ETHANOL REAGENT, ETHANOL CALIBRATOR, ETHANOL QC MATERIAL

K111005 · Medica Corp. · DIC · Nov 15, 2011 · Clinical Toxicology

Device Facts

Record IDK111005
Device NameETHANOL REAGENT, ETHANOL CALIBRATOR, ETHANOL QC MATERIAL
ApplicantMedica Corp.
Product CodeDIC · Clinical Toxicology
Decision DateNov 15, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3040
Device ClassClass 2

Indications for Use

The EasyRA Ethanol Reagent is intended for the quantitative measurement of Alcohol (ETOH) in human urine using Medica's EasyRA Chemistry Analyzer in clinical laboratories. Alcohol measurements are used for the diagnosis and treatment of alcohol intoxication and poisoning. The EasyCal Ethanol Calibrator is intended for the calibration of the Ethyl Alcohol Assay on the EasyRA Chemistry Analyzer for the quantitative determination of ethyl alcohol in urine. The EasyQC Ethanol Quality Control Material is intended for the validation of the Ethyl Alcohol assay, which is used on the EasyRA clinical chemistry analyzer for the quantitative determination of ethyl alcohol in urine.

Device Story

EasyRA Ethyl Alcohol system consists of reagent, calibrator, and QC material for use on Medica EasyRA Chemistry Analyzer. System measures ethyl alcohol concentration in human urine samples. Clinical laboratory personnel operate the analyzer to perform quantitative assays. Output provides alcohol concentration levels to clinicians for diagnosing and treating alcohol intoxication or poisoning. System facilitates patient management by providing objective data on alcohol presence.

Clinical Evidence

No clinical data provided; device is an in-vitro diagnostic reagent/calibrator/QC system.

Technological Characteristics

In-vitro diagnostic reagent, calibrator, and QC material for use with the EasyRA Chemistry Analyzer. Quantitative assay for ethyl alcohol in urine.

Indications for Use

Indicated for quantitative measurement of ethyl alcohol in human urine to assist in the diagnosis and treatment of alcohol intoxication and poisoning in clinical laboratory settings.

