← Product Code [DJG](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG) · K231904

# Rapid Fentanyl (FYL) Test Strip, Rapid Fentanyl (FYL) Test Dipcard (K231904)

_Co-Innovation Biotech Co., Ltd. · DJG · Mar 8, 2024 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG/K231904

## Device Facts

- **Applicant:** Co-Innovation Biotech Co., Ltd.
- **Product Code:** [DJG](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG.md)
- **Decision Date:** Mar 8, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3650
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

The Rapid Fentanyl (FYL) Test Strip is a rapid, screening test for the qualitative detection of Fentanyl (FYL) in human urine at the cut-off concentration of 1 ng/mL. The tests is intended for in vitro diagnostics use. It is intended for prescription use including point of care sites. This assay provides only a preliminary analytical test result. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. Gas Chromatography/ Mass spectrometry (GC/MS) or Liquid chromatography/Mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. The Rapid Fentanyl (FYL) Test Dipcard is a rapid, screening test for the qualitative detection of Fentanyl (FYL) in human urine at the cut-off concentration of 1 ng/mL. The tests is intended for in vitro diagnostics use. It is intended for prescription use including point of care sites. This assay provides only a preliminary analytical test result. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. Gas Chromatography/ Mass spectrometry (GC/MS) or Liquid chromatography/Mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

## Device Story

Rapid Fentanyl (FYL) Test Strip and Dipcard are lateral flow immunochromatographic assays for qualitative fentanyl detection in human urine. Input: human urine sample. Principle: competitive binding antigen-antibody immunochemistry. Output: visual qualitative result (positive/negative) on test strip/dipcard. Used at point-of-care sites by healthcare professionals. No instrument required. Results are preliminary; clinical judgment and confirmatory testing (GC/MS or LC/MS) required for final diagnosis. Benefits: rapid screening for fentanyl presence to assist clinical decision-making.

## Clinical Evidence

Bench testing only. Precision/reproducibility evaluated across 3 lots, 6 operators, and 3 point-of-care sites (1620 observations). Analytical specificity/interference tested against various opioids and common medications; no interference observed. Method comparison performed using 480 clinical urine samples confirmed by LC/MS.

## Technological Characteristics

Lateral flow immunochromatographic assay; competitive binding principle. Form factors: test strip and test dipcard. Storage: 4-30°C. No instrument/platform required. Materials and manufacturing procedures are identical between strip and dipcard formats.

## Regulatory Identification

An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.

## Special Controls

*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- Superbio Fentanyl Urine Detection Kit ([K220046](/device/K220046.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K231904

B Applicant

Co-Innovation Biotech Co., Ltd.

C Proprietary and Established Names

Rapid Fentanyl (FYL) Test Strip, Rapid Fentanyl (FYL) Test Dipcard

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  DJG | Class II | 21 CFR 862.3650 - Opiate Test System | TX - Clinical Toxicology  |

## II Submission/Device Overview:

A Purpose for Submission: New Device

B Measurand: Fentanyl in human urine.

C Type of Test: Qualitative competitive binding, lateral flow immunochromatographic assay.

## III Intended Use/Indications for Use:

A Intended Use(s): See Indications for Use below.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

B Indication(s) for Use:

The Rapid Fentanyl (FYL) Test Strip is a rapid, screening test for the qualitative detection of Fentanyl (FYL) in human urine at the cut-off concentration of 1 ng/mL. The tests is intended for in vitro diagnostics use. It is intended for prescription use including point of care sites. This assay provides only a preliminary analytical test result. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. Gas Chromatography/ Mass spectrometry (GC/MS) or Liquid chromatography/Mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

The Rapid Fentanyl (FYL) Test Dipcard is a rapid, screening test for the qualitative detection of Fentanyl (FYL) in human urine at the cut-off concentration of 1 ng/mL. The tests is intended for in vitro diagnostics use. It is intended for prescription use including point of care sites. This assay provides only a preliminary analytical test result. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. Gas Chromatography/ Mass spectrometry (GC/MS) or Liquid chromatography/Mass spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

For in-vitro diagnostic use only.

