K210293 · AtriCure, Inc. · PZX · Mar 3, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210293
Device Name
AtriClip LAA Exclusion System
Applicant
AtriCure, Inc.
Product Code
PZX · General, Plastic Surgery
Decision Date
Mar 3, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The AtriClip LAA Exclusion System is indicated for the exclusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures. Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.
Device Story
AtriClip LAA Exclusion System is a single-use, sterile, implantable clip preloaded on a disposable applier. Used by cardiac surgeons during open or minimally invasive cardiac procedures to exclude the left atrial appendage (LAA). The clip applies uniform pressure to the LAA, facilitating mechanical closure and electrical isolation. The system includes a handle, shaft, suture anchors, and articulation controls. The device is provided sterile via gamma irradiation. It is intended to reduce the risk of thrombus formation in the LAA. No software or electronic components are involved.
Clinical Evidence
No new clinical data was provided. The submission relies on previously cleared bench and animal data (Kamohara 2005, 2006; Starck 2012) which remain valid for the modified device.
Technological Characteristics
Implantable clip and applier; titanium clip components; polyethylene terephthalate suture material. Dimensions: 35mm, 40mm, 45mm, 50mm lengths. Sterilization: Gamma irradiation (10^-6 SAL). Biocompatibility assessed per ISO 10993. Non-pyrogenic. No software or electronic components.
Indications for Use
Indicated for patients undergoing cardiac surgical procedures requiring exclusion of the left atrial appendage (LAA) under direct visualization.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
K173031 — AtriClip LAA Exclusion System with Preloaded PRO·V Clip · AtriCure, Inc. · Oct 25, 2017
K233407 — AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip (LAA0); AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip (ACH1); AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip (ACH2); AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip (PRO1); AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip (PRO2); AtriClip PRO-V LAA Exclusion System (PROV); AtriClip Flex-V LAA Exclusion System with preloaded V Clip (ACHV) · AtriCure, Inc. · Nov 2, 2023
K243860 — AtriClip PRO-Mini LAA Exclusion System (PROM) · AtriCure, Inc. · Jan 15, 2025
K160454 — AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip · AtriCure, Inc. · Mar 18, 2016
K163261 — AtriClip PRO, PRO2, and PRO V LAA Exclusion System devices with preloaded Gillinov-Cosgrove Clip · AtriCure, Inc. · May 19, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 3, 2021
AtriCure, Inc. Mary Galeano Senior Regulatory Affairs Specialist 7555 Innovation Way Mason, Ohio 45040
Re: K210293
Trade/Device Name: AtriClip LAA Exclusion System Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: PZX Dated: January 29, 2021 Received: February 2, 2021
Dear Mary Galeano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210293
### Device Name
AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip, AtriClip LAA Exclusion System with preloaded V-Clip
### Indications for Use (Describe)
The AtriClip LAA Exclusion System is indicated for the exclusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures.
Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### l. Applicant Information
| Manufacturer: | AtriCure, Inc.<br>7555 Innovation Way<br>Mason, Ohio 45040 |
|--------------------|-----------------------------------------------------------------------------|
| Contact Person: | Mary Galeano, RAC<br>Sr. Regulatory Affairs Specialist<br>P: 513-257-6279 |
| Alternate Contact: | Jonathan McElwee, RAC<br>Sr. Manager, Regulatory Affairs<br>P: 513-795-9127 |
| Date Prepared: | 03/03/2021 |
#### ll. Device Information
| Proprietary Name: | AtriClip™ LAA Exclusion System with preloaded Gillinov-<br>Cosgrove Clip<br>(LAA0, ACH1, ACH2, PRO1, and PRO2)<br>AtriClip™ LAA Exclusion System with preloaded V-Clip<br>(PROV and ACHV) |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Implantable Clip and Clip Applier |
| Classification: | Implantable Clip and Clip Applier<br>Regulatory Class: Class II; per 21 CFR 878.4300<br>Product Code: PZX<br>Classification Panel: Cardiovascular |
| Predicate Device: | The device proposed for modification in this submission is the<br>AtriClip LAA Exclusion System cleared via K191413 on August 21<br>2019. |
#### III. Device Description
The AtriClip LAA Exclusion System consists of a single use, sterile, self-closing, implantable Clip preloaded on a Single Use Clip Applier along with a selection guide. When closed, the Clip applies uniform pressure over the length of the Clip to ensure consistent, reproducible, and secure exclusion of the left atrial appendage (LAA). Preclinical animal studies (Kamohara 2005,2006) demonstrate that complete exclusion with the Clip also results in acute and chronic electrical isolation of the LAA. A human clinical study (Starck 2012) has demonstrated acute electrical isolation. Chronic electrical isolation has not been evaluated in human clinical studies.
