ADVIA CHEMISTRY DRUG CALIBRATOR II, MODEL 10376771
Applicant
Siemens Healthcare Diagnostics
Product Code
DKB · Clinical Toxicology
Decision Date
Jun 1, 2009
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 862.3200
Device Class
Class 2
Indications for Use
The ADVIA Chemistry TDM DRUG Calibrator II is for in vitro diagnostic use in the calibration of Carbamazepine_2 (CARB_2), Gentamicin_2 (GENT_2), Tobramycin_2 (TOBR_2), Valproic Acid_2 (VPA_2), and Vancomycin_2 (VANC_2) methods on the ADVIA Chemistry Systems. For prescription use only.
Device Story
ADVIA Chemistry TDM DRUG Calibrator II is a multi-analyte, liquid, bovine serum-based product; supplied as a kit with 5 levels of calibrators (5.0 mL per vial). Used for calibration of specific drug assays on ADVIA Clinical Chemistry Systems. Calibrator values are assigned using an ADVIA system calibrated with an internal Anchor Calibrator, which is traceable to USP material. Product is intended for professional use in clinical laboratory settings to ensure accurate quantification of therapeutic drugs in patient samples.
Clinical Evidence
No clinical data. Bench testing only. Stability established via real-time data (12 months unopened, 30 days open). Traceability established via internal Anchor Calibrator linked to USP material.
Technological Characteristics
Multi-analyte liquid calibrator; bovine serum matrix; 5 levels; 5.0 mL vials. Traceable to USP material. Stability: 12 months at 4°C (unopened), 30 days at 2-8°C (opened).
Indications for Use
Indicated for in vitro diagnostic calibration of Carbamazepine, Gentamicin, Tobramycin, Valproic Acid, and Vancomycin assays on ADVIA Chemistry Systems. No specific patient population or contraindications defined.
Regulatory Classification
Identification
A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Dade Behring Dimension Drug Calibrator II (DRUG CAL II - DC49D) (K033809)
Related Devices
K982871 — DRUG CALIBRATOR II · Dade Behring, Inc. · Oct 13, 1998
K990255 — DIMENSION DRUG CALIBRATOR II, MODEL DC49B · Dade Behring, Inc. · Feb 16, 1999
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k090900
B. Purpose for Submission:
New device
C. Measurand:
Carbamazepine, Gentamicin, Tobramycin, Valproic Acid, Vancomycin
D. Type of Test:
Calibrator Materials
E. Applicant:
Siemens Healthcare Diagnostics Inc.
F. Proprietary and Established Names:
ADVIA Chemistry DRUG Calibrator II
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| DKB | Class II | 21 CFR§ 862.3200 | Toxicology (91) |
H. Intended Use:
1. Intended use(s):
See Indications for Use below.
