← Product Code [QAI](/submissions/SU/subpart-e%E2%80%94surgical-devices/QAI) · K212558

# MicroPen EVO (K212558)

_Eclipse Medcorp, LLC · QAI · Nov 24, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QAI/K212558

## Device Facts

- **Applicant:** Eclipse Medcorp, LLC
- **Product Code:** [QAI](/submissions/SU/subpart-e%E2%80%94surgical-devices/QAI.md)
- **Decision Date:** Nov 24, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4430
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Eclipse MicroPen® EVO is a microneedling device and accessories intended to be used as a treatment to improve the appearance of wrinkles of the neck for Fitzpatrick skin types II - IV and to improve the appearance of facial acne scars in adults with all Fitzpatrick skin types aged 22 years and older.

## Device Story

Minimally invasive, powered microneedling device; uses reciprocating cartridge head with 14 stainless steel needles to create microscopic epidermal punctures. Operated by licensed healthcare practitioners or individuals directed by practitioners in clinical settings. Device consists of reusable pen body, rechargeable Li-ion battery, and sterile, single-use microneedling cartridge. User adjusts needle penetration depth (up to 2.0 mm) via device controls. Microprocessor-controlled rotary mechanism drives needle puncture rate of 105-128 stamps/second. Intended to improve appearance of facial acne scars and neck wrinkles. Benefits include mechanical skin rejuvenation through controlled micro-injury. Cross-contamination prevented via disposable MicroSleeve sheath.

## Clinical Evidence

Bench testing only. Testing included puncture rate, needle penetration accuracy, needle retention, battery life, cartridge reliability, suction prevention, fluid ingress, cleaning/disinfection validation, and biocompatibility (ISO 10993-1, -5, -10, -11). Sterilization (ISO 11135) and shelf-life/package integrity (ASTM F1980, F1886, F1929, F88) were validated. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2.

## Technological Characteristics

Powered microneedling device; 14 stainless steel needles (32 Ga) in radial arrangement. Rotary operating principle; 6300-7700 RPM; 105-128 stamps/second. Rechargeable Li-ion battery. Depth settings: 0-2.0 mm in 0.25 mm increments. Biocompatible materials per ISO 10993. Sterile, single-use cartridges (Ethylene Oxide). Microprocessor-based control. Cross-contamination barrier: disposable MicroSleeve sheath.

## Regulatory Identification

A microneedling device for aesthetic use is a device using one or more needles to mechanically puncture and injure skin tissue for aesthetic use. This classification does not include devices intended for transdermal delivery of topical products such as cosmetics, drugs, or biologics.

## Special Controls

The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate. Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Accuracy of needle penetration depth and puncture rate; (ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and (iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use. Performance data must demonstrate the sterility of the patient-contacting components of the device. Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life. Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device. Software verification, validation, and hazard analysis must be performed for all software components of the device. The patient-contacting components of the device must be demonstrated to be biocompatible. Performance data must validate the cleaning and disinfection instructions for reusable components of the device. Labeling must include the following: (i) Information on how to operate the device and its components and the typical course of treatment; (ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate; (iii) Validated methods and instructions for reprocessing of any reusable components: (iv) Disposal instructions; and (v) Shelf life. Patient labeling must be provided and must include: (i) Information on how the device operates and the typical course of treatment; (ii) The probable risks and benefits associated with use of the device; and (iii) Post-operative care instructions.

*Classification.* Class II (special controls). The special controls for this device are:(1) The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate.
(2) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Accuracy of needle penetration depth and puncture rate;
(ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and
(iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use.
(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.
(5) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device.
(6) Software verification, validation, and hazard analysis must be performed for all software components of the device.
(7) The patient-contacting components of the device must be demonstrated to be biocompatible.
(8) Performance data must validate the cleaning and disinfection instructions for reusable components of the device.
(9) Labeling must include the following:
(i) Information on how to operate the device and its components and the typical course of treatment;
(ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate;
(iii) Validated methods and instructions for reprocessing of any reusable components;
(iv) Disposal instructions; and
(v) A shelf life.
(10) Patient labeling must be provided and must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) The probable risks and benefits associated with use of the device; and
(iii) Postoperative care instructions.