D Special Instrument Requirements:

N/A

IV Device/System Characteristics:

A Device Description:

Rapid Fentanyl (FYL) Test Strip and Rapid Fentanyl (FYL) Test Dipcard are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of fentanyl at or above the cut-off concentration of 1 ng/mL. Test Strip and Test Dipcard use identical test strips made with same chemical formulation and manufacturing procedures.

Content of the kit:

Test Strip, one test in one pouch or Test Dipcard, one test in one pouch.

Desiccant. The desiccant is only for storage purposes only and is not used in the test procedures.

Leaflet with instructions for use.

B Principle of Operation:

The Rapid Fentanyl (FYL) Test Strip is a competitive immunoassay that is used to screen for the presence of Fentanyl and metabolites in urine. It is chromatographic absorbent device in which, drugs within a urine sample, competitively combined to a limited number of drug monoclonal antibody (mouse) conjugate binding sites.

When the test is activated, the urine is absorbed into test strip by capillary action, mixes with the respective drug monoclonal antibody conjugate, and flows across a pre-coated membrane. When drug within the urine sample is below the detection level of the test, respective drug monoclonal

K231904 - Page 2 of 10

{2}

antibody conjugate binds to the respective drug-protein conjugate immobilized in the Test Region (T) of the test strip. This produces a colored Test line in the Test Region (T) of the strip, which, regardless of its intensity, indicates a negative test result.

When sample drug levels are at or above the detection level of the test, the free drug in the sample binds to the respective drug monoclonal antibody conjugate, preventing the respective drug monoclonal antibody conjugate from binding to the respective drug-protein conjugate immobilized in the Test Region (T) of the device. This prevents the development of a distinct colored band in the test region, indicating a preliminary positive result.

To serve as a procedure control, a colored line will appear at the Control Region (C), of each strip, if the test has been performed properly.

V Substantial Equivalence Information:

A Predicate Device Name(s):

Superbio Fentanyl Urine Detection Kit, Superbio Immunofluorescence Analyzer EASY-11

B Predicate 510(k) Number(s):

K220046

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K231904 | K220046  |
| --- | --- | --- |
|  Device Trade Name | Rapid Fentanyl (FYL) Test Strip, Rapid Fentanyl (FYL) Test Dipcard | Superbio Fentanyl Urine Detection Kit  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications for Use | Qualitative detection of fentanyl in urine, | Same  |
|  Intended Users | Prescription use only. | Same  |
|  Calibrator and Cutoff Values | 1 ng/ml | Same  |
|  Type of Test | Qualitative | Same  |
|  Specimen Type | Human urine | Same  |
|  Methodology | Competitive binding, lateral flow immunochromatographi c assay based on the principle of antigen antibody immunochemistry | Same  |

K231904 - Page 3 of 10

{3}

|  General Device Characteristic Differences |  |   |
| --- | --- | --- |
|  Configurations | Test Strip and Dipcard | Test Card  |
|  Platform required | None | Immunofluorescence Analyzer  |
|  Storage/Operating Temp | 4-30°C | 4-35°C  |

VI Standards/Guidance Documents Referenced:
None

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:
Precision studies were performed using 3 lots of test Strip and test Dipcard. Drug free specimens were spiked with target drug fentanyl at 0, ±75% cutoff, ±50% cutoff, ±25% cutoff and +100% cutoff of drug. The concentrations of the target drugs were confirmed with LC/MS. Each concentration of the urine specimen was divided into aliquots. Each aliquot was blindly labeled by a nonparticipant. Separate sets of blinded coded samples were assigned and randomized prior to testing. The study was conducted by 6 operators/nurses at 3 Point-of-Care sites. Two operators per location tested 3 aliquots at each concentration for each lot per day (3 runs/day) for 10 non-consecutive days using one device lot per location. One operator tested the test strip format, and the second operator tested the test dipcard format. There were 1620 observations by 3 sites at 9 concentrations.