Regulatory Classification

Identification

An alcohol test system is a device intented to measure alcohol (e.g., ethanol, methanol, isopropanol, etc.) in human body fluids (e.g., serum, whole blood, and urine). Measurements obtained by this device are used in the diagnosis and treatment of alcohol intoxication and poisoning.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k111005 B. Purpose for Submission: New assay C. Measurand: Ethanol (Ethyl Alcohol) D. Type of Test: Automated quantitative enzymatic assay E. Applicant: Medica Corporation F. Proprietary and Established Names: EasyRA Ethyl Alcohol Reagent EasyRA Ethyl Alcohol Calibrator EasyRA Ethyl Alcohol QC Material. G. Regulatory Information: 1. Regulation section: 21 CFR 862.3040, Alcohol Test System; 21 CFR 862.3200, Clinical Toxicology Calibrator; 21 CFR 862.3280, Clinical Toxicology Control Material 2. Classification: Class II, (Test System and Calibrator), Class I (control materials) {1} 3. Product code: DIC, DLJ, LAS 4. Panel: Toxicology (91) H. Intended Use: 1. Intended use(s): See indications for use. 2. Indication(s) for use: The EasyRA Ethanol Reagent is intended for the quantitative measurement of Alcohol (ETOH) in human urine using Medica's EasyRA Chemistry Analyzer in clinical laboratories. Alcohol measurements are used for the diagnosis and treatment of alcohol intoxication and poisoning. The EasyCal Ethanol Calibrator is intended for the calibration of the Ethyl Alcohol Assay on the EasyRA Chemistry Analyzer for the quantitative determination of ethyl alcohol in urine The EasyQC Ethanol Quality Control Material is intended for the validation of the Ethyl Alcohol assay, which is used on the EasyRA clinical chemistry analyzer for the quantitative determination of ethyl alcohol in urine. 3. Special conditions for use statement(s): For prescription use only. 4. Special instrument requirements: For use on the EasyRA Analyzer I. Device Description: Reagents supplied are liquid, ready-to-use and consist of the following: Buffer Reagent (R1): Tris-based buffer (50 mM) Sodium azide (0.09%) as preservative. Enzyme Reagent (R2): Alcohol dehydrogenase (ADH) Nicotinamide adenine dinucleotide (NAD, 10 mM), {2} Sodium azide (0.09%) as preservative. Stabilizers The calibrator and control matrix is an aqueous solution. Positive calibrators and controls are spiked with pure ethanol. ## J. Substantial Equivalence Information: 1. Predicate device name(s): Microgenics Corporation Ethanol Reagent, including calibrator and control materials 2. Predicate 510(k) number(s): k923783 3. Comparison with predicate: | Feature | Easy RA Ethanol Reagent | Predicate Device k923783 | | --- | --- | --- | | Intended Use | Clinical chemistry reagent used to provide a quantitative measurement of Alcohol | Same | | Clinical sample matrix | Urine | Serum, plasma, urine | | Reagent type | Liquid ready-for-use | Same | | Linearity range | 10 – 500 mg/dL | Up to 600 mg/dL | | Wavelength | 340 nm | Same | | Reaction type | Enzymatic | Same | | Reagent storage | 2 – 8 °C | Same | | Test Methodology | An enzymatic reaction of ethanol with ADH in the presence of NAD ADH converts ethyl alcohol to acetaldehyde and reduces NAD to NADH. The rate of change in absorption of the NADH chromogen is monitored over time on the EasyRA, to establish the amount of alcohol present. | Same | {3} | Feature | EasyRA EtOH Calibrator | EtOH Calibrator | | --- | --- | --- | | Intended Use | Intended for the calibration of Ethyl Alcohol Enzymatic Assay to quantitate ethyl alcohol. | Same | | Sample | Urine | Serum, plasma, urine | | Reagent type | Liquid ready-for-use | Same | | Reagent storage | 2 – 8 °C | Same | | Test Methodology | The Ethyl Alcohol Calibrator is an aqueous based liquid, and ready to use. A reagent blank measurement functions as a negative calibrator establishing the EtOH zero level. The calibrator contains about 100mg/dL ethanol (lot specific) and is used to establish the reference value for the EtOH urine assay. | The Ethyl Alcohol Calibrators are aqueous based liquids, and ready to use. The calibrator and two controls are prepared by spiking known concentrations of ethanol into the negative calibrator matrix. The negative calibrator contains no EtOH. The second contains about 100mg/dL ethanol (lot specific) and is used to establish the reference value for the EtOH urine assay. | | Feature | EasyRA EtOH Control Materials | EtOH Control Materials | | --- | --- | --- | | Intended Use | The EasyQC Ethanol Quality Control Material is intended for the validation of the Ethyl Alcohol assay . | same | | Sample | Urine | Serum, plasma, urine | | Reagent type | Liquid ready-for-use | Same | | Reagent storage | 2 – 8 °C | 2 – 8 °C | {4} 5 K. Standard/Guidance Document Referenced (if applicable): CLSI Guidelines: EP09-A2-Method Comparison and Bias Estimation Using Patient Samples EP07-A2 Interference Testing in Clinical Chemistry EP05-A2 Evaluation of Precision Performance of Quantitative Measurement Methods EP06-A Evaluation of the Linearity of Quantitative Measurement Procedures L. Test Principle: Medica’s Ethyl Alcohol Enzymatic Assay is a two component ready-to-use liquid reagent based on alcohol dehydrogenase (ADH) unique enzymatic reaction. In the presence of nicotinamide adenine dinucleotide (NAD), ADH converts ethyl alcohol to acetaldehyde and reduces NAD to NADH. The ethyl alcohol concentration is directly proportional to the ADH activity. The rate of NADH formation is measured at 340 nm wavelength. $$ \text{ADH} + \text{NAD} \rightleftharpoons \text{NADH} + \text{Acetaldehyde} $$ M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Within-Run, Run-to-Run, and Total precision were determined following CLSI EP5-A2 by analyzing four levels of urine samples spiked with pure (HPLC grade) ethanol. The study included an EasyRA analyzer and Medica’s EasyRA EtOH reagent and calibrator. For each of the four levels tested, two runs were conducted in the morning and two in the afternoon for 20 consecutive working days. Three calibrations were performed (on one analyzer) during the evaluation. Within-run, day to day, and run to run analyses precision were