## Predicate Devices

- SkinPen Precision System ([K202243](/device/K202243.md))

## Reference Devices

- Eclipse MicroPen EVO ([K203144](/device/K203144.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 24, 2021

Eclipse Medcorp LLC Julie Summerville Senior Director of Product Management 5916 Stone Creek Drive The Colony, Texas 75056

Re: K212558

Trade/Device Name: MicroPen EVO Regulation Number: 21 CFR 878.4430 Regulation Name: Microneedling Device For Aesthetic Use Regulatory Class: Class II Product Code: OAI Dated: September 1, 2021 Received: September 2, 2021

Dear Julie Summerville:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212558

Device Name

MicroPen® EVO

#### Indications for Use (Describe)

The Eclipse MicroPen® EVO is a microneedling device and accessories intended to be used as a treatment to improve the appearance of wrinkles of the neck for Fitzpatrick skin types II - IV and to improve the appearance of facial acne scars in adults with all Fitzpatrick skin types aged 22 years and older.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| 区 Prescription Use (Part 21 CFR 801 Subpart D)  | Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 510(k) Summary<br>[This 510(k) Summary is submitted in accordance with 21 CFR Part<br>807.92(c)] |                                                                                       |                                                                                                                                  |
|--------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Submitted by:                                                                                    | Eclipse Medcorp, LLC.<br>5916 Stone Creek Dr.<br>Suite 120<br>The Colony, TX<br>75056 | Contact Person:<br>Julie Summerville<br>Senior Dir of Product<br>Management972-380-2911 x<br>2405<br>jsummerville@eclipsemed.com |
| Date Prepared:                                                                                   | November 3, 2021                                                                      |                                                                                                                                  |
| Trade Name:                                                                                      | MicroPen EVO™                                                                         |                                                                                                                                  |
| Common Name:                                                                                     | Powered Microneedling Device                                                          |                                                                                                                                  |
| Product Code; Regulation Name & No:                                                              | QAI<br>Microneedling device for aesthetic use                                         | 21 CFR §878.4430                                                                                                                 |

## Device Description:

The Eclipse MicroPen EVO™ is a minimally invasive microneedling device that mechanically creates microscopic punctures in the epidermal layers of the skin by means of micro-needles in a reciprocating cartridge head. The MicroPen EVO is comprised of a reusable pen body, a sterile, single use microneedlingcartridge, a rechargeable battery pack, and a battery charger with power supply. The microneedling cartridge is attached to the pen body and activated with an On/Off button. The depth of needle penetrationcan be adjusted by the user depending on the condition of the skin being treated. Charging is accomplished by placing the MicroPen EVO battery pack on the Charger base.

## Indications for Use:

The Eclipse MicroPen® EVO is a microneedling device and accessories intended to be used as a treatment to improve the appearance of wrinkles of the neck for Fitzpatrick skin types II - IV and to improve the appearance of facial acne scars in adults with all Fitzpatrick skin types aged 22 years and older.

## Predicate Device:

The Eclipse MicroPen EVO predicate device is the SkinPen Precision System by Crown Aesthetics (formerly Bellus Medical), K202243.

## Reference Device:

The Eclipse MicroPen EVO K203144 is a reference device for the Eclipse MicroPen EVO that is subject to this submission.

