Resolv Endoscopic Hemostat System

K234131 · Hemostasis, LLC · QAU · Jul 26, 2024 · General, Plastic Surgery

Device Facts

Record IDK234131
Device NameResolv Endoscopic Hemostat System
ApplicantHemostasis, LLC
Product CodeQAU · General, Plastic Surgery
Decision DateJul 26, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4456
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Hemostasis Resolv Endoscopic Hemostat System is intended for use under the care of a health care professional for hemostasis of upper nonvariceal gastrointestinal bleeding, excluding Forrest Ia classification of bleeding.

Device Story

Sterile, single-use endoscopic system delivering plant-based polysaccharide hemostat powder to GI bleeding sites. System includes powder cartridge, flow dispenser, catheter tubing, and gas source tubing. Connects to external medical CO2 source (0.6–1.0 slpm flow). Operated by physician via endoscope distal controls to aim powder flow; trigger stops flow. Used in clinical settings for non-variceal upper GI bleeds. Powder absorbs moisture to form a mechanical barrier, promoting hemostasis. Maximum 16.7g powder per device; limit 33.4g per patient. Provides immediate bleeding control; reduces rebleeding risk.

Clinical Evidence

Clinical evaluation conducted in India with 59 subjects undergoing endoscopic upper GI procedures. Primary endpoint: hemostasis achievement. Results: 100% hemostasis success rate, 3.4% endoscopically confirmed re-bleed rate, and 5.1% all-cause rebleed rate. Supported by porcine animal model study validating safety and performance requirements per 21 CFR 878.4456.

Technological Characteristics

Plant-based polysaccharide hemostat powder; identical to NexStat Plus. Delivery via catheter/dispenser system using external medical CO2. Flow rate 0.6–1.0 slpm. Sterile (gamma radiation, SAL 10⁻⁶). Biocompatible per ISO 10993-1. Single-use. No software/electronics.

Indications for Use

Indicated for hemostasis of non-variceal upper gastrointestinal bleeding in patients, excluding Forrest Ia classification. Intended for use by healthcare professionals.

