This device is intended to measure alcohol in human breath. Measurements obtained by this device are used in the diagnosis of alcohol intoxication.
Device Story
The BACTRACK® Select Breathalyzer (models S30, S50, S70) is an over-the-counter alcohol screening device. It uses a semiconductor-oxide sensor to detect alcohol in a human breath sample provided by the user. The device requires a 5-second blowing time and provides a digital BAC (Blood Alcohol Content) result on an LED or LCD display. It is powered by AA or AAA batteries. The device is intended for personal use to assist in the diagnosis of alcohol intoxication. Users must perform periodic recalibration (every 6-12 months, or monthly with daily use) to maintain accuracy. The output allows the user to estimate their BAC, which informs decisions regarding alcohol consumption and safety.
Clinical Evidence
Clinical testing demonstrated that over-the-counter purchasers could read and understand instructions, properly operate the device, and obtain results comparable to professional-grade breathalyzers. Bench and laboratory testing confirmed accuracy of ± 0.01% BAC at 0.10% BAC, consistent with the predicate device.
Technological Characteristics
Semiconductor-oxide sensor; plastic case with internal circuit board; replaceable polystyrene mouthpiece; powered by AA or AAA batteries; 3-digit LED/LCD display; measurement range 0.00% to 0.40% BAC; accuracy ± 0.01% BAC at 0.10% BAC; warm-up time 15-25 seconds.
Indications for Use
Indicated for use by the general population for the measurement of alcohol in human breath to aid in the diagnosis of alcohol intoxication. Available for over-the-counter use.
Regulatory Classification
Identification
A breath-alcohol test system is a device intened to measure alcohol in the human breath. Measurements obtained by this device are used in the diagnosis of alcohol intoxication.
Predicate Devices
AL-5000 Sentech Korea Corp. (k061922)
Related Devices
K090067 — BACTRACK SELECT BREATHALYZER, MODEL S80 · Khn Solutions, LLC · Mar 24, 2009
K110614 — PAD (PERSONAL ALCOHOL DETECTOR) BREATH ALCOHOL TEST SYSTEM · Seju Engineering Co., Ltd. · Nov 21, 2011
Submission Summary (Full Text)
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE
A. 510(k) Number:
k083801
B. Purpose for Submission:
New device for US consumer market
C. Manufacturer and Instrument Name:
KHN Solutions, LLC Bactrack Select 30, 50, and 70 Breathalyzers
D. Type of Test or Tests Performed:
Quantitative (electrochemical fuel cell sensor)
E. System Descriptions:
1. Device Description:
The Bactrack Select 30, 50, and 70 breathalyzers are self-contained units with a sensor at the top of the unit for breath sampling. The LCD displays a timer for countdown. There is one start button on the S30 and two buttons, a start button and mode button, on the S50 and S70. The mode button is used to display the number of uses and aids in the re-calibration of the device. The devices display the calculated percent blood alcohol concentration (BAC) up to 2 decimal places (0.08%) for the S30 and S50 and the 3 decimal places (0.082%) for the S70. The LED/LCD display also shows the number of uses prior to each device usage. These devices are powered by two AA batteries.
2. Principles of Operation:
The sensing material in gas sensors is metal oxide (SnO₂). When the SnO₂ is heated to a high temperature in air, oxygen is adsorbed on the crystal surface with a negative charge. Then donor electrons in the crystal surface are transferred to the adsorbed oxygen, resulting in leaving positive charges in a space charge layer. Thus, surface potential is formed to serve as a potential barrier against electron flow. Inside the sensor, electric current flows through the conjunction parts (grain boundary) of SnO₂ micro crystals. At grain boundaries, adsorbed oxygen forms a potential barrier which prevents carriers from moving freely. The electrical resistance of the sensor is attributed to this potential barrier. In the presence of a deoxidizing gas, the surface density of the negatively charged oxygen decreases, so the barrier height in the grain boundary is reduced. The reduced barrier height
{1}
decreases sensor resistance. The relationship between sensor resistance and the concentration of gas is expressed by an equation to determine the alcohol concentration.
3. Modes of Operation:
This device has only one mode of operation.
4. Specimen Identification:
There is no mechanism to identify the specimen.
5. Specimen Sampling and Handling:
The user provides a breath sample by exhaling into the device.