{4}------------------------------------------------

| Technological Characteristics Comparison<br>Chart |                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                    |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                   | Subject device<br>Eclipse MicroPen EVO                                                                                                                                                                                                                                                                                                       | Predicate<br>SkinPen Precision<br>System                                                                                                                                                                                                                                                                                                        | Comparison                                                                                                                                                         |
| 510(k)                                            | K212558                                                                                                                                                                                                                                                                                                                                      | K202243                                                                                                                                                                                                                                                                                                                                         | NA                                                                                                                                                                 |
| Manufacturer                                      | Eclipse MedCorp LLC                                                                                                                                                                                                                                                                                                                          | Crown<br>Aesthetics                                                                                                                                                                                                                                                                                                                             | NA                                                                                                                                                                 |
| Trade Name                                        | MicroPen EVO™                                                                                                                                                                                                                                                                                                                                | SkinPen® Precision System                                                                                                                                                                                                                                                                                                                       | NA                                                                                                                                                                 |
| Product Code                                      | QAI                                                                                                                                                                                                                                                                                                                                          | QAI                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                               |
| Regulation #                                      | 21 CFR Part 878.4430                                                                                                                                                                                                                                                                                                                         | 21 CFR Part 878.4430                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                               |
| Reg Name                                          | Microneedling device for<br>aesthetic use                                                                                                                                                                                                                                                                                                    | Microneedling device for<br>aesthetic use                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                               |
| Device Class                                      | Class II                                                                                                                                                                                                                                                                                                                                     | Class II                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                               |
| Indication for use /<br>Intended use              | The Eclipse MicroPen® EVO is a<br>microneedling device and<br>accessories intended to be used<br>as a treatment to improve the<br>appearance of wrinkles of the<br>neck for Fitzpatrick skin types II -<br>IV and to improve the<br>appearance of facial acne scars<br>in adults with all Fitzpatrick skin<br>types aged 22 years and older. | The SkinPen Precision System is a<br>microneedling device and<br>accessories intended to be used<br>as a treatment to improve the<br>appearance of wrinkles of the<br>neck for Fitzpatrick skin types II -<br>IV and to improve the<br>appearance of facial acne scars in<br>adults with all Fitzpatrick skin<br>types aged 22 years and older. | Same                                                                                                                                                               |
| Intended Users                                    | Rx Only: Licensed healthcare<br>practitioners or individuals<br>directedby practitioners                                                                                                                                                                                                                                                     | Rx Only: licensed healthcare<br>practitioners or individuals<br>directedby practitioners                                                                                                                                                                                                                                                        | Same                                                                                                                                                               |
| Use Location                                      | Face                                                                                                                                                                                                                                                                                                                                         | Face                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                               |
| Power Source<br>(Pen Body)                        | Rechargeable Li-ion battery                                                                                                                                                                                                                                                                                                                  | Rechargeable Li-ion battery                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                               |
| Power Source<br>(Battery Charger)                 | AC Powered                                                                                                                                                                                                                                                                                                                                   | AC Powered                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                               |
| Control Mechanism                                 | Microprocessor; embedded<br>software                                                                                                                                                                                                                                                                                                         | Microprocessor; embedded<br>software                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                               |
| Control Mechanism                                 | Microprocessor; embedded<br>software                                                                                                                                                                                                                                                                                                         | Microprocessor; embedded<br>software                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                               |
| Operating Principle                               | Rotary                                                                                                                                                                                                                                                                                                                                       | Rotary                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                               |
| Single Speed(RPM)                                 | 6300-7700                                                                                                                                                                                                                                                                                                                                    | 6300 - 7700                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                               |
| Puncture Rate                                     | 105-128 stamps/second                                                                                                                                                                                                                                                                                                                        | 105-128 stamps/second                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                               |
| Microneedling<br>Cartridge                        | Sterile, Single Use                                                                                                                                                                                                                                                                                                                          | Sterile, Single Use                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                               |
| No. of Needles                                    | 14                                                                                                                                                                                                                                                                                                                                           | 14                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                               |
| Needle Gauge                                      | 32 Ga                                                                                                                                                                                                                                                                                                                                        | 32 Ga                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                               |
| Needle Material                                   | Stainless Steel                                                                                                                                                                                                                                                                                                                              | Stainless Steel                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                               |
| Needle Shape<br>Geometry                          | Straight, cylindrical body with a<br>conical tapered, sharp point                                                                                                                                                                                                                                                                            | Straight, cylindrical body with<br>aconical tapered, sharp point                                                                                                                                                                                                                                                                                | Same                                                                                                                                                               |
| Arrangement                                       | Needles radially arranged                                                                                                                                                                                                                                                                                                                    | Needles radially arranged                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                               |
| Needle Spacing                                    | 2 mm spacing/3.48 mm² per<br>needle                                                                                                                                                                                                                                                                                                          | 2 mm spacing/3.54 mm² per<br>needle                                                                                                                                                                                                                                                                                                             | Equivalent:<br>subject device<br>has a slightly<br>smaller total<br>surface area of<br>the hub. No<br>affect to<br>geometry,<br>puncture pattern,<br>needle stamp. |
| Penetration Depth                                 | 1.5 mm (Recommended)                                                                                                                                                                                                                                                                                                                         | 1.5 mm (Recommended)                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                               |
| Max Needle Depth<br>Setting                       | 2.25 mm                                                                                                                                                                                                                                                                                                                                      | 2.5 mm                                                                                                                                                                                                                                                                                                                                          | Different: Not<br>significant;<br>treatment depth<br>is 1.5 mm for<br>both devices.                                                                                |
| Penetration Depth<br>Selection                    | 9 depth settings; 0 mm to 2.0<br>mm in 0.25 mm increments                                                                                                                                                                                                                                                                                    | 11 depth settings; 0 mm to 2.5mm in 0.25 mm increments                                                                                                                                                                                                                                                                                          | Similar                                                                                                                                                            |
| Sterility<br>(cartridge)                          | Ethylene Oxide                                                                                                                                                                                                                                                                                                                               | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                               |
| Shelf Life<br>(Cartridge)                         | 2 years                                                                                                                                                                                                                                                                                                                                      | 2 years (min)                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                               |
| Barrier: Cross-<br>Contamination                  | MicroSleeve Sheath(Disposable)                                                                                                                                                                                                                                                                                                               | BioSheath (Disposable)                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                               |