Rapid Fentanyl (FYL) Test Strip:

|  Approximate concentration of sample | % of cutoff | Number of determinations per lot | Test Strip Lot 1 |   | Test Strip Lot 2 |   | Test Strip Lot 3  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  Positive | Negative | Positive | Negative | Positive | Negative  |
|  0ng/ml | Negative | 60 | 0 | 60 | 0 | 60 | 0 | 60  |
|  0.25ng/ml | -75% cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60  |
|  0.5ng/ml | -50% cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60  |
|  0.75ng/ml | -25% cutoff | 60 | 4 | 56 | 2 | 58 | 6 | 54  |
|  1ng/ml | cutoff | 60 | 36 | 24 | 34 | 26 | 32 | 28  |
|  1.25ng/ml | +25% cutoff | 60 | 54 | 6 | 56 | 4 | 56 | 4  |
|  1.5ng/ml | +50% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0  |
|  1.75ng/ml | +75% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0  |
|  2ng/ml | +100% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0  |

K231904 - Page 4 of 10

{4}

Rapid Fentanyl (FYL) Test Dipcard:

|  Approximate concentration of sample | % of cutoff | Number of determinations per lot | Test Dipcard Lot 1 |   | Test Dipcard Lot 2 |   | Test Dipcard Lot 3  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  Positive | Negative | Positive | Negative | Positive | Negative  |
|  0ng/ml | Negative | 60 | 0 | 60 | 0 | 60 | 0 | 60  |
|  0.25ng/ml | -75% cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60  |
|  0.5ng/ml | -50% cutoff | 60 | 0 | 60 | 0 | 60 | 0 | 60  |
|  0.75ng/ml | -25% cutoff | 60 | 4 | 56 | 4 | 56 | 2 | 58  |
|  1ng/ml | cutoff | 60 | 34 | 26 | 34 | 26 | 36 | 24  |
|  1.25ng/ml | +25% cutoff | 60 | 56 | 4 | 58 | 2 | 56 | 4  |
|  1.5ng/ml | +50% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0  |
|  1.75ng/ml | +75% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0  |
|  2ng/ml | +100% cutoff | 60 | 60 | 0 | 60 | 0 | 60 | 0  |

2. Linearity:
Not applicable, this device is intended for qualitative use only

3. Analytical Specificity/Interference:

Cross Reactivity:
The sponsor performed a study to verify potential cross-reactants using the target drug, drug metabolites, and other related compounds. The target drugs, drug metabolites and other components that are likely to interfere in urine samples were added to drug-free urine specimens at a concentration of 100μg/ml. The study was conducted by 8 laboratory professionals. Two Laboratory professionals spiked the compound controls and divided them into individual containers, the other 6 laboratory professionals conducted the testing using the instructions for use package insert. Each sample was tested in 3 replicates using 3 lots of Rapid Fentanyl (FYL) Test Strip and the same protocol for Rapid Fentanyl (FYL) Test Dipcard. One operator conducted each individual lot (and product type; strip and dipcard) separately.

The formula for calculating the percent of cross-reactivity was calculated by dividing the cutoff concentration by the minimum concentration required to obtain a positive result and then multiplying by 100%. The lowest concentration that caused a positive result for each compound are listed below for both the Rapid Fentanyl (FYL) Test Strip and the same protocol for Rapid Fentanyl (FYL) Dipcard:

Rapid Fentanyl (FYL) Test-Strip:

|  Compound | Quantity equivalent to cutoff(ng/mL) conducted on 3 lots | % Cross Reactivity  |
| --- | --- | --- |
|  Norfentanyl | 10,000 | 0.01  |

K231904 - Page 5 of 10

{5}

Rapid Fentanyl (FYL) Dipcard:

|  Compound | Quantity equivalent to cutoff(ng/mL) conducted on 3 lots | % Cross Reactivity  |
| --- | --- | --- |
|  Norfentanyl | 10,000 | 0.01  |
|  Acetyl fentanyl | 1.2 | 83.3  |
|  Acetyl norfentanyl | 10,000 | 0.01  |
|  Acrylfentanyl | 1.5 | 66.7  |
|  Butyryl fentanyl | 1.6 | 62.5  |
|  Carfentanil | 500 | 0.2  |
|  (±)-3-cis-methylfentanyl | 5 | 20  |
|  4-Fluoro-isobutyrylfentanyl | 3 | 33.3  |
|  Furanyl fentanyl | 1.8 | 55.6  |
|  ω-1-Hydroxyfentanyl | 20,000 | 0.01  |
|  (±) β-hydroxythiofentanyl | 2.8 | 35.7  |
|  Isobutyryl fentanyl | 1.5 | 66.7  |
|  Ocfentanil | 1.8 | 55.6  |
|  Para-fluorobutyrylfentanyl (p-FBF) | 3 | 33.3  |
|  Para-fluoro fentanyl | 3 | 33.3  |
|  Sufentanil | 625 | 0.16  |
|  Valeryl fentanyl | 2.5 | 40  |
|  Compound | Quantity equivalent to cutoff(ng/mL) conducted on 3 lots | % Cross Reactivity  |
| --- | --- | --- |
|  Norfentanyl | 10,000 | 0.01  |
|  Acetyl fentanyl | 1.2 | 83.3  |
|  Acetyl norfentanyl | 10,000 | 0.01  |
|  Acrylfentanyl | 1.5 | 66.7  |
|  Butyryl fentanyl | 1.6 | 62.5  |
|  Carfentanil | 500 | 0.2  |
|  (±)-3-cis-methylfentanyl | 5 | 20  |
|  4-Fluoro-isobutyrylfentanyl | 3 | 33.3  |
|  Furanyl fentanyl | 1.8 | 55.6  |
|  ω-1-Hydroxyfentanyl | 20,000 | 0.01  |
|  (±) β-hydroxythiofentanyl | 2.8 | 35.7  |
|  Isobutyryl fentanyl | 1.5 | 66.7  |

K231904 - Page 6 of 10

{6}

|  Ocfentanil | 1.8 | 55.6  |
| --- | --- | --- |
|  Para-fluorobutyryl fentanyl (p-FBF) | 3 | 33.3  |
|  Para-fluoro fentanyl | 3 | 33.3  |
|  Sufentanil | 625 | 0.16  |
|  Valeryl fentanyl | 2.5 | 40  |

## Interference Study on the Effect of Specific Gravity and pH:

The specific Gravity study was conducted using 5 increments of specific gravity levels: 1.000, 1.010, 1.020, 1.030, and 1.040 on 3 different lots of the Rapid Fentanyl (FYL) Test Strip, and 3 different lots of the Rapid Fentanyl (FYL) Test Dipcard with drug free urine spiked with Fentanyl at $0.5\mathrm{ng / ml}$ (50% below cutoff) and $1.5\mathrm{ng / ml}$ (50% above cutoff). All concentrations were confirmed with LC/MS. No discrepant results were observed.

The pH study was conducted on 3 different lots of the Rapid Fentanyl (FYL) Test Strip, and 3 different lots of the Rapid Fentanyl (FYL) Test Dipcard using seven increments of pH level ranging from 3 to 9 in aliquot negative urine spiked with Fentanyl at $0.5\mathrm{ng / ml}$ (50% below cutoff) and $1.5\mathrm{ng / ml}$ (50% above cutoff). All concentrations were confirmed with LC/MS. No discrepant results were observed.

## Interference

The sponsor conducted two different studies for interference, using their Test Strip and Test Dipcard for each. The first interference study was conducted using opioid compounds (a separate study from their Cross-Reactivity study), and the second interference study was conducted using commonly ingested medications/substances. Both are described below:

### A. Interfering Substances (Opioid compounds):

To test the interference of opioid compounds. This study was conducted by 8 laboratory professionals. 2 laboratory professionals spiked the compound controls and divided them into individual containers. The other 6 laboratory professionals conducted the following testing using the English package insert as guide and read the results at 5 minutes.

Potential interfering opioid compounds, each at a concentration of $100~\mu \mathrm{g / mL}$ (except Despropionly fentanyl (4-ANPP) was tested at $50\mu \mathrm{g / ml}$, norcarfentanil was tested at $5\mu \mathrm{g / mL}$ and remifentanil was tested at $10~\mu \mathrm{g / mL}$), were added to drug-free urine specimens spiked with fentanyl to concentrations at $50\%$ below and $50\%$ above cutoff levels at $0.5\mathrm{ng / ml}$ and $1.5\mathrm{ng / ml}$ level and were tested in replicates of 3 on 3 different lots of the Rapid Fentanyl (FYL) Test Strip, and 3 different lots of the Rapid Fentanyl (FYL) Test Dipcard. The compounds showed no interference with the Rapid Fentanyl (FYL) Test Strip or the Rapid Fentanyl (FYL) Test Dipcard.

