial expiration dates are listed on the box and vial labels. In the package insert, the user is directed to check the expiration dates indicated on the box label. The recommendations in the labeling are to store calibrators at 2°C to 8°C.
## Value Assignment and Validation
Nominal values are assigned to production lots using Liquid Chromatography with tandem Mass Spectrometry (LC-MS/MS). The table below shows the measurand values contained in each level of calibrator materials. These values are also provided in the labeling.
| Analyte | Negative Calibrator (ng/mL) | Cutoff Calibrator (ng/mL) | High Calibrator (ng/mL) |
| --- | --- | --- | --- |
| Amphetamine | 0.0 | 50.0 | 200.0 |
| Benzoylecgonine | 0.0 | 5.0 | 50.0 |
| Morphine | 0.0 | 10.0 | 80.0 |
| Phencyclidine | 0.0 | 1.0 | 20.0 |
Three lots of each level of calibrator was validated using the CEDIA® Multi-Drug OFT Assays, a reference set of calibrators, and two MGC 240 Benchtop Clinical Chemistry Analyzers set to qualitative mode. The validation protocol, acceptance criteria, and data were reviewed and found to be acceptable.
d. Detection limit:
Not applicable
{4}
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The values for each analyte are stated in the labeling
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
{5}
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
6
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.