← Product Code [PSN](/submissions/SU/subpart-e%E2%80%94surgical-devices/PSN) · K062736

# SIASCOPE V (K062736)

_Astron Clinica Limited · PSN · Sep 14, 2007 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/PSN/K062736

## Device Facts

- **Applicant:** Astron Clinica Limited
- **Product Code:** [PSN](/submissions/SU/subpart-e%E2%80%94surgical-devices/PSN.md)
- **Decision Date:** Sep 14, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4580
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The SIAscope is a non-invasive skin analysis system, which provides color bitmaps called 'SIAscans' that show the relative location of blood, collagen and pigment.

## Device Story

SIAscope V is a non-invasive skin analysis system used by clinical practitioners to image skin conditions. The device utilizes Spectrophotometric Intracutaneous Analysis (SIAscopy) to investigate the skin's interior structure. It captures input via a 3.2-megapixel color digital camera while illuminating the skin with visible and near-infrared LEDs. The device transforms remitted light intensity measurements into synthesized graphical outputs called 'SIAscans'. These scans display the relative distribution of blood, collagen, and pigment. The device connects to a PC via USB 2.0 for processing and display. By providing these visual maps, the device assists clinicians in evaluating skin conditions, potentially aiding in clinical decision-making compared to traditional visual inspection or biopsy. The system is designed for ease of use with an updated user interface compared to its predecessor.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological comparison and design specifications.

## Technological Characteristics

Non-invasive skin analysis system. Materials: Glass, ABS, and TPE plastics. Illumination: Visible and near-infrared LEDs. Imager: 3.2-megapixel CMOS sensor. Connectivity: USB 2.0 to PC. Power: USB-powered. Form factor: Handheld, 250g. Software: DERMETRICS™ platform with unchanged SIAscopy algorithms.

## Regulatory Identification

A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

## Predicate Devices

- SIAscope II ([K023729](/device/K023729.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with medicine and healthcare. The caduceus features a staff with two snakes entwined around it.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 10, 2017

Astron Clinica Limited c/o Smith Associates Mr. E.J. Smith 1676 Village Green, Suite A Crofton. Maryland 21114

Re: K062736

Trade/Device Name: SIAscope V Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: II Product Code: PSN Dated: June 15, 2007 Received: June 18, 2007

Dear Mr. Smith:

This letter corrects our substantially equivalent letter of September 14, 2007.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ); medical device reporting (reporting of medical

{1}------------------------------------------------

Page 2 - Mr. E. J. Smith

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known): K062736

Device Name: SIAscope V

Indications for Use:

The SIAscope is a non-invasive skin analysis system, which provides color bitmaps called 'SIAscans' that show the relative location of blood, collagen and pigment'.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sig: Off)                                        | Page 1 of 1 |
|------------------------------------------------------------|-------------|
| Division of General, Restorative, and Neurological Devices |             |
| 10(k) Number                                               | L062736     |

3

{3}------------------------------------------------

62736

SEP 1 4 2007

# 510(k) Summary

#### September 6, 2007

Company Name & Address: Astron Clinica Ltd The Mount Toft Cambridge, United Kingdon CB3 7RL

Contact Name:

E.J. Smith 1676 Village Green, Suite A Crofton, MD 21114 Phone: 888-729-9674 Fax: 410-793-0448 Email: ESmith9746@aol.com

Date Summary Prepared:

August 31, 2006

Device Identification:

| Classification Name:               | Light, Surgical, Floor Standing |
|------------------------------------|---------------------------------|
| Common/Usual Name:                 | Skin Analysis System            |
| Proprietary Name:                  | SIAscope                        |
| Establishment Registration Number: | 3004519907                      |
| Regulation Number:                 | 21 CFR 878.4580                 |
| Product Code:                      | FSS                             |
| Classification Panel:              | General & Plastic Surgery       |
| Device Classification:             | Class II                        |

#### Intended Use

The SIAscope is a non-invasive skin analysis system, which provides color bitmaps called 'SIAscans' that show the relative location of blood, collagen and pigment.

CAUTION : This device should only be used to scan the patient's skin.

{4}------------------------------------------------

# KO62736

#### Device Overview

The SIAscope is a quick, non-invasive device for imaging skin conditions. It uses a technique known as 'Spectrophotometric Intracutaneous Analysis' (SIAscopy) to identify and display graphically the separate components of the skin. SIAscopy uses a digital camera and light (both visible and near-infrared) to investigate the skin's interior structure.

The SlAscope operates by illuminating the skin with LEDs and measuring the intensity of remitted light. The intensity of the illumination is similar to the emission from a typical hand held torch or remote control unit.

#### Description of SIAscans

Unlike biopsy, SIAscopy provides an array of readings of the skin that are displayed in graphical form creating a synthesized image called a SIAscan.

#### Predicate Product Comparison

The SIAscope V operates in a very similar fashion to the predicate device. Where alterations have been make this have been design to improve the imaging performance, reliability and safety of the device over and above the predicate device.

Although the design of the user interface has changed, these changes have been made to make the device easier and more intuitive to use for the clinical practitioner. However, the software algorithms and theory of SIAscopy is the same as found in the predicate device.

We assert that the SIAscope V substantially equivalent to the SIAscope II (K023729)

| Parameter         | SIAscope V                                                                                                                                                                  | SIAscope II (K023729)                                                                                                                                           |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code      | FSS                                                                                                                                                                         | FSS                                                                                                                                                             |
| Intended Use      | SIAscope is a non-invasive skin<br>analysis system, which provides<br>color bitmaps called 'SIAscans' that<br>show the relative location of blood,<br>collagen and pigment. | SIAscope is a non-invasive skin<br>analysis system, which provides a<br>synthesized 'image' showing the<br>relative location of blood, collagen<br>and pigment. |
| Illumination For: | Skin                                                                                                                                                                        | Skin                                                                                                                                                            |
| Illumination:     | LED                                                                                                                                                                         | LED                                                                                                                                                             |
| Imager Chip:      | CMOS                                                                                                                                                                        | CCD                                                                                                                                                             |

### Summary comparison of features and specification SIAscope V vs. SIAscope II

{5}------------------------------------------------

# 长662736

| Power Source:                      | USB via PC                                                             | AC                          |
|------------------------------------|------------------------------------------------------------------------|-----------------------------|
| External Materials:<br>(Caseworks) | Glass, ABS & TPE Plastics                                              | Glass, Aluminium            |
| PC Connection:                     | USB 2.0                                                                | Proprietary Umbilical Cable |
| Weight:                            | 250grams                                                               | 1Kg                         |
| Calibration Method:                | Factory Calibrated                                                     | Factory Calibrated          |
| Viewfinder:                        | Colour (3.2 MegaPixel)                                                 | Monochrome (0.7 MegaPixel)  |
| Software:                          | DERMETRICS™<br>- New User Interface<br>- SIAscopy algorithms unchanged | SIAscope II                 |

·

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/PSN/K062736](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/PSN/K062736)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