The following opioid compounds showed no interference with the Rapid Fentanyl (FYL) Test Strip or the Rapid Fentanyl (FYL) Test Dipcard:

6-Acetyl morphine, Alfentanil, Amphetamine, Buprenorphine, Buprenorphineglucuronide, Codeine, Despropionyl fentanyl (4-ANPP),

K231904 - Page 7 of 10

{7}

Dextromethorphan, Dihydrocodeine, EDDP, EMDP, Fluoxetine, Heroin, Hydrocodone, Hydromorphone, Ketamine, Levorphanol, Meperidine, Methadone, Morphine, Morphine-3-glucuronide, Naloxone, Naltrexone, Norbuprenorphine, Norcarfentanil, Norcodeine, Norketamine, Normeperidine, Normorphine, Noroxycodone, Oxycodone, Oxymorphone, Pentazocine (Talwin), Pipamperone, Remifentanil, Risperidone, Tapentadol, Thioridazine, Tilidine, Tramadol, Tramadol-O-Desmethyl, Tramadol-NDesmethyl, Trazodone, Dextromethorphan, Meperidine, and Pentazocine

B. Interfering Substances (Commonly ingested medications/substances):

This study was conducted by 8 laboratory professionals. 2 laboratory professionals spiked the compound controls and divided them into individual containers. The other 6 laboratory professionals conducted the following testing using the English package insert as guide and read the results at 5 minutes.

Potential interfering substances, each at a concentration of 100 µg/mL (or specified concentrations noted in the tables below), were added to drug-free urine specimens spiked with fentanyl to concentrations at 50% below and 50% above cutoff levels at 0.5 ng/ml and 1.5 ng/ml level and were used to test 3 different lots of the Rapid Fentanyl (FYL) Test Strip, and 3 different lots of the Rapid Fentanyl (FYL) Test Dipcard. The following substances showed no interference with the Rapid Fentanyl (FYL) Test Strip or the Rapid Fentanyl (FYL) Test Dipcard:

Acetaminophen, Acetone (1000 mg/dL), Acetophenetidin, Acetylsalicylic acid, Albumin (100 mg/dL), Albuterol, Aminopyrine, Amitriptyline (35 µg/ml), Amobarbital, Amoxicillin, Ampicillin, Apomorphine, Ascorbic acid, Aspartame, Atropine, Benzilic acid, Benzoic acid, Benzoylecgonine, Bilirubin, Boric Acid (1%), Bupropion (50 µg/ml), Caffeine, Carbamazepine, Chloral hydrate, Chloramphenicol, Chlorothiazide, Chlorpromazine, Cholesterol, Clomipramine (50 µg/ml), Clonidine, Cortisone, Cotinine, Creatinine, Cyclobenzaprine (10 µg/ml), Deoxycorticosterone, Desipramine (50 µg/ml), Dextromethorphan, Diclofenac, Diflunisal, Digoxin, Diphenhydramine, DL-Tryptophan, DL-Tyrosine, Doxepin (50 µg/ml), Ecgonine methyl ester, Ephedrine, Erythromycin, Ethanol (1%), Fenoprofen, Fluphenazine, Furosemide, Galactose (10 mg/dL), Gamma Globulin (500 mg/dL), Gentisic acid, Glucose (3000 mg/dL), Hemoglobin, Hydralazine, Hydrochlorothiazide, Hydrocortisone, Hydroxytyramine, Ibuprofen, Imipramine (30 µg/ml), Isoproterenol, Isoxsuprine, Ketoprofen, Labetalol, Lidocaine (50 µg/ml), Loperamide, Maprotiline (50 µg/ml), Meperidine, Meprobamate, Methapyrilene (10 µg/ml), Methaqualone (50 µg/ml), Methoxyphenamine, Metronidazole (300 µg/ml), N-Acetylprocainamide, NaCl (4000 mg/dL), Nalidixic acid, Naloxone, Naltrexone, Naproxen, Niacinamide, Nicotine (10 µg/ml), Nifedipine, Norethindrone, Nortriptyline (25 µg/ml), Noscapine, OHydroxyhippuric acid, Octopamine, Oxalic acid (100 mg/dL), Oxazepam, Oxolinic acid, Oxymetazoline, Papaverine, Penicillin G, Perphenazine, Phencyclidine, Phenelzine, Phenobarbital, Prednisone, Propoxyphene (50 µg/ml), Propranolol, Pseudoephedrine, Quinine, Ranitidine, Riboflavin (7.5 mg/dL), Ketamine, Salicylic acid, Secobarbital, Serotonin (5-Hydroxytyramine), Sulfamethazine, Sulindac, Tetrahydrocortisone 3-(β-D-glucuronide), Tetrahydrocortisone 3-acetate, Tetrahydrozoline, Thiamine, Thioridazine, Triamterene, Trifluoperazine, Trimethoprim, Tyramine, Urea (2000 mg/dL), Uric acid, Valproic acid (250 µg/ml), Venlafaxine, Verapamil, Zomepirac, β-Estradiol.