6. Calibration:
The sponsor recommends a recalibration interval of every 200 tests or every six to twelve months if less than 200 tests have been performed. These intervals are based on calibration stability data collected internally. The unit must be sent back to the factory for calibration.
Traceability:
This device is traceable to a commercially available certified ethyl alcohol reference solution.
7. Quality Control:
There are no external quality controls available for these types of devices for over-the-counter use.
8. Software:
FDA has reviewed applicant's Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X ☐ or No ☐
F. Regulatory Information:
1. Regulation section:
21 CFR 862.3050
{2}
2. Classification:
Class I, reserved
3. Product code:
DJZ
4. Panel:
Toxicology
G. Intended Use:
1. Indication(s) for Use:
This device is intended to measure alcohol in human breath. Measurements obtained by this device are used in the diagnosis of alcohol intoxication.
2. Special Conditions for Use Statement(s):
For over the counter use.
H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers:
AL-5000 Sentech Korea Corp.
k061922
{3}
2. Comparison with Predicate Device:
| | AL-5000 Sentech Korea Corp. (k061922) | BACTRACK® Select Breathalyzer: S30, S50, S70 |
| --- | --- | --- |
| Similarities | | |
| Indication Of Use | This device is intended to measure alcohol in the human breath. Measurements obtained by this device are used as an aid in the detection of alcohol intoxication. | SAME |
| Mode | Breath Alcohol Concentration | SAME |
| Practitioner Use | Over the Counter | SAME |
| Blowing Time | 5 Seconds | SAME |
| Type Of Sensor | Semiconductor-Oxide Sensor | SAME |
| Anatomical Site | Mouth | SAME |
| Mouthpiece | Replaceable, polystyrene | SAME |
| Warm Up Time | 15-25 seconds | SAME |
| Operating Temperature | 10 - 40°C (50 - 104°F) | SAME |
| Measurement Range | 0.000% to 0.400% | SAME |
| Differences | | |
| Display | 3 Digit LED | S50: Identical to AL-5000 S30: 3 Digit LCD S70: 4 Digit LCD |
| Power Source | 2 AA Alkaline | SAME except 2 AAA for Model S30 |
| Battery Life | “Over 200 tests” | SAME. 200-300 tests for all 3 models |
| Size | 116 x 59 x27 mm | S50: Identical to AL-5000 S30: 103 x 34 x 25 mm S70: 122 x 58 x 22 mm |
| Weight | 105 g | S50: Identical to AL-5000 S30: 53/77 g (without, with batteries) S70: 67/115g (without, with batteries) |
{4}
I. Special Control/Guidance Document Referenced (if applicable):
1. EN 61326:1997 +A1:1998+A2:2001+A3:2003 Electrical equipment for measurement, control and laboratory use. EMC Requirements. General requirements.
2. EN 61000-4-2:1995 +A1:1998 +A2:2001 Testing and measurement techniques Electrostatic Discharge
3. EN 61000-4-3:2002+A1:2002 Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test.
4. EN 61326:1997 +A1:1998+A2:2001+A3:2003 Class B. Electrical equipment for measurement, control and laboratory use. EMC Requirements. General requirements
J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy:
The sponsor performed a consumer study to determine if consumers could correctly use and interpret the device using only the supplied instructions for use and compare to the results obtained by an evidential breath alcohol tester (the LifeLoc FC10). Seventy-six participants took their breath alcohol reading with the Bactrack S30, S50 and S70 Breathalyzers and recorded the result. Immediately afterward, the participants were administered a breath alcohol test using the professional device operated by a trained individual.
For the S30, the breath alcohol paired concentrations ranged from BAC of 0.000 to 0.272% (by the evidential device). Linear regression analysis of the data showed a slope of 0.8519, a y-intercept of 0.0173 and a correlation coefficient of 0.861.
For the S50, the breath alcohol paired concentrations ranged from BAC of 0.000 to 0.272 (by the evidential device). Linear regression analysis of the data showed a slope of 1.019, a y-intercept of 0.0061 and a correlation coefficient of 0.9519.
For the S70, the breath alcohol paired concentrations ranged from BAC of 0.000 to 0.272 (by the evidential device). Linear regression analysis of the data showed a slope of 1.0667, a y-intercept of 0.0034 and a correlation coefficient of 0.9244.