evaluated, where: - Svr = Estimate of repeatability standard deviation (within-run precision) - Sdd = Estimate of between-day standard deviation - Srr = Estimate of between-run standard deviation - St = Estimate of within-device precision standard deviation {5} Results are shown in the tables below: | Grand mean | 51.2 mg/dL | | --- | --- | | Swr | 1.22 | | CVwr | 2.40% | | Sdd | 1.36 | | CVdd | 2.70% | | Srr | 0.00 | | CVrr | 0.00% | | St | 1.83 | | CVt | 3.57% | | Grand mean | 100.69 mg/dL | | --- | --- | | Swr | 1.32 | | CVwr | 1.31% | | Sdd | 2.46 | | CVdd | 2.44% | | Srr | 1.02 | | CVrr | 1.01% | | St | 2.97 | | CVt | 2.95% | | Grand mean | 197.63 mg/dL | | --- | --- | | Swr | 2.32 | | CVwr | 1.17% | | Sdd | 4.27 | | CVdd | 2.16% | | Srr | 1.28 | | CVrr | 0.65% | | St | 5.02 | | CVt | 2.54% | | Grand mean | 296.11 mg/dL | | --- | --- | | Swr | 4.58 | | CVwr | 1.55% | | Sdd | 4.47 | | CVdd | 1.51% | | Srr | 4.74 | | CVrr | 1.60% | | St | 7.96 | | CVt | 2.69% | {6} b. Linearity/assay reportable range: Pure ethanol (HPLC grade) stock solution was used to spike a drug free urine pool to ethanol values of 500mg/dL (High) and 10mg/dL (Low). These solutions were subsequently mixed in pre-defined ratios using accurate volumetric techniques to make seven more solutions of known ethanol concentrations. The nine solutions tested covered the entire claimed linearity range. Linearity was determined using the guidelines in CLSI EP6-A. Acceptance criteria are defined by the sponsor as +/-10% relative to the expected concentration (volumetrically determined) or within +/- 3 mg/dL – whichever is greater. Results are shown below. Results: | Sample # | Assigned value mg/dL | Found value (mg/dL) | Percent recovery (found/assigned) | % Difference (Specification ±10% or ±3.0 mg/dL whichever is greater) | | --- | --- | --- | --- | --- | | 1 | 10.0 | 10.8 | 108% | -25.0% (0.8 mg/dL) | | 1 | | 10.4 | 104% | -25.0% (0.4 mg/dL) | | 2 | 71.25 | 74.4 | 104% | -0.9% | | 2 | | 71.9 | 101% | -0.9% | | 3 | 132.5 | 138.3 | 104% | 0.6% | | 3 | | 138.5 | 105% | 0.6% | | 4 | 193.75 | 200.8 | 104% | 0.8% | | 4 | | 199.3 | 103% | 0.8% | | 5 | 255.0 | 259.8 | 102% | 0.7% | | 5 | | 260.9 | 102% | 0.7% | | 6 | 316.25 | 320.8 | 101% | 0.5% | | 6 | | 319.2 | 101% | 0.5% | | 7 | 377.5 | 379.8 | 101% | 0.2% | | 7 | | 377.5 | 100% | 0.2% | | 8 | 438.75 | 444.1 | 101% | -0.2% | | 8 | | 444.7 | 101% | -0.2% | | 9 | 500.0 | 503.1 | 101% | -0.5% | | 9 | | 496.5 | 99% | -0.5% | The claimed assay range is 10-500 mg/dL. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Calibrator and control materials were previously cleared under k032461. Stability claims and calibrator value assignments are the same as in that submission. {7} Control materials are further value assigned for Medica's Easy RA Assay. The value assignments are based on +/-10% of the median value obtained from multiple runs and multiple analyzers. ## d. Detection limit: Ethanol solutions above and below the lower claimed assay limit of 10 mg/dL were prepared by spiking pure ethanol into ethanol-free samples. At the claimed assay lower limit of 10 mg/dL, the percent recovery was within 7%, and CV was within 10% based on evaluation of five solutions with ethanol concentrations below 10 mg/dL tested in duplicate for each of five days. The lower assay limit of 10 mg/dL was established in the linearity studies (see section M1b). ## e. Analytical specificity: Three interference studies were performed to evaluate potentially interfering substances including those with a similar chemical structure to ethanol; endogenous substances, and urine preservatives typically used in clinical laboratories. The interference studies were performed with drug-free urine samples spiked with ethanol to 20 mg/dL and to 100 mg/dL by weight; each interfering solution tested was made by adding the corresponding volume of interferent from a stock solution. Data were collected with two EasyRA chemistry analyzers. Analytes and the highest concentrations tested are listed below: | Acetaldehyde | 2000mg/dL | | --- | --- | | Acetone | 2000mg/dL | | n-Butanol | 2000mg/dL | | Ethylene glycol | 2000mg/dL | | Isopropanol | 2000mg/dL | | Methanol | 2000mg/dL | | n-Propanol | 2000mg/dL | | Ascorbic Acid | 1000 mg/dL | | Conjugated Bilirubin | 10 mg/dL | | Total Bilirubin | 15 mg/dL | | Creatinine | 500 mg/dL | | Glucose | 1200 mg/dL | | Hemoglobin | 100 mg/dL | | Human Albumin | 300 mg/dL | | Oxalic Acid | 200 mg/dL | | Sodium Chloride | 2300 mg/dL | | Gamma Globulin | 500 mg/dL | | Rivoflavin | 7.5 mg/dL | | pH | 3.0 and 9.0 units | {8} Changes in recovery of ethanol (i.e., interference) from these compounds were < 10%. Specific results are listed in the package insert. Cross-reactivities of > 10% were observed for acetaldehyde and n-propanol and this is listed in the package insert. f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device A study was performed to demonstrate the correlation between Medica's EtOH assay on the EasyRA chemistry analyzer and the EtOH predicate assay. A total of 68 patient urine specimens were tested using Medica's EtOH Reagent and calibrant on Medica's EasyRA analyzer and analyzed in duplicate on the Cobas-Mira using the Microgenics EtOH reagent and calibrant. Samples ranged from 13 to 460 mg/dL. The following are summary results of linear regression with 95% Confidence Intervals for slope and intercept: Slope: 0.9914 [+/- .0097]; y- intercept: 0.3602 [+/- 1.55] mg/dL; coefficient of variation (r²): 0.9982; Std. Error: 4.4750 b. Matrix comparison: This test is only for use with urine. 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: See a, above. c. Other clinical supportive data (when a. and b. are not applicable): Data regarding patient demographics and selection criteria were provided in the method comparison evaluation in the 510(k). 4. Clinical cut-off: Not applicable {9} Expected values/Reference range: The package insert states the following: Ethyl alcohol is not present in detectable concentrations in healthy adults who have not consumed ethanol. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 10
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...