{4}------------------------------------------------
The Clip is available in the following lengths to accommodate difference sizes of LAA: 35 mm, 40 mm, 45 mm, and 50 mm. The Clip Appliers are disposable device with a handle, shaft, suture anchors, articulation controls and deployment loop which contains the Clip.
## Intended Use/ Indications for Use
The AtriClip LAA Exclusion System is indicated for the exclusion of the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures.
Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.
### IV. Comparison of Technological Characteristics with the Predicate Device (AtriClip LAA Exclusion System - K191413)
| # | Feature | Previously Cleared AtriClip™ LAA<br>Exclusion System | Modified AtriClip™ LAA Exclusion<br>System | Equivalence<br>Comparison |
|------|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| 1 | Proprietary Name | AOD1: AtriClip LAA Exclusion System with<br>Preloaded Gillinov-Cosgrove Clip<br><br>AOD2: AtriClip LAA Exclusion System with<br>Preloaded PROV Clip | AOD1: AtriClip LAA Exclusion System with<br>Preloaded Gillinov-Cosgrove Clip<br><br>AOD2: AtriClip LAA Exclusion System with<br>PROV Clip | Same |
| | | 2 | Clip Applier<br>Product Code(s) | |
| 3 | Device<br>Classification | 21 CFR 878.4300,<br>product code: PZX | 21 CFR 878.4300,<br>product code: PZX | Same |
| 4 | Regulatory Class | Class II | Class II | Same |
| 5 | FDA Division | Cardiovascular | Cardiovascular | Same |
| 6 | Performance<br>Standards | No performance standards applicable to<br>this product have been promulgated by<br>FDA | No performance standards applicable to<br>this product have been promulgated by<br>FDA | Same |
| 7 | Intended Use | The AtriClip LAA Exclusion System is<br>indicated for the exclusion of the heart's left<br>atrial appendage, performed under direct<br>visualization and in conjunction with other<br>cardiac surgical procedures.<br>Direct visualization, in this context, requires<br>that the surgeon is able to see the heart<br>directly, with or without assistance from a<br>camera, endoscope, etc., or other<br>appropriate viewing technologies. | The AtriClip LAA Exclusion System is<br>indicated for the exclusion of the heart's left<br>atrial appendage, performed under direct<br>visualization and in conjunction with other<br>cardiac surgical procedures.<br>Direct visualization, in this context, requires<br>that the surgeon is able to see the heart<br>directly, with or without assistance from a<br>camera, endoscope, etc., or other<br>appropriate viewing technologies. | Same |
| # | Feature | Previously Cleared AtriClip™ LAA<br>Exclusion System | Modified AtriClip™ LAA Exclusion<br>System | Equivalence<br>Comparison |
| 8 | Device<br>Description in the<br>Instructions for<br>Use | The AtriClip LAA Exclusion System<br>contains the Gillinov-Cosgrove LAA Clip<br>(Clip) for exclusion of the heart's left atrial<br>appendage (LAA). Preclinical animal<br>studies (Kamohara 2005, 2006)<br>demonstrate that complete exclusion with<br>the Clip also results in acute and chronic<br>electrical isolation of the LAA. A human<br>clinical study (Starck 2012) has<br>demonstrated acute electrical isolation.<br>Chronic electrical isolation has not been<br>evaluated in human clinical studies.<br>The Clip is pre-loaded on a disposable<br>Clip applier. The AtriClip LAA Exclusion<br>System with preloaded Gillinov-Cosgrove<br>Clip does not contain natural rubber latex<br>components. | The AtriClip LAA Exclusion System<br>contains the Gillinov-Cosgrove LAA Clip<br>(Clip) for exclusion of the heart's left atrial<br>appendage (LAA). Preclinical animal<br>studies (Kamohara 2005, 2006)<br>demonstrate that complete exclusion with<br>the Clip also results in acute and chronic<br>electrical isolation of the LAA. A human<br>clinical study (Starck 2012) has<br>demonstrated acute electrical isolation.<br>Chronic electrical isolation has not been<br>evaluated in human clinical studies.<br>The Clip is pre-loaded on a disposable<br>Clip applier. The AtriClip LAA Exclusion<br>System with preloaded Gillinov-Cosgrove<br>Clip does not contain natural rubber latex<br>components. | Same |