2. Indication(s) for use:
The ADVIA Chemistry TDM DRUG Calibrator II is for in vitro diagnostic use in the calibration of Carbamazepine_2 (CARB_2), Gentamicin_2 (GENT_2), Tobramycin_2 (TOBR_2), Valproic Acid_2 (VPA_2), and Vancomycin_2 (VANC_2) methods on the ADVIA Chemistry Systems.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
ADVIA Clinical Chemistry Systems
{1}
I. Device Description:
ADVIA Chemistry TDM DRUG Calibrator II is a multi-analyte, liquid, bovine serum based product containing multiple analytes. The kit consists of 2 vials of each of 5 calibrator levels which are ready for use (no preparation is required). The volume per vial is 5.0 mL. Tobramycin, Carbamazepine, Valproic Acid, Vancomycin and Gentamicin analytes are value assigned for ADVIA Chemistry systems.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Dade Behring Dimension DRUG Calibrator II
2. Predicate 510(k) number(s):
k033809
3. Comparison with predicate:
| SIMILARITIES | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| Intended Use | For in vitro diagnostic use in the calibration of Carbamazapine_2 (CARB_2), Gentamicin_2 (GENT_2), Tobramycin_2 (TOB_2), Valproic Acid_2 (VPA_2), and Vancomycin_2 (VANC_2) methods on the ADVIA Chemistry systems. | DRUG CAL II is an in vitro diagnostic product for the calibration of the following methods packaged in the Flex reagent cartridges: acetaminophen (ACTM), carbamazepine (CRBM), digitoxin (DGTX), gentamicin (GENT), lidocaine (LIDO), N-acetylprocainamide (NAPA), procainamide (PROC), tobramycin (TOBR), valproic acid (VALP), and vancomycin (VANC). |
| Form | Liquid | Liquid |
| Traceability | USP | USP |
| Matrix | Bovine | Bovine |
| Number of Levels | Five, 5.0 mL each vial | Five, 5.0 mL each vial |
| Stability | 12 months – shelf-life
30 days open vial | 12 months – shelf-life
30 days open vial |
| Packaging | Ten vials | Ten vials |
{2}
| DIFFERENCES | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| Measured Analytes (value assigned) | Carbamazepine (CARB_2), | carbamazepine (CRBM) |
| | Tobramycin (TOBR_2) | tobramycin (TOBR) |
| | Valproic Acid_2 (VPA_2) | acetaminophen (ACTM) |
| | Vancomycin_2 (VANC_2). | digitoxin (DGTX) |
| | Gentamicin_2 (GENT_2) | gentamicin (GENT) |
| | | lidocaine (LIDO) |
| | | N-acetylprocainamide (NAPA) |
| | | procainamide (PROC) |
| | | valproic acid (VALP) |
| | | vancomycin (VANC). |
# K. Standard/Guidance Document Referenced (if applicable):
- CEN 13640: Stability Testing of In Vitro Diagnostic Reagents
Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators (1999)
- Format for Traditional and Abbreviated 510(k)s - Guidance for Industry and FDA Staff (2005)
# L. Test Principle:
Not Applicable
# M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility: Not Applicable
b. Linearity/assay reportable range: Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: All measurands in the calibrator (carbamazepine, gentamicin, tobramycin, vancomycin, valproic acid) are traceable to an internal Anchor Calibrator.
{3}
The Anchor Calibrator is in turn traceable to USP material gravimetrically spiked into human serum.
## Stability:
The stability of the calibrators is established through real-time data. Testing is conducted at multiple time points and must pass pre-defined acceptance criteria. The unopened product refrigerated at 4°C is stable for 12 months. The opened product is stable for 30 days, refrigerated at 2 - 8 °C.
## Value Assignment:
Values are assigned to production lots by using an ADVIA Clinical Chemistry System calibrated with the Anchor Calibrator and adjusted with a ratio derived from the Anchor Calibrator assigned value (target value) and the mean recovery of the Anchor Calibrator values (observed). The table below shows the target values for one specific lot of calibrator material
| Analyte Concentration (ug/mL) | | | | | Level 5 |
| --- | --- | --- | --- | --- | --- |
| Analyte | Level 1 | Level 2 | Level 3 | Level 4 | |
| Carbamazepine | 0.0 | 2.9 | 6.1 | 12.8 | 21.9 |
| Gentamicin | 0.0 | 1.6 | 3.3 | 6.5 | 12.7 |
| Tobramycin | 0.0 | 1.5 | 3.0 | 6.0 | 12.6 |
| Vancomycin | 0 | 5.2 | 10.5 | 21.1 | 44.4 |
| Valproic Acid | 0 | 18.8 | 37.5 | 75.0 | 157.5 |
d. Detection limit: Not Applicable
e. Analytical specificity: Not Applicable
f. Assay cut-off: Not applicable
2. Comparison studies:
a. Method comparison with predicate device: Not Applicable
b. Matrix comparison: Not Applicable
3. Clinical studies:
a. Clinical Sensitivity: Not applicable
b. Clinical specificity:
{4}
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not Applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
5
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.