K231904 - Page 8 of 10

{8}

4. Assay Reportable Range:
Not applicable, this device is intended for qualitative use only.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The device is traceable to a commercially available standard that was verified by LC/MS-MS

6. Detection Limit:
Not applicable.

7. Assay Cut-Off:
Analytical performance of the device around the claimed cutoff is described in the precision section VII.A1. above (1ng/ml).

B Comparison Studies:

1. Method Comparison with Predicate Device:
Four hundred eighty clinical urine samples collected from several hospitals and drug relief reformatories. All samples were unaltered clinical specimens, and the concentrations were confirmed with LC/MS. The samples were divided among the 6 nurses. All aliquots were blindly labeled by a nonparticipant and were randomized prior to testing.

The study was conducted by 6 nurses at three Point-of-Care sites. Each of the three sites were responsible for conducting testing on one lot of 80 samples for both the Rapid Fentanyl (FYL) Test Strip, and the Rapid Fentanyl (FYL) Test Dipcard (160 samples at each of the 3 sites=480). Each nurse only tested one format of Rapid Fentanyl Test, one tested with strip format another tested with Dipcard.

Comparison data of Rapid Fentanyl (FYL) Test (Strip)

|  Format | LC/MS |   |   |   |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  Neg. (drug free) | Neg. (<-50% cutoff) | Near cutoff neg. (-50% cutoff to cutoff) | Near cutoff pos. (cutoff to +50% cutoff) |   | Pos. (> +50% cutoff)  |
|  Lot 1 | Positive | 0 | 0 | 1 | 8 | 31 | 80  |
|   |  Negative | 27 | 5 | 7 | 1 | 0  |   |
|  Lot 2 | Positive | 0 | 0 | 1 | 8 | 31 | 80  |
|   |  Negative | 27 | 5 | 7 | 1 | 0  |   |
|  Lot 3 | Positive | 0 | 0 | 1 | 8 | 31 | 80  |
|   |  Negative | 27 | 5 | 7 | 1 | 0  |   |

K231904 - Page 9 of 10

{9}

Comparison data of Rapid Fentanyl (FYL) Test (Dipcard)

|  Format | LC/MS |   |   |   |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  Neg. (drug free) | Neg. (<-50% cutoff) | Near cutoff neg. (-50% cutoff to cutoff) | Near cutoff pos. (cutoff to +50% cutoff) |   | Pos. (> +50% cutoff)  |
|  Test Dipcard Lot 1 | Positive | 0 | 0 | 1 | 8 | 31 | 80  |
|   |  Negative | 27 | 5 | 7 | 1 | 0  |   |
|  Test Dipcard Lot 2 | Positive | 0 | 0 | 1 | 8 | 31 | 80  |
|   |  Negative | 27 | 5 | 7 | 1 | 0  |   |
|  Test Dipcard Lot 3 | Positive | 0 | 0 | 1 | 8 | 31 | 80  |
|   |  Negative | 27 | 5 | 7 | 1 | 0  |   |

2. Matrix Comparison:

Not applicable. This device is intended to be used with urine samples only.

C Clinical Studies:

1. Clinical Sensitivity: Not applicable.
2. Clinical Specificity: Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable.

D Clinical Cut-Off: Not applicable.

E Expected Values/Reference Range: Not applicable.

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K231904 - Page 10 of 10

---

**Source:** [https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG/K231904](https://fda.innolitics.com/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/DJG/K231904)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