{5}
After the study, participants were asked questions about the ease of use, instructions and about the ability to understand and interpret the test results displayed by the device. The results are presented below:
| Question | Strongly Disagree | Disagree | Neither Agree or Disagree | Agree | Strongly Agree |
| --- | --- | --- | --- | --- | --- |
| I found the device easy to use | 0 | 0 | 0 | 40% | 60% |
| I understood how to use the device after reading the instructions | 0 | 0 | 0 | 45% | 55% |
| I was able to easily understand and interpret the test results displayed by the device | 0 | 1% | 0 | 39% | 60% |
b. Precision/Reproducibility:
This device was not tested by the National Highway Traffic Safety Administration (NHTSA). The sponsor performed their own testing using the S50 and collected performance data using the NHTSA requirements for precision and accuracy as a guide. These requirements (referred to as Model Specifications) consist of 20 trials at a Blood Alcohol Concentration (BAC) of 0.000, 20 trials at a BAC of 0.008 and 20 trials at a BAC of 0.032. BACs are simulated in breath by a Breath Alcohol Sample Simulator (BASS), which provides an alcohol-in air test sample with known alcohol concentrations, flow rate and air composition. The acceptance criteria for the Model Specifications are: not more than one negative result at 0.032 BAC, not more than one positive result at 0.008 BAC, and not more than one negative greater than zero and no positives at 0.000 BAC. The Bactrack S50, being representative of the S30 and S70 technological characteristics, had no negatives at 0.032 BAC, no positives at 0.008 BAC and no positives at 0.000 BAC. Negatives are defined as being less than 0.020 BAC. Positives are defined as being greater than or equal to 0.020 BAC.
c. Linearity:
These devices will report concentrations from 0.00 to 0.400% BAC. However, the DOT Model Specifications require accuracy and precision testing up to a concentration of 0.032 only; therefore, true linearity over the entire measuring range of the device was not evaluated. These devices met all of the applicable NHTSA requirements for precision and accuracy as described above.
d. Carryover:
Carryover studies are not required by NHTSA and were not performed using these devices.
{6}
e. Interfering Substances:
The sponsor chose to duplicate the NHTSA requirements for cigarette smoke interference. The sponsor evaluated the possibility of cigarette smoke interference with the devices. An alcohol-free person who smokes cigarettes smoked approximately one half of a cigarette. Within one minute after smoking, the subject took a breath alcohol reading. The subject then repeated this procedure for 5 trials. The acceptance criterion for this trial is: no positive results. The Bactrack S50, being representative of the S30 and S70 technological characteristics, produced all negative results in this study.
2. Other Supportive Instrument Performance Data Not Covered Above:
a. Limit of Detection:
The DOT Model Specifications do not specifically address the detection limit of breath alcohol devices. However, the devices must be tested at a BAC of zero (blank reading) to assess the possibility of false positives. This consists of 20 trials under normal laboratory conditions at a BAC of 0.000. Non-alcoholic human breath is to be used as the sample. For devices capable of providing a reading of greater than 0.000 BAC and less than 0.020 BAC, the acceptance criterion is: not more than one such result. The Bactrack S50, being representative of the S30 and S70 technological characteristics, had no false positives in this trial.
b. Temperature:
The DOT Model Specifications require testing 10 and 40° C to assess any possible effects of temperature. At 10° C, 20 trials are required at 0.008 BAC and 20 trials are required at 0.032 BAC. Acceptance criteria are: not more than one positive result at 0.008 BAC, and not more than one non-positive result at 0.032 BAC. The Bactrack S50, being representative of the S30 and S70 technological characteristics, had no positive results at 0.008 BAC and no non-positive results at 0.032 BAC.
At 40° C, the requirements are identical. Again the Bactrack S50, being representative of the S30 and S70 technological characteristics, had no positive results at 0.008 BAC and no non-positive results at 0.032 BAC.
c. Vibration:
The DOT Model Specifications require vibration testing to assess any possible vibrational effects.
{7}
Twenty trials are required at 0.008 BAC and 0.032 BAC. Acceptance criteria are: not more than one positive result at 0.008 BAC, and not more than one non-positive result at 0.032 BAC. The Bactrack S50, being representative of the S30 and S70 technological characteristics, had no positive results at 0.008 BAC and no non-positive results at 0.032 BAC.
d. Cutoff:
For the purposes of performance testing, a BAC cutoff of 0.020 was used to distinguish positive from negative samples.
**K. Proposed Labeling:**
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
**L. Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.