| 9 | Software | This product does not contain software. | This product does not contain software. | Same |
| 10 | Shelf Life | 3 years. | 3 years. | Same |
| 11 | Biocompatibility | Passed. Materials have been assessed<br>based on ISO 10993 and are commonly<br>employed in tissue contacting devices. | Passed. Materials have been assessed<br>based on ISO 10993 and are commonly<br>employed in tissue contacting devices. | Same |
| 12 | Sterilization | Gamma Irradiation. | Gamma Irradiation. | Same |
| 13 | Sterility<br>Assurance Level | 10-6 SAL | 10-6 SAL | Same |
| 14 | Pyrogen | Non-Pyrogenic. | Non-Pyrogenic. | Same |
| Clip | | | | |
| 15 | Suture Material | Polyethylene Terephthalate | Polyethylene Terephthalate | Same |
| 16 | Preloaded Clip | AOD1: Preloaded Gillinov-Cosgrove Clip<br>AOD2: Preloaded V-Clip | AOD1: Preloaded Gillinov-Cosgrove Clip<br>AOD2: Preloaded V-Clip | Same |
| 17 | Clip Opening | AOD1: Opens 12.5mm ± 1.5mm as<br>measured between the beams of the<br>AOD1 clip.<br>AOD2: Opens AOD2 Clip to a minimum of<br>12mm at the distal end and 4mm at the<br>proximal end of the clip. | AOD1: Opens 12.5mm ± 1.5mm as<br>measured between the beams of the<br>AOD1 clip.<br>AOD2: Opens AOD2 Clip to a minimum of<br>12mm at the distal end and 4mm at the<br>proximal end of the clip. | Same |
| 19 | Biocompatibility | Passed. Materials have been assessed<br>based on ISO 10993 and are commonly<br>employed in tissue contacting devices. | Passed. Materials have been assessed<br>based on ISO 10993 and are commonly<br>employed in tissue contacting devices. | Same |
| 20 | Sterilization | Gamma Irradiation. | Gamma Irradiation. | Same |
AtriCure added an alternate raw material supplier of the titanium for the AOD1 clip.
{5}------------------------------------------------
- The devices have the same intended use, and;
- No changes were made in operating principle, or specifications of performance.
- · The contraindications, warnings, and precautions remain the same
- The results of the verification and validation testing:
- o Demonstrated equivalency in performance
- Device biocompatibility remains unchanged o
- Did not raise any new issues of safety o
{6}------------------------------------------------
#### V. Performance Data
A review of the risk analysis concluded there is no overall change to the risk profile of the proposed AtriClip LAA Exclusion System versus the previously cleared AtriClip LAA Exclusion System as the modifications to the proposed AtriClip LAA Exclusion System do not add or remove any features of the device or change the clinical application.
The proposed changes do not include any chanqe to design or performance specifications of the AtriClip LAA Exclusion System. Additionally, the proposed changes did not modify the intended use, therefore the previously submitted bench and animal data remain valid for the AtriClip LAA Exclusion System.
#### VI. Conclusions
AtriCure has demonstrated that the modifications made to the AtriClip LAA Exclusion System are substantially equivalent in fundamental design, technology, function, device materials, packaging, sterilization, operating principle, and intended use / indication for use as the previously cleared AtriClip LAA Exclusion System) per K191413.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.