# Technological Characteristics and Comparison to Predicate:

{5}------------------------------------------------

{6}------------------------------------------------

## Performance Data:

In combination with the general controls of the FD&C Act, the Eclipse MicroPen EVO™ microneedling device for aesthetic use has been subjected to performance testing and adheres to the following special controls and standards:

- Puncture rate; needle penetration depth and accuracy; needle retention; battery life; cartridge reliability, and suction prevention testing;
- Fluid ingress testing for cross contamination prevention;
- Cleaning and disinfection validation for reusable components;
- Biocompatibility evaluation conducted in accordance with ISO 10993-1 on the final, finished microneedling cartridge: (1) Cytotoxicity (ISO 10993-5:2009); (2) Sensitization and (3) Irritation/Intracutaneous Reactivity (ISO 10993-10: 2010); (4) Acute Systemic Toxicity (ISO 10993-11:2017); (5) Material Mediated Pyrogenicity (ISO 10993-11:2017, USP 41 NF 36:2018, <151> Pyrogen Test. The results of these tests demonstrated the device to be biocompatible with no evidence of material mediated pyrogenicity.
- Sterilization, Shelf Life/Package Integrity in accordance with the following standards:
	- o Ethylene oxide sterilization per ISO 11135-2014; ISO 11737-2: 2009; ISO 11737-2: 2009; ISO 10993-7:2008
	- o ASTM-F1980 Std Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
	- ASTM-F1886-2016 Std Test Method for Determining Integrity of Seals for Medical Packaging by Visual Inspection;
	- o ASTM-F1929-2015 Std Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
	- o ASTM-F88 Std Test Method for Seal Strength of Flexible Barrier Materials; ANSI/AAMI/ISO 11607-1: 2006(R) 2010/ A1-2014, Packaging for Terminally Sterilized Medical Devices Part 1: Requirements for Materials, Sterile Barrier System and Packaging Systems
- Electrical Safety and Electromagnetic Compatibility: IEC- 60601-1:2005 + A1: 2012 Medical electrical equipment– Part 1: General Requirements for Basic Safety and Essential Performance
- EN/IEC 60601-1-2: 2015 /IEC 60601-1-2:2014–Medical electrical equipment—Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility - Requirement and tests

# Substantial Equivalence:

The MicroPen EVO is substantially equivalent to the Crown Aesthetics SkinPen®Precision System predicate device. The devices are under the same product code (QAI), both have the same intended use/ indication for use, same number of needles, gauge, shape and arrangement, same material, recommended penetration depth, speed, puncture rate and sterilization method. The only technological differencesare the maximum needle depth setting: the MicroPen EVO is 0.5 mm shorter (2.0 mm) than the predicate (2.5 mm); and the penetration depth selection: the MicroPen EVO has 9 depth settings (0-2.0 mm in 0.25 mm increments) and the predicate device has 11 depth settings (0-2.5 mm in 0.25 mm increments). These differencesare minor and are not significant since both devices recommend the same treatment depth (1.5 mm).

{7}------------------------------------------------

There is a small and insignificant difference between the subject and predicate in total surface area of the hub, however theneedle spacing is the same (2 mm) and there is no effect on geometry, puncture pattern, needle stamp. These minor differences do not raise new questions of safety and effectiveness. Further, the results of performance testing support substantial equivalence of the Eclipse MicroPen EVO to the predicate device and demonstrate the MicroPen EVO can perform its intended function.

## Conclusion:

The Eclipse MicroPen EVO is considered to be substantially equivalent to the predicate device based on the intended use, technological characteristics, and the results of device testing submitted.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QAI/K212558](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QAI/K212558)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
