K181518 · Sebacia, Inc. · QCY · Sep 6, 2018 · General, Plastic Surgery
Device Facts
Record ID
K181518
Device Name
Sebacia Microparticles
Applicant
Sebacia, Inc.
Product Code
QCY · General, Plastic Surgery
Decision Date
Sep 6, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Sebacia Microparticles are indicated for use as an accessory to 1064 nm lasers to facilitate photothermal heating of sebaceous glands for the treatment of mild to moderate inflammatory acne vulgaris.
Device Story
Sebacia Microparticles are a non-sterile, single-use topical suspension containing silica-core, gold-shell, polyethylene glycol-coated microparticles. Applied to the face and massaged into the skin, the particles act as an accessory to commercially available 1064 nm Nd:YAG lasers. The laser energy is absorbed by the microparticles, inducing localized selective photothermolysis of sebaceous glands. Used in a clinical setting by a healthcare provider, the device allows for effective acne treatment using lower laser fluences and fewer treatment sessions compared to laser-only therapy. The procedure aims to reduce inflammatory lesion counts, providing a therapeutic benefit for patients with acne vulgaris.
Clinical Evidence
Randomized, controlled, parallel-group clinical study (n=168) comparing Sebacia Microparticles + 1064 nm laser vs. laser alone. Primary endpoint: % reduction in inflammatory lesion count (ILC) at Week 12. Results: SM + Laser (mean 40.17%, median 53.33%) vs. Laser Alone (mean 41.89%, median 45.45%). Secondary endpoints (e.g., IGA clear/almost clear) were supportive. SM + Laser group required fewer treatment sessions (median 3.0 vs 6.0) and lower mean fluence (29.44 J/cm² vs 48.93 J/cm²). Safety profile was positive with no severe or serious treatment-related adverse events.
Technological Characteristics
Microparticles consist of a silica core, gold shell, and polyethylene glycol coating. Designed for use with 1064 nm Nd:YAG lasers meeting specific parameters: fluence 20-35 J/cm², 30 ms pulse duration, and 5-15 mm spot size. Biocompatibility tested per ISO 10993-1. Non-sterile, single-use topical suspension. Shelf-life: 2 years.
Indications for Use
Indicated for patients with mild to moderate inflammatory acne vulgaris.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
K251149 — AviClear Laser System · Cutera, Inc. · Jul 11, 2025
K230660 — AviClear Laser System · Cutera, Inc. · Jun 6, 2023
K242035 — Accure Laser System · Accure Acne, Inc. · Oct 11, 2024
K213461 — AviClear Laser System · Cutera, Inc. · Mar 24, 2022
K040081 — PALOMAR LUXV HANDPIECE · Palomar Medical Technologies, Inc. · Apr 1, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Sebacia, Inc. % John Smith, Partner Hogan Lovells US LLP 555 13th Street. NW Washington, District of Columbia 20004
# Re: K181518
Trade/Device Name: Sebacia Microparticles Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: OCY Dated: June 8, 2018 Received: June 8, 2018
# Dear John Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Neil R Ogden -S 2018.09.06 23:23:49 -04'00'
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below
510(k) Number (if known)
Device Name
Sebacia Microparticles
Indications for Use (Describe)
Sebacia Microparticles are indicated for use as an accessory to 1064 nm lasers to facilitate photothermal heating of sebaceous glands for the treatment of mild to moderate inflammatory acne vulgaris.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 C.F.R. § 801 Subpart D)
□ Over-The-Counter Use (21 C.F.R. § 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
#### 510(k) SUMMARY
### SEBACIA, INC.
#### SEBACIA MICROPARTICLES
#### Submitter
Sebacia, Inc. 2905 Premiere Parkway Suite 150 Duluth, GA 30097
Contact: Todd J. Meyer COO & General Counsel Phone: 678-417-7626 Email: tmeyer@tif.net
Date Prepared: June 7. 2018
#### Device Information
| Device Name: | Sebacia Microparticles |
|-----------------------|-------------------------------------------------------------------------------------|
| Common Name: | Powered Laser Surgical Instrument |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery and in Dermatology |
| Classification Panel: | General & Plastic Surgery |
| Device Class: | Class II |
| Product Code: | QCY |
| Regulation: | 21 C.F.R. § 878.4810 |
#### Predicate Devices
Sebacia Microparticles are an accessory to legally marketed 1064 nm medical lasers intended for dermatologic use, and specifically for the treatment of mild to moderate inflammatory acne vulgaris, such as the 1064 nm Nd:YAG Cutera CoolGlide Aesthetic Lasers (K153671).
The predicate device is the 1064 nm Nd:YAG Cutera CoolGlide Aesthetic Lasers (K153671). This device is one example of a legally marketed device meeting the specifications for use with the Sebacia Microparticles.