Regulatory Classification

Identification

A hemostatic device for intraluminal gastrointestinal use is a prescription device that is endoscopically applied to the upper and/or lower gastrointestinal tract and is intended to produce hemostasis via absorption of fluid or by other physical means.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible. (2) Performance data must support the sterility and pyrogenicity of the device. (3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life. (4) In vivo performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The testing must evaluate the following: (i) The ability to deliver the hemostatic material to the bleeding site; (ii) The ability to achieve hemostasis in a clinically relevant model of gastrointestinal bleeding; and (iii) Safety endpoints, including thromboembolic events, local and systemic toxicity, tissue trauma, gastrointestinal tract obstruction, and bowel distension and perforation. (5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated: (i) Materials characterization of all components must demonstrate the device meets established specifications, which must include compositional identity and purity, characterization of impurities, physical characteristics, and reactivity with fluids. (ii) Performance testing must demonstrate the mechanical integrity and functionality of the system used to deliver the device and demonstrate the device meets established specifications, including output pressure for propellant-based systems. (6) Labeling must include: (i) Information identifying and explaining how to use the device and its components; and (ii) A shelf life.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION July 26, 2024 Hemostasis LLC Lakshmi Ganesh Bollina Regulatory Affairs Specialist 5000 Township Parkway St. Paul, Minnesota 55110 Re: K234131 Trade/Device Name: Resolv Endoscopic Hemostat System Regulation Number: 21 CFR 878.4456 Regulation Name: Hemostatic Device For Intraluminal Gastrointestinal Use Regulatory Class: Class II Product Code: QAU Dated: June 25, 2024 Received: June 26, 2024 Dear Lakshmi Ganesh Bollina: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K234131 - Lakshmi Ganesh Bollina Page 2 Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Tek N. Lamichhane - S Digitally signed by Tek N. Lamichhane Date: 2024.07.26 14:56:04 -04'00" Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2} FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K234131 | | | Device Name Resolv Endoscopic Hemostat System | | | Indications for Use (Describe) The Hemostasis Resolv Endoscopic Hemostat System is intended for use under the care of a health care professional for hemostasis of upper nonvariceal gastrointestinal bleeding, excluding Forrest Ia classification of bleeding. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {3} 510(k) Summary K234131 Resolv™ Endoscopic Hemostat System Date Prepared: 25 July 2024 Submitter: Hemostasis, LLC 5000 Township Parkway St. Paul, MN 55110 Telephone: 651-855-1466 Fax: 651-855-1465 Contact: Mr. Lakshmi Ganesh Bollina Regulatory Affairs Specialist 5000 Township Parkway St. Paul, MN 55110 Telephone: 651-855-1466 Proprietary Name: Resolv™ Endoscopic Hemostat System Common/Usual Name: Endoscopic Hemostat Classification: Hemostatic device for intraluminal gastrointestinal use, Product Code QAU, Class II, 21 CFR 878.4456 Predicate Device: EndoClot Polysaccharide Hemostatic System K190677 Reference Device: NexStat® Plus Topical Hemostat K122886 Establishment Registration Number: 3007225047 Description: The Resolv™ Endoscopic Hemostat System is a sterile prescription device intended for use under the care of a health care professional for hemostasis of nonvariceal upper gastrointestinal bleeding, excluding Forrest Ia classification of bleeding. The device consists of a hemostat powder and dispenser components provided in a tray with a Tyvek lid acting as the sterile barrier. The hemostat powder is further contained within in a capped cartridge and packaged in a foil pouch. The dispenser components include a flow dispenser, two catheter tubing sets and a gas source tubing set. The gas source tubing set is connected to a medical CO₂ gas source that is flow and pressure controlled. The endoscope distal end controls are used to aim the flow and placement of the hemostat powder. The trigger is released to stop the flow of powder. The device contains a maximum of 16.7 grams of hemostat powder and no more than 2 devices (33.4 grams) may be used on a patient either in a single procedure or in conjunction with rebleeding up to 10 days after the initial procedure. If 2 devices (33.4 grams) have been used in the initial procedure an alternate modality must be used should rebleeding occur. The hemostat powder is identical to the commercially available topical hemostat, NexStat® Plus which received 510(k) clearance with K122886 as a topical wound dressing, comprised of plant-based polysaccharides. {4} # Indications for Use: The Hemostasis Resolv™ Endoscopic Hemostat System is intended for use under the care of a health care professional for hemostasis of non-variceal upper gastrointestinal bleeding, excluding Forrest Ia classification of bleeding. # Substantial Equivalence: The Resolv™ Endoscopic Hemostat System is substantially equivalent to the primary predicate EndoClot® Polysaccharide Hemostatic System, 510(k) Number K190677 (QAU), clearance date 01/29/2021, for the following reasons: - Both have the same FDA Classification, Intended use and target population. - Both are sterile, single use devices that deliver a plant-based polysaccharide powder to the GI bleeding site. - Both contain materials proven biocompatible for their intended use in accordance with ISO 10993-1. The primary differences appear to be the gas source, flow rate and digestibility in the GI Tract. - Gas Source: Resolv™ utilizes the hospital supply CO₂, while the EndoClot® PHS utilizes air drawn from OR and filtered. - Gas Flowrate: The flow is set within the range of 0.6 to 1.0 slpm for Resolv™, while for Endoclot the flow is based on gas pressure defined in Endoclot IFU. - Digestibility in the GI Tract leading to release of sugar molecules: The Resolv hemostat powder is 13% digestible where Endoclot PHS powder is fully digested. In addition, the powder hemostat used in the Resolv™ is identical to the Class II topical hemostat, NexStat® Plus, in terms of material composition, mechanism of action, manufacturing and sterilization. Due to the differences in Intended Use and delivery system, Resolv™ Endoscopic Hemostat System has undergone additional biocompatibility testing, sterilization validation, packaging testing and in vivo and non-clinical performance testing in accordance with its intended use special controls specified in 21 CFR 878.4456 for a Hemostatic device for intraluminal gastrointestinal use. # Summary of Non-Clinical Performance Data ## Biocompatibility: Biocompatibility testing was performed using ISO 10993 – Biological Evaluation of Medical Devices and FDA guidance document Use of International Standard ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process” dated Sept.8, 2023. The Hemostasis Resolv™ Endoscopic Hemostat System complies with the biocompatibility requirements for its intended use. ## Sterilization, Packaging, Shelf Life: The Hemostasis Resolv™ Endoscopic Hemostat System is sterilized using a validated gamma radiation method to assure a sterility assurance level (SAL) of 10⁻⁶. The packaged device has been successfully tested for package integrity and device functionality over the labeled shelf life of 24 months. ## Performance Bench Testing: Design verification testing was performed for the Resolv™ Endoscopic Hemostat System to demonstrate physical and functional requirements were met. {5} # Animal Performance Testing: In vivo animal studies have been conducted to support the substantial equivalence claims of Resolv™ Endoscopic Hemostat system as compared to the predicate device Endoclot®. The objective of the study was to validate the efficacy and safety of Resolv™ Endoscopic Hemostat system in a porcine animal model for bleeding control in the upper GI tract. The results of the animal study demonstrated that the Resolv™ did meet the in vivo safety and performance requirements of the special controls in 21 CFR 878.4456 by effectively delivering the hemostat to the bleeding site and achieving hemostasis and meeting all the safety endpoints and is substantially equivalent to the predicate. # Clinical Performance Testing: A human clinical evaluation was conducted in India to support the safety and effectiveness of the use of Resolv™ Endoscopic Hemostat System. Resolv™ was studied in 59 subjects undergoing endoscopic upper GI procedures. Hemostasis was achieved in 100% of patients with an endoscopically confirmed re-bleed rate of 3.4% and an all-cause rebleed rate of 5.1%. | Hemostat System | Forrest Bleeding Classification | Number of Subjects | Hemostasis % | Rebleed Rate % | | | --- | --- | --- | --- | --- | --- | | | | | | Endoscopically Confirmed | All Cause | | Resolv | Ib | 59 | 100% | 3.4% | 5.1% | Real world evidence provided for the Resolv Endoscopic Hemostat System support that the Resolv is substantially equivalent to the predicate in all aspects of safety and efficacy. # Conclusion: Utilizing FDA's Guidance for Industry and FDA Staff "Format for Traditional and Abbreviated 510(k)s" issued on September 13, 2019, a comparison of key characteristics demonstrates that the proposed device Resolv™ Endoscopic Hemostat System is substantially equivalent to the predicate device in terms of intended use and performance characteristics. The Resolv™ Endoscopic Hemostat System is as safe, as effective, and performs as well as the predicate device. The substantial equivalence analysis supports that the Resolv™ Endoscopic Hemostat System presents no new concerns about safety and effectiveness and is suitable for its intended use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...