← Product Code [KCY](/submissions/SU/subpart-f%E2%80%94therapeutic-devices/KCY) · K112874

# EZ VEIN INFLATABLE TOURNIQUET (K112874)

_Dominion Medical Devices, LLC · KCY · Oct 17, 2011 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-f%E2%80%94therapeutic-devices/KCY/K112874

## Device Facts

- **Applicant:** Dominion Medical Devices, LLC
- **Product Code:** [KCY](/submissions/SU/subpart-f%E2%80%94therapeutic-devices/KCY.md)
- **Decision Date:** Oct 17, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5910
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The EZvein™ inflatable secondary tourniquet is used in conjunction with a nonpneumatic tourniquet, to enhance the presentation of veins for access (blood draw or IV administration.)

## Device Story

EZ Vein is an inflatable secondary tourniquet; constructed of soft fabric and flexible plastic forming an inflatable bladder. Used by healthcare professionals in clinical settings to assist venous access. Applied after a conventional non-pneumatic tourniquet (e.g., latex tubing) is placed to block venous return. EZ Vein is positioned over the target vein via an integrated 'Vein Access Window'; inflation helps move blood from deep tissues into veins, causing distension for easier visualization and access. Velcro strips secure the device around appendages; an extension is provided for lower limbs or large arms. Device includes a self-release mechanism to prevent over-inflation. Benefits include more accurate and rapid venous access for the clinician.

## Clinical Evidence

Bench testing only. Testing demonstrated that the Velcro strip self-releases between 195 mmHg and 220 mmHg to prevent over-inflation.

## Technological Characteristics

Inflatable bladder constructed of soft fabric and flexible plastic. Features an integrated Vein Access Window and Velcro attachment strips. Includes an extension for varied limb sizes. Operates as a secondary pneumatic tourniquet. Self-release mechanism limits pressure to 195-220 mmHg.

## Regulatory Identification

A pneumatic tourniquet is an air-powered device consisting of a pressure-regulating unit, connecting tubing, and an inflatable cuff. The cuff is intended to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery.

## Predicate Devices

- Degania Silicone Tourniquet ([K875092](/device/K875092.md))
- DeRoyal's Pneumatic Tourniquet Cuff ([K953953](/device/K953953.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 1 7 2011

715-549-5380

Submitted on behalf of:

| Company Name: | Dominion Medical Devices, LLC    |              |
|---------------|----------------------------------|--------------|
| Address:      | 818 N. Oak<br>Guthrie, OK 73044  |              |
|               | Telephone:                       | 405-755-5308 |
|               | Fax:                             | 405-232-9659 |
| by:           | Elaine Duncan, M.S.M.E., RAC     |              |
|               | President, Paladin Medical, Inc. |              |
|               | PO Box 560                       |              |
|               | Stillwater, MN 55082             |              |
|               | Telephone:                       | 715-549-6035 |

Fax:

CONTACT PERSON: Elaine Duncan DATE PREPARED: October 12, 2011 TRADE NAME: EZ Vein inflatable tourniquet COMMON NAME: inflatable tourniquet CLASSIFICATION NAME: pneumatic tourniquet PRO CODE: КСУ

SUBSTANTIALLY EQUIVALENT TO: The EZ Vein™ inflatable tourniquet is substantially equivalent to both the non-pneumatic tourniquet such as the Degania Silicone Tourniquet (K875092) and a pneumatic accessory device, like the DeRoyal's Pneumatic Tourniquet Cuff (K953953). The similarities are the inflatable cuff technology.

DESCRIPTION of the DEVICE: The EZ Vein inflatable secondary tourniquet is made with soft fabric and flexible plastic to form an inflatable bladder. The integral Velcro attachment strips secure the EZVein around an appendage in the same way as would a blood pressure cuff. Dimensions have been chosen to wrap around most anyone's arm, but an extension is included to allow the EZVein to work on lower limbs or with large arms. The Vein Access Window allows the health care professional to position the EZVein over the target vein for access for IV drip, blood draw or injection. The EZVein cuff is applied after the initial conventional non-inflatable tourniquet (such as latex tubing) is applied. The initial tourniquet, as is conventional, blocks the return of venous blood through the major veins to the heart, but the application of the EZVein inflatable cuff helps move the blood in deep tissues into the veins to help them to distend for easier visualization and access. This may aid the medic desiring venous access to more accurately and rapidly achieve the desired aim.

510(K) Submission:

*** Confidential ***

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows handwritten text. The top line reads "K112874". The bottom line reads "2 of 2".

## 510(k) Summary-Continued

INDICATIONS FOR USE: The EZvein™ inflatable secondary tourniquet is used in conjunction with a non-pneumatic tourniquet, to enhance the presentation of veins for access (blood draw or IV administration.)

SUMMARY of PERFORMANCE TESTING: Conventional pneumatic tourniquets are more typically used in surgery to reduce bleeding and higher compressures may be required. In contrast, the EZVein™ cuff is not intended to be used without a primary non-pneumatic tourniquet, and has been tested to demonstrate that the Velcro™ strip self-releases between 195 mmHg and 220 mmHg, in order to ensure that the EZVein™ cannot be over-inflated.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

OCT 1 7 2011

Dominion Medical Devices, LLC % Paladin Medical, Inc. Elaine Duncan, M.S.M.E., RAC P.O. Box 560 Stillwater, Minnesota 55082

Re: K112874

Trade/Device Name: EZ Vein Inflatable Tourniquet Regulation Number: 21 CFR 878.5910 Regulation Name: Pneumatic tourniquet Regulatory Class: Class I Product Code: KCY Dated: September 29, 2011 Received: September 30, 2011

Dear Ms. Duncan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{3}------------------------------------------------

Page 2 - Elaine Duncan, M.S.M.E., RAC

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Sincerely yours,

Evin S. Keith

For Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known):

Device Name:

Indications for Use:

The EZvein™ inflatable secondary tourniquet is used in conjunction with a nonpneumatic tourniquet, to enhance the presentation of veins for access (blood draw or IV administration.)

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C) ·

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

NihReQh for mxm
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K112874

Page 1 of 1

Page 21 of 115

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-f%E2%80%94therapeutic-devices/KCY/K112874](https://fda.innolitics.com/submissions/SU/subpart-f%E2%80%94therapeutic-devices/KCY/K112874)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