#### Device Description
Sebacia Microparticles are a non-sterile, single-use product that is topically applied to the face for the purpose of supporting and supplementing the performance of legally marketed 1064 nm medical lasers for dermatologic use. The liquid suspension contains microparticles comprised of a silica core wrapped in a gold shell and coated with polyethylene glycol. The Sebacia Microparticles are topically applied to the face followed by gentle agitation of the skin surface using a mechanical massager and irradiated with a commercially available 1064 nm laser that heats the microparticles resulting in localized selective photothermolysis. The legally marketed 1064 nm medical laser used with the Sebacia Microparticles must meet the following specifications:
{4}------------------------------------------------
- o Fluence ranging from 20 J/cm2 to 35 J/cm²
- o Pulse duration of 30 ms
- At least 5-15mm spot size
# Intended Use / Indications for Use
Sebacia Microparticles are indicated for use as an accessory to 1064 nm lasers to facilitate photothermal heating of sebaceous glands for the treatment of mild to moderate inflammatory acne vulgaris.
## Performance Data
# Nonclinical Testing
Sebacia Microparticles were tested ex vivo for photothermal heating ability, biocompatibility (in accordance with ISO 10993-1) and shelf life. All results met predetermined performance criteria. The established shelf-life for the product is two (2) years based on real-time stability testing data. Three animal studies in a porcine model were performed to evaluate formulation and laser parameters, as well as safety, prior to the commencement of clinical testing. The results of animal testing demonstrate that Sebacia Microparticles meet prospective product requirements for functionality and safety.
### Clinical Studies
A randomized, controlled, parallel group, clinical study with blinded assessment was conducted in the U.S. to evaluate the Sebacia Microparticles when used with a 1064 nm Nd:Yag laser in facial mild to moderate inflammatory acne vulgaris. A total of 168 subjects received treatment, with approximately half of the subjects receiving the Sebacia Microparticles with laser ("SM + Laser") treatment and the other half receiving Laser Alone. Results are summarized in the table below:
| | SM + Laser<br>(n = 86) | Laser Alone<br>(n = 82) |
|-------------------------------------------------------|------------------------|-------------------------|
| PRIMARY ENDPOINT | | |
| % reduction in ILC (Baseline to Week 12); mean | 40.17 | 41.89 |
| % reduction in ILC (Baseline to Week 12); median | 53.33 | 45.45 |
| SECONDARY ENDPOINTS | | |
| % with ≥ 40% ILC reduction at Week 12 | 64.4 | 56.5 |
| % with IGA clear/almost clear at Week 12 | 30.1 | 31.9 |
| absolute reduction in ILC (Baseline to Wk 12); mean | 8.4 | 8.5 |
| absolute reduction in ILC (Baseline to Wk 12); median | 11.0 | 9.0 |
| TREATMENT PARAMETERS | | |
| laser fluence (J/cm²); mean | 29.44 | 48.93 |
| laser fluence (J/cm²); median | 30.00 | 49.79 |
| number of treatment sessions (All Patients); mean | 2.9 | 5.3 |
| number of treatment sessions (All Patients); median | 3.0 | 6.0 |
| number of treatment sessions (Completers); mean | 3.0 | 5.8 |
| number of treatment sessions (Completers); median | 3.0 | 6.0 |
# EFFECTIVENESS AND TREATMENT PARAMETER RESULTS
ILC = inflammatory lesion count
Percentage reduction in inflammatory lesion count from Baseline to Week 12 was assessed, revealing similar acne treatment effects for each group. Analyses of the secondary endpoints were supportive of the primary effectiveness analyses with results indicative of similar acne treatment effects for
{5}------------------------------------------------
SM + Laser when compared to Laser Alone. Summary results for subject satisfaction revealed that more subjects reported "Complete Improvement" or "Marked Improvement" in the SM + Laser treatment group than in the Laser Alone group. Analyses of overall mean fluence and the number of treatment sessions clearly demonstrated that use of fewer treatments and lower laser fluences in the SM + Laser Group resulted in equivalent reduction in mild to moderate inflammatory acne lesions when compared to the reduction in the Laser Alone Group.
Treatment with the Sebacia Microparticles also demonstrated a positive safety profile. The most frequent adverse events reported in both treatment groups were primarily general illnesses and complaints that were unrelated to study treatment. Furthermore, over 90% of the AEs reported in each treatment group were unrelated to study treatment. All reported AES, regardless of study treatment were of mild to moderate intensity. There were no severe AEs nor were there any serious and/or unanticipated adverse events related to study treatment.
# Conclusions
Sebacia Microparticles are indicated for use as an accessory to 1064 nm lasers to facilitate photothermal heating of sebaceous glands for the treatment of mild to moderate inflammatory acne vulgaris. In sum, the non-clinical and clinical data support the substantial equivalence of the Sebacia Microparticles to the predicate device for the proposed indications.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.