← Product Code [KMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/KMF) · K073663

# GELSPRAY WOUND DRESSING (K073663)

_Biocure, Inc. · KMF · Jan 30, 2008 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/KMF/K073663

## Device Facts

- **Applicant:** Biocure, Inc.
- **Product Code:** [KMF](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/KMF.md)
- **Decision Date:** Jan 30, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5090
- **Device Class:** Class 1
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

GelSpray™ Liquid Bandage is indicated for use to cover minor cuts, scrapes and minor irritations of the skin.

## Device Story

GelSpray™ Liquid Bandage is a two-part hydrogel system for topical wound coverage. Device consists of polyvinyl alcohol (PVA) macromer pre-polymers packaged in a double-barrel syringe with a common plunger and integrated static mixer. Upon application, components undergo redox-initiated in-situ polymerization to form a hydrogel dressing containing approximately 80% water. Device is applied directly to wound beds to cover minor cuts, scrapes, and irritations. System ensures uniform mixing and controlled delivery. Benefits include rapid curing and biocompatible wound protection. Used in clinical or home settings; operated by healthcare providers or patients.

## Clinical Evidence

Bench testing only. Biocompatibility and design validation testing performed to demonstrate performance, safety, and technical characteristics equivalent to predicate devices.

## Technological Characteristics

Two-part hydrogel system; polyvinyl alcohol (PVA) macromer; redox-reaction curing; 80% water content; double-barrel syringe delivery system with static mixer; sterile; Class I device.

## Regulatory Identification

A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.

## Predicate Devices

- 3M Liquid Bandage ([K053409](/device/K053409.md))
- Johnson & Johnson Liquid bandage ([K062602](/device/K062602.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 510k Summary

GelSpray™ Liquid Bandage 510K Summary

Submitter: BioCure, Inc. 2975 Gateway Drive Suite 100 Norcross, GA 30071

14 30 

Contact: Mr. Sameer Shums

Name of device: Liquid Bandage Proprietary/Trade Name: GelSpray™ Liquid bandage Classification name: Liquid bandage

## 510(k) Numbers and Product Codes of equivalent devices.

3M company 3M Liquid Bandage; Bandage, Liquid Code: KMF (21CFR 880.5090); K053409; S.E. June 19, 2006

Johnson & Johnson Johnson & Johnson Liquid bandage: bandage, Liguid Code: KMF (21CFR880.5090); K062602; S.E. December 11, 2006

## Indications for Use and Intended Population

GelSpray™ Liquid Bandage is indicated for use to cover minor cuts, scrapes and minor irritations of the skin.

#### Device Description

The GelSpray™ Liquid Bandage is a two part hydrogel system, capable of rapid curing when applied to a wound bed. The GelSpray™ Liquid Bandage consists of a macromer derived from polyvinyl alcohol (PVA). When mixed, the pre-polymers utilize a redox reaction to cure, resulting in the final dressing consisting predominantly of water (approximately 80%). The pre-polymer components are initially packaged and sterilized in a two part syringe with a common plunger for delivery. A static mixer is incorporated in the delivery system to ensure a uniform and adequate mix of the two pre-polymers and to allow a controlled delivery of the dressing to its intended site.

The product will be supplied sterile and packaged in a pre-filled, double barrel syringe. The target quantity will be providing the user with 4.5mL minimum.

#### Similarities and Differences to Predicates

{1}------------------------------------------------

K07 3663
Page 2 of 2

The Intended Use of GelSpray™ Liquid Bandage and the predicate devices are the same and unchanged, primarily other than product names.

GelSpray™ Liquid Bandage is supplied in a two barreled syringe, with applicator, and is un-polymerized prior to use. At the time of application, the two components in each barrel of the syringe are mixed and begin to polymerize immediately on application. Both GelSpray™ Liquid bandage and the predicate devices utilize in-situ polymerization. GelSpray™ Liquid bandage has similar properties to the predicate devices, however uses GelSpray™ Liquid Bandage uses PVA and the predicate devices use cyanoacrylates. GelSpray™ Liquid Bandage and the predicate devices have similar biocompatibility profiles.

Biocompatibility testing and design validation testing provided in the 510K have demonstrated that GelSpray™ Liquid bandage has the same performance, safety and technical characteristics such that no new issues of safety or effectiveness are raised.

There are more similarities than differences when comparing BioCure's GelSpray™ Liquid Bandage to the predicate devices.

### Conclusion

There are more similarities than differences between the predicate devices and BioCure GelSpray™ Liquid Bandage. The predicate devices and GelSpray™ Liquid Bandage have the nearly the same intended use, warnings, precautions and contraindications. When used in accordance with the instructions for use, by qualified personnel, the BioCure GelSpray™ Liquid Bandage is safe and effective, as indicated, for the intended use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Public Health Service

JAN 30 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

BioCure, Incorporated % Generic Devices Consulting, Inc. Mr. John Greenbaum President 20310 Southwest 48th Street Southwest Ranches, Florida 33332

Re: K073663

Trade/Device Name: GelSpray™ Liquid Bandage Regulation Number: 21 CFR 880.5090 Regulation Name: Liquid bandage Regulatory Class: Class I Product Code: KMF Dated: December 20, 2007 Received: December 26, 2007

Dear Mr. Greenbaum:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

# Page 2 – Mr. John Greenbaum

This letter will allow you to begin marketing your device as described in your Section 510(k) rms lotter with cation. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Millican

- Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{4}------------------------------------------------

Indications for Use

Page 1 of 1

510(k) Number(if known): K073663

Device Name: GelSpray™ Liquid Bandage

Indications For Use:

GelSpray™ Liquid Bandage is indicated for use to cover minor cuts, scrapes and minor irritations of the skin.

Prescription Use X (Per 21 CRF 801.109) Over-The-Counter Use

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Optional Format 1-2-96)
**(Division Sign-Off)**
Division of General, Restorative,
and Neurological Devices

510(k) Number 1073663

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/KMF/K073663](https://fda.innolitics.com/submissions/SU/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/KMF/K073663)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
